This document is a solicitation for the upgrade of the JERON Nurse Call System at the Clement J. Zablocki VA Medical Center in Milwaukee, Wisconsin. The Department of Veterans Affairs is seeking a contractor to remove and replace existing JERON 680 system components with JERON 790 system equipment and integrate it into the existing VAMC 790 Nurse Call system. The project covers upgrades to the 1AN Clinic, 3C, 4AN, 5C, 6C, 7C, 9A, 9C and Spinal Cord Injury Center (Building 144) areas. Key requirements include project management services, software programming, service and installation, and training. The solicitation has a 180-day period of performance from the date of award. Pricing is requested for each individual building/area upgrade on a fixed-price basis. The procurement is unrestricted, but includes an evaluation factor for service-disabled veteran-owned and veteran-owned small businesses. Offers are due by July 11, 2024.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| P03 - JA REDACTED_Redacted.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
36C25224Q0559
PAGE 1 OF
1. REQUISITION NO.
2. CONTRACT NO.
3. AWARD/EFFECTIVE DATE
4. ORDER NO.
5. SOLICITATION NUMBER
6. SOLICITATION ISSUE DATE
a. NAME
b. TELEPHONE NO. (No Collect Calls)
8. OFFER DUE DATE/LOCAL
TIME
9. ISSUED BY
CODE
10. THIS ACQUISITION IS
UNRESTRICTED OR
SET ASIDE:
% FOR:
SMALL BUSINESS
HUBZONE SMALL
BUSINESS
SERVICE-DISABLED
VETERAN-OWNED
SMALL BUSINESS
WOMEN-OWNED SMALL BUSINESS
(WOSB) ELIGIBLE UNDER THE WOMEN-OWNED
SMALL BUSINESS PROGRAM
EDWOSB
8(A)
NAICS:
SIZE STANDARD:
11. DELIVERY FOR FOB DESTINA-
TION UNLESS BLOCK IS
MARKED
SEE SCHEDULE
12. DISCOUNT TERMS
13a. THIS CONTRACT IS A
RATED ORDER UNDER
DPAS (15 CFR 700)
13b. RATING
14. METHOD OF SOLICITATION
RFQ
IFB
RFP
15. DELIVER TO
CODE
16. ADMINISTERED BY
CODE
17a. CONTRACTOR/OFFEROR
CODE
FACILITY CODE
18a. PAYMENT WILL BE MADE BY
CODE
TELEPHONE NO.
UEI:
EFT:
PHONE:
FAX:
17b. CHECK IF REMITTANCE IS DIFFERENT AND PUT SUCH ADDRESS IN OFFER 18b. SUBMIT INVOICES TO ADDRESS SHOWN IN BLOCK 18a UNLESS BLOCK BELOW IS CHECKED
SEE ADDENDUM
19.
20.
21.
22.
23.
24.
ITEM NO.
SCHEDULE OF SUPPLIES/SERVICES
QUANTITY
UNIT
UNIT PRICE
AMOUNT
(Use Reverse and/or Attach Additional Sheets as Necessary)
25. ACCOUNTING AND APPROPRIATION DATA
26. TOTAL AWARD AMOUNT (For Govt. Use Only) 27a. SOLICITATION INCORPORATES BY REFERENCE FAR 52.212-1, 52.212-4. FAR 52.212-3 AND 52.212-5 ARE ATTACHED. ADDENDA
ARE
ARE NOT ATTACHED.
27b. CONTRACT/PURCHASE ORDER INCORPORATES BY REFERENCE FAR 52.212-4. FAR 52.212-5 IS ATTACHED. ADDENDA
ARE
ARE NOT ATTACHED
28. CONTRACTOR IS REQUIRED TO SIGN THIS DOCUMENT AND RETURN _______________
29. AWARD OF CONTRACT: REF. ___________________________________ OFFER
COPIES TO ISSUING OFFICE. CONTRACTOR AGREES TO FURNISH AND
DATED ________________________________. YOUR OFFER ON SOLICITATION
DELIVER ALL ITEMS SET FORTH OR OTHERWISE IDENTIFIED ABOVE AND ON ANY
(BLOCK 5), INCLUDING ANY ADDITIONS OR CHANGES WHICH ARE
ADDITIONAL SHEETS SUBJECT TO THE TERMS AND CONDITIONS SPECIFIED
SET FORTH HEREIN IS ACCEPTED AS TO ITEMS:
30a. SIGNATURE OF OFFEROR/CONTRACTOR 31a. UNITED STATES OF AMERICA (SIGNATURE OF CONTRACTING OFFICER) 30b. NAME AND TITLE OF SIGNER (TYPE OR PRINT) 30c. DATE SIGNED 31b. NAME OF CONTRACTING OFFICER (TYPE OR PRINT) 31c. DATE SIGNED
AUTHORIZED FOR LOCAL REPRODUCTION
(REV. NOV 2021)
PREVIOUS EDITION IS NOT USABLE
Prescribed by GSA - FAR (48 CFR) 53.212
7. FOR SOLICITATION
INFORMATION CALL:
STANDARD FORM 1449
SOLICITATION/CONTRACT/ORDER FOR COMMERCIAL PRODUCTS AND COMMERCIAL SERVICES
OFFEROR TO COMPLETE BLOCKS 12, 17, 23, 24, & 30
None 36C25224Q0559 Matthew Hohl 414-844-4816 07- -2024
11:59PM
CDT
Department of Veterans Affairs
GLAC (69D)
115 South 84th Street, Suite 101 Milwaukee WI 53214-1476 X X 334290 800 Employees N/A X Attn: TBD Clement J. Zablocki VA Medical Center 5000 W National Ave Milwaukee, WI 53295
Department of Veterans Affairs
GLAC-MILWAUKEE (69D)
Great Lakes Acquisition Center - HC 115 South 84th Street, Suite 101 Milwaukee WI 53214-1476
Department of Veterans Affairs Financial Services Center PO Box 149971 Austin TX 78714-9971 See CONTINUATION Page Nurse Call System (Brand Name Only) See Continuation Pages See CONTINUATION Page X X Matthew Hohl
SECTION B - CONTRACT ADMINISTRATION DATA
B.1 CONTRACT ADMINISTRATION
All contract administration matters will be handled by the following individuals:
| A. CONTRACTOR: | ________________________________ | (Contractor’s Name) | |
| ________________________________ | (Address) | ||
| ________________________________ | (City-State-Zip) | ||
| ________________________________ | (Point of Contact/Title) | ||
| ________________________________ | (Phone/FAX Number) | ||
| ________________________________ | (E-Mail Address) | ||
| ________________________________ | (SAM UEI Number) |
B. GOVERNMENT: Matthew Hohl, CONTRACTING SPECIALIST
DEPARTMENT OF VETERANS AFFAIRS
GREAT LAKES ACQUISITION CENTER
115 S. 84TH ST., SUITE 100
MILWAUKEE WI 53214-1476
PHONE NUMBER: 414-844-4816
EMAIL: Matthew.Hohl@VA.GOV
B.2 CONTRACTOR REMITTANCE ADDRESS
All payments by the Government to the contractor will be made in accordance with:
52.232-32, Payment by Electronic Funds Transfer - System for Award Management
B.3 INVOICES
Invoices shall be submitted in arrears in accordance with:
1. 852.232-72 ELECTRONIC SUBMISSION OF PAYMENT REQUESTS (NOV 2018)
1. UPON ACCEPTANCE OF ITEMS
B.4 GOVERNMENT INVOICE ADDRESS
All invoices from the contractor shall be submitted electronically in accordance with 852.232-72 ELECTRONIC SUBMISSION OF PAYMENT REQUESTS (NOV 2018)
FACSIMILE, E-MAIL, AND SCANNED DOCUMENTS ARE NOT ACCEPTABLE FORMS OF SUBMISSION FOR PAYMENT REQUESTS.
For assistance setting up e-Invoice, the below information is provided:
* Vendors must register online at http://www.tungsten-network.com/veteransaffairs/
* Tungsten Network Setup Information: 877-489-6135
* Tungsten Network email: VA.Registration@tungsten-network.com
* FSC Vendor Support Section Contact: 1-877-353-9791
* FSC Vendor Support Section email: vafsccshd@va.gov
Vendors can inquire on the status of payments by accessing VIS at: https://www.vis.fsc.va.gov/login.aspx?ReturnUrl=%2fDefault.aspx
INVOICES: Invoices shall be submitted in arrears:
| a. Quarterly _______ |
| b. Semi-Annually _______ |
| c. Other X At Receipt of Acceptance Supplies |
GOVERNMENT INVOICE ADDRESS: All invoices from the contractor shall be mailed to the following address:
| DEPARTMENT OF VETERANS AFFAIRS |
| FINANCIAL SERVICES CENTER |
| PO BOX 149971 |
| AUSTIN, TX 78714-9975 |
Vendor inquires: Toll Free Number 1-877-353-9791
B.5 BASIS OF AWARD
The Government will make the award to the lowest price technically acceptable offer.
B.6 SECURITY & PRIVACY CONTROL
The Certification & Accreditation (C&A) requirements of VHA Handbook 6500.6 do not apply to this requirement; therefore, a Security Accreditation package is not required. No C&A or MOU/ISA is required. Sensitive information exposure as part of this contract involves applicable security controls within the facility as part of the VA Information Security Program.
If there are any questions related to privacy, please have a member of your staff contact the VA Privacy Service at 202-461-6309. If there are any questions related to the information security, please have a member of your staff contact the Office of Cyber Security at 304-262-7733.
ACKNOWLEDGMENT OF AMENDMENTS: The offeror acknowledges receipt of amendments to the Solicitation numbered and dated as follows:
| AMENDMENT NO |
| DATE |
B.7 Statement of Work Purpose The Clement J. Zablocki VA Medical Center’s 1AN Clinic, 3C, 4AN, 5C, 6C, 7C, 9A, 9C and Spinal Cord Injury (SCI) Center (Building 144) requires an upgrade of the JERON Nurse Call System. All existing JERON 680 system components within the work area shall be removed and replaced with JERON 790 system equipment and integrated into existing VAMC 790 Nurse call system(s).
Project Scope:
Project Management Services
1. Contractor shall provide project management services to coordinate all aspects of installation. The contractor shall provide an installation schedule within 30 days of award that outlines a project completion within 180 days of award.
a. Project Milestones – Contractor schedule shall include the following milestones.
i. Specifications and revised drawings
ii. Equipment Delivery
iii. Installation of New Equipment
iv. Server Configuration
1. Test compatibility with existing JERON 790 System
v. Interface configuration
vi. Training Completion
vii. System Verification and Testing
1. Upon installation of hardware in each room, the Nurse Call Station shall be tested to verify system operation.
2. Verify that alarms are successfully sent to Vocera when buttons are pressed.
2. Programming- Contractor shall provide all necessary software on the JERON Nurse call system to ensure the systems are functioning and can communicate with the centralized database and Vocera.
a. All licensing necessary will be provided by the contractor.
b. Contractor shall program integration of new JERON 790 hardware into existing JERON Nurse Call System.
3. Service- Contractor shall provide all labor, time, materials, and equipment necessary to assemble, test, and install the equipment at the VA during hours of operation
a. Conduct system verification and testing as specified in Project Milestones Nurse Call system 1st Floor A-wing North Outpatient Clinic
1. 1AN Existing Hardware to be removed for Upgrade- refer to the 1AN Floorplans for locations.
a. Twenty-four (24) Nurse Call Pull Cords
b. Twenty-four (24) room zone lights in existing locations
c. Fifteen (15) Room Zone Light Switches
d. Four (4) corridor zone lights in existing locations
e. Three (3) Staff and Nurse Duty stations
2. 1AN New System Hardware for Installation
a. All the following required equipment shall be tied into the existing infrastructure installed during headend project and added to the existing server. The Installation Floor Plan drawing for 1AN contains the locations of all hardware to be installed. Contractor shall furnish and install the following new equipment- refer to the 1AN Floorplans for locations:
i. Eighteen (18) Combination Room Controller and Prism Dome Light (model 7953
1. Case shall be Flame retardant UL 94-5VA PC/ABS blend enclosure.
2. Shall have a two-gang backbox with a two-gang trim ring.
3. Shall have the following dimensions (H x W x D): 4.8” x 4.9” x 3.7” (12.1 x 12.4 x 9.3 cm)
4. Shall have eight (8) ultra-bright LED indicators:
a. Two per Dome Light segment, capable of producing up to eight distinct colors (white, red, yellow, orange, green, blue, magenta, and pink)
5. Shall have an electrical rating of 40V at 1.2 W.
6. Shall have the following terminations:
a. Three (3) RJ-45
b. One (1) plug on two-conductor screw terminal.
7. Shall have a gasket included.
8. Shall have the following certifications:
a. UL 1069 Listed
ii. Eighteen (18) Procedure, Presence, and Bed Management Station (model 7940)
1. Case shall be Flame retardant UL 94-5VA PC/ABS blend enclosure.
2. Shall have a one-gang backbox or two-gang backbox with a one-gang trim ring.
3. Shall have the following dimensions (H x W x D): 4.6” x 3” x 1” (11.7 x 7.6 x 2.5 cm)
4. Shall have the following controls/indicators:
a. Four (4) Call/Event/Present push buttons with LED indicators
b. One (1) Cancel Button
c. One (1) Low Ambient Light LED
5. Shall have an electrical rating of 5V @ 25mW.
6. Shall have the following terminations:
a. One (1) plug on two-conductor screw terminal
7. Shall have the following certifications:
a. UL 1069 Listed
iii. Two (2) pullcord and pushbutton bathroom stations (Model 7958)
1. Case shall be Flame retardant UL 94-5VA PC/ABS blend enclosure
2. The pull cord shall have the length of 6’ (1.8 m), non-porous coated braided polyester cord.
3. Shall have a one-gang backbox or two-gang backbox with one-gang trim ring.
4. Shall have the following dimensions (H x W x D): 4.6” x 3” x 1” (11.7 x 7.6 x 2.6 cm).
5. Shall have the following controls/indicators:
a. One (1) configurable call button with call placed LED
b. One (1) 6’ (1.8m) blue PVC coated pullcord
c. One (1) cancel button
d. One (1) low ambient light LED
6. Shall have an electrical rating of 5V at 25Mw.
7. Shall have the following terminations:
a. One (1) plug on two-conductor screw terminal
8. Shall have a gasket included.
9. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL Canadian Standard C22.22 NO.205
iv. Three (3) Zone Corridor Lights
v. Nurse Gateways and Nurse Call Power Supply with Integrated Battery Backup (Model 7993 and Model 7995)
vi. One (1) additional software license for the NET addition of one (1) gateway 3rd Floor C-wing Inpatient Unit
1. 3C Existing Hardware requiring removal for upgrade
a. One hundred fifty (150) pull cord patient stations in patient rooms and bathrooms
b. Twenty-six (26) Staff Emergency Buttons
c. Twelve (12) Staff and Duty Stations
d. Five (5) Staff Stations
e. Three (3) JERON Nurse Consoles
f. Eleven (11) Corridor zone lights
g. Fifty-one (51) room zone lights in existing locations
h. One (1) 680 Main controller with 790 conversion card (gateway) and software licensing for the 790 EIS Reporting System.
2. 3C New System Hardware requiring install during the upgrade
a. All of the following required equipment shall be tied into the existing infrastructure installed during headend project and added to the existing server.
b. All pullcords in patient’s rooms and bathrooms will be replaced and upgraded with the bed-level and floor-level pushbuttons (Model 7920-PC and Model 7921-PC).
c. Vendor shall perform all additional labor to replace and center the headwall buttons (Item 2.d.i) over the location of the patient bed in each room. Additionally, the vendor shall perform all labor to add all floor-level buttons (Item 2.d.ii).
d. The Installation Floor Plan drawing for 3C contains the locations of all hardware to be installed. Contractor shall furnish and install the following new equipment:
i. Ninety (90) JERON Bed-level push button patient stations, (Model 7920-PC)
1. Located 38” to 44” from the floor
ii. Ninety (90) JERON Floor-level push button patient stations, (Model 7921-PC)
1. Located 12” from the floor
iii. Thirty-three (33) New Staff Emergency Buttons (Model 7930)
1. Shall be Flame-retardant UL 94-5VA PC/ABS blend enclosure.
2. Shall have one-gang back box or two-gang gang backbox with one-gang trim ring
3. Shall have the following dimensions (H x W x D): 4.6” x 3” x 1” (11.7 x 7.6 x 2.5 cm)
4. Shall have the following controls/indicators:
a. One (1) configurable call button with call placed LED
i. Staff Emergency
b. One (1) Cancel Button
c. One (1) Low ambient light LED
5. Shall have an electrical rating of 5V at 25mW
6. Shall have the terminations:
a. One (1) Plug on two-conductor screw terminal
7. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL Canadian Standard C22.22 NO.205
iv. Four (4) New Staff and Duty stations (Model: 7927)
1. Shall be flame retardant UL 94-5VA PC/ABS blend enclosure
2. Shall have a two-gang backbox with a two-gang trim ring.
3. Shall the following dimensions (H x W x D): 4.8” x 4.8” x 1.2” (12.2 x 12.2 x 3 cm)
4. Shall have the following controls/indicators:
a. One (1) Green “Routine” LED Call Indicator
b. One (1) Amber “Urgent” LED Call Indicator
c. One (1) Red “Emer” LED Call Indicator
d. One (1) Blue “Code” LED Call Indicator
e. One (1) Call button with call placed LED
f. One (1) Combinator tone mute/tone volume/cancel button
g. One (1) Monitor LED
h. One (1) Low ambient light LED
5. Shall have a 2.75” (7 CM); 8 Ohm speaker
6. Shall have an electret microphone
7. Shall have a 5V @ 85mW
8. Shall have a 1-RJ45 termination
9. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL Canadian Standard C22.22 NO.125
v. Two (2) New Staff Stations (Model 7912)
1. Shall be flame retardant UL 94-5VA PC/ABS blend enclosure
2. Shall have a two-gang backbox with a two-gang trim ring.
3. Shall the following dimensions (H x W x D): 4.8” x 4.8” x 1.2” (12.2 x 12.2 x 3 cm)
4. Shall have the following controls/indicators:
a. One (1) Call button with call placed LED
b. One (1) Cancel button
c. One (1) Monitor LED
d. One (1) Low ambient light LED
5. Shall have a 2.75” (7 CM); 8 Ohm speaker
6. Shall have an electret microphone
7. Shall have a 5V @ 85mW
8. Shall have a 1-RJ45 termination
9. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL Canadian Standard C22.22 NO.125
vi. Three (3) New JERON Nurse Consoles (Model 7965)
vii. Fifty (50) Room Zone Lights (Model 7953)
viii. Eleven (11) Area/Corridor Zone Lights (Model 7973)
ix. One (1) New gateway (Model 7993)
x. One (1) Power supply for gateway (Model 7995)
xi. One (1) additional software license for the NET addition of one (1) gateway
4th Floor A-wing North Pain Clinic
3. 4AN New System Hardware requiring install during the upgrade
a. All the following required equipment shall be tied into the existing infrastructure installed during headend project and added to the existing server. The Installation Floor Plan drawing for 4AN contains the locations of all hardware to be installed. Contractor shall furnish and install the following new equipment:
i. Eight (8) Callcord, 1/4” Plug (Model 8833/8835)
1. Shall have a pendant and cord, grey molded thermoplastic pendent with grey vinyl cord
2. Shall have the minimum length of 10 feet
3. Shall have the following controls:
a. Call button: Momentary contact
4. Shall have a ¼ inch (0.64cm) connector plug
5. Shall have a removable white nylon strap with spring-type, plated metal bed clip
6. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL-Canadian Standard C22.22 NO.125
ii. Eight (8) Dual Auxiliary Alarm Input Stations (Model 7941)
1. Shall have one-gang back box or two-gang gang backbox with one-gang trim ring
2. Shall have the following dimensions (H x W x D) 4.6” x 3” x 1” (11.7 x 7.6 x 2.5 cm)
3. Shall have the following controls/indicators:
a. Two (2) ¼” jack inputs with call placed LEDs
b. One (1) Cancel button
c. One (1) Low ambient light LED
4. Shall have an electrical rating of 5V at 25mW
5. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL-Canadian Standard C22.22 NO.205
iii. Ten (10) Single call/staff emergency buttons (Model 7930)
1. Shall be Flame-retardant UL 94-5VA PC/ABS blend enclosure.
2. Shall have one-gang back box or two-gang gang backbox with one-gang trim ring
3. Shall have the following dimensions (H x W x D): 4.6” x 3” x 1” (11.7 x 7.6 x 2.5 cm)
4. Shall have the following controls/indicators:
a. One (1) configurable call button with call placed LED
i. Staff Emergency
b. One (1) Cancel Button
c. One (1) Low ambient light LED
5. Shall have an electrical rating of 5V at 25mW
6. Shall have the terminations:
a. One (1) Plug on two-conductor screw terminal
7. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL Canadian Standard C22.22 NO.205
iv. Two (2) Staff Stations (Model 7912)
1. Shall be flame retardant UL 94-5VA PC/ABS blend enclosure
2. Shall have a two-gang backbox with a two-gang trim ring.
3. Shall the following dimensions (H x W x D): 4.8” x 4.8” x 1.2” (12.2 x 12.2 x 3 cm)
4. Shall have the following controls/indicators:
a. One (1) Call button with call placed LED
b. One (1) Cancel button
c. One (1) Monitor LED
d. One (1) Low ambient light LED
5. Shall have a 2.75” (7 CM); 8 Ohm speaker
6. Shall have an electret microphone
7. Shall have a 5V @ 85mW
8. Shall have a 1-RJ45 termination
9. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL Canadian Standard C22.22 NO.125
v. One (1) Staff and Duty stations (Model: 7927)
1. Shall be flame retardant UL 94-5VA PC/ABS blend enclosure
2. Shall have a two-gang backbox with a two-gang trim ring.
3. Shall the following dimensions (H x W x D): 4.8” x 4.8” x 1.2” (12.2 x 12.2 x 3 cm)
4. Shall have the following controls/indicators:
a. One (1) Green “Routine” LED Call Indicator
b. One (1) Amber “Urgent” LED Call Indicator
c. One (1) Red “Emer” LED Call Indicator
d. One (1) Blue “Code” LED Call Indicator
e. One (1) Call button with call placed LED
f. One (1) Combinator tone mute/tone volume/cancel button
g. One (1) Monitor LED
h. One (1) Low ambient light LED
5. Shall have a 2.75” (7 CM); 8 Ohm speaker
6. Shall have an electret microphone
7. Shall have a 5V @ 85mW
8. Shall have a 1-RJ45 termination
9. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL Canadian Standard C22.22 NO.125
vi. One (1) Pullcord and Pushbutton Station (Model 7932)
1. Shall be flame retardant UL 94-5VA PC/ABS blend enclosure
2. Shall have a 6 feet (1.8 m) non-porous coated braided polyester cord
3. Shall have a one-gang backbox or two-gang backbox with one-gang trim ring
4. Shall have the following dimensions (H x W x D): 4.6” x 3”x 1” (11.7 x 7.6 x 2.5 cm)
5. Shall have the following controls/indicators:
a. Two (2) Configurable call buttons with call placed LED
b. One (1) 6’ (1.8 m) blue PVC coated pullcord
c. One (1) Cancel button
d. One (1) low ambient light LED
6. Shall have a 5V @25mV electrical rating
7. Shall have one (1) plug on two-conductor screw terminal termination
8. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL Canadian Standard C22.22 NO.125
vii. Eleven (11) Dome Room Zone Lights (Model 7953)
viii. Nurse Gateways and Nurse Call Power Supply with Integrated Battery Backup (Model 7993 and Model 7995)
ix. One (1) additional software license for the NET addition of one (1) gateway 5th Floor C-wing Inpatient Unit
1. 5C Existing Hardware requiring removal for upgrade.
a. Fifteen (15) patient stations in patient rooms
b. Fifteen (15) Pillow speakers in patient rooms
c. Thirty (30) Dual Auxiliary Alarms in patient rooms
d. Fifteen (15) CODE buttons
e. Fifteen (15) Staff Emergency Buttons
f. Nine (9) Staff and Duty Stations
g. Sixteen (16) room zone lights in existing locations
h. Five (5) corridor zone lights in existing locations
i. Thirty-two (32) bathroom pull cords.
j. One (1) 680 Main controller with 790 conversion card (gateway) and software licensing for the 790 EIS Reporting System.
2. 5C New System Hardware requiring install during the upgrade.
a. All the following required equipment shall be tied into the existing infrastructure installed during headend project and added to the existing server. The Installation Floor Plan drawing 5C contains the locations of all hardware to be installed. Contractor shall furnish and install the following new equipment:
i. Fifteen (15) JERON Model 7923 patient stations, that shall meet the following criteria:
1. Shall have a three-gang backbox with three-gang trim ring.
2. Shall have the following dimensions (H x W x D) 4.6” x 6.8” x 1.2” (11.7 x 17.2 x 3 cm)
3. Shall have the following controls/indicators:
a. One (1) DIN jacks with call placed LEDs.
b. One (1) ¼” jack with call placed LED.
c. One (1) Feature bed status and call
d. Two (2) Configurable call buttons with call placed LEDs.
e. One (1) Cancel button with remote call cancel LED.
f. One (1) Monitor LED
g. Two (2) Low ambient light LED
4. Speaker shall have width of 2.75” (7cm) with 8 Ohm.
5. Microphone shall be Electret.
6. Shall have an electrical rating of 5V at 200mW.
7. Shall have the following terminations:
a. Two (2) RJ45
b. Three (3) Plug on IDC multi-conductions.
8. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL-Canadian Standard C22.22 NO.205
ii. Fifteen (15) New Standard pillow speakers (Model 7901/7902)
1. Case shall be Flame-retardant ABS plastic.
2. Shall have the following dimensions (H x W x D) 8.4” x 3” x 1.25” (21.3 x 7.6 x 3 cm)
3. Shall have the following control buttons:
a. Nurse CALL button, TV power, Channel Up/Down arrows, Closed Captioning, Mute, Sleep
b. Two Auxiliary buttons
iii. Thirty (30) New AUX input stations (Model 7941)
1. Shall have one-gang back box or two-gang gang backbox with one-gang trim ring.
2. Shall have the following dimensions (H x W x D) 4.6” x 3” x 1” (11.7 x 7.6 x 2.5 cm)
3. Shall have the following controls/indicators:
a. Two (2) ¼” jack inputs with call placed LEDs.
b. One (1) Cancel button.
c. One (1) Low ambient light LED
4. Shall have an electrical rating of 5V at 25mW.
5. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL-Canadian Standard C22.22 NO.205
iv. Fifteen (15) New CODE buttons (Model 7949)
v. Fifteen (15) New Staff Emergency Buttons (Model 7930)
vi. Fifteen (15) Room Zone Lights (Model 7953)
vii. Five (5) Area/Corridor Zone Lights (Model 7973)
viii. Thirty-two (32) New bath and shower stations (Model 7958)
ix. One (1) New gateway (Model 7993)
x. One (1) Power supply for gateway (Model 7995)
xi. One (1) additional software license for the NET addition of one (1) gateway 6th Floor C-wing Inpatient Unit
1. 6C Existing Hardware requiring removal for upgrade.
a. Twenty-eight (28) patient stations in patient rooms
b. Twenty-eight (28) Pillow speakers in patient rooms
c. Fifty-six (56) Dual Auxiliary Alarms in patient rooms
d. Twenty-eight (28) CODE buttons
e. Twenty-eight (28) Staff Emergency Buttons
f. Seventeen (17) Staff and Duty Stations
g. Thirty-one (31) room zone lights in existing locations
h. Six (6) corridor zone lights in existing locations
i. Sixty-one (61) bathroom pull cords.
j. One (1) 680 Main controller with 790 conversion card (gateway) and software licensing for the 790 EIS Reporting System.
2. 6C New System Hardware requiring install during the upgrade.
e. All the following required equipment shall be tied into the existing infrastructure installed during headend project and added to the existing server. The Installation Floor Plan drawing for 6C contains the locations of all hardware to be installed. Contractor shall furnish and install the following new equipment:
i. Twenty-eight (28) JERON Model 7923 patient stations, that shall meet the following criteria:
1. Shall have a three-gang backbox with three-gang trim ring.
2. Shall have the following dimensions (H x W x D) 4.6” x 6.8” x 1.2” (11.7 x 17.2 x 3 cm)
3. Shall have the following controls/indicators:
a. One (1) DIN jacks with call placed LEDs.
b. One (1) ¼” jack with call placed LED.
c. One (1) Feature bed status and call
d. Two (2) Configurable call buttons with call placed LEDs.
e. One (1) Cancel button with remote call cancel LED.
f. One (1) Monitor LED
g. Two (2) Low ambient light LED
4. Speaker shall have width of 2.75” (7cm) with 8 Ohm.
5. Microphone shall be Electret.
6. Shall have an electrical rating of 5V at 200mW.
7. Shall have the following terminations:
a. Two (2) RJ45
b. Three (3) Plug on IDC multi-conductions.
8. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL-Canadian Standard C22.22 NO.205
ii. Twenty-eight (28) New Standard pillow speakers (Model 7901)
1. Case shall be Flame-retardant ABS plastic.
2. Shall have the following dimensions (H x W x D) 8.4” x 3” x 1.25” (21.3 x 7.6 x 3 cm)
3. Shall have the following control buttons:
a. Nurse CALL button, TV power, Channel Up/Down arrows, Closed Captioning, Mute, Sleep
b. Two Auxiliary buttons
iii. Fifty-six (56) New AUX input stations (Model 7941)
1. Shall have one-gang back box or two-gang gang backbox with one-gang trim ring.
2. Shall have the following dimensions (H x W x D) 4.6” x 3” x 1” (11.7 x 7.6 x 2.5 cm)
3. Shall have the following controls/indicators:
a. Two (2) ¼” jack inputs with call placed LEDs.
b. One (1) Cancel button.
c. One (1) Low ambient light LED
4. Shall have an electrical rating of 5V at 25mW.
5. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL-Canadian Standard C22.22 NO.205
iv. Twenty-eight (28) New CODE buttons (Model 7949)
v. Twenty-eight (28) New Staff Emergency Buttons (Model 7930)
vi. Two (2) New Staff and Duty stations (Model: 7927)
vii. Thirty (30) Room Zone Lights (Model 7953)
viii. Six (6) Area/Corridor Zone Lights (Model 7973)
ix. Fifty-nine (59) New bath and shower stations (Model 7958)
x. One (1) New gateway (Model 7993)
xi. One (1) Power supply for gateway (Model 7995)
xii. One (1) additional software license for the NET addition of one (1) gateway 7th Floor C-wing Inpatient Unit
3. 7C Existing Hardware requiring removal for upgrade
a. Thirty (30) patient stations in patient rooms
b. Thirty (30) Pillow speakers in patient rooms
c. Sixty (60) Dual Auxiliary Alarms in patient rooms
d. Thirty-one (31) CODE buttons
e. Thirty-one (31) Staff Emergency Buttons
f. Eight (8) Staff and Duty Stations
g. Thirty-one (31) room zone lights in existing locations
h. Five (5) corridor zone lights in existing locations
i. Sixty-two (62) bathroom pull cords
j. One (1) 680 Main controller with 790 conversion card (gateway) and software licensing for the 790 EIS Reporting System.
4. 7C New System Hardware requiring install during the upgrade
a. All of the following required equipment shall be tied into the existing infrastructure installed during headend project and added to the existing server. The Installation Floor Plan drawing for 7C contains the locations of all hardware to be installed. Contractor shall furnish and install the following new equipment:
i. Thirty (30) JERON Model 7923 patient stations, that shall meet the following criteria:
1. Shall have a three-gang backbox with three-gang trim ring
2. Shall have the following dimensions (HxWxD) 4.6” x 6.8” x 1.2” (11.7 x 17.2 x 3 cm)
3. Shall have the following controls/indicators:
a. One (1) DIN jacks with call placed LEDs
b. One (1) ¼” jack with call placed LED
c. One (1) Feature bed status and call
d. Two (2) Configurable call buttons with call placed LEDs
e. One (1) Cancel button with remote call cancel LED
f. One (1) Monitor LED
g. Two (2) Low ambient light LED
4. Speaker shall have width of 2.75” (7cm) with 8 Ohm
5. Microphone shall be Electret
6. Shall have an electrical rating of 5V at 200mW
7. Shall have the following terminations:
a. Two (2) RJ45
b. Three (3) Plug on IDC multi-conductions
8. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL-Canadian Standard C22.22 NO.205
ii. Thirty (30) New Standard pillow speakers (Model 7901/7902)
1. Case shall be Flame-retardant ABS plastic
2. Shall have the following dimensions (H x W x D) 8.4” x 3” x 1.25” (21.3 x 7.6 x 3 cm)
3. Shall have the following control buttons:
a. Nurse CALL button, TV power, Channel Up/Down arrows, Closed Captioning, Mute, Sleep
b. Two Auxiliary buttons
iii. Sixty (60) New AUX input stations (Model 7941)
1. Shall have one-gang back box or two-gang gang backbox with one-gang trim ring
2. Shall have the following dimensions (H x W x D) 4.6” x 3” x 1” (11.7 x 7.6 x 2.5 cm)
3. Shall have the following controls/indicators:
a. Two (2) ¼” jack inputs with call placed LEDs
b. One (1) Cancel button
c. One (1) Low ambient light LED
4. Shall have an electrical rating of 5V at 25mW
5. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL-Canadian Standard C22.22 NO.205
iv. Thirty (30) New CODE buttons (Model 7949)
v. Thirty-one (31) New Staff Emergency Buttons (Model 7930)
vi. Five (5) New Staff and Duty stations (Model: 7927)
vii. Twenty-nine (29) Room Zone Lights (Model 7953)
viii. Five (5) Area/Corridor Zone Lights (Model 7973)
ix. Fifty-six (56) New bath and shower stations (Model 7958)
x. One (1) New gateway (Model 7993)
xi. One (1) Power supply for gateway (Model 7995)
xii. One (1) additional software license for the NET addition of one (1) gateway 9th Floor A-wing Inpatient Unit
1. 9A Existing Hardware to be removed for Upgrade
a. Forty (40) Patient stations, staff emergency buttons, CODE buttons, pillow speakers, bath stations, and shower stations. Each bed has two (2) dual AUX input jacks for misc. inputs from medical equipment to nurse call system.
b. Twenty-seven (27) room zone lights in existing locations
c. Nine (9) corridor zone lights in existing locations
d. Four (4) Staff and Nurse Duty stations
e. Two (2) Touchscreen Nurse Consoles
f. One (1) 680 Main controller with 790 conversion card (gateway) and software licensing for the 790 EIS Reporting System.
2. 9A New System Hardware for Installation
a. All the following required equipment shall be tied into the existing infrastructure installed during headend project and added to the existing server. The Installation Floor Plan drawing for 9A contains the locations of all hardware to be installed. Contractor shall furnish and install the following new equipment:
i. Forty (40) patient stations with separate (40) staff emergency and (40) CODE buttons to the side of the station, one (1) zone light outside of each patient room (total of 27 rooms) and shall meet the following criteria:
1. Shall have a three-gang backbox with three-gang trim ring
2. Shall have the following dimensions (H x W x D) 4.6” x 6.8” x 1.2” (11.7 x 17.2 x 3 cm)
3. Shall have the following controls/indicators:
a. One (1) DIN jacks with call placed LEDs
b. One (1) ¼” jack with call placed LED
c. One (1) Feature bed status and call
d. Two (2) Configurable call buttons with call placed LEDs
e. One (1) Cancel button with remote call cancel LED
f. One (1) Monitor LED
g. Two (2) Low ambient light LED
4. Speaker shall have width of 2.75” (7cm) with 8 Ohm
5. Microphone shall be Electret
6. Shall have an electrical rating of 5V at 200mW
7. Shall have the following terminations:
a. Two (2) RJ45
b. Three (3) Plug on IDC multi-conductions
8. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL-Canadian Standard C22.22 NO.205
ii. Forty (40) standard pillow speakers (Model 7901/7902)
1. Case shall be Flame-retardant ABS plastic
2. Shall have the following dimensions (H x W x D) 8.4” x 3” x 1.25” (21.3 x 7.6 x 3 cm)
3. Shall have the following control buttons:
a. Nurse CALL button, TV power, Channel Up/Down arrows, Closed Captioning, Mute, Sleep
b. Two Auxiliary buttons
iii. Eighty (80) Dual Auxiliary Alarm Input Stations (Model 7941)
1. Shall have one-gang back box or two-gang gang backbox with one-gang trim ring
2. Shall have the following dimensions (H x W x D) 4.6” x 3” x 1” (11.7 x 7.6 x 2.5 cm)
3. Shall have the following controls/indicators:
a. Two (2) ¼” jack inputs with call placed LEDs
b. One (1) Cancel button
c. One (1) Low ambient light LED
4. Shall have an electrical rating of 5V at 25mW
5. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL-Canadian Standard C22.22 NO.205
iv. Fifty-four (54) pullcord and pushbutton shower stations (Model 7958)
1. Case shall be Flame retardant UL 94-5VA PC/ABS blend enclosure
2. The pull cord shall have the length of 6’ (1.8 m), non-porous coated braided polyester cord.
3. Shall have a one-gang backbox or two-gang backbox with one-gang trim ring.
4. Shall have the following dimensions (H x W x D): 4.6” x 3” x 1” (11.7 x 7.6 x 2.6 cm).
5. Shall have the following controls/indicators:
a. One (1) configurable call button with call placed LED
b. One (1) 6’ (1.8m) blue PVC coated pullcord
c. One (1) cancel button
d. One (1) low ambient light LED
6. Shall have an electrical rating of 5V at 25Mw.
7. Shall have the following terminations:
a. One (1) plug on two-conductor screw terminal
8. Shall have a gasket included.
9. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL Canadian Standard C22.22 NO.205
v. Two (2) Staff and Nurse Duty Stations (Model: 7927)
1. Case shall be flame retardant UL 94-5 VA PC/ABS blend enclosure
2. Two-gang backbox with a two-gang trim ring
3. Shall have the following dimensions (H x W x D) 4.8” x 4.8” x 1.2” (12.2 x 12.2 x 3 cm)
4. Shall have the following controls/indicators:
a. One (1) Green “Routine” LED call indicator
b. One (1) Amber “Urgent” LED call indicator
c. One (1) Red “Emer” LED call indicator
d. One (1) Blue “Code” LED call indicator
e. One (1) Call button
f. One (1) Combination tone mute/ cancel button
g. One (1) Low ambient light LED
h. One (1) Monitor LED
5. Speaker shall have width of 2.75” (7cm) with 8 Ohm
6. Microphone shall be Electret
7. Shall have an electrical rating of 5V at 85mW
8. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL-Canadian Standard C22.22 NO.205
vi. Nine (9) Zone Light Corridor Lights
vii. Two (2) Nurse Consoles (Model 7965)
1. Shall be Flame-retardant UL 94-5VA PC/ABS blend enclosure with metal base
2. Shall have the following dimensions (H x W x D): 8.3” x 9.2” x 3.0” (21 x 24.1 x 7.6 cm).
3. Shall have a desk mount footprint of 75.9in2 (489.7 cm2) or wall mount (wall mount bracket, model 9769).
4. Shall have a 5.7” (14.5 cm) display, 320 x 240 resolution color touchscreen.
5. Speaker shall have width of 2.75” (7cm) with 8 Ohm
6. Microphone shall be Electret
7. Shall have an electrical rating of 40V at 5.8W.
8. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL-Canadian Standard C22.22 NO.205
viii. Two (2) additional single call/staff emergency buttons (Model 7930)- refer to the 9A Floor Plan drawings for Installation Locations.
1. Shall be Flame-retardant UL 94-5VA PC/ABS blend enclosure.
2. Shall have one-gang back box or two-gang gang backbox with one-gang trim ring
3. Shall have the following dimensions (H x W x D): 4.6” x 3” x 1” (11.7 x 7.6 x 2.5 cm)
4. Shall have the following controls/indicators:
a. One (1) configurable call button with call placed LED
i. Staff Emergency
b. One (1) Cancel Button
c. One (1) Low ambient light LED
5. Shall have an electrical rating of 5V at 25mW
6. Shall have the terminations:
a. One (1) Plug on two-conductor screw terminal
7. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL Canadian Standard C22.22 NO.205
ix. Nurse Call Gateways and Nurse Call Power Supply with Integrated Battery Backup (Model 7993 AND Model 7995)
x. One (1) additional software license for the NET addition of one (1) gateway 9th Floor C-wing Inpatient Unit
1. 9C Existing Hardware to be removed for Upgrade
a. Forty (40) Patient stations, staff emergency buttons, CODE buttons, pillow speakers, bath stations, and shower stations. Each bed has two (2) dual AUX input jacks for misc. inputs from medical equipment to nurse call system. Each patient room has a zone light outside of each room.
b. Twenty-eight (28) room zone lights in existing locations
c. Nine (9) Zone lights in existing locations
d. Ten (10) Staff and Nurse Duty stations
e. Three (3) Touchscreen Nurse Consoles
f. One (1) 680 Main controller with 790 conversion card (gateway) and software licensing for the 790 EIS Reporting System.
2. 9C New System Hardware for Installation
a. All the following required equipment shall be tied into the existing infrastructure installed during headend project and added to the existing server. The Installation Floor Plan drawing for 9C contains the locations of all hardware to be installed. Contractor shall furnish and install the following new equipment:
i. Forty (40) patient stations with separate (40) staff emergency and (40) CODE buttons to the side of the station, one (1) zone light outside of each patient room (total of 28 rooms) and shall meet the following criteria:
1. Shall have a three-gang backbox with three-gang trim ring
2. Shall have the following dimensions (H x W x D) 4.6” x 6.8” x 1.2” (11.7 x 17.2 x 3 cm)
3. Shall have the following controls/indicators:
a. One (1) DIN jacks with call placed LEDs
b. One (1) ¼” jack with call placed LED
c. One (1) Feature bed status and call
d. Two (2) Configurable call buttons with call placed LEDs
e. One (1) Cancel button with remote call cancel LED
f. One (1) Monitor LED
g. Two (2) Low ambient light LED
4. Speaker shall have width of 2.75” (7cm) with 8 Ohm
5. Microphone shall be Electret
6. Shall have an electrical rating of 5V at 200mW
7. Shall have the following terminations:
a. Two (2) RJ45
b. Three (3) Plug on IDC multi-conductions
8. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL-Canadian Standard C22.22 NO.205
ii. Forty (40) standard pillow speakers (Model 7901/7902)
1. Case shall be Flame-retardant ABS plastic
2. Shall have the following dimensions (H x W x D) 8.4” x 3” x 1.25” (21.3 x 7.6 x 3 cm)
3. Shall have the following control buttons:
a. Nurse CALL button, TV power, Channel Up/Down arrows, Closed Captioning, Mute, Sleep
b. Two Auxiliary buttons
iii. Eighty (80) Dual Auxiliary Alarm Input Stations (Model 7941), one on each side of the bed:
1. Shall have one-gang back box or two-gang gang backbox with one-gang trim ring
2. Shall have the following dimensions (H x W x D) 4.6” x 3” x 1” (11.7 x 7.6 x 2.5 cm)
3. Shall have the following controls/indicators:
a. Two (2) ¼” jack inputs with call placed LEDs
b. One (1) Cancel button
c. One (1) Low ambient light LED
4. Shall have an electrical rating of 5V at 25mW
5. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL-Canadian Standard C22.22 NO.205
iv. Fifty-nine (59) pullcord and pushbutton shower stations (Model 7958)
1. Case shall be Flame retardant UL 94-5VA PC/ABS blend enclosure
2. The pull cord shall have the length of 6’ (1.8 m), non-porous coated braided polyester cord.
3. Shall have a one-gang backbox or two-gang backbox with one-gang trim ring.
4. Shall have the following dimensions (H x W x D): 4.6” x 3” x 1” (11.7 x 7.6 x 2.6 cm).
5. Shall have the following controls/indicators:
a. One (1) configurable call button with call placed LED
b. One (1) 6’ (1.8m) blue PVC coated pullcord
c. One (1) cancel button
d. One (1) low ambient light LED
6. Shall have an electrical rating of 5V at 25Mw.
7. Shall have the following terminations:
a. One (1) plug on two-conductor screw terminal
8. Shall have a gasket included.
9. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL Canadian Standard C22.22 NO.205
v. Six (6) Staff and Nurse Duty Stations (Model: 7927). One of the Staff and Nurse Duty stations will be placed in the CLC dining rooms (see 9B Floor Plan for location).
1. Case shall be flame retardant UL 94-5 VA PC/ABS blend enclosure
2. Two-gang backbox with a two-gang trim ring
3. Shall have the following dimensions (H x W x D) 4.8” x 4.8” x 1.2” (12.2 x 12.2 x 3 cm)
4. Shall have the following controls/indicators:
a. One (1) Green “Routine” LED call indicator
b. One (1) Amber “Urgent” LED call indicator
c. One (1) Red “Emer” LED call indicator
d. One (1) Blue “Code” LED call indicator
e. One (1) Call button
f. One (1) Combination tone mute/ cancel button
g. One (1) Low ambient light LED
h. One (1) Monitor LED
5. Speaker shall have width of 2.75” (7cm) with 8 Ohm
6. Microphone shall be Electret
7. Shall have an electrical rating of 5V at 85mW
8. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL-Canadian Standard C22.22 NO.205
vi. Nine (9) Zone Corridor Lights
vii. Three (3) Nurse Consoles (Model 7965)
1. Shall be Flame-retardant UL 94-5VA PC/ABS blend enclosure with metal base
2. Shall have the following dimensions (H x W x D): 8.3” x 9.2” x 3.0” (21 x 24.1 x 7.6 cm).
3. Shall have a desk mount footprint of 75.9in2 (489.7 cm2) or wall mount (wall mount bracket, model 9769).
4. Shall have a 5.7” (14.5 cm) display, 320 x 240 resolution color touchscreen.
5. Speaker shall have width of 2.75” (7cm) with 8 Ohm
6. Microphone shall be Electret
7. Shall have an electrical rating of 40V at 5.8W.
8. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL-Canadian Standard C22.22 NO.205
viii. Five (5) additional single call/staff emergency buttons (Model 7930) – refer to the 9C Floor Plan drawings for Installation Locations.
1. Shall be Flame-retardant UL 94-5VA PC/ABS blend enclosure.
2. Shall have one-gang back box or two-gang gang backbox with one-gang trim ring
3. Shall have the following dimensions (H x W x D): 4.6” x 3” x 1” (11.7 x 7.6 x 2.5 cm)
4. Shall have the following controls/indicators:
a. One (1) configurable call button with call placed LED
i. Staff Emergency
b. One (1) Cancel Button
c. One (1) Low ambient light LED
5. Shall have an electrical rating of 5V at 25mW
6. Shall have the terminations:
a. One (1) Plug on two-conductor screw terminal
7. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL Canadian Standard C22.22 NO.205
ix. Nurse Gateways and Nurse Call Power Supply with Integrated Battery Backup (Model 7993 and Model 7995)
x. One (1) additional software license for the NET addition of one (1) gateway Spinal Cord Injury Clinic (SCI-Building. 144)
1. SCI Existing Hardware to be removed for Upgrade
a. Building 144 Ground Floor
i. Nine (9) Staff emergency buttons
ii. Five (5) Bathroom pull cords
iii. Twelve (12) Room zone lights in existing locations
iv. Five (5) Corridor zone lights in existing locations
v. One (1) Code button
vi. One (1) Staff and Nurse Duty Stations
vii. One (1) Nurse Console
b. Building 144 First Floor
i. Thirty-eight (38) Patient stations, staff emergency buttons, CODE buttons, pillow speakers, bath stations, and shower stations. Each bed has two (2) dual AUX input jacks for misc. inputs from medical equipment to nurse call system. Each patient room has a zone light outside of each room.
ii. One (1) Patient station, staff emergency button, code button and pillow speaker and two (2) pull cords in apartment rehab room. (See attached floor plans)
iii. Thirty-nine (39) room zone lights in existing locations
iv. Two (2) Zone lights in existing locations
v. Four (4) Staff and Nurse Duty stations
vi. One (1) Nurse Console
vii. Two (2) bathroom pull cords
viii. Four (4) separate Staff Emergency Buttons
ix. One (1) 680 Main controller with 790 conversion card (gateway) and software licensing for the 790 EIS Reporting System.
2. SCI New System Hardware for Installation
a. All the following required equipment shall be tied into the existing infrastructure installed during headend project and added to the existing server. The Installation Floor Plan drawing for SCI contains the locations of all hardware to be installed. Contractor shall furnish and install the following new equipment:
b. Building 144 Ground Floor
i. One (1) Nurse Console (Model 7965)
1. Shall be Flame-retardant UL 94-5VA PC/ABS blend enclosure with metal base
2. Shall have the following dimensions (H x W x D): 8.3” x 9.2” x 3.0” (21 x 24.1 x 7.6 cm).
3. Shall have a desk mount footprint of 75.9in2 (489.7 cm2) or wall mount (wall mount bracket, model 9769).
4. Shall have a 5.7” (14.5 cm) display, 320 x 240 resolution color touchscreen.
5. Speaker shall have width of 2.75” (7cm) with 8 Ohm
6. Microphone shall be Electret
7. Shall have an electrical rating of 40V at 5.8W.
8. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL-Canadian Standard C22.22 NO.205
ii. Eight (8) single call/staff emergency buttons (Model 7930) – refer to the Building 144/SCI Floor Plan drawings for Installation Locations.
1. Shall be Flame-retardant UL 94-5VA PC/ABS blend enclosure.
2. Shall have one-gang back box or two-gang gang backbox with one-gang trim ring
3. Shall have the following dimensions (H x W x D): 4.6” x 3” x 1” (11.7 x 7.6 x 2.5 cm)
4. Shall have the following controls/indicators:
a. One (1) configurable call button with call placed LED
b. Staff Emergency
c. One (1) Cancel Button
d. One (1) Low ambient light LED
5. Shall have an electrical rating of 5V at 25mW
6. Shall have the terminations:
a. One (1) Plug on two-conductor screw terminal
7. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL Canadian Standard C22.22 NO.205
iii. Three (3) pullcord and pushbutton shower stations (Model 7958)
1. Case shall be Flame retardant UL 94-5VA PC/ABS blend enclosure
2. The pull cord shall have the length of 6’ (1.8 m), non-porous coated braided polyester cord.
3. Shall have a one-gang backbox or two-gang backbox with one-gang trim ring.
4. Shall have the following dimensions (H x W x D): 4.6” x 3” x 1” (11.7 x 7.6 x 2.6 cm).
5. Shall have the following controls/indicators:
a. One (1) configurable call button with call placed LED
b. One (1) 6’ (1.8m) blue PVC coated pullcord
c. One (1) cancel button
d. One (1) low ambient light LED
6. Shall have an electrical rating of 5V at 25Mw.
7. Shall have the following terminations:
a. One (1) plug on two-conductor screw terminal
8. Shall have a gasket included.
9. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL Canadian Standard C22.22 NO.205
iv. One (1) CODE Blue button (Model 7949)
1. Shall be flame retardant UL 94-5VA PC/ABS blend enclosure
2. Shall have a one-gang backbox or two-gang backbox with one-gang trim ring.
3. Shall have the following dimensions (H x W x D): 4.6” x 3” x 1” (11.7 x 7.6 x 2.5 cm)
4. Shall have the following controls/indicators:
a. One (1) Code Blue button with call placed LED
b. One (1) Cancel Button
c. One (1) Low Ambient light LED
5. Shall have the following terminations:
a. One (1) Plug on two-conductor screw terminal
6. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL Canadian Standard C22.22 NO.205
v. Six (6) Corridor Zone Lights (Model 7973)
vi. Ten (10) Room Zone Lights (Model 7953)
vii. Nurse Gateways and Nurse Call Power Supply with Integrated Battery Backup (Model 7993 and Model 7995)
viii. One (1) additional software license for the NET addition of one (1) gateway
c. Building 144 First Floor
i. Thirty-nine (39) patient stations with ¼ inch TV/entertainment jack for the sip and puff attachment with separate (39) staff emergency and (39) CODE buttons to the side of the station, one (1) zone light outside of each patient room (total of 33 rooms) and shall meet the following criteria:
1. Shall have a three-gang backbox with three-gang trim ring
2. Shall have the following dimensions (H x W x D) 4.6” x 6.8” x 1.2” (11.7 x 17.2 x 3 cm)
3. Shall have the following controls/indicators:
a. One (1) DIN jacks with call placed LEDs
b. One (1) ¼” jack with call placed LED
c. One (1) Feature bed status and call
d. Two (2) Configurable call buttons with call placed LEDs
e. One (1) Cancel button with remote call cancel LED
f. One (1) Monitor LED
g. Two (2) Low ambient light LED
4. Speaker shall have width of 2.75” (7cm) with 8 Ohm
5. Microphone shall be Electret
6. Shall have an electrical rating of 5V at 200mW
7. Shall have the following terminations:
a. Two (2) RJ45
b. Three (3) Plug on IDC multi-conductions
8. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL-Canadian Standard C22.22 NO.205
ii. Thirty-Nine (39) standard pillow speakers (Model 7901/7902)
1. Case shall be Flame-retardant ABS plastic
2. Shall have the following dimensions (H x W x D) 8.4” x 3” x 1.25” (21.3 x 7.6 x 3 cm)
3. Shall have the following control buttons:
a. Nurse CALL button, TV power, Channel Up/Down arrows, Closed Captioning, Mute, Sleep
b. Two Auxiliary buttons
iii. Seventy-six (76) Dual Auxiliary Alarm Input Stations (Model 7941), one on each side of the bed:
1. Shall have one-gang back box or two-gang gang backbox with one-gang trim ring
2. Shall have the following dimensions (H x W x D) 4.6” x 3” x 1” (11.7 x 7.6 x 2.5 cm)
3. Shall have the following controls/indicators:
a. Two (2) ¼” jack inputs with call placed LEDs
b. One (1) Cancel button
c. One (1) Low ambient light LED
4. Shall have an electrical rating of 5V at 25mW
5. Shall have the following certifications:
a. UL 1069 Listed
b. C-UL-Canadian Standard C22.22 NO.205
iv. Sixty-nine (79) pullcord and pushbutton shower stations (Model 7958)
1. Case shall be Flame retardant UL 94-5VA PC/ABS blend enclosure
2. The pull cord shall have the length of 6’ (1.8 m), non-porous coated braided polyester cord.
3. Shall have a one-gang backbox or two-gang backbox with one-gang trim ring.
4. Shall have the following dimensions (H x W x D): 4.6” x 3” x 1” (11.7 x 7.6 x 2.6 cm).
5. Shall have the following controls/indicators:
a. One (1) configurable call button with call placed LED
b. One (1) 6’ (1.8m) blue PVC coated pullcord
c. One (1) cancel button
d. One (1) low ambient light LED
6. Shall have an electrical rating of 5V at 25Mw.
7. Shall have the following terminations:
a. One (1) plug on two-conductor screw terminal
8. Shall have a gasket included.
9. Shall have the…