This document contains a special notice and related federal contract opportunity regarding a mobile USP 797 and 800 clean room requirement for the Robley Rex VA Medical Center in Louisville, Kentucky. The special notice seeks to identify any verified SDVOSB businesses capable of providing a mobile clean room meeting USP 797 and 800 standards to support the Inpatient Pharmacy during ongoing construction renovations. Key specifications for the clean room include maintaining ISO class 7 and 5 standards, a 6-zone design with hazardous and non-hazardous compounding areas, negative pressure monitoring, and completion by June 1, 2020. The opportunity will be sole sourced due to its emergent nature, and interested SDVOSB firms should express capability to the contracting officer by January 17, 2020. The mobile clean room must ensure 99% uptime and include interior components, wall panels, flooring, refrigeration, monitoring equipment, and installation.

36C24920Q0061 36C24920Q0061.docx

View the file

Other files

File Type Posted
36C24920Q0061-001.docx DOCX document

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Contract Opportunity Special Notice

CLASSIFICATION CODE

SUBJECT

CONTRACTING OFFICE'S

ZIP-CODE

SOLICITATION NUMBER

ARCHIVE

DAYS AFTER THE RESPONSE DATE

RECOVERY ACT FUNDS

NAICS CODE

CONTRACTING OFFICE

ADDRESS

POINT OF CONTACT

(POC Information Automatically Filled from User Profile Unless Entered)

DESCRIPTION

See Attachment

AGENCY'S URL

URL DESCRIPTION

AGENCY CONTACT'S EMAIL

ADDRESS

EMAIL DESCRIPTION

GENERAL INFORMATION

ADDITIONAL INFORMATION

* = Required Field Contract Opportunity Special Notice Robley Rex VA Medical Center Emergency Requirement Mobile Compound Center USP 797 & 800 Compliant Clean Room 37129 36C24920Q0061 N 423450 Department of Veterans Affairs Network Contracting Office 9 1639 Medical Center Parkway Suite 400 Murfreesboro TN 37129 Contracting Officer's email address The Robley Rex VA Medical Center located at 800 Zorn Avenue in Louisville, Kentucky has a unique requirement to support the Inpatient Pharmacy during the existing construction renovation where the USP 797 and 800 clean rooms are utilized for numerous low-risk intravenous (IV) medications, medium-risk parenteral nutrition (TPN) solutions and hazardous solutions daily. In order to maintain continuous patient care services, the medical center will require the use of a mobile USP 797 and 800 clean room.

This notice seeks to identify any verified SDVOSB businesses with the capability and capacity to provide a mobile USP 797 and 800 clean room as well as to state the government’s intent to sole source this procurement action due to its emergent nature to the Pharmacy.

Specifications for the mobile USP 797 and 800 compliant clean are as follows:

Vendor shall furnish all parts and expertise necessary to install cleanroom and maintain USP 797 compliance. A 6-zone cleanroom strategy while maintaining total exhausted separation of the hazardous drugs areas from the rest of the interior will be utilized. Design will include a centralized work area with a hazardous compounding area on one end and non-hazardous compounding on the opposing end (i.e., maximized workspace design). A pass-thru CVE will be utilized to reduce exposure risk to hazardous drugs. The cleanroom system will include provision and installation of all represented interior components including external exhausting, exhaust filtration, exhaust alarm, air handling, HVAC control, HEPA filtration, ceiling system, lighting, wall panels, doors, auto open actuators, shelving units, engineering controls (isolators), containment ventilated enclosure (CVE), exhausted refrigerators, switches, gauges, eyewash and sinks with foot actuation, office desks, carts, chairs, waste bins, outlets and flooring. Vendor will provide a mobile system that ensures ISO class 7 in all anterooms, and ISO class 5 in designated compounding areas.

Entry Air Lock and work area will incorporate an airlock. There will be a common entry area, washing and gowning zones, a clean side, CVE for HD staging and a segmented HD anteroom. All HD handling zones will be maintained as negative pressure areas. Pressure will be monitored by supplied pressure monitoring equipment. Refrigeration will be included in both ante areas. The HD room will incorporate an updraft exhaust system with monitoring system for assure adequate vacuum is maintained.

Paneling will be smooth clear polycarbonate exoskeleton with corner coving with seam free connections. Exterior wall and ceiling panels will be insulated, having ¾” fire treated plywood laminated to class A 0.090 smooth glass FRP. Panel junction seams filled with dual durometer profiles for clean-ability. Ceilings will incorporate hung, drop ceiling PVC panels. Flooring will be insulated, with seamless vinyl materials, cleanroom grade.

Contractor shall develop a project implementation strategy with Robley Rex VAMC to include design of a USP 797/800 compliant (v2019) mobile cleanroom including needs assessment for site location preparation. Contractor shall provide mobile cleanroom that maintains a minimum of 99% up time performance rate. All work shall be completed by June 1, 2020.

Items included:

Area
Description
Common
Exhaust system and connections for refrigeration equipment
1
Common
Plumbing connection for quoted Cleanroom grade sinks
1
Common
Soap, Alcohol and paper towel dispensers
1
Common
HEPA filtration of building exhausted air
1
Common
Temperature monitoring system
6
Common
Modular wall panel and furring system s300, for interior and perimeter interior furring
1
Common
Container
1
HD ante
Ceiling, Cleanroom, 4x2, tiles per sq. ft
1
HD ante
Door, swinging, auto, closer, AE-ADO5600, cut sheet
2
HD ante
Door, swinging, louver, adjustable, stainless, activar
3
HD ante
Door, Swinging
2
HD buffer
CACI
1
HD buffer
Shelves, Rolling, SSRS
4
HD buffer
Ceiling, Cleanroom, 4x2, Grid System, Per sq.ft.
Y
Office area
Chairs – Office Furnishings, etc.
1
Office area
Tacky Mat, AE-TM583 these are sold in four packs
1
Office area
Doors, Auto, Hands free switch. Same switch for swing or slide, 2 switches per door
3
Common
Monitor, pressure, magnahelic gauge
4
Common
Ceiling, HPEA filtration exhaust system for air to be exhausted 4000CFM
1
Option, Common
Installation of interior equipment and components quoted above
1
Option, Common
Mechanical system, design, provision, installation and balancing for quoted area only (sqft)
1
Option, Common
Supervision – Installation
4
Option, PP Ante +
3-compartment dispenser, ¼” acrylic, 12” long x 20” wide x 18” high
1
Option, PP Ante +
BootieMatic
2
Option, PP Ante +
Locker, Cleaning Supplies, Vented
1
Option, Supervision, Balance and Test
Differential Balancing
Y
Option, Supervision, Certification
Supervision. Certification not included
PP Ante +
Bench, crossover, AE-SSC18x16
1
PP Ante +
Option, work area and furnishings shown
1
PP Ante +
Paper Towels
1
PP Ante +
Sink, Eyewash
1
PP Ante +
Soap
1
PP Ante +
Refrigerator, STR1R-1G built to exhaust heat of rejection above ceiling or out the building
2
PP Ante +
Waste Containers – Standard
3
PP Ante +
Ceiling, HEPA filter module, terminal, ducted. Each
5
PP Ante +
Flooring, vinyl, sheet, Armstrong, meditone, per sq ft
320
PP Buffer ++
CAI Neutral pressure, recirculates room air
1
PP Buffer ++
Cart, AE-CRT36x24
2
PP Buffer ++
Chair, Cleanroom, AE-EC525
2
PP Buffer ++
Low wall return, external, AE-LWREXT. Each
8
Staging
Stagethru800TM
1
Staging
Ceiling, lights, cleanroom
5
Staging
Low wall return, internal
Y