RFI 36C24827Q0152.pdf

PDF 129 KB Posted

Download PDF, 129 KB

Attached to
6515--Fresenius 5008X Machines
Solicitation number
36C24827Q0152
Issued by

About this file

Sources Sought Notice Summary

This is a Sources Sought Notice issued by the Department of Veterans Affairs, Network Contracting Office 8 (NCO 8), VA Caribbean Healthcare System for market research purposes regarding Fresenius 5008X dialysis machines. The solicitation number is 36C24827Q0152, and responses must be submitted via email to Contract Specialist Eunice Bones at eunice.bones@va.gov no later than 5:00 PM Eastern Time on November 30, 2026. The place of performance is located at VA Caribbean Healthcare System in San Juan, Puerto Rico (zip code 00921). This is classified under NAICS code 334510 (Electromedical and Electrotherapeutic Apparatus Manufacturing) with a size standard of 1,250 employees and Product Service Code 6515. No Recovery Act funds are associated with this notice, and no set-aside is indicated.

The VA seeks an equal or equivalent product to the Fresenius 5008X CAREsystem with a strong preference for domestic-origin products. Required capabilities include FDA-cleared high-volume online hemodiafiltration (HVHDF) capability achieving convection volumes of at least 23 liters per session, automatic fluid substitution control, real-time fluid status monitoring options, on-demand sterile non-pyrogenic substitution fluid generation compliant with dialysis water standards, compatibility with high-flux dialyzers, FDA clearance for Hemodialysis and High Volume Hemodiafiltration therapy modes, compliance with IEC 60601 medical electrical safety standards and Class II device requirements, technical support availability to Puerto Rico through a 10-year lifecycle, and availability of replacement parts and consumables. Prospective respondents must submit a qualifications letter including company information, SAM registration, socio-economic category, capability statement, authorized distributor letter, product brochure, and evidence of past or current relevant contracts. Responses will not receive replies, and the government will not reimburse preparation costs.

View the file

On GovTribe

Work with this file

  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Sources Sought Notice

*= Required Field Sources Sought Notice

SUBJECT* Fresenius 5008X Machines

GENERAL INFORMATION

CONTRACTING OFFICE’S ZIP CODE* 00921

SOLICITATION NUMBER* 36C24827Q0152

RESPONSE DATE/TIME/ZONE 11-30-2026 17:00 EASTERN TIME, NEW YORK,

USA

ARCHIVE 30 DAYS AFTER THE RESPONSE DATE

RECOVERY ACT FUNDS N

SET-ASIDE

PRODUCT SERVICE CODE* 6515

NAICS CODE* 334510

CONTRACTING OFFICE ADDRESS Department of Veterans Affairs

Network Contracting Office 8 (NCO 8)

VA Caribbean Healthcare System

10 Casia St.

San Juan, PR 00921

POINT OF CONTACT*

Contract Specialist

Eunice Bones eunice.bones@va.gov

407-646-4143

PLACE OF PERFORMANCE

ADDRESS VA Caribbean Healthcare System (VACHS)

10 Casia St.

San Juan PR

POSTAL CODE 00921

COUNTRY USA

ADDITIONAL INFORMATION

AGENCY’S URL https://www.caribbean.va.gov

URL DESCRIPTION Agency Website

AGENCY CONTACT’S EMAIL ADDRESS eunice.bones@va.gov

EMAIL DESCRIPTION Contract Specialist Email

DESCRIPTION

This is a Sources Sought Notice only. It is neither a solicitation nor a request for a quote, and it does not obligate the Government to award a contract. Quotes received in response to this notice will not receive a response. The Government will not pay or reimburse any costs of preparing this response. The purpose of this notice is for market research purposes only and to make appropriate acquisition decisions to gain knowledge of potential qualified businesses and their Small Business Administration (SBA) size classification in relation to North American Industry Classification System (NAICS) 334510, Electromedical and Electrotherapeutic Apparatus Manufacturing with a size standard of 1,250 employees. The Network Contracting Office (NCO) 8, VA Caribbean Healthcare System, located in San Juan, Puerto Rico, is seeking a business that can provide an EQUAL to the Fresenius 5008X CAREsystem.

NOTE: The following list includes items manufactured by Fresenius Medical Care. We are requesting information about an EQUAL product with the preference at this time for domestic products. The domestic product must meet or exceed the salient characteristics provided as follows:

BRANDNAME

Product Number Description

M204441 5008X CARESystem + CLiC + CDX

F50044946 Support plate

F50044953 Acid connector guide 1 Gal

F50035686 Kit Insulated Water Inlet Tube 5008X

CAREkit-7 FX100 CORAL F00012971 + CliC 03-5300-3C

CAREkit-8 FX120 CORAL F00012972 + CLiC 03-5300-

3C

08-4080-BB bibag 900 grams/bag 12 bags/case

04-3000-9 5008X HD/HDF Medication Line

04-9009-4 Fluid Sampling Accessory 5008X

28-0600-1 CitraSure Disinfectant

Including Clinical Training

Including 2 Biomed tokens for training

EQUAL

• Must deliver FDA-cleared high-volume online hemodiafiltration (HVHDF), no exception.

• Must be able to achieve convection volumes meeting or exceeding ≥ 23 L per session via hemodiafiltration therapy.

• Must have automatic fluid substitution control to allow achieving safely convection volumes of ≥ 23 L per session.

• Must provide option for real-time fluid status monitoring to minimize intradialytic hypotension.

• Must generate sterile, non-pyrogenic substitution fluid on demand and comply with dialysis water standards.

• Must be compatible with high-flux dialyzers to enhance uremic toxin clearance.

• Must demonstrate FDA clearance for the specific device, Hemodialysis and High

Volume Hemodiafiltration therapy modes.

• Must have technical support available to Puerto Rico, through the expected 10-year lifecycle.

• Must have availability of replacement parts and consumables deliverable to Puerto

Rico.

• Must comply with IEC 60601 medical electrical safety standards and Class II device requirements.

• Must be domestic-origin product.

Concerns may respond by submitting a qualifications letter including the following information:

Company Name, Address, Point of Contact, Phone Number, FAX, Email, System for Award

Management (SAM) Unique Identifier Code (UEI), Socio-Economic Category i.e.,(SDVOSB, VOSB, HUBZone, etc.) and Capability Statement specific to the technical requirements, authorized distributor letter, and include a product brochure, if possible. Provide evidence that your firm can provide the requested items, including past and/or current contracts for the same requirements, and any other compelling evidence to help the Government determine your company's capabilities. To be eligible for an award, prospective offerors must be registered on the SAM website, https://sam.gov/content/home.

Responses to this notice shall be sent via email no later than 5:00 PM EST on November 30, 2026. Telephone and facsimile responses will not be considered. Please email Contract

Specialist Eunice Bones at eunice.bones@va.gov. When responding via email, ensure the subject line is titled with RFI# 36C24827Q0152-Fresenius 5008X Machines.

No further information is available at this time.

We thank you for your time and interest.

https://sam.gov/content/home

File details come from the government source that posted it. Updated .