36C24823Q1958_4.docx

DOCX document 17 KB Posted

Attached to
H166--FY24: Laboratory Refrigerators Calibration & Certification Federal contract opportunity
Solicitation number
36C24823Q1958
Issued by
Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 8

About this file

This document is an amendment to a previous combined solicitation from the Department of Veterans Affairs seeking laboratory refrigerator calibration and certification services. The solicitation seeks calibration of temperature sensors and chart recorders for various refrigerator and freezer units, with manufacturers and models provided. The response due date is extended to October 10, 2023 at 1:00 PM local time in Puerto Rico. Questions and answers are included, clarifying that calibration should follow NIST standards and manufacturer specifications, with sensors and chart recorders to be calibrated for each applicable unit. The place of performance is listed as the VA Caribbean Healthcare System facilities in San Juan, Puerto Rico.

View the file

Other files for this federal contract opportunity

Other files attached to H166--FY24: Laboratory Refrigerators Calibration & Certification, newest first.
File Type Posted
36C24823Q1958_3.docx DOCX document
36C24823Q1958_2.docx DOCX document
RFQ 36C24823Q1958 Attachment C- Clauses and Provisions-Final.pdf PDF
RFQ 36C24823Q1958 Attachment A- Price Cost Schedule.pdf PDF
36C24823Q1958_1.docx DOCX document
RFQ 36C24823Q1958 Attachment B- PWS.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Amendment to a Previous Combined Solicitation

Amendment to a Previous Combined Solicitation

SUBJECT*
FY24: Laboratory Refrigerators Calibration & Certification

GENERAL INFORMATION

CONTRACTING OFFICE’S ZIP CODE*
00921-3201
SOLICITATION NUMBER*
36C24823Q1958
RESPONSE DATE/TIME/ZONE
10-10-2023 1PM SAN JUAN, PUERTO RICO
ARCHIVE
60 DAYS AFTER THE RESPONSE DATE
RECOVERY ACT FUNDS
N
SET-ASIDE
SBA
PRODUCT SERVICE CODE*
H166
NAICS CODE*
541380
CONTRACTING OFFICE ADDRESS
Department of Veterans Affairs

Network Contracting Office 8 (NCO 8) VA Caribbean HCS (Admin Bldg) #10 Calle Casia Suite #4M440 San Juan PR 00921-3201

POINT OF CONTACT*

Contract Specialist Aurea Gonzalez Aurea.Gonzalez@va.gov

PLACE OF PERFORMANCE

ADDRESS
VA Caribbean Healthcare System

10 Casia Street

San Juan PR

POSTAL CODE
00921
COUNTRY
USA

ADDITIONAL INFORMATION

AGENCY’S URL

URL DESCRIPTION

AGENCY CONTACT’S EMAIL ADDRESS

EMAIL DESCRIPTION

DESCRIPTION

AMENDMENT No. 3 SOLICITATION No. 36C24823Q1958 Laboratory Refrigerators Calibration and Certification Services

Due date for Proposal/Offer submittal is hereby extended from 10/4/2023 at 1:00PM AST (Puerto Rico Local Time) to 10/10/2023 at 1:00 PM AST (Puerto Rico Local Time)

RESPONSE TO VENDORS QUESTIONS AND REQUESTS FOR INFORMATION

Q1. Are due dates for calibrations and certifications for each asset available?

R. This is a performance-based requirement, it is the prospective contractor who shall proposed their best approach to fulfil the requirement. The equipment listed has different dates and providing them all at this time may induce to error, a detailed list of dates will be provided upon award. A minimum of three (3) are needed immediately upon award.

In addition, it is the facility preference that the calibration of all units be realigned, around the same time to avoid having multiple due dates.

Q2. Does the VA wish to have the refrigerators and freezers calibrated along with the temp sensors and chart recorders? Or just the temperature sensor and chart recorders?

R. Each unit have their internal temperature sensor, and some also have the chart records as listed in the table included in the PWS. What gets calibrated are the unit temperature sensors and chart recorders.

Q3. Can you please provide Tentative Schedules for calibration for various equipment (mentioned in SOW) or give us a general idea?

R. Facility would like to calibrate all units at once (around the same time) to avoid having multiple due dates. Therefore, calibrations shall be coordinated as soon as contract gets awarded.

Q4. For each freezer and refrigerator:

a. What is/are the temperature point or points you will require for the refrigerator or freezer?

b. Do you require a 5- or 9-point profile?

c. If you do not require a profile, we will need to know what temperature point or points you need us to check along with how many probes.

R. Contractor shall comply with NIST Standard to calibrate sensors & chart recorders. Typically, the calibration report shows the “as found” values and “after calibration” values. If instrument calibration adjustments are needed, then Contractor shall follow NIST standard and instruments manufacturer data to adjust. If Contractor founds that instrument is not taking calibration, then they need to report it in order for government to replace part.

Q5. The Thermo Co thermometers is there a model number?

R. See list below:

MANUFACTURER
MODEL
SERIAL
Thermo Co
ACC10033S
34291
Thermo Co
ACC6113XXS
6029
Thermo Co
ACC6113XXS
6164
Thermo Co
ACC10033S
34456

Q6. You list the refrigerators but says the calibration is 1 chart recorder and 1 temp sensor (hence qty 2) are you looking for a calibration of the oven or the recorder/sensor inside it? They have different calibration. If it’s the chart recorder and sensor, we need the Mfgr/model of those (versus the ovens).

R. Sensor and chart recorder for each unit, as applicable. Detailed information for sensor and chart recorder are specified by refrigerator or freezer manufacturer and model information which was already provided in original table.