Notice Type: Sources Sought
This is a SPECIAL NOTICE ANNOUNCEMENT ONLY. The Department of Veteran Affairs (VA), VISN 8 Network Contracting Office for the Malcom Randall VA Medical Center, 1601 S.W. Archer Road Gainesville, Florida 32608 is seeking potential sources for research and information purposes only.
The Government will not entertain quotes and no award will be made from this special notice
Place of Performance:
Department of Veterans Affairs Malcom Randall VA Medical Center 1601 S.W. Archer Road Gainesville, FL 32608
Purpose and Objectives:
The intent of this Special Notice is to identify potential sources that supply/provide the specified description of Rodent Imaging System supplies/services listed below in the statement of need:
Statement of Need:
Technical Specifications:
Equipment must be compatible with existing MR Solutions cryogen-free MRI
Must allow for sequential PET/MR
Must fit small to large bore MR
Must be able to plug and play
High Pet Resolution: 0.7mm
Must have high sensitivity
Must be able to use for mouse to whole body rat imaging
Equipment must come with Motorized bed
MR Bore Compatibility ≥ 17 cm
Magnet strength: up to 9.4T
Removable from MR Yes
Detector Ring Design Stationary
Inner Diameter 120,0
Transaxial. FOV (mm) 80,0
Axial FOV (mm) 50,40 aFOV (mm) with motorized bed 300,00
Crystal Material LYSO/LYSO
Crystal Pitch (mm) 1,68
Crystal Dimension (mm) 1.605 x 1.605
Crystal Thickness (mm) 10 (4 top, 6 bottom)
# of Modules 8
Total # of Crystals 13 928
PMT Silicon PM
Spatial Resolution with 3D OSEM (mm) 0,7
Spatial Resolution with FBP (mm) 1.4
Depth of Interaction Yes – True DOI
DOI: Base on hardware configuration (Not on Monte-Carlo simulation)
Yes, dual layer of crystals with ½ pixel offset between
Reconstruction Algorithms FBP, 2D-OSEM, 3D-OSEM
3-D Corrections Applied Attenuation correction, Random, Scatter, Dead Time, Decay, Positron Range, Normalization, Quantification, Energy Calibration, Energy Discrimination
Absolute Sensitivity 4,56%
Energy Resolution 13,10%
NEC mouse phantom @ peak 170 Kcps
NEC mouse phantom @ 3.7MBq 51 Kcps
NEC rat phantom @ peak 42 Kcps
NEC rat phantom @ 10 MBq 29 Kcps
Timing Resolution (FWHM, ns) 1 ns
Uniformity 6%
Spill Over Ratio (Air) 9.2%
Spill Over Ratio (Water) 7.5%
Attenuation Correction MR AC
Gating (PET) Cardiac + Respiratory
List Mode Acquisition Yes
Weight 15kg
Responses must be emailed to:
Please list your socieconomic category in your response.
POC: Scott Brickey Scott Brickey@va.gov
Alternate POC: Robert Bish Robert.bish@va.gov
Contracting Office Address:
Department of Veterans Affairs VHA Regional Procurement Office (RPO) - East Network Contracting Office (NCO) 8 300 East University Ave Gainesville, FL 32601
Point of Contact:
Scott Brickey Scott.Brickey@va.gov Contracting Officer mailto:Scott.Brickey@va.gov