This is a Sources Sought Notice issued by the Department of Veterans Affairs Network Contracting Office 6 on behalf of Garner VA Clinic to conduct market research regarding the availability of qualified businesses capable of providing a NIBP/EKG Simulator Kit, specifically the Pronk Technologies SC-5+FLEXSLIMKIT or equivalent products with comparable salient characteristics.
The required simulator kit must include NIBP/IBP simulator, Pulse Ox simulator, multi-parameter patient simulator, six NIBP adapters, cuff jacket, one IBP adapter (Philips), one IBP extension cable, and carrying/storage case. Key technical specifications include: NIBP simulations for adult, neonatal, hypertensive, and hypotensive conditions; 12-lead ECG simulation with eight selectable heart rates (30-240 bpm) and ten selectable arrhythmias; eight selectable respiration rates (15-120 bpm); SpO2 simulations with saturation levels in 1% increments (10-100%) and pulse rates in 1 bpm increments (25-240); four IBP channels with eight dynamic and seven static pressure options; cardiac output simulations (3, 5, and 7 l/m); and temperature simulations (25, 37, and 40 degrees Celsius). The place of performance is Garner VA Clinic located at 2700 Benson Road, Garner, NC 27529. Responses must be submitted no later than July 23, 2026 at 12:00 PM Eastern Time to Byron Brown, Contracting Officer, at Byron.Brown4@va.gov. Capability statements should include company name and address, business size designation, socioeconomic status (8(a), HUBZone, Service Disabled Veteran-Owned, Veteran-Owned Small Business, or other small/large business classifications), qualifications, and relevant information demonstrating the ability to meet requirements. Product Service Code is 6515 and NAICS Code is 334510; the solicitation number is 36C24626Q0815, with ZIP code 23667 for the contracting office located in Hampton, VA.
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Sources Sought Notice Sources Sought Notice Sources Sought NIBP/EKG Simulator Kit
| SUBJECT* |
| NIBP/EKG Simulator Kit |
GENERAL INFORMATION
| CONTRACTING OFFICE’S ZIP CODE* |
| 23667 |
| SOLICITATION NUMBER* |
| 36C24626Q0815 |
| RESPONSE DATE/TIME/ZONE |
| 07-23-2026 12pm EASTERN TIME, NEW YORK, USA |
| ARCHIVE |
| 15 DAYS AFTER THE RESPONSE DATE |
| RECOVERY ACT FUNDS |
| N |
SET-ASIDE
| PRODUCT SERVICE CODE* |
| 6515 |
| NAICS CODE* |
| 334510 |
CONTRACTING OFFICE ADDRESS
Department of Veterans Affairs Network Contracting Office 6 100 Emancipation Drive Hampton VA 23667
POINT OF CONTACT*
Contracting Officer Byron Brown Byron.Brown4@va.gov None
PLACE OF PERFORMANCE
| ADDRESS |
| Garner VA Clinic |
2700 Benson Road
Garner NC
| POSTAL CODE |
| 27529 |
COUNTRY
ADDITIONAL INFORMATION
AGENCY’S URL
URL DESCRIPTION
AGENCY CONTACT’S EMAIL ADDRESS
EMAIL DESCRIPTION
DESCRIPTION
The Department of Veterans Affairs, Network Contracting Office (NCO) 6 on behalf of Garner VA Clinic is conducting Market Research through this sources sought notice to help determine the availability of qualified Businesses, capable of providing NIBP/EKG Simulator Kit – Pronk Technologies SC-5+FLEXSLIMKIT, but not limited to, the following:
Salient characteristics are those characteristics that make an item equal to the brand name item we specify for the VA needs.
Salient Characteristics
NIBP/EKG Simulator Kit – Pronk Technologies SC-5+FLEXSLIMKIT
· Kit must include: NIBP/IBP simulator, Pulse Ox simulator, Multi-parameter patient simulator, six (6) NIBP adapters, cuff jacket, one (1) IBP adapter (Philips), one (1) IBP extension cable and carrying/storage case.
· NIBP Simulations:
· Adult (120/80 mmHg), Neonatal (70/40 mmHg), Hypertensive (190/120 mmHg), and Hypotensive (80/40 mmHg) NIBP Simulations
· Digital Manometer
· Over Pressure (Peak Detect) Test mode measures the peak pressure produced by the patient monitor. The readout latches the highest pressure valve sensed, in 0.1 mmHg steps
· Leak Rate Test Mode
· ECG/Respiration Simulations:
· 12-lead ECG Simulation
· 8 Selectable Heart Rates: 30, 60, 90, 120, 140, 160, 200, 240 bpm
· 10 Selectable Arrhythmias: NSR w/ PVCs, VTACH, VFIB, Asys, Bigeminy, Pacer, Trigeminy, ST+, 2nd Deg. Block, Square Wave
· 8 Selectable Respiration Rates: 15, 30, 45, 60, 70, 80, 100, 120 bpm
· Banana Plug Interface
· SpO2 Simulations:
· Saturations in 1% increments (10-100%)
· Pulse rates in 1 BPM increments (25-240)
· Create custom preset simulations for faster Alarm Testing
· Perfusion Index levels in 1% increments (10-100%)
· Factory Preset Simulations:
· Saturation: 85%, Heart Rate: 80 bpm, Perfusion Index: approx. 2.0
· Saturation: 95%, Heart Rate: 40 bpm, Perfusion Index: approx. 2.0
· Saturation: 98%, Heart Rate: 80 bpm, Perfusion Index: approx. 2.0
· Saturation: 98%, Heart Rate: 140 bpm, Perfusion Index: approx. 2.0
· Saturation: 99%, Heart Rate: 80 bpm, Perfusion Index: approx. 2.0
· Masimo rainbow SET® Compatible
· Automatic detection of SpO2 Manufacturer
· Color Display – OLED Technology
· IBP Simulations:
· 4 IBP Channels
· 8 Dynamic: 120/80, 60/30, 30/10, 20/10, 125/15, 60/5, 8/2, 4/1 mmHg
· 7 Static: 0, 25, 50, 100, 150, 200, 250 mmHg
· Cardiac Output Simulations:
· 3, 5 and 7 l/m. No adapter cable required
· Temperature Simulations:
· YSI 400 and 700
· 25, 37, and 40 degrees Celsius
Responses to this sources sought notice should include the following: name and address of company, business size (large or small): Please identify your companies’ socioeconomic categories-8(a), HUBZone, Services Disabled Veteran-Owned/Veteran-Owned Small Business, other small, small or large; provide a brief description of qualifications and any other pertinent information which demonstrates the ability to meet the above requirement. Subject line must reflect: Patient Queue Management System. All capability statements shall be submitted in writing no later than July 23, 2026 by 12PM to the attention of: Byron Brown, emailed to the address of Byron.Brown4@va.gov
Please title the subject line for all responses as: NIBP/EKG Simulator Kit