36C24626Q0697.docx
DOCX document 19 KB Posted
- Solicitation number
- 36C24626Q0697
About this file
This is a Special Notice for a sole source procurement issued by the U.S. Department of Veterans Affairs Network Contracting Office (NCO) 6. The Department of Veterans Affairs intends to procure two FerrAlert Halo II Plus Ferromagnetic Portal Detectors, Model H22+ (adjustable width 25-54 inches), for the Richmond VAMC Radiology Department to replace end-of-life aging equipment. The procurement is a firm-fixed-price, one-time buy from Veteran Healthcare Supply Solutions (VHSS), located at 2615 Port Industrial Drive, Jacksonville, FL 32226. The contracting office is located in Hampton, VA 23667, with Contracting Officer Keicha Passley serving as the point of contact (Keicha.Passley@va.gov, 757-722-9961 X7191).
Interested parties must submit information demonstrating capability to meet the requirement by 8:00 a.m. EST on July 8, 2026, with no solicitation to be made available. The procurement applies NAICS code 339112 (Surgical and Medical Instrument Manufacturing) with a 1,000-employee size standard. Vendors must be OEM authorized dealers, distributors, or resellers verified by authorization letters from the OEM, with new items only—no gray market or remanufactured items acceptable. The ferromagnetic detectors must comply with VA MRI Directive 1105.05 and the American College of Radiology's MR Safety Manual, include both audio and visual alarm signals, feature a minimum 83-inch height, accommodate stretcher and ADA access, support 100-240 volts AC power, provide pre-warning on magnet room approach, and activate only on entry to the magnet room without door-operation alarms to minimize alarm fatigue.
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Special Notice Special Notice
| SUBJECT* |
| FERRALERT® HALO II PLUS-MRI Entry Metal Detectors |
GENERAL INFORMATION
| CONTRACTING OFFICE’S ZIP CODE* |
| 23667 |
| SOLICITATION NUMBER* |
| 36C24626Q0697 |
| RESPONSE DATE/TIME/ZONE |
| 07-07-2026 8:00 EASTERN TIME, NEW YORK, USA |
| ARCHIVE |
| 90 DAYS AFTER THE RESPONSE DATE |
| RECOVERY ACT FUNDS |
| N |
| PRODUCT SERVICE CODE* |
| 6515 |
| NAICS CODE* |
| 339112 |
CONTRACTING OFFICE ADDRESS
Department of Veterans Affairs Network Contracting Office 6 100 Emancipation Drive Hampton VA 23667
POINT OF CONTACT*
Contracting Officer Keicha Passley Keicha.Passley@va.gov 757-722-9961 X7191
ADDITIONAL INFORMATION
AGENCY’S URL
URL DESCRIPTION
| AGENCY CONTACT’S EMAIL ADDRESS |
| keicha.passley@va.gov |
EMAIL DESCRIPTION
DESCRIPTION
The U.S. Department of Veterans Affairs Network Contracting Office (NCO) 6, intends to enter into a sole source procurement with Veteran Healthcare Supply Solutions (VHSS), 2615 Port Industrial Drive Jacksonville, FL 32226. The procurement is for Ferralert Halo II Plus – Ferromagnetic Portable Detector Model H22+. For the Richmond VAMC Radiology Department to replace end of life/aging detectors. This acquisition is for a firm-fixed-price, one-time buy.
Equipment to be purchased: Two (2) FerrAlert Halo II Plus: Ferromagnetic Portal Detector, Model H22+ (Adjustable width 25 - 54").
NAICS code 339112 Surgical and Medical Instrument Manufacturing applies to this procurement. The size standard is 1,000 Employees.
The proposed contract action is for supplies for which the Government intends to procure from one source under the authority of FAR Overhaul 12.102(a). This notice of intent is NOT a request for quotation. Interested parties who believe they can meet this requirement are to submit information to: Keicha.Passley@va.gov , Contract Specialist. Telephone inquiries will not be accepted. Responses must be received no later than 8:00 am, EST on July 8, 2026. All information shall be furnished at no cost or obligation to the Government.
Gray market items are Original Equipment Manufacturers (OEM) goods sold through unauthorized channels in direct competition with authorized distributors. This procurement is for new OEM items only. No remanufacture or gray market items will be acceptable as replacement parts.
Vendors claiming to provide the above products shall be an OEM authorized dealer, authorized distributors, or authorized reseller for the proposed alternate equipment necessary for this procurement, verified by an authorization letter or other documents from the OEM, such that the OEM’s warranty and service are provided and maintained by the OEM. All software licensing, warranty and service associated with the equipment and/or services contracts for maintenance of equipment shall be in accordance with the OEM terms and conditions.
No solicitation will be made available. Information received will normally be considered solely for the purpose of determining whether to conduct a competitive procurement. A determination by the government not to complete this proposed contract based upon responses to the notice is solely within the discretion of the Government.
*Please see Salient Characteristics below
Salient Characteristics:
(A) Functions to be performed;
a. Ferromagnetic detectors that are attached to the entry way outside of the MRI scanner room that detect ferromagnetic threats that may enter Zone IV.
b. The primary purpose of ferromagnetic detectors is to prevent ferromagnetic objects from entering Zone III (controlled access area) and Zone IV (MRI scanner room), where they can become projectiles due to the powerful magnetic field.
c. These detectors enhance patient and staff safety by identifying potentially hazardous ferromagnetic materials, such as certain medical implants, tools, and personal items.
(B) Performance required;
a. System must satisfy National VA MRI Directive 1105.05 pg. 13.
b. System must satisfy American College of Radiology manual on MR Safety.
c. Must provide a pre warning on approach to the magnet room.
d. System must reduce alarm fatigue which is recommended by The Joint Commission.
i. System must not have alarms on the door which include the opening and closing of the door, to prevent constant alarm sounding which lead to alarm fatigue.
ii. System must only alarm on entry into the magnet and not alarm on exit.
iii. Non-ferromagnetic material must not activate the alarms.
e. System must have both audio and visual signals that will activate on detection of ferromagnetic materials.
(C) Essential physical characteristics; Dimensions (if applicable)
a. The system must have a minimum height of 83”.
b. The width of the system should be specific to the width of the current entryway door into the magnet room, which must be wide enough for stretcher, ADA compliant.
c. The power supply must be a minimum of 100 volts ac power and a maximum of 240 volts ac power.
| *= Required Field |