4610--WATER PURIFICATION SYSTEM HAMPTON VAMC/DIALYSIS
Solicitation number
36C24626Q0335
Issued by

About this file

This is a Sources Sought Notice issued by the Department of Veterans Affairs Network Contracting Office for a water purification system procurement at Hampton VAMC (Veterans Affairs Medical Center) in Hampton, Virginia. The solicitation number is 36C24626Q0335, with a response deadline of February 17, 2026 at 10:00 AM Eastern Time. The product service code is 4610, and the NAICS code is 333310. The contracting office zip code is 23667.

VAMC Hampton requires a water purification system with specific technical capabilities, including a minimum production capacity of 200 liters per hour with integrated deionization capabilities. The specified model is MEDICA-R 200 with a make-up rate of 200 l/hr at 15°C, a maximum delivery flow rate of 21 l/min at 45 psi (3 bar) on 115V, 60Hz electrical service. The system must achieve inorganic resistivity greater than 10 MO-cm at 25°C and meet specified organic (TOC) standards. This is a pre-solicitation notice; responses will be archived 30 days after the response date. No Recovery Act funds are designated for this procurement, and no set-asides are indicated.

View the file

Other files

File Type Posted
36C24626Q0335 Revised.docx DOCX document

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Sources Sought Notice Sources Sought Notice

SUBJECT*
WATER PURIFICATION SYSTEM HAMPTON VAMC/DIALYSIS

GENERAL INFORMATION

CONTRACTING OFFICE’S ZIP CODE*
23667
SOLICITATION NUMBER*
36C24626Q0335
RESPONSE DATE/TIME/ZONE
02-17-2026 10am EASTERN TIME, NEW YORK, USA
ARCHIVE
30 DAYS AFTER THE RESPONSE DATE
RECOVERY ACT FUNDS
N

SET-ASIDE

PRODUCT SERVICE CODE*
4610
NAICS CODE*
333310

CONTRACTING OFFICE ADDRESS

Department of Veterans Affairs Network Contracting Office 6 100 Emancipation Drive Hampton VA 23667

POINT OF CONTACT*

PLACE OF PERFORMANCE

ADDRESS

POSTAL CODE

COUNTRY

ADDITIONAL INFORMATION

AGENCY’S URL

URL DESCRIPTION

AGENCY CONTACT’S EMAIL ADDRESS

EMAIL DESCRIPTION

DESCRIPTION

VAMC Hampton is looking for a Water Purification system with the following:

-Salient Characteristics
-Water Purification System that can produce not less than 200 liters of water per hour, fitted with deionization capabilities
-MODEL MEDICA-R 200
-Make-up rate 15C 200 l/hr
-Delivery flow rate – maximum115V, 60Hz 21 l/min @ 45psi (3bar) Inorganics (resistivity @25°C) >10 MΩ-cm (1)
-Organics (TOC) - typical <30ppb -500 PPB (1) Bacteria - typical <10 CFU/ml (2)
-Particles 0.2 μm filter Silica <0.05 mg/l
-Notes: 1 with Ion-exchange cylinder installed (Nuclear or Hypex grade resin), 2 subject to correct operating and maintenance procedures
-Feedwater Requirement:
-Source: Tap water as detailed below Conductivity: <1000 μS/cm
-Contaminant:
-Measure:Range:Pretreatment:
-Calcium: Ca ppm as CaCO3<250None3
->250Softener or use very low RO recovery3
-Free chlorine: Cl2ppm<0.1None
-0.1 – 0.520 inch carbon block
->0.5Cylinder of carbon sized correctly to obtain <0.1 ppm
-Chloramine: Cl2 ppm<0.1None
-0.1 – 0.520 inch high efficiency carbon (5)
->0.5Cylinder of carbon sized correctly to obtain <0.1 ppm
-Silica: SiO2 ppm<30None
->3020 inch cartridge depth filter or use very low RO recovery (4)
-Fouling index: FI<10None
-10 to 2020 inch cartridge depth filter
->20Back washable media filter with a minimum flow rate of 20 l/min
-Iron/manganese: Fe/Mn ppm<0.05None
->0.0520 inch cartridge depth filter (5)
->0.1Back-washable Fe filter (6)
-Organics: TOC ppm C<2None
-2 to 320 inch carbon block (4)
->3Carbon sized Cylinder correctly for TOC demand (4) Temperature: 4 - 40°C (Recommended 15 - 25°C)
--Flowrate: (requirement at 15°C) 20 l/min
-Drain requirements: (gravity fall with air gap) 45 l/min
-Feedwater pressure: 60psi (4bar) maximum, 30psi (2bar) minimum
-Notes: 3) Check LSI, increase frequency of acid cleaning, 4) Increase frequency of alkaline cleaning,
-5) Please use ELGA Part Numbers LA683 (Filter Housing) and LC172 (Carbon Filter), 6) Increase frequency of acid cleaning Wall mount canister filter will need (Wall backing required)
-Must include Installation, Commissioning, Training
-Electrical Requirements Mains Input: 115V, 60Hz
-System control voltage (not including pumps and UV) 24V dc Power consumption (peak demand): 2000VA
-Electrical protection rating 20 amps Noise level during recirculation <70dBA
-***Note: System requires approximately 4' x 6' of floor space for tanks, filtration Unit, and filters.
--The floor drain must be a minimum of 3" and meet all applicable building codes and system requirements
--Actual Unit measurement 32" Wide x35"Deep x72"High