This sources sought notice requests information from qualified vendors to provide semi-annual certification and repair services for chemical fume hoods, biological safety cabinets, and laminar flow equipment at the Durham VA Medical Center in Durham, North Carolina. Interested parties should respond in writing by 10:00 AM on September 7, 2021 with company and point of contact information, socio-economic status, FSS contract details if applicable, tax ID and DUNS numbers, NAICS code, and capabilities statement. Responses for SDVOSB or VOSB firms must show VetBiz certification. The NAICS code for this requirement is 811310 with a size standard of $8 million. Respondents must be registered in SAM.

View the file

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

DEPARTMENT OF VETERANS AFFAIRS

Durham VA Health Care System 508 Fulton Street Durham, NC 27705

SUBJECT*
Semi Annual Chemical Fume Hoods, Bio Cabinets, Laminar Flow Certifications and Repairs

GENERAL INFORMATION

CONTRACTING OFFICE’S ZIP CODE*
23667
SOLICITATION NUMBER*
36C24621Q1333
RESPONSE DATE/TIME/ZONE
09-07-2021 10 AM EASTERN TIME, NEW YORK, USA
ARCHIVE
90 DAYS AFTER THE RESPONSE DATE
RECOVERY ACT FUNDS
N

SET-ASIDE

PRODUCT SERVICE CODE*
J059
NAICS CODE*
811310

CONTRACTING OFFICE ADDRESS

Department of Veterans Affairs Network Contracting Office 6 100 Emancipation Drive Hampton VA 23667

POINT OF CONTACT*

Contracting Officer Rodney Wilson rodney.wilson2@va.gov

PLACE OF PERFORMANCE

ADDRESS
Department of Veterans Affairs

Durham VA Medical Center

508 Fulton Street

Durham NC

POSTAL CODE
27705

COUNTRY

ADDITIONAL INFORMATION

AGENCY’S URL

URL DESCRIPTION

AGENCY CONTACT’S EMAIL ADDRESS
rodney.wilson2@va.gov

EMAIL DESCRIPTION

DESCRIPTION

SOURCES SOUGHT NOTICE ONLY

This notice does NOT constitute a request for proposal, request for quote, or invitation for bid. No formal solicitation document exists at this time. We are seeking information for market research purposes only. The Government will not pay for any information solicited

The VHA Network Contracting Office 6 (NCO 6) is conducting a market survey to find qualified vendors who can provide Semi Annual Chemical Fume Hoods, Bio Cabinets, Laminar Flow Certifications and Repairs Services for the Department of Veterans Affairs, Durham VA Medical Center (VAMC), 508 Fulton Street, Durham, NC 27705:

DESCRIPTION OF SERVICE

Items: CLINs (Please See Attachment) Performance Work Statement: (Please See Attachment)

All interested parties should respond, in writing via email to rodney.wilson2@va.gov. No telephone inquiries will be accepted. In response, please include the following:

· Company/individual name

· Address

· Point of contact with phone number

· Company Socio-Economic SIZE (example…large, small, SDVOSB, VOSB, WOSB etc.)

· Information describing your interest

· Indicate FSS/GSA contract number

· FSS Category you qualify for

· Tax ID number

· DUNS number

· NAICS code (if different than below)

· Capabilities statement

If your firm is a Service-Disabled Veteran Owned or Veteran Owned Small Business, you must be CERTIFIED in VetBiz (see internet site: http://vip.vetbiz.gov) to be recognize as such.

The NAICS 811310 is applicable to this acquisition; with a size standard of $8.

Interested firms are reminded that in accordance with FAR 4.1201(a) Prospective contractors shall complete electronic annual representations and certifications at the System for Award Management (SAM) as a part of required registration in conjunction with FAR 4.1102(a) that states, Prospective contractors shall be registered in the (SAM) database prior to award of a contract or agreement. Therefore, due to time constraints, vendors will be required to be registered in System for Award Management (SAM) at http://www.sam.gov to be considered for an award.

Sources Sought Notice Sources Sought Notice Your response is required by 10:00 AM, Tuesday September 7, 2021.

*= Required Field
Sources Sought Notice

Sources Sought Notice

SCOPE OF WORK

Semi-Annual Chemical Fume Hoods / Biological Cabinets / Laminar Flow, Testing, Certification and Repair Services

Hours of Operation: Services shall be performed during the VA normal business hours, 7:30 a.m. through 4:00 p.m., Monday through Friday, excluding Federal holidays, and any other day specifically declared by the President of the United States. Federal holidays are available at the Federal Holiday OPM Site. For working outside the “normal working hours” as defined in this contract, the vendor shall give 3-day notice to the Contracting Officer so that arrangements can be provided for the employees. This notice is separate from any notices required for utility shutdown described later in this section. All work to be coordinated around an operational medical center as to minimize impact to patient care.

This Scope of Work is for One (1) base year and four (4) option years

Scope of Work: The contractor shall provide all labor, personnel, supervision, equipment, tools, and all materials necessary to perform preventive maintenance, repair and certification on the chemical fume hoods/biological safety cabinets and Laminar Flow devices and assess the Primary Environmental Controls for the associated spaces where aforementioned equipment is located in the Durham Veterans Affairs Health Care System. The frequency of the preventative maintenance and certification is semi-annual.

Testing Standards: All testing of biosafety cabinets to be performed in accordance with criteria established by the National Sanitation Foundation (NSF) International standard number # 49-2002e, Annex F, or the prevailing NSF standard at time of inspection. All fume hood testing to be performed in accordance with criteria established by ANSI/ASHRAE 110, or the prevailing standard at time of inspection. All testing of laminar flow units to be done in accordance with criteria established by IEST-RP-CC002.2, Federal Standard 209E, ISO 14644 and the OEM specifications. All testing of animal isolator racks/units to be performed in accordance with criteria established by IEST-RP-CC002.2, Federal Standard 209E and the OEM specifications. Hoods are to be certified semiannually. Primary Environmental Controls test to be performed per ANSI/AIHA Z9.5-2003 American National Standards for Laboratory Ventilation.

Preventive Maintenance Procedures: The contractor shall contact the COR for approval at least two weeks prior to their arrival with a list of the fume hoods and/or biological safety cabinets that are to be tested and certified so the COR can coordinate with the affected services. Contractor shall report to the COR upon arrival at the facility and inform COR of the fume hoods and/or biological safety cabinets that are to be tested and certified for the day. In the event of a fume hood failure, the contractor shall inform the COR who will then arrange for the repair of the exhaust system and the rescheduling of the fume hood testing and certification. Contractor shall be responsible for fume hood HEPA filter replacement as needed. Contractor shall decontaminate hoods as required prior to repair or filter replacement. The contractor shall be responsible for the proper operation and repairs of the sash, motor, belts, fan unit, electrical connections of each certifiable device where applicable. The contractor shall also insure proper calibration on existing primary environmental control monitoring devices.

Face Velocity Tests. Set hood to the designed opening sash position as follows:

Height: Height shall be based on the dimension from the bottom most part of the sash to the work surface located in a straight plane directly beneath the sash for vertical sashes, or the opening between horizontal tracks on horizontal sashes.

Width: Width shall be based in the interior dimensions of the design sash opening.

The average face velocity shall be determined by the method described in the ANSI/ASHRAE 110-1995 Method of Testing Performance of Laboratory Fume Hoods. Face velocity measurements shall be made by dividing the hood opening into equal area grids with sides measuring no more than 12 in. (30.5 cm). The tip of the probe shall be positioned in the plane of the sash opening and fixed (not handheld) at the approximate center of each grid. Grid measurements around the perimeter of the hood opening shall be made at approximately 6 in. (15.2 cm) from the top, bottom, and sides of the opening enclosure. The average face velocity shall be the average of the grid velocity measurements. Each grid velocity shall be the average of at least 10 measurements made over at least 10 seconds. The plane of the sash shall be located at the midpoint of the sash frame depth. Manufacturer’s specifications of each piece of equipment shall also be used for testing.

Face Velocity Measurement. A 1.0 ft2 imaginary grid pattern shall be formed by equally dividing the design hood opening into vertical and horizontal openings. Velocity readings shall be taken with a calibrated anemometer. For VAV hoods, measurements should be made at 25, 50, and 100% sash openings. A hood is considered certified when the average face velocity at the manufacturer recommended working sash height is between 80 – 120 feet per minute.

Routine performance tests shall be conducted in accordance with the SOW. A hood that is found to be operating with an average face velocity more than 10% below the designated average face velocity shall be labeled as OUT OF SERVICE and corrective actions shall be taken by VA Engineering staff to adjust and/or replace belts. Each hood shall be posted with a notice giving the date of the routine performance test, the measured average face velocity and initialed by the technician. If it is taken OUT OF SERVICE, it shall be posted with an out-of-service notice.

Repair Procedures: Any necessary repairs performed during the inspections are to be included in the scope of work at additional charge. The scope of these repairs will be limited to repair or replacement of blower motors, motor speed controls and switches. All labor, travel and any decontamination necessary to perform a repair are to be included at additional charge. All germicidal and illumination lamps and filters are consumables and are specifically included in the scope of this contract. Blowers/fans external to the equipment, i.e. roof top exhaust fans, primary environmental controls are excluded from the coverage of the contract. The contractor will assist the COR with descriptions and part numbers for any filters found to require replacement. Any filters removed by the contractor must be properly decontaminated and disposed of in accordance to all applicable EPA regulations. The decontaminated filters shall be the responsibility of the contractor to dispose of and documentation of proper disposal shall be provided to the COR.

Personnel Qualifications: Before allowing employees to start work under the terms of this contract, the Contractor shall ensure that employees have the required qualifications and, if applicable, have the required current and valid professional certifications and licenses in compliance with federal, state, and local environmental requirements or laws. The NEBB/TABB/AABC Certified Professional has a responsibility to assure that technicians performing the work are properly trained and possess sufficient skills. The Contractor shall provide the CO and COR with information on the qualifications or certifications for employees and sub-Contractors. All service technicians performing under this service contract must be able to document/provide certificates of training for Original Equipment Manufacturer (OEM) service training received from The Baker Company, Nuaire, Labconco, Kewaunee Scientific and Germ Free. Licensed journeyman tradespersons shall be licensed for their respective trades in accordance with city, county, state or federal requirements. Contactor shall ensure all certificates; licenses and bonds are kept current and valid. All licenses and certifications shall be submitted to the CO and COR annually and any time as requested. Associated training, licensing and certifications’ expenses shall be borne by the Contractor.

Equipment Calibration: All fume hood performance testing instrumentation shall be inspected annually and maintained with a current annual certificate, traceable to the National Institute of Standards and Technology (NIST). The testing instrumentation report provided to the VA shall include the instrument type, manufacturer, model number, serial number, date of calibration, and Tracer Gas purity certification. Contractor shall also provide equipment certifications with this report.

Emergency Service/Repairs – Contractor must return calls for service within two-hours if requested by the COR and provided the service call is initiated within normal business hours of 7:30AM -4:00PM, M-F, excluding federal holidays. Emergency service/repairs performed outside of the SOW that are not covered in this contract and will be contracted for on a time & materials basis as needed/required. The contractor shall respond to emergency repair requests within 8-hours to restore equipment back to first-class operating condition. Contractor shall make allowances for revisits for clean room and hood certification due to failure to certify. Contractor shall schedule certification of the equipment once repairs have been completed.

DELIVERABLES: At the completion of each day’s testing, the contractor shall provide the COR written reports for the hoods that were inspected indicating whether the unit has been passed and certified or failed with description of the deficiencies noted.

Certifications: The contractor shall provide hard copy and electronic inspection reports to the COR as soon as practical or within 30-days of completion of the testing. Inspection placards/stickers will be affixed to the equipment as soon as an inspection/certification is completed. Certification of Compliance Test Reports will be provided for each Pharmacy IV Clean Room inspected. Contractor will maintain an electronic database of all test reports for not less than five years after the date of completion of the inspection. Service reports shall be signed by both service provider and COR at completion of work. An electronic version of same shall be sent to the COR. In addition, stickers shall be affixed to equipment indicating date certified and due date of next inspection.

The vendor’s employees shall not enter the work site without appropriate identification. They may also be subject to inspection of their personal effects when entering or leaving the project site.

Execute work to interfere as little as possible with normal functioning of the hospital including operations of utility services, fire protection systems and any existing equipment, and with work being done by others. Use of equipment and tools that transmit vibrations and noises through the building structure, are not permitted in buildings that are occupied jointly by patients or medical personnel, and Contractor’s personnel, except as permitted by COR where required by limited working space.

No utility service such as water, gas, steam sewers or electricity, or fire protection systems and communications systems may be interrupted without prior approval of COR. Electrical work shall be accomplished with all affected circuits or equipment de energized. When an electrical outage cannot be accomplished, work on any energized circuits or equipment shall not commence without the Medical Center Director’s prior knowledge and written approval. The vendor shall submit a request to interrupt any such services to COR, in writing, 48 hours in advance of proposed interruption. Request shall state reason, date, exact time of, and approximate duration of such interruption.

VA reserves the right to stop work or shut down the project site and order the vendor’s employees off the premises in the event of a national or Durham VA emergency. The vendor may return to the site only with the written approval of the Contracting Officer.

Site and Building Access: Maintain free and unobstructed access to facility emergency services and for fire, police and other emergency response forces in accordance with NFPA 241.

Means of Egress: Do not block exiting for occupied buildings, including paths from exits to roads. Minimize disruptions and coordinate with COR.

All Buildings will be occupied during performance of work. Contractor shall take all measures and provide all material necessary for protecting existing equipment and property in affected areas against dust and debris, so that equipment and affected areas to be used in the Medical Centers operations will not be hindered. Coordinate alteration work in areas occupied by Department of Veterans Affairs so that Medical Center operations will continue during the installation period.

The contractor shall maintain neat and orderly work areas and shall clean area of debris and rubbish daily and dispose of waste materials offsite per hospital policies. The contractor shall provide a dumpster if it is determined to be necessary, and shall work with the COR in advance for a spot-on site to establish the dumpster.

All work shall be performed in full accordance with applicable local and Federal regulations. All materials and installation shall conform to the recommendations of National Fire Protection Association (NFPA), plumbing and mechanical codes. No departures from specification requirements will be permitted without written approval. The Supplier shall use good safety practices while working. All hardware should be installed in such a manner to minimize damage and maximize patient and employee safety.

New work shall be installed and connected to existing work nearly, safely and professionally. Disturbed or damaged work shall be replaced or repaired to its prior conditions.

All items furnished shall be free from defects that would adversely affect the performance, maintainability and appearance of individual components and overall assembly.

Where the Government determines that the vendor has installed equipment not conveniently accessible for operation and maintenance, equipment shall be removed and reinstalled or remedial action performed as directed at no additional cost to the Government.

Contractors, contractor personnel, subcontractors, and subcontractor personnel shall be subject to the same Federal laws, regulations, standards, and VA Directives and Handbooks as VA and VA personnel regarding information and information system security.

The contractor will not be allotted any parking permits for onsite parking. All vehicles must have a parking permit assigned by the COR for off-site parking located up to 4 blocks from the facility. If work is needed to be performed outside of normal working hours, the contractor will be responsible to checking in with the VA Police to request permission to park on site as designated by the VA Police.

ATTACHMENT

“A”

EQUIPMENT LIST

With EXAMPLE of

SUGGESTED SCHEDULE

Quantities and locations of equipment subject to change

FEBRUARY

BLDG
ROOM
MFG
MODEL#
SERIAL#
TYPE
EE#
14
215
* PLAS LABS/SCIENTIFIC
FIC-24-2/SP
26900
BSC
28013
14
111
Chemical Fume Hood
TBD
FH
8339
14
121
Chemical Fume Hood
TBD
FH
22278
14
B02
Chemical Fume Hood
TBD
FH
34866
14
117
* PLAS LABS/SCIENTIFIC
FIC-24-2/SP
27000
BSC
28104
1
E3008
ALC COLLEGEDALE
G3R11-6036
10-100388-7
FH
66110
1
E3006
LABCONCO
G3R11-6036
10-100388-6
FH
66112
1
E4004
ALC COLLEGEDALE
G3R11-6036
10-100388-15
FH
66114
1
E3016
ALC COLLEGEDALE
G3R11-6036
10-100388-10
FH
66113
1
E3016
Chemical Fume Hood
FH
66104
1
E4016
ALC COLLEGEDALE
G3R11-6036
10-100388-9
FH
66115

MARCH

BLDG
R00M
MFG
MODEL#
SERIAL#
TYPE
EE#
1
E1018
NUAIRE
NU408FM600
10185OT
BSC
19963
1
EG017
BAKER CO
VBM600
SG18929V
BSC
2105
1
E3014
NUAIRE
NU408FM400
72550W
BSC
17507
14
228
NUAIRE
NU-607-400
86911102203
BSC
34866
14
228
NUAIRE
NU-124-600
13712 SN
FH
23120
14
108
LAB PRODUCTS
59030
M129770701
BSC
47983
1
C10018
Chemical Fume Hood
FH
13436
1
F3157
Chemical Fume Hood
FH
16453
1
F3232
Chemical Fume Hood
FH
16455
1
F3236
Chemical Fume Hood
FH
16456
GVHCC**
1A134
Laminar Flow
LF
75191
GVHCC**
2C420
Laminar Flow
LF
75407
GVHCC**
2D100
Laminar Flow
LF
87735
GVHCC**
2D100
Laminar Flow
LF
87736

APRIL

BLDG
ROOM
MFG
MODEL#
SERIAL#
TYPE
EE#
1
F3179
HAMILTON IND
SAFEAIRE
FH
16453
1
F3205
HAMILTON IND
SAFEAIR II
FH
16454
1
F3177
BAKER CO
SG-250
SL44799V
BSC
19272
1
N3018
NUAIRE
NU425400
17725TY
BSC
19279
1
N3021
LABCONCO
Protector
FH
19280
1
N3018
NUAIRE
NU425400
17724TY
BSC
19282
1
N3018
NUAIRE
NU425400
17723TY
BSC
19283
1
F3183
AIR CLEAN SYSTEMS
AC648LFUVC
41435
BSC
47854
1
EG017
LABCONCO
LOGIC
091117518B
BSC
60835
1
E1010
ALC COLLEGEDALE
G3R11-6036
10-100388-1
FH
66106
1
E2006
ALC COLLEGEDALE
G3R11-6036
10-100388-18
FH
66107
1
E3016
ALC COLLEGEDALE
G3R11-6036
10-100388-10
FH
66113
1
E2008
ALC COLLEGEDALE
G3R11-6036
10-100388-47
FH
66116
1
F3168
BAKER CO
SG404
119235
BSC
88070
1
F3203
BAKER CO
SG400
SL43654V
BSC
16457
1
F3157
BAKER CO
SG400
SL43653V
BSC
16461
1
F3157
BAKER CO
SG400
SL43650V
BSC
16462
1
F3183
BAKER CO
SG600
SL43678V
BSC
16463

MAY

BLDG
ROOM
MFG
MODEL#
SERIAL#
TYPE
EE#
1
N3022
LABCONCO
PROTECTOR
FH
19276
1
N3020
LABCONCO
PROTECTOR
FH
19277
1
N3017
LABCONCO
PROTECTOR
FH
19278
1
N3019
LABCONCO
PROTECTOR
FH
19281
1
F3206
JEWETT
JGP4-1B
04K000445
FH
37527
1
F3206
JEWETT
JGP4-1B
04K000447
FH
37528
1
F3206
JEWETT
JGP4-1B
04K000446
FH
37529
1
E2016
Chemical Fume Hood
FH
66111
1
A5017
Laminar Flow
LF
11793
15
203
Laminar Flow
LF
20487

JUNE

BLDG
ROOM
MFG
MODEL#
SERIAL#
TYPE
EE#
1
E2016
BAKER CO
SG400
SL42862V
BSC
10568
1
F3185
BAKER CO
SG400
SL43651V
BSC
16458
1
F3176
BAKER CO
BSL-3
SL43679V
BSC
16459
1
A2007
BAKER CO
SG600
SL48214V
BSC
17508
1
E2016
BAKER CO
SG400
5L-41353
BSC
18494
1
E1006
BAKER CO
VBM600
SG11762V
BSC
21386
1
E1006
BAKER CO
VBM600
SG13658V
BSC
21387
1
F3183
LABCONCO
Logic+
121168356-C
BSC
76033
1
E1006
LABCONCO
4FT
10-100388-12
FH
66104
1
E1018
LABCONCO
4FT
10-100388-3
FH
66105
1
E2011
LABCONCO
4FT
10-100388-8
FH
66108
1
E2014
LABCONCO
4FT
10-100388-5
FH
66109
1
E2016
LABCONCO
4FT
10-100388-11
FH
66111
14
215
BAKER CO
SG-600
SL38333V
BSC
19300
14
215
NUAIRE
NU602400
62901
BSC
20089
14
117
BAKER CO
NCB6
SR10008V
BSC
22152
15
102
BAKER CO
SG-600
SL34479V
BSC
19288
15
215
BAKER CO
VBM600
SG23255V
BSC
25796
15
103
BAKER CO
SG-400
SL30357V
BSC
49290
15
415
BAKER CO
SG-403A
86883
BSC
82072
15
104
INDUSTRIAL SCIENTIFIC
SUPREME AIRE
104-15
FH
85115
15
313
HAMILTON IND
SAFEAIRE
313-15
FH
85127
15
523
HAMILTON IND
SAFEAIRE II
523-15
FH
85152
15
123
KEWAUNEE
SUPREME AIRE
145251
FH
80218

AUGUST

BLDG
ROOM
MFG
MODEL#
SERIAL#
TYPE
EE#
14
215
* PLAS LABS/SCIENTIFIC
FIC-24-2/SP
26900
BSC
28013
14
117
* PLAS LABS/SCIENTIFIC
FIC-24-2/SP
27000
BSC
28014
14
111
Chemical Fume Hood
TBD
FH
8339
14
121
Chemical Fume Hood
TBD
FH
22278
14
B02
Chemical Fume Hood
TBD
FH
34866
1
E3008
ALC COLLEGEDALE
G3R11-6036
10-100388-7
FH
66110
1
E4004
ALC COLLEGEDALE
G3R11-6036
10-100388-15
FH
66114
1
E3016
ALC COLLEGEDALE
G3R11-6036
10-100388-10
FH
66113
1
E3016
Chemical Fume Hood
FH
66104
1
E3006
LABCONCO
G3R11-6036
10-100388-6
FH
66112
1
E4015
ALC COLLEGEDALE
G3R11-6036
10-100388-9
FH
66115

SEPTEMBER

BLDG
ROOM
MFG
MODEL#
SERIAL#
TYPE
EE#
1
E1018
NUAIRE
NU408FM600
10185OT
BSC
19963
1
EG017
BAKER CO
VBM600
SG18929V
BSC
2105
1
E3014
NUAIRE
NU408FM400
72550W
BSC
17507
14
228
NUAIRE
NU-607-400
86911102203
BSC
34866
14
228
NUAIRE
NU-124-600
13712 SN
FH
23120
14
108
LAB PRODUCTS
59030
M129770701
BSC
47983
1
C10018
Chemical Fume Hood
FH
13436
1
F3157
Chemical Fume Hood
FH
16453
1
F3232
Chemical Fume Hood
FH
16455
1
F3236
Chemical Fume Hood
FH
16456
GVHCC**
1A134
Laminar Flow
LF
75191
GVHCC**
2C420
Laminar Flow
LF
75407
GVHCC**
2D100
Laminar Flow
LF
87735
GVHCC**
2D100
Laminar Flow
LF
87736

OCTOBER

BLDG
ROOM
MFG
MODEL#
SERIAL#
TYPE
EE#
1
F3179
HAMILTON IND
SAFEAIRE
FH
16453
1
F3205
HAMILTON IND
SAFEAIR II
FH
16454
1
F3177
BAKER CO
SG-250
SL44799V
BSC
19272
1
N3018
NUAIRE
NU425400
17725TY
BSC
19279
1
N3021
LABCONCO
Protector
FH
19280
1
N3018
NUAIRE
NU425400
17724TY
BSC
19282
1
N3018
NUAIRE
NU425400
17723TY
BSC
19283
1
F3183
AIR CLEAN SYSTEMS
AC648LFUVC
41435
BSC
47854
1
EG017
LABCONCO
LOGIC
091117518B
BSC
60835
1
E1010
ALC COLLEGEDALE
G3R11-6036
10-100388-1
FH
66106
1
E2006
ALC COLLEGEDALE
G3R11-6036
10-100388-18
FH
66107
1
E3016
ALC COLLEGEDALE
G3R11-6036
10-100388-10
FH
66113
1
E2008
ALC COLLEGEDALE
G3R11-6036
10-100388-47
FH
66116
1
F3168
BAKER CO
SG404
119235
BSC
88070
1
F3203
BAKER CO
SG400
SL43654V
BSC
16457
1
F3157
BAKER CO
SG400
SL43653V
BSC
16461
1
F3157
BAKER CO
SG400
SL43650V
BSC
16462
1
F3183
BAKER CO
SG600
SL43678V
BSC
16463

NOVEMBER

BLDG
ROOM
MFG
MODEL#
SERIAL#
TYPE
EE#
1
N3022
LABCONCO
PROTECTOR
FH
19276
1
N3020
LABCONCO
PROTECTOR
FH
19277
1
N3017
LABCONCO
PROTECTOR
FH
19278
1
N3019
LABCONCO
PROTECTOR
FH
19281
1
F3206
JEWETT
JGP4-1B
04K000445
FH
37527
1
F3206
JEWETT
JGP4-1B
04K000447
FH
37528
1
F3206
JEWETT
JGP4-1B
04K000446
FH
37529
1
E2016
Chemical Fume Hood
FH
66111
1
A5017
Laminar Flow
LF
11793
15
203
Laminar Flow
LF
20487

DECEMBER

BLDG
ROOM
MFG
MODEL#
SERIAL#
TYPE
EE#
1
E2016
BAKER CO
SG400
SL42862V
BSC
10568
1
F3185
BAKER CO
SG400
SL43651V
BSC
16458
1
F3176
BAKER CO
BSL-3
SL43679V
BSC
16459
1
A2007
BAKER CO
SG600
SL48214V
BSC
17508
1
E2016
BAKER CO
SG400
5L-41353
BSC
18494
1
E1006
BAKER CO
VBM600
SG11762V
BSC
21386
1
E1006
BAKER CO
VBM600
SG13658V
BSC
21387
1
F3183
LABCONCO
Logic+
121168356-C
BSC
76033
1
E1006
LABCONCO
4FT
10-100388-12
FH
66104
1
E1018
LABCONCO
4FT
10-100388-3
FH
66105
1
E2011
LABCONCO
4FT
10-100388-8
FH
66108
1
E2014
LABCONCO
4FT
10-100388-5
FH
66109
1
E2016
LABCONCO
4FT
10-100388-11
FH
66111
14
215
BAKER CO
SG-600
SL38333V
BSC
19300
14
215
NUAIRE
NU602400
62901
BSC
20089
14
117
BAKER CO
NCB6
SR10008V
BSC
22152
15
102
BAKER CO
SG-600
SL34479V
BSC
19288
15
215
BAKER CO
VBM600
SG23255V
BSC
25796
15
103
BAKER CO
SG-400
SL30357V
BSC
49290
15
415
BAKER CO
SG-403A
86883
BSC
82072
15
104
INDUSTRIAL SCIENTIFIC
SUPREME AIRE
104-15
FH
85115
15
313
HAMILTON IND
SAFEAIRE
313-15
FH
85127
15
523
HAMILTON IND
SAFEAIRE II
523-15
FH
85152
15
123
KEWAUNEE
SUPREME AIRE
145251
FH
80218

* Two (2) Chambers

** Greenville Health Care Campus, Greenville, NC 27834

SECTION 01 00 00

GENERAL REQUIREMENTS

TABLE OF CONTENTS

1.1 GENERAL INTENTION24
1.2 SHUTDOWNS25
1.3 SPECIFICATIONS AND DRAWINGS FOR CONTRACTOR25
1.4 PROJECT SECURITY REQUIREMENTS25
1.5 ROOF AND VERTICAL WORKSAFETY28
1.6 FIRE SAFETY28
1.7 HAZARDOUS MATERIALS MANAGEMENT31
1.8 OPERATIONS AND STORAGE AREAS32
1.9 ALTERATIONS36
1.10 INFECTION PREVENTION MEASURES38
1.11 DISPOSAL AND RETENTION41
1.12 PROTECTION OF EXISTING VEGETATION, STRUCTURES, EQUIPMENT, UTILITIES, AND IMPROVEMENTS42
1.13 RESTORATION43
1.14 LAYOUT OF WORK44
1.15 USE OF ROADWAYS45
1.16 TEMPORARY USE OF MECHANICAL AND ELECTRICAL EQUIPMENT45
1.17 TEMPORARY USE OF EXISTING ELEVATORS46
1.18 TEMPORARY TOILETS47
1.19 AVAILABILITY AND USE OF UTILITY SERVICES48
1.20 TESTS49
1.21 INSTRUCTIONS50
1.22 RELOCATED EQUIPMENT and ITEMS51
1.23 HOURS OF WORK52
1.24 PARKING52
1.25 TUBERCULOSIS (TB) TESTING52
1.26 TOBACCO USE53
1.27 CONTRACTOR BADGES53
1.28 PROJECT QUALITY CONTROL54
1.29 WARRANTY MANAGEMENT56
1.30 STORED MATERIALS64
PART 1 - GENERAL96
1.1DESCRIPTION96
This section specifies requirements for Contractor Quality Control (CQC) for Design-Bid-Build (DBB) or Design-Build (DB) contract projects. This section can be used for both project types.96
1.2APPLICABLE PUBLICATIONS96
A.The publication listed below form a part of this specification to the extent referenced. The publications are referred to within the text by the basic designation only.96
B.ASTM International (ASTM)96
1.3SUBMITTALS97
Government approval is required for all submittals. CQC inspection reports shall be submitted under this Specification section and follow the [Applicable CQC Control Phase (Preparatory, Initial, or Follow-Up)]: [Applicable Specification section] naming convention.97
PART 2 PRODUCTS – NOT USED97
PART 3 - EXECUTION97
3.1GENERAL REQUIREMENTS97
Establish and maintain an effective quality control (QC) system. that complies with the FAR Clause 52.246.12 titled “Inspection of Contract”. QC consists of plans, procedures, and organization necessary to produce an end product which complies with the Contract requirements. The QC system covers all design and contract operations, both onsite and offsite, and be keyed to the proposed design and contract sequence. The project superintendent will be held responsible for the quality of work and is subject to removal by the Contracting Office or Authorized designee for non-compliance with the quality requirements specified in the Contract. In this context the highest-level manager responsible for the overall contract activities at the site, including quality and production is the project superintendent. The project superintendent maintains a physical presence at the site at all times and is responsible for all contract and related activities at the site, except as otherwise acceptable to the Contracting Officer.97
3.2CQC PLAN:98
A.Submit the CQC Plan no later than 15 days after receipt of Notice to Proceed (NTP) proposed to implement the requirements of the FAR Clause 52.246.12 titled “Inspection of Contract”. Design and/or contract will be permitted to begin only after acceptance of the CQC Plan or acceptance of an Interim plan applicable to the particular feature of work to be started. Work outside of the accepted Interim CQC Plan will not be permitted to begin until acceptance of a CQC Plan or another Interim CQC Plan containing the additional work scope is accepted.98
B.Content of the CQC Plan: Include, as a minimum, the following to cover all contract operations, both onsite and offsite, including work by subcontractors, fabricators, suppliers, and purchasing agents:98
C.Acceptance of Plan: Acceptance of the Contractor’s plan is required prior to the start of design and contract. Acceptance is conditional and will be predicated on satisfactory performance during the design and contract. The Government reserves the right to require the Contractor to make changes in the CQC Plan and operations including removal of personnel as necessary, to obtain the quality specified.100
D.Notification of Changes: After acceptance of the CQC Plan, notify the Contracting Officer or Authorized designee in writing of any proposed change. Proposed changes are subject to acceptance by the Government prior to implementation by the Contractor.100
1.3COORDINATION MEETING:100
After the Precontract Conference Post-award Conference before start of design or contract, and prior to acceptance by the Government of the CQC Plan, meet with the Contracting Officer or Authorized designee to discuss the Contractor’s quality control system. Submit the CQC Plan a minimum of 5 business days prior to the Coordination Meeting. During the meeting, a mutual understanding of the system details must be developed, including the forms for recording the CC operations, design activities (if applicable), control activities, testing, administration of the system for both onsite and offsite work, and the interrelationship of Contractor’s Management and control with the Government’s Quality Assurance. Minutes of the meeting will be prepared by the Government, signed by both the Contractor and Contracting Officer or Authorized designee and will become a part of the contract file. There can be occasions when subsequent conferences will be called by either party to reconfirm mutual understandings or address deficiencies in the CQC system or procedures which can require corrective action by the Contractor.100
1.4QUALITY CONTROL ORGANIZATION:101
A.Personnel Requirements: The requirements for the CQC organization are a Safety and Health Manager, CQC System Manager, a Design Quality Manager (if applicable), and sufficient number of additional qualified personnel to ensure safety and Contract compliance. The Safety and Health Manager shall satisfy the requirements of Specification 01 35 26 Safety Requirements and reports directly to a senior project (or corporate) official independent from the CQC System Manager. The Safety and Health Manager will also serve as a member of the CQC Staff. Personnel identified in the technical provisions as requiring specialized skills to assure the required work is being performed properly will also be included as part of the CQC organization. The Contractor’s CQC staff maintains a presence at the site at all times during progress of the work and have complete authority and responsibility to take any action necessary to ensure Contract compliance. The CQC staff will be subject to acceptance by the Contracting Officer or Authorized designee. Provide adequate office space, filing systems, and other resources as necessary to maintain an effective and fully functional CQC organization. Promptly complete and furnish all letters, material submittals, shop drawings submittals, schedules and all other project documentation to the CQC organization. The CQC organization is responsible to maintain these documents and records at the site at all times, except as otherwise acceptable to the Government.101
B.CQC Personnel: In addition to CQC personnel specified elsewhere in the contract, provide as part of the CQC organization specialized personnel to assist in the CQC System Manager for the following areas, as applicable: electrical, mechanical, civil, architectural, materials technician, and low voltage systems. These individuals or specified technical companies are employees of the prime or subcontractor; be physically present at the contract site during work on the specialized personnel’s areas of responsibility. These individuals can perform other duties but need to be allowed sufficient time to perform the specialized personnel’s assigned quality controls duties as described in the CQC Plan. A single person can cover more than one area provided that the single person is qualified to perform QC activities in each designated and that workload allows.101
C.Organizational Changes: Maintain the CQC staff at full strength at all times. When it is necessary to make changes to the CQC staff, revise the CQC Plan to reflect the changes and submit the changes to the Contracting Officer or Authorized designee for acceptance.102
1.5SUBMITTALS AND DELIVERABLES: Submittals have to comply with the requirements in Section 01 33 23 Shop Drawings, Product Data, and Samples. The CQC organization is responsible for certifying that all submittals and deliverables are in compliance with the contract requirements.102
1.6CONTROL:102
A.CQC is the means by which the Contractor ensures that the contract, to include that of subcontractors and suppliers, complies with the requirements of the contract. At least three phases of control are required to be conducted by the CQC System Manager for each definable feature of the contract work as follows:102
B.Initial Phase: This phase is accomplished at the beginning of a definable feature of work. Accomplish the following:103
C.Follow-Up Phase: Perform daily checks to assure control activities,104
including control testing, are providing continued compliance with contract requirements until the completion of the particular feature of work. Record the checks in the CQC documentation. Conduct final Follow-Up checks and correct all deficiencies prior to the start of additional features of work which may be affected by the deficient work. Do not build upon nor conceal non-conforming work. Coordinate scheduled work with Special Inspections required by Section 01 45 35 Special Inspections, the Statement of Special Inspections, and the Schedule of Special Inspections104
D.Additional Preparatory and Initial Phases on the same definable features of work if: the quality ongoing work is unacceptable; if there are changes in the applicable CQC staff, onsite production supervision or work crew; if work on a definable feature is resumed after a substantial period of inactivity, or if other problems develop.105
1.7TESTS105
A.Testing Procedure: Perform specified or required tests to verify that control measures are adequate to provide a product which conforms to contract requirements. Upon request, furnish to the Government duplicate samples of test specimens for possible testing by the Government. Testing includes operation and acceptance test when specified. Procure the services of a Department of Veteran Affairs approved testing laboratory or establish an approved testing laboratory at the project site. Perform the following activities and record and provide the following data:105
B.Testing Laboratories: All testing laboratories must be validated through the procedures contained in Specification section 01 45 29 Testing Laboratory Services.106
C.Onsite Laboratory: The Government reserves the right to utilize the Contractor’s control testing laboratory and equipment to make assurance tests, and to check the Contractor’s testing procedures, techniques, and test results at no additional cost to the Government.106
1.8COMPLETION INSPECTION106
A.Punch-Out Inspection: Conduct an inspection of the work by the CQC system Manager near the end of the work, or any increment of the work established by the specifications. Prepare and include in the CQC documentation a punch list of items which do not conform to the approved drawings and specifications. Include within the list of deficiencies the estimated date by which the deficiencies will be corrected. Make a second inspection the CQC System Manager or staff to ascertain that all deficiencies have been corrected. Once this is accomplished, notify the Government that the facility is ready for the Government Pre-Final Inspection.106
B.Pre-Final Inspection: The Government will perform the Pre-Final Inspection to verify that the facility is complete and ready to be occupied. A Government Pre-Final Punch List may be developed as a result of this inspection. Ensure that all items on this list have been corrected before notifying the Government, so that a Final Acceptance Inspection with the customer can be scheduled. Correct any items noted on the Pre-Final Inspection in a timely manner. These inspections and any deficiency corrections required by this paragraph need to be accomplished within the time slated for completion of the entire work or any particular increment of the work if the project is divided into increments by separate contract completion dates.106
C.Final Acceptance Inspection: The Contractor’s QC Inspection personnel, plus the superintendent or other primary management person, and the Contracting Officer’s Authorized designee is required to be in attendance at the Final Acceptance Inspection. Additional Government personnel can also be in attendance. The Final Acceptance Inspection will be formally scheduled by the Contracting Officer’s or Authorized designee based upon results of the Pre-Final Inspection. Notify the Contracting Officer through the Resident Engineer office at least 14 days prior to the Final Acceptance Inspection and include the Contractor’s assurance that all specific items previously identified to the Contractor as being unacceptable, along with all remaining work performed under the contract, will be complete and acceptable by the date schedule for the Final Acceptance Inspection. Failure of the Contractor to have all contract work acceptably complete for this inspection will be cause for the Contracting Officer to bill the Contractor for the Government’s additional inspection cost in accordance with FAR Clause 52.246-12 titled “Inspection of Contract”.107
1.9DOCUMENTATION107
A.Quality Control Activities: Maintain current records providing factual evidence that required QC activities and tests have been performed. Include in these records the work of subcontractors and suppliers on an acceptable form that includes, as a minimum, the following information:107
B.Verification Statement: Indicate a description of trades working on the project; the number of personnel working; weather conditions encountered; and any delays encountered. Cover both conforming and deficient features and include a statement that equipment and materials incorporated in the work and workmanship comply with the Contract. Furnish the original and one copy of these records in report form to the Government daily with 1 week after the date covered by the report, except that reports need not be submitted for days on which no work is performed. As a minimum, prepare and submit on report for every 7 days of no work and on the last day of a no work period. All calendar days need to be accounted for throughout the life of the contract. The first report following a day of no work will be for that day only. Reports need to be signed and dated by the CQC System Manager. Include copies of test reports and copies of reports prepared by all subordinate QC personnel within the CQC System Manager Report.108
1.10SAMPLE FORMS109
Templates of various quality control reports can be found on the Whole Building Design Guide website at https://www.wbdg.org/FFC/NAVGRAPH/quality_control_reports.pdf109
1.11NOTIFICATION OF NONCOMPLIANCE: The Contracting Officer or Authorized designee will notify the Contractor of any detected noncompliance with the foregoing requirements. The Contractor should take immediate corrective action after receipt of such notice. Such notice, when delivered to the Contractor at the work site will be deemed sufficient for the purpose of notification. If the Contractor fails or refuses to comply promptly, the Contracting Officer can issue an order stopping all or part of the work until satisfactory corrective action has been taken. No part of the time lost due to such stop orders will be made the subject of claim for extension of time or for excess costs or damages by the Contractor.109

Page 1 of

Page 1 of xxiii Page xxiii of 125

SECTION 01 00 00

GENERAL REQUIREMENTS

1.1 GENERAL INTENTION

A. Contractor shall completely prepare site for building operations, including demolition and removal of existing materials, and furnish labor and materials and perform work to repair fire damper in room C5050 at the Durham VA Medical Center as required by NFPA 80, drawings, and specifications.

B. All employees of general contractor and subcontractors shall comply with VA security management program and obtain permission of the VA police, be identified by project and employer, and restricted from unauthorized access.

C. Prior to commencing work, general contractor shall provide proof that a OSHA certified “competent person” (CP) (29 CFR 1926.20(b)(2) will maintain a presence at the work site whenever the general or subcontractors are present.

D.Training:
1.All employees of general contractor and subcontractors shall have the 10-hour OSHA certified Safety course, Asbestos awareness safety courses, and /or other relevant competency training, as determined by VA COR or Safety Office with input from the ICRA team.

2. Supervisors of general contractor and subcontractors shall have the 30-hour OSHA certified Safety course, ASHE Health Care: Subcontractor Certification, ASHE Health Care Workshop Certification and /or other relevant competency training, as determined by VA CP with input from the ICRA team.

3.Submit training records of all such employees for approval before the start of work.
E.All employees of the contractor and subcontractors serve at the discretion of the Chief of Engineering. Chief of Engineering has the right to remove any person from the job site at any time

1.2 SHUTDOWNS

A. Contractor shall maintain all services to occupied portions of building, except as scheduled and approved by COR. To ensure such executions, Contractor shall furnish the COR with a schedule of shutdown dates on which the Contractor intends to accomplish work on adjacent floors and areas. In addition, Contractor shall notify the COR at least two weeks in advance of the proposed date of starting work in each specific area or riser which will affect adjacent occupied areas. More extensive or critical shutdowns will require additional notice.

1.3 SPECIFICATIONS AND DRAWINGS FOR CONTRACTOR

A. AFTER AWARD OF CONTRACT, the Contractor can download PDF files of the contract drawings and specifications for his use and distribution to Subcontractors.

1.4 PROJECT SECURITY REQUIREMENTS

A.Security Plan:
1.Prior to start of work, prepare a plan detailing project specific security measure. The security plan defines both physical and administrative security procedures that will remain effective for the entire duration of the project.
2.The General Contractor is responsible for assuring that all sub-contractors working on the project and their employees also comply with these regulations.
B.Security Procedures:
1.General Contractor’s employees shall not enter the project site without appropriate badge. They may also be subject to inspection of their personal effects when entering or leaving the project site or DURVAMC campus.
2.For working outside the “regular hours” as defined in the contract, The General Contractor shall give 3 business days’ notice to the Contracting Officer so that security arrangements can be provided for the employees. This notice is separate from any notices required for utility shutdown described later in this section.
3.No photography of VA premises is allowed without written permission of the Contracting Officer.
4.VA reserves the right to close-down or shut down the project site and order General Contractor’s employees off the premises in the event of a national emergency. The General Contractor may return to the site only with the written approval of the Contracting Officer.
C.Key Control:
1.The General Contractor shall provide duplicate keys and lock combinations to the COR for the purpose of security inspections of every area of project including toolboxes and parked machines and take any emergency action.
2.DURVAMC will issue keys as needed through the Engineering Program Support Assistant. Contractor should request a key through the COR. Contractor will need photo ID and is required to sign-out keys.
One key per contract can be issued. Additional keys can be requested and are subject to approval by the Chief of Engineering.

Lost keys must be reported immediately to VA Police and to the COR. Contractor can be billed the cost to re-core and re-key all affected areas.

Keys must be returned at the completion of the project, or prior to vacating the project site, whichever comes first.

Contractor will comply with Space Key Control policy and other regulations presented by Engineering Support Assistant at the time of key issue.

D.Document Control:
1.Before starting any work, the General Contractor/Sub Contractors shall submit an electronic security memorandum describing the approach to following goals and maintaining confidentiality of “sensitive information”.
2.The General Contractor is responsible for safekeeping of all drawings, project manual and other project information. This information shall be shared only with those with a specific need to accomplish the project.
3.Certain documents, sketches, videos or photographs and drawings may be marked “Law Enforcement Sensitive” or “Sensitive Unclassified”. Secure such information in separate containers and limit the access to only those who will need it for the project. Return the information to the Contracting Officer upon request.
4.These security documents shall not be removed or transmitted from the project site without the written approval of Contracting Officer.
5.All paper waste or electronic media such as CD’s and diskettes shall be shredded and destroyed in a manner acceptable to the VA.
6.Notify Contracting Officer and Site Security Officer immediately when there is a loss or compromise of “sensitive information”.
7.All electronic information shall be stored in specified location following VA standards and procedures using an Engineering Document Management Software (EDMS).
a.Security, access, and maintenance of all project drawings, both scanned and electronic shall be performed and tracked through the EDMS system.
b.“Sensitive information” including drawings and other documents may be attached to e-mail provided all VA encryption procedures are followed.
E.Motor Vehicle Restrictions
1.Vehicle authorization request shall be required for any vehicle entering the site and such request shall be submitted in writing 24 hours before the date and time of access. Access shall be restricted to picking up and dropping off materials and supplies.
2.See 1.25 Parking for parking restrictions.

1.5 ROOF AND VERTICAL WORKSAFETY

A. Contractor is responsible for providing safe working conditions at all times. Work performed on roof and vertical elevations of building shall comply with OSHA regulations and VA Safety Office.

1.6 FIRE SAFETY

A.Applicable Publications: Publications listed below form part of this Article to extent referenced. Publications are referenced in text by basic designations only.
1.American Society for Testing and Materials (ASTM):
E84-2009Surface Burning Characteristics of Building Materials
2.National Fire Protection Association (NFPA):
10-2018Standard for Portable Fire Extinguishers
30-2018Flammable and Combustible Liquids Code
51B-2019Standard for Fire Prevention During Welding, Cutting and Other Hot Work
70-2020National Electrical Code
72-2019National Fire Alarm and Signaling Code
80-2019Standard for Fire Doors and Other Opening Protectives
101-2018Life Safety Code
241-2019Standard for Safeguarding Project, Alteration, and Demolition Operations
3.Occupational Safety and Health Administration (OSHA):
29 CFR 1910Safety and Health Regulations for Industrial

4. The VA Fire Protection Design Manual (2015)

B.Fire Safety Plan: Establish and maintain a fire protection program in accordance with 29 CFR 1926. Prior to start of work, prepare a plan detailing project-specific fire safety measures, including periodic status reports, and submit to COR and Facility Safety Officer for review for compliance with contract requirements in accordance with Section 01 33 23, SHOP DRAWINGS, PRODUCT DATA AND SAMPLES Prior to any worker for the contractor or subcontractors beginning work, they shall undergo a safety briefing provided by the general contractor’s competent person per OSHA requirements. This briefing shall include information on the project limits, VAMC safety guidelines, means of egress, break areas, work hours, locations of restrooms, use of VAMC equipment, etc. Documentation shall be provided to the COR that individuals have undergone contractor’s safety briefing.
C.Site and Building Access: Maintain free and unobstructed access to facility emergency services and for fire, police and other emergency response forces in accordance with NFPA 241.
D.Separate temporary facilities, such as trailers, storage sheds, and dumpsters, from existing buildings and new project by distances in accordance with NFPA 241. For small facilities with less than 6 m (20 feet) exposing overall length, separate by 3m (10 feet).
E.Temporary Partitions:
1.Install two-hour fire-rated temporary partitions as shown on drawings to maintain integrity of existing exit stair enclosures, exit passageways, fire-rated enclosures of hazardous areas, horizontal exits, smoke barriers, vertical shafts and openings enclosures.
2.Close openings in smoke barriers and fire-rated project to maintain fire ratings.
F.Temporary Heating and Electrical: Install, use and maintain installations in accordance with 29 CFR 1926, NFPA 241 and NFPA 70.
G.Means of Egress: Do not block exiting for occupied buildings, including paths from exits to roads. Minimize disruptions and coordinate with COR and facility Safety Officer.