Salient Characteristics for Flow Cytometer, VAMC Richmond Virginia

Quantity: 1

A. Functions to be performed:

Laboratory analyzer for the detection and phenotypical characterization of leukemia and lymphoma. Also, monitoring of T cell counts for HIV patients.

To replace discontinued Becton, Dickinson and Company FACSCalibur and loader placed into service 06/2006.

S/N E97500402 and F57400321

652-EE6014838 and 652-EE6021010

B. Performance required:

o Must include auto loader.

o Must include 10 fluorescent channels, 3 laser configuration.

o Must have uncompensated data range of 0-262,143 o Must have automated optical alignment.

o Must have minimum 5 Liter tank capacity.

o Must be equipped with internal barcode reader.

o Must have integrated bi-directional LIS interface.

o Must support barcode format Code 128.

o Must have preconfigured IVD clinical software templates for CD3/CD8/CD45/CD4 and 6 color TBNK panels.

o Computer loaded with supporting and analytical software.

C. Essential physical characteristics:

• Must be a bench top analyzer with an approximate footprint of 42”x 23”x 23”

• 100-240 + 10% VAC

• Normal operating noise level under 60 dBA over 8 hours.

• Under 500 BTU/hour heat dissipation at ambient temperature.

• Must include operating computer with keyboard, color printer and LCD flat panel.

• Must not require wet cart.

Instrument installation, setup and training will be provided by the manufacturer.

Deliver to:

Wayne Stevens, PLMS (113) Immunology Laboratory 3A-156 1201 Broad Rock Boulevard Richmond, Va. 23249 804-675-5889