36C24618Q0469-000.docx

DOCX document 23 KB Posted

Attached to
590-18-2-985-0054 GreenLight XPS Laser System (VA-18-00011410)
Solicitation number
36C24618Q0469
Issued by

About this file

36C24618Q0469 36C24618Q0469_1.docx

View the file

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

FedBizOpps Sources Sought Notice

CLASSIFICATION CODE

SUBJECT

CONTRACTING OFFICE'S

ZIP-CODE

SOLICITATION NUMBER

RESPONSE DATE (MM-DD-YYYY)

ARCHIVE

DAYS AFTER THE RESPONSE DATE

RECOVERY ACT FUNDS

SET-ASIDE

NAICS CODE

CONTRACTING OFFICE

ADDRESS

POINT OF CONTACT

(POC Information Automatically Filled from User Profile Unless Entered)

DESCRIPTION

See Attachment

AGENCY'S URL

URL DESCRIPTION

AGENCY CONTACT'S EMAIL

ADDRESS

EMAIL DESCRIPTION

ADDRESS

POSTAL CODE

COUNTRY

ADDITIONAL INFORMATION

GENERAL INFORMATION

PLACE OF PERFORMANCE

* = Required Field FedBizOpps Sources Sought Notice Rev. March 2010 590-18-2-985-0054 GreenLight XPS Laser System

(VA-18-00011410)

23667 36C24618Q0469 02-28-2018 N 339113 Kim Pauley Department of Veterans Affairs Network Contracting Office 6 100 Emancipation Drive Hampton VA 23667 Department of Veterans Affairs Surgical Department 100 Emancipation Drive Hampton, VA 23667

USA

This is a sources sought for an XPS Green Light Laser Therapy System and accompanying supplies. The equipment is brand name or equal, list below with item detail. To be considered to receive a solicitation for quote please submit a product brochure for clinical consideration as an equal product if it is not brand name. This is not a solicitation, responses to this notice will be used as Market Research only. Please respond to kim.pauley@va.gov no later than February 28, 2018 at 2:00 PM EST if your company can supply this product. Please provide your Dun and Bradstreet number and socio-economic status if you are interested in this acquisition. It is the intent of the government to solicit and award this contract if possible to an SDVOSB. Carefully consider the brand name or equal product prior to submitting your interest in this acquisition.

Built on the GreenLight HPS™ system, the GreenLight XPS™ system increases the power and area of the laser beam by 50%, providing a wider tissue vaporization effect without sacrificing depth of vaporization or coagulation.1 Its 532nm wavelength vaporizes prostate tissue without charring and limits penetration, leaving a thin rim of coagulated tissue that aids in hemostasis. It has been found to cost less than TURP while providing similar clinical outcomes. The system can be used for patients on anticoagulants, in urinary retention, or who have large glands >100 ml.

MoXyTM Active Cooling Cap technology uses saline flow to minimize fiber tip devitrification, significantly reducing power degradation throughout the duration of the procedure.

The FiberLifeTM automatic safety system detects conditions when the fiber cap may overheat and briefly interrupts the laser beam, ensuring improved fiber reliability.

TruCoagTM uses pulsating light to cauterize ruptured vessels, reducing bleeding faster and in more situations compared to previous generations Console: XPS

· Laser type: solid state, frequency doubled

· Wavelength: 532 nm

· Max. power output: 180 W

· Nominal optical hazard distance (NOHD): 33.9 meters (MPE = 1x10-3 W/cm2)

· Eye protection: OD ≥ 6

· Working distance: 0.5−3.0 mm

· Vaporization beam mode: quasi-CW (15 kHz−25 kHz)

· Coagulation beam mode: pulse modulated by 12 Hz, 25% duty cycle

· Compatible fibers: MoXyTM fiber (750 µm core diameter), HPS fibers (600 µm core diameter)

· Optical penetration: 0.8 mm

· Coagulation depth: 1−2 mm

· Coagulation mode: yes; TruCoagTM 5−40 W

· FiberLifeTM: yes

· MoJoTM Software enabled: yes

· Current draw: 20 A

· Cooling water: internal

· Dimensions: 20” W x 36” D x 43.5” H (50.8 cm x 91.4 cm x 110.5 cm)

· Weight: approximately 420 lb (190 kg)

· Length of power cord: 15 ft (4.6 m) Fiber: MoXy

· Laser compatibility: XPS

· Firing angle: ~70−80°

· Power range: ≤ 180 W

· Beam area at working distance: 0.44 mm2

· Energy limit: 650 kJ

· Fiber core diameter: 750 µm

· Fiber tip O.D.: 2.3 mm

· Active Cooling CapTM: Yes

· Use environment: most 22−26 FR continuous flow endoscope/cystoscope