About this file

This Statement of Work (SOW) details the procurement and installation of two ultrasonic washers for the VA Maryland Health Care System's Sterile Processing Service in Baltimore, MD. The requirement includes delivering new equipment with specific salient characteristics such as a touchscreen operating panel, automated water usage and detergent injection, specific external and chamber dimensions, programmable temperature settings, and options for manual and automatic tray adjustments. The contractor must remove existing Steris Innowave PCF and Pro washers, coordinate with Biomedical Engineering and Infection Control, relocate plumbing connections, and ensure proper installation.

The solicitation includes five evaluation volumes: Technical Capability (equipment specifications and Censitrak system connectivity), Quality of Proposed Solution and Delivery Terms (project timeline and CAD drawings), Maintenance Availability (service support details), Price, and Past Performance. Offerors must submit three recent and relevant references from hospital settings comparable to VA Capitol Health Care Network (VISN 5), focusing on experience in critical care areas within the last five years. The contract requires a minimum one-year warranty on parts and labor, onsite repairs, phone support, and routine maintenance, with additional warranty years recommended as added value.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

B.3 STATEMENT OF WORK

STATEMENT OF WORK

SPS Ultrasonic Washer Replacement with Installation VA Maryland Health Care System

Baltimore, MD

PURPOSE:

The overall purpose of this requirement is to procure two (2) ultrasonic washers to be deployed in the Sterile Processing Service (SPS) at the VA Maryland Health Care System (VAMHCS), Baltimore, MD location. The requirement will account for the procurement of two ultrasonic washers, the necessary accessories, training, extended installation, and system validation.

SCOPE:

The contractor shall deliver and install the systems with all specified salient characteristics. Only new equipment will be accepted. The contractor shall provide shipping, transport, installation, and all equipment necessary to meet the requirements and capabilities listed within the scope.

The contractor shall deliver the cart washer and washer-disinfectors to the rooms of choice at the Baltimore VA medical center. The assembly and validation of the system must be performed by a representative who has been service trained by the Original Equipment Manufacturer (OEM). The contractor is responsible for identifying any missing parts or components not included in the order for the specified equipment to be successfully installed and validated to their full potential. The contractor shall include accessories and consumables that have been detailed in the provided quotes.

SALIENT CHARACTERISTICS

Ultrasonic washers must have:

• Touchscreen operating panel

• Water usage, per load, shall be automatically varied by the washer in response to the size of the load using a frequency-controlled wash pump

• Automated, adjustable detergent injection

• External dimension of ultrasonic washers must be equal to or cannot exceed 42” W x 48” H x 34” L

• Must have minimum net usable chamber dimensions of 32” W x 15” L x 16” D

• Programmable temperature setting for chamber solution

• Service access shall be through an easily opened panel(s) on front of the machine

• Pre-programmed water fill settings

• Options for manual and automatic raising and lowering of tray

REMOVAL AND EXTENDED INSTALLATION

The contractor will perform installation as the VAMHCS deems fit through discussions with Biomedical Engineering and Sterile Processing. The contractor will provide for the removal of the existing ultrasonic washers that are currently in use in Baltimore VA’s Sterile Processing Department. These devices are as follows, one (1) Steris Innowave PCF and one (1) Steris Innowave Pro. Size Dimensions for the devices are as follows:

Innowave PCF: 34.5" x 42.6" x 27.8" (W x H x L) and Innowave Pro: 49.6" x 43.6" x 36.2" (W x H x L). The contractor will disconnect all water lines and utilities and remove the devices and any accessories from the space and from the facility to dispose of them.

Contractors will need to coordinate with Biomedical Engineering and Infection Control staff to establish ICRA Plans, if any are necessary and will install all necessary ICRA barriers per VAMHCS Infection control guidance and VHA requirements. The contractor will be responsible for all labor and materials to relocate plumbing connections as necessary to connect to new equipment. Contractors will use existing drains and will cap off any unused lines. The contractor will be responsible for providing all labor and materials necessary for reconnecting the DI/RO water, hot and cold water and compressed air to all equipment. Contractor will provide necessary labor and material to disconnect and reconnect power to the 2 ultrasonic washers. Contractors will be responsible for providing labor and materials to ensure the flooring is level to equipment manufacture specifications.

DELIVERY

The contractor shall deliver equipment and any necessary accessories to the Baltimore VA Warehouse as required in the quantities specified. Any delivery must be coordinated with the COR prior to shipping. The contract must give the COR a minimum of 3 weeks’ notice prior to shipping. Delivered materials will arrive in the manufacturer’s original sealed containers with the associated purchase order number listed on the outside of the box. The contractor shall remove all related shipping debris and packing from the VAMHCS facilities. Delivery address will be:

Baltimore VA Medical Center 10 North Greene St.

Baltimore, MD 21201 ATTN: Biomedical Engineering

SITE CONDITIONS

There shall be no smoking on hospital property at any time. Contractor personnel on-site may be required to wear masks to protect against the spread of COVID-19 at any time in accordance with local health guidelines. Due to the sensitive location of the equipment installation, personnel on-site may be required to don and off disposable protective clothing upon entering and exiting the space.

INSPECTION AND ACCEPTANCE

The contractor shall conduct a joint inspection with the COR upon delivery of equipment.

The contractor shall provide the VA COR with a list of missing and/or damaged items found upon inspection and have those items replaced at no cost to the Government. The

COR shall ensure all work is completed satisfactorily prior to acceptance. Disputes shall be resolved by the VA Contracting Officer.

OPERATOR TRAINING

Onsite education and training by a vendor service trained representative will be provided to VAMHCS clinical staff, Facilities and Engineering staff, and the Biomedical Engineering staff. The scheduling of operator training shall be coordinated with the COR after delivery and installation.

PROTECTION OF PROPERTY

The contractor shall protect all items from damage. The contractor shall take precautions against damage to the building(s), grounds and furnishings. The contractor shall repair or replace any items related to building(s) or grounds damaged accidentally or on purpose due to actions by the Contractor.

SECURITY REQUIREMENTS

The contractor must have a photo ID available when performing work at the VA Medical Center and be willing to present it to VA staff. The contractor is responsible for providing any personal protective equipment (PPE) that is deemed necessary to complete tasks stated in this requirement. The contractor must follow the VA Maryland Health Care Systems current COVID-19 guidance for entry to any medical center or Community Based Outpatient Clinic (CBOC).

WARRANTY

At a minimum the cart washer and washer-disinfectors and accessories must include a 1-year warranty on all parts and labor at no cost to the Government. Additional years are highly recommended as added value to the government. The warranty shall begin upon completion of validation testing. The warranty shall include onsite repairs and over-the-phone support. Contractor shall provide routine maintenance service program during warranty period. All repairs shall be performed by a vendor or contractor which has been certified by the vendor to perform the repairs. The warranty shall include all travel and shipping costs associated with any warranty repair.

EVALUATION FACTORS

SPS Washer-Disinfectors and Cart Washer Replacement with Extended Installation VA Maryland Health Care System

Baltimore, MD

Volume I - Technical Capability

The offeror shall provide technical information and specifications regarding the proposed washer-disinfector and cart washer as described in the Statement of Work with technical information required to include the following:

• Provide equipment specifications and cut sheets

• Explain how the proposed equipment connectivity software interfaces with Censitrak instrument tracking system, to include interfaced information and equipment required, server hardware, etc.

Volume II – Quality of Proposed Solution and Delivery Terms

The offeror should provide a complete package with how the proposed equipment will fit into the existing Sterile Processing space. The offeror should clearly articulate the estimated scope and timeline for equipment delivery, extended installation for all 3 phases with technical information required to include the following:

• Microsoft Project or equivalent project timeline, to include estimated equipment lead times

• Detailed scope of work for the extended installation to include what is provided by the offeror and what is the responsibility of the VA Maryland Health Care System

• Provide CAD Drawings or equivalent of the equipment placement in the SPS area

Volume III – Maintenance Availability

The offeror should provide information pertaining to their service costs and capabilities for supporting the cart washer and the 3 washer disinfectors with technical information required to include the following:

• Number of parts depots in the service area and number of service technicians covering the service area

• Provide information on contracted time to arrive on site for repair and average time to repair

• Provide information about parts availability and backordered part information

Volume IV – Price

Price - The offeror shall complete the pricing schedule provided in the solicitation. This will be used to determine selection should all other factors be considered equal.

Volume V – Past Performance

The offeror submits three (3) past performance references that are recent and relevant which focus on experience within the last five (5) years in the hospital settings that are comparable in size as VA Capitol Health Care Network (VISN 5). References shall include the contract number for the project, a brief description, and a point of contact to include an email address. The information shall be provided in 5 pages or less.

Past Performance- Acceptable or Unacceptable Rating The Government shall make an objective assessment of the Offeror's current/past performance. It will conduct an acceptable performance assessment based on aspects of the offeror’s past performance that are recent and relevant, which focuses on experience within the last 5 years in hospital settings that are comparable in size as VA Capitol Health Care Network (VISN 5) or larger ones which contain Operating Rooms, PACUs and other critical care areas. They have to be relevant, similar in complexity, and magnitude requirements of solicitation. Performances that are not recent, relevant, similar in complexity or magnitude will be evaluated with unacceptable performance in this area.

File details come from the government source that posted it. Updated .