36C24426Q0066.docx
DOCX document 20 KB Posted
- Solicitation number
- 36C24426Q0066
About this file
This Special Notice is a sole-source procurement by the Department of Veterans Affairs Network Contracting Office 04 for a service agreement with Hillrom to maintain 48 Centrella Smart+ Beds at the Erie VA Medical Center. The contract will cover a one-year base period with four one-year option periods, focusing on preventive and corrective maintenance services for medical equipment.
The procurement requires Original Equipment Manufacturer (OEM) certified labor, replacement parts, and comprehensive maintenance services, including annual preventative maintenance, emergency and routine repairs, remote technical support, and detailed service reporting. Key performance requirements include 24/7 technical support, on-site emergency response within 24 hours, routine repairs within 2 business days, and comprehensive equipment inspections covering electrical safety, functional controls, software updates, and calibration. The solicitation is classified under NAICS code 811210 for Electronic and Precision Equipment Repair and Maintenance, with responses due by 11/04/2025 at 10:00 AM Eastern Time via email to the Contracting Officer, Thomas Cossentino.
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Special Notice Special Notice
| SUBJECT* |
| INTENT TO SOLE SOURCE HILL ROM SERVINITY SERVICE AGREEMENT |
GENERAL INFORMATION
| CONTRACTING OFFICE’S ZIP CODE* |
| 19104 |
| SOLICITATION NUMBER* |
| 36C24426Q0066 |
| RESPONSE DATE/TIME/ZONE |
| 11-04-2025 10:00am EASTERN TIME, NEW YORK, USA |
| ARCHIVE |
| 10 DAYS AFTER THE RESPONSE DATE |
| RECOVERY ACT FUNDS |
| N |
| PRODUCT SERVICE CODE* |
| J065 |
| NAICS CODE* |
| 811210 |
| CONTRACTING OFFICE ADDRESS |
| Department of Veterans Affairs |
Network Contracting Office 4
3900 Woodland Ave.
Philadelphia PA 19104
POINT OF CONTACT*
Contracting Officer Thomas A. Cossentino thomas.cossentino@va.gov email only email only
ADDITIONAL INFORMATION
AGENCY’S URL
URL DESCRIPTION
| AGENCY CONTACT’S EMAIL ADDRESS |
| thomas.cossentino@va.gov |
EMAIL DESCRIPTION
DESCRIPTION
Notice of Intent to Sole Source
The Department of Veterans Affairs, Network Contracting Office 04 (NCO4), intends to negotiate on a sole source basis with Hillrom in accordance with FAR 12.102(a).
The Erie VA Medical Center (Erie VAMC) requires a service agreement providing Original Equipment Manufacturer (OEM) certified labor, OEM replacement parts, tools, test equipment, travel, and shipping, required to perform annual preventative maintenance (PM) and corrective maintenance (CM) services for 48 Centrella Smart+ Beds at the Erie VAMC. Services shall conform to the Hillrom SERVinity Complete Protection program.
The anticipated contract will include a one-year base performance period, and four one-year option periods.
This Notice of Intent is not a request for competitive proposals. Responsible sources may identify their interest, accompanied by supporting literature that clearly demonstrates capability to meet the performance requirements detailed below, to the Contracting Officer no later than 11/04/2025 at 10:00 AM Eastern Time. All responses must be sent by email to the Contracting Officer, Thomas Cossentino, at thomas.cossentino@va.gov. Responses must include “ISS 36C24426Q0066” in the subject line. Telephone responses or inquiries will not be accepted. Determination of procurement strategy based on responses to this notice is solely within the discretion of the Contracting Officer. The NAICS Code for this procurement is 811210, Electronic and Precision Equipment Repair and Maintenance.
PERFORMANCE REQUIREMENTS:
Covered Equipment: Forty-eight (48) Hill-Rom Centrella Smart+ Beds with MAX surfaces and their Ready Connect modules. Exclusions: Any third-party accessories, consumables such as pads, batteries, bladders, and hoses.
Services shall conform to the Hill-Rom Complete Protection program and the following specific requirements:
· Preventive Maintenance (PM): performed annually, in accordance with OEM specifications, technical service manuals, and ANSI/AAMI standards.
· Corrective Maintenance (CM): On-call service, including emergency repairs, on-site labor, OEM parts replacement, and loaner provision if equipment cannot be repaired on-site.
· Parts and Labor: All replacement parts shall be OEM authorized, new or remanufactured to OEM standard, and installed by certified service personnel. Substitutions require COR approval.
· Remote Support: 24/7 remote technical support, remote diagnostics, firmware/software updates, asset tracking.
· OEM Service Reports: Contract shall provide service history reports generated through Hill-Rom’s Computerized Maintenance Management System (CMMS). Reports will document preventive and corrective maintenance performed and will be delivered in PDF to Biomedical Engineering.
· Service Levels: Emergency response within 24 hours; Routine repairs within 2 business days
Specific Tasks:
Preventative Maintenance (PM) The contractor shall perform annual PM on each covered Centrella Smart+ bed in accordance with Hill-Rom OEM service manuals. At a minimum, PM shall include the following tasks:
1. Inspect Internal Components (mattress)
a. Inspect air blowers, mattress bladders and hoses
b. Check the mattress interior for moisture or stains
2. Inspect Bed Movement Functions
a. Inspect brakes, steering, and 5th wheel
b. Verify powered drive system
3. Inspecting / Adjusting Movement Limits
a. Verify/adjust:
i. Trendelenburg
ii. Auto Contour
iii. Chair
iv. Hi-Low
v. Head articulation
vi. Foot retraction/articulation
vii. SlideGuard feature
4. Inspect Safety Features
a. Inspect side rails
b. Verify CPR feature
c. Verify alerts/communications:
i. SideCom Communication System
ii. Bed Exit
iii. SafeView System
iv. Wireless Connectivity
5. Scale System / Calibration
a. Check and calibrate the scale system to ensure accurate patient weight display
6. Battery Backup
a. Confirm battery activates when the bed is unplugged and has sufficient power
b. Confirm caregiver controls function on battery backup
7. Electrical Safety Certification:
a. Measure electrical leakage and ground resistance
b. Inspect power cord and line filter for damage
8. Functional Controls
a. Verify caregiver and patient controls operate correctly
9. Improve Appearance
a. Replace labels and plastic parts
10. Cleaning/Lubrication/Adjustments
a. Clean, lubricate moving parts, and make necessary mechanical adjustment per OEM guidance
11. Documentation / Certification
a. Record and provide PM documentation and checklists suitable for accreditation review per OEM.
b. Must provide OEM certified PM checklist plus OEM Service Report generated via Hill-Rom’s CMMS within three (3) business days of each PM.
12. Software/Firmware
a. Apply OEM release firmware/software updated during PM
Corrective Maintenance (CM)
1. Emergency Repair
a. On-Site within 24 hours (Contracting Officer authorization required if outside contract coverage).
2. Routine Repair
a. On-Site within 2 business days.
3. On-site repairs include OEM parts and labor; if a unit cannot be repaired on-site, the Contractor will provide a loaner.
4. Only OEM authorized parts shall be used.
Technical Support
1. 24/7 remote technical support.
2. Remote diagnostics and firmware updates.
Service Documentation:
The contractor shall provide to Biomedical Engineering within three (3) business days after each PM or CM event:
· OEM Service Report from CMMS
· OEM Certified PM Checklist
· Date/time of service
· Equipment serviced
· Action taken and parts replaced
· Test/calibration results (before/after)
· Signatures of contractor technician and VA Biomedical acceptance End of Document
| *= Required Field |
| Special Notice |
Special Notice
File details come from the government source that posted it. Updated .