36C24222Q1158.docx

DOCX document 16 KB Posted

Attached to
H366--BATH VAMC Quarterly Hood Testing - Sources Sought Notice
Solicitation number
36C24222Q1158
Issued by

View the file

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Sources Sought Notice Sources Sought Notice

SUBJECT*
BATH VAMC Quarterly Hood Testing - Sources Sought Notice

GENERAL INFORMATION

CONTRACTING OFFICE’S ZIP CODE*
14227
SOLICITATION NUMBER*
36C24222Q1158
RESPONSE DATE/TIME/ZONE
10:00 EASTERN TIME, NEW YORK, USA
ARCHIVE
60 DAYS AFTER THE RESPONSE DATE
RECOVERY ACT FUNDS
N

SET-ASIDE

PRODUCT SERVICE CODE*
H366
NAICS CODE*
621511
CONTRACTING OFFICE ADDRESS
Department of Veterans Affairs

Network Contracting Office (NCO) 2 2875 Union Road Suite 3500 Cheektowaga NY 14227

POINT OF CONTACT*

Contracting Specialist Zielinski, Daniel daniel.zielinski@va.gov 716-862-7461 x22380

PLACE OF PERFORMANCE

ADDRESS
Bath VA Medical Center

76 Veterans Avenue

Bath NY

POSTAL CODE
14810
COUNTRY
USA

ADDITIONAL INFORMATION

AGENCY’S URL
https://www.va.gov
URL DESCRIPTION
Department of Veterans Affairs
AGENCY CONTACT’S EMAIL ADDRESS
daniel.zielinski@va.gov
EMAIL DESCRIPTION
Contracting Specialist

DESCRIPTION

This is a “REQUEST FOR INFORMATION” only. This Sources Sought Notice is not a request for formal proposals or quotes. It is neither a solicitation announcement nor a request for proposals or quotes and does not obligate the Government to award a contract. Requests for a solicitation will not receive a response. Responses to this Sources Sought must be in writing.

The Bath VA Medical Center located at 76 Veterans Avenue, Bath, NY 14810 has a requirement for semi-annual inspection and certification of two (2) compounding aseptic isolators, semi-annual inspection, and certification of two (2) laminar flow hoods, and quarterly airflow testing. This service requirement shall include:

1. Contractor shall provide proof of technicians completing the following three (3) certifications

a. Controlled Environment Testing Association (CETA) registered Cleanroom Certification Professional for Sterile Compounding Facilities.

b. National Environmental Balancing Bureau (NEBB) Certified for testing, adjusting, and balancing of environmental systems.

c. Testing, Adjusting, and Balancing Bureau (TABB) Certified for testing, adjusting, and balancing of environmental systems.

2. Contractor shall perform semi-annual inspection and certification of two (2) Compounding Aseptic Isolators (CAI) used for compounding category 1 sterile preparations for treatment of patients at the Bath VA Medical Center in accordance with CETA CAG-002-2206 guidelines (or most up to date applicable CAG guideline) and United States Pharmacopeia Convention (USP) 797 guidelines in order to maintain an International Standards Organization (ISO) class 5 environment.

3. Contractor shall provide semi-annual inspection and certification of two (2) laminar flow hoods used for the processing of laboratory specimens in accordance with National Sanitation Foundation (NSF) Standard 49 guidelines.

a. Maintain minimum airflow velocity of 100ft/min (0.5 m/s) through work access opening.

4. Contractor shall perform quarterly air flow testing.

a. Contractor shall perform all testing and certification in accordance with American Society of Heating, Refrigerating, and Air-Conditioning Engineers (AHSARE) guidelines.