23-11-2 VA Brockton Annual Inspection 2023_Redacted.pdf
PDF 750 KB Posted
- Attached to
- Amendment 0001 - Medical Gas Inspections
- Solicitation number
- 36C24124Q0482
About this file
This document is an inspection report for the medical gas systems at the VA Boston Healthcare System - Brockton Campus. The report was conducted in October 2023 to determine if the medical gas systems function as designed, identify areas for improvement, and note what could be upgraded to meet current NFPA 99 code criteria.
The report provides a detailed inventory of the medical gas components, including the number of alarm zones, zone valves, and outlets/inlets in each building and department. It also summarizes the maintenance and compliance deficiencies found, such as leaks, incorrect labeling, and non-compliant piping configurations. The report includes the results of various tests performed, including master alarm, oxygen particulate, medical air moisture, alarm, zone valve, outlet/inlet, and source testing. Recommendations are provided for upgrading the system to meet newer NFPA 99 code requirements. The report concludes with the credentials of the inspectors who conducted the evaluation.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
2023 Medical Gas Testing and Evaluation
Testing Completed October 30, 2023
November 2, 2023
VA Boston Healthcare System – Brockton Campus 940 Belmont Street
Brockton, MA 02301
TABLE OF CONTENTS BY SECTION
Table of Contents Section Introduction
Purpose of Inspection and Evaluation ............................................................... 1.1
Methodology to Complete Inspection .............................................................. 1.2
Signature Page .................................................................................................. 1.3
Inventory of Medical Gas Components
Summary Maintenance Deficiencies
Summary Compliance Deficiencies
Summary General Comments
Newer Health Care Facility Code Recommendations
Master Alarm Testing Results
Oxygen Particulate Testing Results
Medical Air Moisture Testing Results
Alarm Testing Results
Zone Valve Testing Results
Outlet/Inlet Testing Results
Source Testing Results
Verification Reports Completed
Repairs Completed
Credentials
PURPOSE OF INSPECTION AND EVALUATION
The following inspection report is designed using the NFPA 99, 2012 edition code, which is the current Center for Medicare & Medicaid Services standard.
The purpose of an inspection and evaluation of a medical gas system is to determine if the system functions as designed, to suggest areas for improvement, and to note what could be upgraded to meet current NFPA 99 criteria so that patient/employee safety is always being maximized.
Your facility was inspected by representatives using all state-of-the-art testing equipment carrying all credentials necessary to properly inspect medical gas systems.
It is important to understand that the findings in this report may or may not be looked at as a patient or employee safety issue, but each item needs to be evaluated by hospital risk management personnel to determine if there are safety issues.
NFPA 99, 2012 EDITION PARAGRAPH 5.1.1.4 STATES THAT AN EXISTING SYSTEM
THAT IS NOT IN STRICT COMPLIANCE WITH THE PROVISIONS OF THIS CODE
SHALL BE PERMITTED TO BE CONTINUED IN USE AS LONG AS THE AUTHORITY
HAVING JURISDICTION HAS DETERMINED THAT SUCH USE DOES NOT
CONSTITUTE A DISTINCT HAZARD TO LIFE.
This report also may satisfy any preventative maintenance requirement from all regulatory agencies. Please note that our methodology constantly changes to meet any changes in NFPA and other regulatory requirements.
THIS REPORT REFLECTS THE DEFICIENCIES THAT WERE OBSERVED AT THE TIME
OF THE INSPECTION. OTHER DEFICIENCIES MAY ALSO EXIST THAT WERE NOT
OBSERVED DURING THIS INSPECTION.
METHODOLOGY TO COMPLETE INSPECTION
Medical Gas Sources – Medical air compressor, vacuum pumps, gas manifolds, and any other device used to deliver medical gases to patients.
1. Inventoried by location.
2. Leak checked and documented.
3. Recommendations to comply with NFPA 99, 2012 edition Health Care Facilities Code.
Master Alarm Panels – Alarm panels located at two separate locations throughout the facility. These panels monitor conditions directly from the medical gas sources.
2. Tested by removing existing pressure switches and other sensing devices to test the actuation points. No interruption of service is necessary if each device is mounted on code compliant demand fittings.
3. Verify proper points are being monitored at each panel.
4. Verify each panel is in the proper locations.
5. Verify proper labeling.
Area Alarm Panels – Alarm panels that are located throughout the facility that monitor medical gas pressures for the area only.
2. Tested by lowering pressure in each zone for each gas and documenting the alarm point.
3. Verify proper location.
4. Verify proper labeling.
Zone Valve Boxes – Accessible valves typically found in a corridor covered by a plastic window used mostly for emergency purposes.
2. Checked for leakage from packing, flange, and gauge port using oxygen safe leak detector.
3. Verify proper labeling.
4. Verify proper locations.
Medical Gas Outlets/Inlets – Connection points for each gas located mostly in patient areas. (Any outlets/inlets located in Airborne or Droplet precaution rooms will not be tested.)
1. Inventoried by location in each room or area inspected left to right and then top to bottom.
2. Complete an operational pressure test on each outlet and document pass/fail. (Test Parameters per NFPA 99, 2012 edition)
3. Check function of outlet/inlet and latching mechanism.
4. Check for leakage with equipment and without equipment attached.
5. Document connection type and outlet/inlet manufacturer.
Oxygen Particulate Analysis – Particulate is often found in existing medical gas systems due to improper installation methods that were used in the past.
1. Samples are taken on each floor in each building. (See particulate analysis section)
Medical Air Dewpoint Analysis – Moisture is removed from medical air systems at the source. Periodic Analysis of moisture throughout the facility is important to prevent contamination.
1. Samples are taken on each floor in each building. (See Medical Air Dewpoint Analysis Section)
2. A sample will be taken at the medical air source to compare building samples to.
Emergency Oxygen Supply Connection Box – This box is typically mounted on the outside of a facility to facilitate the back feeding of the oxygen system if there is an interruption in service.
2. Check existing location for proper compliance.
3. Check existing piping arrangement to insure proper order of components.
4. Check to make sure box is secured.
Testing to ensure all medical gas booms, and columns are leak free – Most operating rooms and procedure rooms have some sort of device to help clinical personnel use the medical gases in the middle of the room by the patient. These devices are either medical gas booms or columns. The hoses and threaded connections inside these devices need to be checked periodically to ensure that there are no leaks.
1. All hoses and secondary equipment were removed from the outlets on each boom and column.
2. A test pressure for each system was observed and documented.
3. All medical gas valves were shut over a period of 10-minutes.
4. After the test period of 10-minutes had past the pressure was observed once again.
5. Any pressure drop was recorded.
6. Recommendations will be made to find existing leaks and to make repairs.
Reporting
1. All data was collected and broken down into (3) categories:
a. Items that are maintenance issues and need repair, adjustment, calibration, relabeling, or replacement. These items may or may not be an immediate patient safety issue but can be taken care of easily and without much coordination.
b. Design issues that do not meet NFPA 99, 2012 edition Health Care Facilities Code and need to be updated in the future. These items may or may not be patient safety issues but need to be addressed by hospital risk management personnel as soon as possible.
c. General comments that don’t fit in either category A and/or B.
Department Floor Area
Alarm Zone
Valves Outlet/ Inlets
Outlet/Inlets Tested
Miscellaneous Data
Urgent Care 1 1 3 33 21 CT Scan, MRI 1 0 3 6 6
Radiology 2 2 6 15 15 Neuromodulation
Suite 4 1 3 15 15
Sleep Lab 4 1 3 24 24 Neurology 4 2 6 10 10
Dental 5 1 1 11 11
Master Alarm Panels
1 1
Bld 40 1
Total Count: 10 25 114 102 89% Tested
INVENTORY OF MEDICAL GAS COMPONENTS
Building 3
Alarm Zone
Valves Outlet/ Inlets
Outlet/Inlets Tested
Miscellaneous Data
4-1-B 1 1 2 48 48 4-1-C 1 N/A N/A N/A N/A Construction 4-2-B 2 1 2 44 41 4-2-C 2 N/A N/A N/A N/A Construction
Total Count: 2 4 92 89 96% Tested
Building 4
Alarm Zone
Valves Outlet/ Inlets
Outlet/Inlets Tested
Miscellaneous Data
Spinal Cord Injury 1 5 15 93 93
Total Count: 5 15 93 93 100% Tested
Building 8
Alarm Zone
Valves Outlet/ Inlets
Outlet/Inlets Tested
Miscellaneous Data
Building 3 10 25 114 102
Building 4 2 4 92 89
Building 8 5 15 93 93
Master Alarms 2
Total Count: 19 44 299 284 94% Tested
VA Boston Healthcare System - Brockton Campus Total Inventory
SUMMARY OF MAINTENANCE DEFICIENCIES
Master Alarms
1. Master alarm panel located in the Building 40 Boiler Plant has the oxygen high and low signals flipped. The oxygen low comes in as the high alarm, and the oxygen high comes in as the low alarm.
CT Scan, MRI
1. CT Scan (B128) – Vacuum inlet Leaks.
Radiology
1. Ultrasound (B232) – Vacuum inlet leaks.
2. X-Ray (B246) – Vacuum inlet leaks.
3. X-Ray (B247) – Vacuum inlet leaks.
Sleep Lab
1. Room A418 – 2nd oxygen outlet drops 20 psig at 3.5 scfm.
2. Room A419 A:
a. 1st oxygen outlet drops 12 psig at 3.5 scfm.
b. 2nd oxygen outlet drops 10 psig at 3.5 scfm.
3. Room A420 A – 2nd oxygen outlet drops 16 psig at 3.5 scfm.
4. Room A423 – 2nd oxygen outlet drops 12 psig at 3.5 scfm.
5. Room A424 – 1st oxygen outlet drops 14 psig at 3.5 scfm.
Neurology
1. Room A435:
a. 1st oxygen outlet is recessed, and the adapter cannot be completely attached to get proper flow.
b. 2nd oxygen outlet is recessed, and the adapter cannot be completely attached to get proper flow.
2. Room A436 – 1st oxygen outlet is recessed, and the adapter cannot be completely attached to get proper flow.
Dental
1. Area alarm panel located right of Room B515A has an audible leak behind the panel.
4-1-B
1. Room B115 – 1st vacuum inlet leaks.
2. Room B116 – 1st vacuum inlet leaks.
3. Room B122 – 1st vacuum inlet leaks.
4-2-B
1. Room B239 – Vacuum inlet leaks.
Spinal Cord Injury
1. Room A104 – Moisture analysis taken from this location was +38.8°F which is above NFPA 99, 2012 edition criteria.
2. Room A105 – 1st vacuum inlet leaks.
3. Room B111 – 1st vacuum inlet leaks.
Sources
1. Oxygen Supply System (Inside) – Threaded check valve on main line piping located inside Basement Storage Room leaks on the house pressure side.
2. Medical air compressor - Moisture analysis taken at the compressor was +38.2°F which is above NFPA 99, 2012 edition criteria.
SUMMARY OF COMPLIANCE DEFICIENCIES
Neuromodulation Suite
1. Area alarm panel located at the Nurses’ Station needs to be relabeled “Bays 1-4, Procedure Room”.1
2. Zone valves located across from A400 needs to be relabeled “Bays 1-4, Procedure Room”. 2
Sleep Lab
1. Area alarm located right of Storage Room A416 does not have separate visual indicators for each condition monitored.3
Neurology
1. Area alarms located left of A435 do not have separate visual indicators for each condition monitored.4
1. Bulk Oxygen Supply (Outside) – Does not have permanent guard posts installed around the bulk pad site.5
2. Oxygen Supply System (Inside):
a. Piping located inside Room C007A Mechanical is painted. 6
b. Oxygen check valve located on the main line inside Room C007A Mechanical has threaded connections. 7
c. After the source valve and before the main line valve, located inside Room
C007A Mechanical, there is a threaded plug on the pipeline.8
d. Service valve located inside Room C007A Mechanical needs to be labeled.9
3. Medical Air Compressor:
a. At the time of inspection, the operating line pressure was 47 psig which does not fall within the standard of 50-55psig for medical air systems. 10
b. Service valve needs to be labeled.11
4. Vacuum Pump (New):
a. Source valve needs to be relabeled with the name for the specific medical gas, rooms or areas served, and caution to not close or open the valve except in emergency.12
b. After the source valve there is a threaded cap on the pipeline.13
c. All turns, offsets, and other changes in direction to the exhaust piping are soft-soldered and not brazed.14
5. Vacuum Pump (Old):
a. Source valve needs to be relabeled with the name for the specific medical gas, rooms or areas served, and caution to not close or open the valve except in emergency.15
b. High/low pressure switch did not have a gas-specific demand check installed to facilitate service testing or replacement.16
c. High/low pressure switch is not located immediately downstream of the source valve.17
d. Final line gauge did not have a gas-specific demand check installed to facilitate service testing or replacement.18
e. Piping for the vacuum exhaust is not installed with hard-drawn seamless copper in accordance with ASTM B 819.19
f. All turns, offsets, and other changes in direction to the exhaust piping are not brazed.20
g. Multiple service valves do not consist of three pieces permitting in-line serviceability.21
h. Multiple service valves need to be labeled.22
1. All area alarms in Building 4 do not have separate visual indicators for each condition monitored.23
1. Bulk Oxygen Supply (Outside):
a. Oxygen piping needs protection where it enters the ground on the bulk pad.24
b. Bulk tank enclosure does not have two points of entry/exit.25
c. Does not have permanent guard posts installed around the bulk pad site.26
2. Vacuum Pump:
a. All turns, offsets, and other changes in direction to the exhaust piping are not brazed.27
b. Vacuum exhaust is located less than 10 feet from a window.28
1. Bulk Oxygen Supply (Outside):
a. Bulk tank enclosure does not have two points of entry/exit.29
b. Does not have permanent guard posts installed around the bulk pad site.30
c. Currently there is not a minimum three feet of workspace around the storage container, vaporizers and the cabinet opening or front side of the pressure regulating manifold for system maintenance and operation.31
2. Oxygen Supply System (Inside) – Oxygen check valve located on the main line inside the Basement Storage Room has threaded connections.32
3. Emergency Oxygen Supply System (Inside) - Oxygen check valve located on the EOSC piping inside the Basement Storage Room has threaded connections.33
4. Medical Air Compressor:
a. At the time of inspection, the operating line pressure was 47 psig which does not fall within the standard of 50-55psig for medical air systems. 34
b. There is no source valve installed at the medical air compressor.35
c. All turns, offsets, and other changes in direction to the main line and exhaust piping are not brazed.36
d. Piping to the pressure switch is not installed with hard-drawn seamless copper in accordance with ASTM B 819.37
5. Vacuum Pump:
a. Source valve does not consist of three pieces permitting in-line serviceability.38
b. All turns, offsets, and other changes in direction to the exhaust piping are not brazed.39
c. Piping to the pressure switch is not installed with hard-drawn seamless copper in accordance with ASTM B 819.40
d. Low vacuum switch did not have a gas-specific demand check installed to facilitate service testing or replacement.41
e. There is no pressure/vacuum indicator adjacent to the alarm-initiating device for source main line pressure.42
1 NFPA 99, 2012 edition paragraph 5.1.9.1 (7)
2 NFPA 99, 2012 edition paragraph 5.1.4.8.8
3 NFPA 99, 2012 edition paragraph 5.1.9.1 (1)
4 NFPA 99, 2012 edition paragraph 5.1.9.1 (1)
5 NFPA 99, 2012 edition paragraph 5.1.3.5.12.2 and NFPA 55, 2016 edition paragraph 4.11.1
6 NFPA 99, 2012 edition paragraph 5.1.11.1.3
7 NFPA 99, 2012 edition paragraph 5.1.4.11 (4)
8 NFPA 99, 2012 edition paragraph 5.1.10.10
9 NFPA 99, 2012 edition paragraph 5.1.11.2.6
10 NFPA 99, 2012 edition paragraph 5.1.11
11 NFPA 99, 2012 edition paragraph 5.1.11.2.6
12 NFPA 99, 2012 edition paragraph 5.1.11.2.1 and 5.1.11.2.3
13 NFPA 99, 2012 edition paragraph 5.1.10.10
14 NFPA 99, 2012 edition paragraph 5.1.10.3.1 and 5.1.10.4
15 NFPA 99, 2012 edition paragraph 5.1.11.2.1 and 5.1.11.2.3
16 NFPA 99, 2012 edition paragraph 5.1.8.2.3
17 NFPA 99, 2012 edition paragraph 5.1.9.2.5
18 NFPA 99, 2012 edition paragraph 5.1.8.2.3
19 NFPA 99, 2012 edition paragraph 5.1.10.1.4
20 NFPA 99, 2012 edition paragraph 5.1.10.3.1 and 5.1.10.4
21 NFPA 99, 2012 edition paragraph 5.1.4.3 (5)
22 NFPA 99, 2012 edition paragraph 5.1.11.2.6
23 NFPA 99, 2012 edition paragraph 5.1.9.1 (1)
24 NFPA 99, 2012 edition paragraph 5.1.10.11.5.3 (2)
25 NFPA 99, 2012 edition paragraph 5.1.3.3.2 (3)
26 NFPA 99, 2012 edition paragraph 5.1.3.5.12.2 and NFPA 55, 2016 edition paragraph 4.11.1
27 NFPA 99, 2012 edition paragraph 5.1.10.3.1 and 5.1.10.4
28 NFPA 99, 2012 edition paragraph 5.1.3.7.7.2 (2)
29 NFPA 99, 2012 edition paragraph 5.1.3.3.2 (3)
30 NFPA 99, 2012 edition paragraph 5.1.3.5.12.2 and NFPA 55, 2016 edition paragraph 4.11.1
31 NFPA 99, 2012 edition paragraph 5.1.3.5.12.2 (6)
32 NFPA 99, 2012 edition paragraph 5.1.4.11 (4)
33 NFPA 99, 2012 edition paragraph 5.1.4.11 (4)
34 NFPA 99, 2012 edition paragraph 5.1.11
35 NFPA 99, 2012 edition paragraph 5.1.4.4
36 NFPA 99, 2012 edition paragraph 5.1.10.3.1 and 5.1.10.4
37 NFPA 99, 2012 edition paragraph 5.1.10.1.4
38 NFPA 99, 2012 edition paragraph 5.1.4.3 (5)
39 NFPA 99, 2012 edition paragraph 5.1.10.3.1 and 5.1.10.4
40 NFPA 99, 2012 edition paragraph 5.1.10.1.4
41 NFPA 99, 2012 edition paragraph 5.1.8.2.3
42 NFPA 99, 2012 edition paragraph 5.1.8.2.2 (1)
SUMMARY OF GENERAL COMMENTS
1. Building 3:
a. Neuromodulation Suite – If patients in the Procedure Room are being subjected to moderate sedation, deep sedation, or general anesthesia in any of the rooms, then there are specific requirements per NFPA 99 that need to be adhered to. The hospital shall determine what level of sedation/analgesia that is being administered and make the appropriate changes.
b. The exhaust for the old vacuum pump could not be located outside the building and, therefore, was not inspected for proper NFPA 99, 2012 edition compliance.
2. Building 4:
a. Departments 4-1-C and 4-2-C were both under construction during the 2023 annual inspection and could not be inspected.
b. The oxygen enters the building in an area of asbestos contamination and because of this, the final line gauge/pressure switch have never been located.
3. The bulk oxygen signals are not tested during the annual medical gas inspection and evaluation. These signals are tested by your bulk oxygen supplier and should be reported with their annual preventative maintenance program.
These signals include for Building 4:
a. Oxygen main liquid level low.
b. Oxygen reserve in use.
c. Oxygen reserve liquid level low.
d. Oxygen reserve low pressure.
These signals include for Buildings 3 and 8
a. Oxygen main liquid level low.
b. Oxygen reserve in use.
c. Oxygen reserve low pressure.
NEWER CODE RECOMMENDATIONS FOR HEALTHCARE FACILITIES
Effective July 5, 2016, Center for Medicare & Medicaid Services adopted NFPA 99, 2012 edition of the Health Care Facilities Code. The following chart shows the different changes that may affect your facility.
Any items listed below that are included in these future editions may become applicable upon when CMS and/or the state adopts newer editions of the NFPA 99 Health Care Facilities Code.
Item Description NFPA 99, 2015 edition
NFPA 99,
2018 edition
NFPA 99,
2021 edition
The outdoor central supply system, since it is greater than 18.6 m2 (200 ft2), needs to be provided with a minimum of two entry/exits.
5.1.3.3.2 (4)
The cryogenic fluid central supply system does not have an auxiliary source connection installed on the patient side of the source valve.
5.1.3.5.7 5.1.3.5.7 5.1.4.10
The auxiliary source connection valve does not have a removable plug or cap installed.
5.1.3.5.7.1 5.1.3.5.7.1 5.1.4.10.1
The auxiliary source connection valve is not secured in its normally closed position.
5.1.3.5.7.2 5.1.3.5.7.2 5.1.4.10.2
Vacuum exhaust is located less than 25 feet from any door, window, air intake, or other openings in buildings or places of public assembly.
5.1.3.7.6.2 (2) 5.1.3.7.7.2 (2) 5.1.3.7.7.2 (2)
WAGD exhaust is located less than 25 feet from any door, window, air intake, or other openings in buildings or places of public assembly.
5.1.3.8.5 and
5.1.3.7.6.2 (2)
5.1.3.8.5 and
5.1.3.7.7.2 (2)
5.1.3.8.5 and
5.1.3.7.7.2 (2)
There are no alarms installed at the master alarms allowing the temporary oxygen supply to be monitored while in use.
5.1.3.5.13.2 (8)
Medical air intake shall be labeled in accordance with 5.1.11.1 with any method that would distinguish it as a medical air intake.
5.1.3.6.3.11 (I)
Vacuum exhaust shall be labeled in accordance with 5.1.11.1 with any method that would distinguish it as a vacuum exhaust.
5.1.3.7.7.4
WAGD exhaust shall be labeled in accordance with 5.1.11.1 with any method that would distinguish it as a WAGD exhaust.
5.1.3.8.5 and 5.1.3.7.7.4
Specific source, service, and zone valves do not have threaded purge ports on the patient side and source side.
5.1.4.1.6 (8)
Since certain rooms have the station outlets/inlets installed in a down-facing position, the outlets/inlets need to be DISS connectors.
5.1.5.17 5.1.5.17
Specific zone valves throughout the facility need to be relabeled so that the label is not affixed to the removable cover and is visible from outside the zone valve box assembly.
5.1.11.2.7 5.1.11.2.7 5.1.11.2.8
The sleep lab outlets downstream of the flow control device needs to be labeled to include a warning not to use the outlet for ventilating patients.
5.1.11.3.1.1 5.1.11.3.1.1 5.1.11.3.1.1
Source equipment needs to be labeled with the emergency contact information for the department or individual responsible for maintaining the equipment.
5.1.11.5 (4) 5.1.11.5.1
Where vacuum systems are used to serve WAGD systems in accordance with 5.1.10.2.3.1, labeling for the medical surgical vacuum source shall indicate that it serves both systems.
5.1.11.5.2
When clinical spaces are converted to nonclinical spaces, medical gas inlets and outlets that are not accessible for maintenance and testing shall be either removed or decommissioned.
5.1.14.3.5
VA Brockton- Building 3 Area Alarm Testing
Floor Monitors Department Location Labeled Pass/Fail Comment
1 All Sources Outside Urgent Care Telephone Room A-136A, Building 3 Yes Fail 1 All Sources Boiler Plant Building 40 Yes Fail
1 C-Wing Rooms 1-11 Urgent Care Across from Exam 1 Yes Pass 2 B213, B219 Radiology Right of B219 Yes Pass 2 B232, B246, B247 Radiology Left of B232 Yes Pass Main Line 4 Bays 1-4, Procedure Room Neuromodulation Suite Nurses' Station Yes Pass Relabel 4 Rooms A418-A424 Sleep Lab Right of Storage Room A416 Yes Pass No separate visual indicators 4 Room A435 Neurology Left of A435 Left Side Yes Pass No separate visual indicators 4 Room A436 Neurology Left of A435 Right Side Yes Pass No separate visual indicators 5 Operatory 1-11 Dental Right of Room B515A Yes Pass Audible leak behind panel
Master Alarm Testing Results
Building # Sources Line Pressure High Pressure Low Pressure Lag
3 Oxygen 52 psig 61 psig 41 psig N/A High/low flip flopped at boiler panel 3 Medical Air 47 psig 60 psig 42 psig Pass Line pressure not 50-55psig 3 Vacuum (New) -19" Hg N/A -11" Hg Pass 3 Vacuum (Old) -22" Hg N/A Not Tested N/A
4 Oxygen FLG not located Not Tested Not Tested N/A 4 Vacuum 19" Hg N/A -13" Hg N/A
8 Oxygen 52 psig 62 psig 40 psig N/A 8 Vacuum No FLG N/A Not Tested N/A 8 Medical Air 44 psig 55 psig 40 psig N/A Line pressure not 50-55psig
VA Brockton- Building 4
Floor Monitors Department Location Labeled Pass/Fail Comment
Rooms B101, B102, B114- B120, B122, B126, B134, B138, B139 4-1-B Nurses' Station Yes Pass No separate visual indicators
Rooms C101, C102, C114- C116, C118-C120, C122, C126, C133, C137, C138 4-1-C Nurses' Station Yes N/A Construction Area
Rooms B201-B202, B214- B220, B222, B226, B234, B238, B239 4-2-B Nurses' Station Yes Pass No separate visual indicators
Rooms C201-C202, C214-
C220, C222, C237 4-2-C Nurses' Station Yes N/A Construction Area
VA Brockton- Building 8
Floor Monitors Department Location Labeled Pass/Fail Comment 1 Rooms A104- A105 Spinal Cord Injury Nurses' Station Yes Pass 1 Room A171 Spinal Cord Injury Nurses' Station Yes Pass 1 Rooms B107-B112 Spinal Cord Injury Nurses' Station Yes Pass 1 Rooms B117-B124 Spinal Cord Injury Nurses' Station Yes Pass 1 Rooms B128, B130-B131 Spinal Cord Injury Nurses' Station Yes Pass
Zone Valve Testing
Floor Department Controls Gases Location Ports Labeled Notes 1 Urgent Care C-Wing Rooms 1-11 Medical Air Right of Waiting Room Yes Yes 1 Urgent Care C-Wing Rooms 1-11 Oxygen Right of Waiting Room Yes Yes 1 Urgent Care C-Wing Rooms 1-11 Vacuum Right of Waiting Room Yes Yes 1 Ct Scan, MRI B128, B143 Medical Air R/O B131 Yes Yes 1 Ct Scan, MRI B128, B143 Vacuum R/O B131 Yes Yes 1 Ct Scan, MRI B128, B143 Oxygen R/O B131 Yes Yes 2 Radiology B213, B219 Oxygen R/O B219 No Yes 2 Radiology B213, B219 Medical Air R/O B219 No Yes 2 Radiology B213, B219 Vacuum R/O B219 No Yes 2 Radiology B232, B246, B247 Medical Air Across from B232 Yes Yes 2 Radiology B232, B246, B247 Vacuum Across from B232 Yes Yes 2 Radiology B232, B246, B247 Oxygen Across from B232 Yes Yes
Neuromodulation
Suite Bays 1-4, Procedure Room Oxygen Across from A400 No Yes Relabel
Neuromodulation
Suite Bays 1-4, Treatment Room Medical Air Across from A400 No Yes Relabel
Neuromodulation
Suite Bays 1-4, Treatment Room Vacuum Across from A400 No Yes Relabel 4 Sleep Lab Rooms A418-A424 Medical Air Left of Storage Room A416 No Yes Not in Service 4 Sleep Lab Rooms A418-A424 Oxygen Left of Storage Room A416 No Yes 4 Sleep Lab Rooms A418-A424 Vacuum Left of Storage Room A416 No Yes 4 Neurology Rooms A435 Medical Air Left of Room A435 No Yes Not in Service 4 Neurology Rooms A435 Oxygen Left of Room A435 No Yes 4 Neurology Rooms A435 Vacuum Left of Room A435 No Yes 4 Neurology Room A436 Medical Air Right of Room A436 No Yes Not in Service 4 Neurology Room A436 Oxygen Right of Room A436 No Yes 4 Neurology Room A436 Vacuum Right of Room A436 No Yes 5 Dental Operatory 1-11 Oxygen Right of B515A No Yes
Floor Department Controls Gases Location Ports Labeled Notes
1 4-1-B Rooms B101, B102, B114-B120, B122, B126, B134, B138, B139 Oxygen Left of Room B113 No Yes
1 4-1-B Rooms B101, B102, B114-B120, B122, B126, B134, B138, B139 Vacuum Right of Room B113 No Yes
1 4-1-C
Rooms C101, C102, C114-C116, C119, C120, C122, C126, C133, C137, C138 Vacuum Left of C113 N/A N/A Construction Area
1 4-1-C
Rooms C101, C102, C114-C116, C119, C120, C122, C126, C133, C137, C139 Oxygen Left of C113 N/A N/A Construction Area
2 4-2-B Rooms B201-B202, B214-B220, B222, B226, B234, B238, B239 Vacuum Right of B213 No Yes
2 4-2-B Rooms B201-B202, B214-B220, B222, B226, B234, B238, B239 Oxygen Right of B213 No Yes
2 4-2-C Rooms C201-C202, C214-C220, C222, C237 Vacuum Right of C211A N/A N/A Construction Area
2 4-2-C Rooms C201-C202, C214-C220, C222, C237 Oxygen Right of C211A N/A N/A Construction Area
Floor Department Controls Gases Location Ports Labeled Notes 1 Spinal Cord Injury Rooms A104-A105 Oxygen Left of Room A105 No Yes 1 Spinal Cord Injury Rooms A104-A105 Medical Air Left of Room A105 No Yes 1 Spinal Cord Injury Rooms A104-A105 Vacuum Left of Room A105 No Yes 1 Spinal Cord Injury Rooms A171 Oxygen Right of Room A171 No Yes 1 Spinal Cord Injury Rooms A171 Medical Air Right of Room A171 No Yes 1 Spinal Cord Injury Rooms A171 Vacuum Right of Room A171 No Yes 1 Spinal Cord Injury Rooms B107-B112 Oxygen Left of Room B106 No Yes 1 Spinal Cord Injury Rooms B107-B112 Medical Air Left of Room B106 No Yes 1 Spinal Cord Injury Rooms B107-B112 Vacuum Left of Room B106 No Yes 1 Spinal Cord Injury Rooms B117-B124 Oxygen Right of Room B124 No Yes 1 Spinal Cord Injury Rooms B117-B124 Medical Air Right of Room B124 No Yes 1 Spinal Cord Injury Rooms B117-B124 Vacuum Right of Room B124 No Yes 1 Spinal Cord Injury Rooms B128, B130-B131 Oxygen Left of Room B128 No Yes 1 Spinal Cord Injury Rooms B128, B130-B131 Medical Air Left of Room B128 No Yes 1 Spinal Cord Injury Rooms B128, B130-B131 Vacuum Left of Room B128 No Yes
Outlets/Inlet Testing
Department Floor Room # Outlet Type Outlet/Inlet Pass/Fail Notes Urgent Care 1 Room 1 (C142) DISS - Series B Oxygen N/A Covid Precaution
DISS - Series B Vacuum N/A Covid Precaution DISS - Series B Medical Air N/A Covid Precaution
Urgent Care 1 Room 2 (C141) DISS - Series B Oxygen N/A Covid Precaution DISS - Series B Vacuum N/A Covid Precaution DISS - Series B Medical Air N/A Covid Precaution
Urgent Care 1 Room 3 (C140) DISS - Series B Medical Air N/A Covid Precaution DISS - Series B Oxygen N/A Covid Precaution DISS - Series B Vacuum N/A Covid Precaution
Urgent Care 1 Room 4 (C139) DISS - Series B Oxygen N/A Covid Precaution DISS - Series B Medical Air N/A Covid Precaution DISS - Series B Vacuum N/A Covid Precaution
Urgent Care 1 Room 5 (C136) DISS - Series B Medical Air Pass DISS - Series B Oxygen Pass DISS - Series B Vacuum Pass
Urgent Care 1 Room 6 (C135) DISS - Series B Oxygen Pass DISS - Series B Medical Air Pass DISS - Series B Vacuum Pass
Urgent Care 1 Room 7 (C134) DISS - Series B Medical Air Pass DISS - Series B Oxygen Pass DISS - Series B Vacuum Pass
Urgent Care 1 Room 8 (C133) DISS - Series B Oxygen Pass DISS - Series B Medical Air Pass DISS - Series B Vacuum Pass
Urgent Care 1 Room 9 (C132) DISS - Series B Medical Air Pass DISS - Series B Oxygen Pass DISS - Series B Vacuum Pass
Urgent Care 1 Room 10 (C131) DISS - Series B Oxygen Pass DISS - Series B Medical Air Pass DISS - Series B Vacuum Pass
Urgent Care 1 Room 11 (C130) DISS - Series B Medical Air Pass DISS - Series B Oxygen Pass DISS - Series B Vacuum Pass
CT Scan, MRI 1 CT Scan (B128) DISS - 560 Oxygen Pass DISS - 560 Vacuum Fail Leaks DISS - 560 Medical Air Pass
CT Scan, MRI 1 MRI (B143) DISS - 560 Oxygen Pass DISS - 560 Vacuum Pass
Department Floor Room # Outlet Type Outlet/Inlet Pass/Fail Notes DISS - 560 Medical Air Pass
Radiology 2 X-Ray 1 (B213) DISS-Series B Oxygen Pass DISS-Series B Medical Air Pass DISS-Series B Vacuum Pass
Radiology 2 X-Ray 3 (B219) DISS-Series B Oxygen Pass DISS-Series B Medical Air Pass DISS-Series B Vacuum Pass
Radiology 2 Ultrasound (B232) DISS - 560 Oxygen Pass DISS - 560 Vacuum Fail Leaks DISS - 560 Medical Air Pass
Radiology 2 X-Ray (B246) DISS - 560 Oxygen Pass DISS - 560 Vacuum Fail Leaks DISS - 560 Medical Air Pass
Radiology 2 X-Ray (B247) DISS - 560 Oxygen Pass DISS - 560 Vacuum Fail Leaks DISS - 560 Medical Air Pass
Neuromodulation Suite 4 Bay 1 QC-Power X Oxygen Pass QC-Power X Medical Air Pass QC-Power X Vacuum Pass
Neuromodulation Suite 4 Bay 2 QC-Power X Oxygen Pass QC-Power X Medical Air Pass QC-Power X Vacuum Pass
Neuromodulation Suite 4 Bay 3 QC-Power X Oxygen Pass QC-Power X Medical Air Pass QC-Power X Vacuum Pass
Neuromodulation Suite 4 Bay 4 QC-Power X Oxygen Pass QC-Power X Medical Air Pass QC-Power X Vacuum Pass
Neuromodulation Suite 4 Procedure Room QC-Power X Oxygen Pass Anesthesia machine QC-Power X Medical Air Pass QC-Power X Vacuum Pass
Sleep Lab 4 Room A418 DISS - Hill Rom Hose Vacuum Pass DISS - Hill Rom Hose Oxygen Pass DISS - Hill Rom Hose Medical Air N/A Not in service DISS - Hill Rom Hose Medical Air N/A Not in service DISS - Hill Rom Hose Vacuum Pass DISS - Hill Rom Hose Oxygen Fail Drops 20 psig @ 3.5 scfm/Unable to repair
Sleep Lab 4 Room A419 A DISS - Hill Rom Hose Vacuum Pass
Department Floor Room # Outlet Type Outlet/Inlet Pass/Fail Notes DISS - Hill Rom Hose Oxygen Fail Drops 12 psig @ 3.5 scfm/Unable to repair DISS - Hill Rom Hose Medical Air N/A Not in service DISS - Hill Rom Hose Medical Air N/A Not in service DISS - Hill Rom Hose Oxygen Fail Drops 10 psig @ 3.5 scfm/Unable to repair DISS - Hill Rom Hose Vacuum Pass
Sleep Lab 4 Room A420 A DISS - Hill Rom Hose Vacuum Pass DISS - Hill Rom Hose Oxygen Pass DISS - Hill Rom Hose Medical Air N/A Not in service DISS - Hill Rom Hose Medical Air N/A Not in service DISS - Hill Rom Hose Oxygen Fail Drops 16 psig @ 3.5 scfm/Unable to repair DISS - Hill Rom Hose Vacuum Pass
Sleep Lab 4 Room A421 DISS - Hill Rom Hose Vacuum Pass DISS - Hill Rom Hose Oxygen Pass DISS - Hill Rom Hose Medical Air N/A Not in service DISS - Hill Rom Hose Medical Air N/A Not in service DISS - Hill Rom Hose Oxygen Pass DISS - Hill Rom Hose Vacuum Pass
Sleep Lab 4 Room A423 DISS - Hill Rom Hose Vacuum Pass DISS - Hill Rom Hose Oxygen Pass DISS - Hill Rom Hose Medical Air N/A Not in service DISS - Hill Rom Hose Medical Air N/A Not in service DISS - Hill Rom Hose Oxygen Fail Drops 12 psig @ 3.5 scfm/Unable to repair DISS - Hill Rom Hose Vacuum Pass
Sleep Lab 4 Room A424 DISS - Hill Rom Hose Medical Air N/A Not in service DISS - Hill Rom Hose Oxygen Fail Drops 14 psig @ 3.5 scfm/Unable to repair DISS - Hill Rom Hose Vacuum Pass DISS - Hill Rom Hose Vacuum Pass DISS - Hill Rom Hose Oxygen Pass DISS - Hill Rom Hose Medical Air N/A Not in service
Neurology 4 Room A435 DISS - 560 Medical Air N/A Not in service DISS - 560 Oxygen Fail Cannot latch DISS - 560 Vacuum Pass DISS - 560 Medical Air N/A Not in service DISS - 560 Oxygen Fail Cannot latch DISS - 560 Vacuum Pass DISS - 560 Medical Air N/A Not in service DISS - 560 Oxygen Pass
Department Floor Room # Outlet Type Outlet/Inlet Pass/Fail Notes Neurology 4 Room A436 DISS - 560 Medical Air N/A Not in service
DISS - 560 Oxygen Fail Cannot latch DISS - 560 Vacuum Pass DISS - 560 Oxygen Pass DISS - 560 Medical Air N/A Not in service DISS - 560 Vacuum Pass
Dental 5 B514, Operatory 1 DISS - Series B Oxygen Pass Dental 5 B513, Operatory 2 DISS - Series B Oxygen Pass Dental 5 B511, Operatory 3 DISS - Series B Oxygen Pass Dental 5 B510, Operatory 4 DISS - Series B Oxygen Pass Dental 5 B509, Operatory 5 DISS - Series B Oxygen Pass Dental 5 B508, Operatory 6 DISS - Series B Oxygen Pass Dental 5 B507, Operatory 7 DISS - Series B Oxygen Pass Dental 5 B506, Operatory 8 DISS - Series B Oxygen Pass Dental 5 B505, Operatory 9 DISS - Series B Oxygen Pass Dental 5 B525D, Operatory 10 DISS Nitrogen N/A Panel, Not in service
DISS - Series B Oxygen Pass DISS Nitrogen N/A Not in Service
Dental 5 B525C, Operatory 11 DISS - Series B Oxygen Pass DISS Nitrogen N/A Not in Service DISS Nitrogen N/A Panel, Not in service
Department Floor Room # Outlet Type Outlet/Inlet Pass/Fail Notes 4-1-B 1 Room B101 DISS - Disc Unit Oxygen Pass
DISS - Disc Unit Vacuum Pass DISS - Disc Unit Vacuum Pass DISS - Disc Unit Oxygen Pass
4-1-B 1 Room B102 DISS - Disc Unit Oxygen Pass DISS - Disc Unit Vacuum Pass DISS - Disc Unit Vacuum Pass DISS - Disc Unit Oxygen Pass
4-1-B 1 Room B114 DISS - Disc Unit Vacuum Pass DISS - Disc Unit Oxygen Pass
4-1-B 1 Room B115 DISS - Disc Unit Oxygen Pass DISS - Disc Unit Vacuum Fail Leaks DISS - Disc Unit Vacuum Pass DISS - Disc Unit Oxygen Pass
4-1-B 1 Room B116 DISS - Disc Unit Oxygen Pass DISS - Disc Unit Vacuum Fail Leaks DISS - Disc Unit Vacuum Pass DISS - Disc Unit Oxygen Pass
4-1-B 1 Room B117 DISS - Disc Unit Oxygen Pass DISS - Disc Unit Vacuum Pass DISS - Disc Unit Vacuum Pass DISS - Disc Unit Oxygen Pass
4-1-B 1 Room B118 DISS - Disc Unit Oxygen Pass DISS - Disc Unit Vacuum Pass DISS - Disc Unit Vacuum Pass DISS - Disc Unit Oxygen Pass
4-1-B 1 Room B119 DISS - Disc Unit Oxygen Pass DISS - Disc Unit Vacuum Pass DISS - Disc Unit Vacuum Pass DISS - Disc Unit Oxygen Pass
4-1-B 1 Room B120 DISS - Disc Unit Oxygen Pass DISS - Disc Unit Vacuum Pass DISS - Disc Unit Vacuum Pass DISS - Disc Unit Oxygen Pass
4-1-B 1 Room B122 DISS - Disc Unit Oxygen Pass DISS - Disc Unit Vacuum Fail Leaks DISS - Disc Unit Vacuum Pass DISS - Disc Unit Oxygen Pass
Department Floor Room # Outlet Type Outlet/Inlet Pass/Fail Notes 4-1-B 1 Room B126 DISS - Disc Unit Oxygen Pass
DISS - Disc Unit Vacuum Pass 4-1-B 1 Room B134 DISS - Disc Unit Vacuum Pass
DISS - Disc Unit Oxygen Pass 4-1-B 1 Room B138 DISS - Disc Unit Oxygen Pass
DISS - Disc Unit Vacuum Pass 4-1-B 1 Room B139 DISS - Disc Unit Oxygen Pass
DISS - Disc Unit Vacuum Pass DISS - Disc Unit Vacuum Pass DISS - Disc Unit Oxygen Pass
4-1-C 1 Room C101 DISS Oxygen N/A Construction area DISS Vacuum N/A Construction area DISS Vacuum N/A Construction area DISS Oxygen N/A Construction area
4-1-C 1 Room C102 DISS Vacuum N/A Construction area DISS Oxygen N/A Construction area DISS Oxygen N/A Construction area DISS Vacuum N/A Construction area
4-1-C 1 Room C114 DISS Vacuum N/A Construction area DISS Oxygen N/A Construction area
4-1-C 1 Room C115 DISS Oxygen N/A Construction area DISS Vacuum N/A Construction area DISS Vacuum N/A Construction area DISS Oxygen N/A Construction area
4-1-C 1 Room C116 DISS Oxygen N/A Construction area DISS Vacuum N/A Construction area DISS Vacuum N/A Construction area DISS Oxygen N/A Construction area
4-1-C 1 Room C117 DISS Vacuum N/A Construction area DISS Vacuum N/A Construction area
4-1-C 1 Room C118 DISS Vacuum N/A Construction area DISS Vacuum N/A Construction area
4-1-C 1 Room C119 DISS Oxygen N/A Construction area DISS Vacuum N/A Construction area DISS Vacuum N/A Construction area DISS Oxygen N/A Construction area
4-1-C 1 Room C120 DISS Oxygen N/A Construction area DISS Vacuum N/A Construction area
Department Floor Room # Outlet Type Outlet/Inlet Pass/Fail Notes DISS Vacuum N/A Construction area DISS Oxygen N/A Construction area
4-1-C 1 Room C122 DISS Oxygen N/A Construction area DISS Vacuum N/A Construction area DISS Vacuum N/A Construction area DISS Oxygen N/A Construction area
4-1-C 1 Room C126 DISS Oxygen N/A Construction area DISS Oxygen N/A Construction area
4-1-C 1 Room C133 DISS Oxygen N/A Construction area DISS Oxygen N/A Construction area
4-1-C 1 Room C137 DISS Vacuum N/A Construction area DISS Oxygen N/A Construction area
4-1-C 1 Room C138 DISS Oxygen N/A Construction area DISS Vacuum N/A Construction area DISS Vacuum N/A Construction area DISS Oxygen N/A Construction area
4-2-B 2 Room B201 QC Oxygen N/A Not in Service DISS - Disc Unit Vacuum Pass DISS - Disc Unit Vacuum Pass
QC Oxygen N/A Not in Service 4-2-B 2 Room B202 DISS - Disc Unit Oxygen Pass
DISS - Disc Unit Vacuum Pass DISS - Disc Unit Vacuum Pass DISS - Disc Unit Oxygen Pass
4-2-B 2 Room B214 DISS - Disc Unit Vacuum Pass DISS - Disc Unit Oxygen Pass
4-2-B 2 Room B215 DISS - Disc Unit Oxygen Pass DISS - Disc Unit Vacuum Pass DISS - Disc Unit Vacuum Pass DISS - Disc Unit Oxygen Pass
4-2-B 2 Room B216 DISS - Disc Unit Oxygen Pass DISS - Disc Unit Vacuum Pass DISS - Disc Unit Vacuum Pass DISS - Disc Unit Oxygen Pass
4-2-B 2 Room B217 DISS - Disc Unit Oxygen Pass DISS - Disc Unit Vacuum Pass DISS - Disc Unit Vacuum Pass DISS - Disc Unit Oxygen Pass
Department Floor Room # Outlet Type Outlet/Inlet Pass/Fail Notes 4-2-B 2 Room B218 DISS - Disc Unit Oxygen Pass
DISS - Disc Unit Vacuum Pass DISS - Disc Unit Vacuum Pass DISS - Disc Unit Oxygen N/A In use
4-2-B 2 Room B219 DISS - Disc Unit Oxygen Pass DISS - Disc Unit Vacuum Pass DISS - Disc Unit Vacuum Pass DISS - Disc Unit Oxygen Pass
4-2-B 2 Room B220 DISS - Disc Unit Oxygen Pass DISS - Disc Unit Vacuum Pass DISS - Disc Unit Vacuum Pass DISS - Disc Unit Oxygen Pass
4-2-B 2 Room B222 DISS - Disc Unit Oxygen Pass DISS - Disc Unit Vacuum Pass DISS - Disc Unit Vacuum Pass DISS - Disc Unit Oxygen Pass
4-2-B 2 Room B226 DISS - Disc Unit Oxygen Pass DISS - Disc Unit Vacuum Pass
4-2-B 2 Room B234 DISS - Disc Unit Oxygen Pass DISS - Disc Unit Vacuum Pass
4-2-B 2 Room B238 DISS - Disc Unit Vacuum N/A Blocked DISS - Disc Unit Oxygen N/A Blocked
4-2-B 2 Room B239 QC Oxygen N/A Not in Service DISS - Disc Unit Vacuum Pass DISS - Disc Unit Vacuum Fail Leaks
QC Oxygen N/A Not in Service 4-2-C 2 Room C201 DISS Oxygen N/A Construction area
DISS Vacuum N/A Construction area DISS Vacuum N/A Construction area DISS Oxygen N/A Construction area
4-2-C 2 Room C202 DISS Oxygen N/A Construction area DISS Vacuum N/A Construction area DISS Vacuum N/A Construction area DISS Oxygen N/A Construction area
4-2-C 2 Room C214 DISS Oxygen N/A Construction area DISS Vacuum N/A Construction area
4-2-C 2 Room C215 DISS Oxygen N/A Construction area DISS Vacuum N/A Construction area
Department Floor Room # Outlet Type Outlet/Inlet Pass/Fail Notes DISS Vacuum N/A Construction area DISS Oxygen N/A Construction area
4-2-C 2 Room C216 DISS Vacuum N/A Construction area DISS Vacuum N/A Construction area DISS Oxygen N/A Construction area DISS Oxygen N/A Construction area
4-2-C 2 Room C217 DISS Oxygen N/A Construction area DISS Vacuum N/A Construction area DISS Vacuum N/A Construction area DISS Oxygen N/A Construction area
4-2-C 2 Room C218 DISS Oxygen N/A Construction area DISS Vacuum N/A Construction area DISS Vacuum N/A Construction area DISS Oxygen N/A Construction area
4-2-C 2 Room C219 DISS Oxygen N/A Construction area DISS Vacuum N/A Construction area DISS Vacuum N/A Construction area DISS Oxygen N/A Construction area
4-2-C 2 Room C220 DISS Oxygen N/A Construction area DISS Vacuum N/A Construction area DISS Vacuum N/A Construction area DISS Oxygen N/A Construction area
4-2-C 2 Room C222 DISS Oxygen N/A Construction area DISS Vacuum N/A Construction area DISS Vacuum N/A Construction area DISS Oxygen N/A Construction area
4-2-C 2 Room C237 DISS Vacuum N/A Construction area DISS Oxygen N/A Construction area
Department Floor Room # Outlet Type Outlet/Inlet Pass/Fail Notes Spinal Cord Injury 1 Room A104 DISS - 560 Medical Air Pass
DISS - 560 Vacuum Pass DISS - 560 Oxygen Pass DISS - 560 Medical Air Pass DISS - 560 Vacuum Pass DISS - 560 Oxygen Pass
Spinal Cord Injury 1 Room A105 DISS - 560 Medical Air Pass DISS - 560 Vacuum Fail Leaks DISS - 560 Oxygen Pass DISS - 560 Medical Air Pass DISS - 560 Vacuum Pass DISS - 560 Oxygen Pass
Spinal Cord Injury 1 Room A171 DISS - 560 Oxygen Pass DISS - 560 Medical Air Pass DISS - 560 Vacuum Pass DISS - 560 Oxygen Pass DISS - 560 Medical Air Pass DISS - 560 Vacuum Pass
Spinal Cord Injury 1 Room B107 DISS - 560 Medical Air Pass DISS - 560 Vacuum Pass DISS - 560 Oxygen Pass
Spinal Cord Injury 1 Room B108 DISS - 560 Medical Air Pass DISS - 560 Vacuum Pass DISS - 560 Oxygen Pass DISS - 560 Medical Air Pass DISS - 560 Vacuum Pass DISS - 560 Oxygen Pass
Spinal Cord Injury 1 Room B109 DISS - 560 Medical Air Pass DISS - 560 Vacuum Pass DISS - 560 Oxygen Pass DISS - 560 Medical Air Pass DISS - 560 Vacuum Pass DISS - 560 Oxygen Pass
Spinal Cord Injury 1 Room B110 DISS - 560 Medical Air Pass DISS - 560 Vacuum Pass DISS - 560 Oxygen Pass DISS - 560 Medical Air Pass
Department Floor Room # Outlet Type Outlet/Inlet Pass/Fail Notes DISS - 560 Oxygen Pass
Spinal Cord Injury 1 Room B111 DISS - 560 Medical Air Pass DISS - 560 Vacuum Fail Leaks DISS - 560 Oxygen Pass DISS - 560 Medical Air Pass DISS - 560 Vacuum Pass DISS - 560 Oxygen Pass
Spinal Cord Injury 1 Room B112 DISS - 560 Medical Air Pass DISS - 560 Vacuum Pass DISS - 560 Oxygen Pass DISS - 560 Medical Air Pass DISS - 560 Vacuum Pass DISS - 560 Oxygen Pass
Spinal Cord Injury 1 Room B117 DISS - 560 Medical Air Pass DISS - 560 Vacuum Pass DISS - 560 Oxygen Pass
Spinal Cord Injury 1 Room B118 DISS - 560 Medical Air Pass DISS - 560 Vacuum Pass DISS - 560 Oxygen Pass
Spinal Cord Injury 1 Room B119 DISS - 560 Medical Air Pass DISS - 560 Vacuum Pass DISS - 560 Oxygen Pass
Spinal Cord Injury 1 Room B120 DISS - 560 Medical Air Pass DISS - 560 Vacuum Pass DISS - 560 Oxygen Pass
Spinal Cord Injury 1 Room B121 DISS - 560 Medical Air Pass DISS - 560 Vacuum Pass DISS - 560 Oxygen Pass
Spinal Cord Injury 1 Room B122 DISS - 560 Medical Air Pass DISS - 560 Vacuum Pass DISS - 560 Oxygen Pass
Spinal Cord Injury 1 Room B123 DISS - 560 Medical Air Pass DISS - 560 Vacuum Pass DISS - 560 Oxygen Pass
Spinal Cord Injury 1 Room B124 DISS - 560 Medical Air Pass DISS - 560 Vacuum Pass
Spinal Cord Injury 1 Room B128 DISS - 560 Medical Air Pass
Department Floor Room # Outlet Type Outlet/Inlet Pass/Fail Notes DISS - 560 Vacuum Pass DISS - 560 Oxygen Pass DISS - 560 Medical Air Pass DISS - 560 Vacuum Pass DISS - 560 Oxygen Pass
Spinal Cord Injury 1 Room B130 DISS - 560 Medical Air Pass DISS - 560 Vacuum Pass DISS - 560 Oxygen Pass DISS - 560 Medical Air Pass DISS - 560 Vacuum Pass DISS - 560 Oxygen Pass
Spinal Cord Injury 1 Room B131 DISS - 560 Medical Air Pass DISS - 560 Vacuum Pass DISS - 560 Oxygen Pass DISS - 560 Medical Air Pass
Floor Source Manufacturer Location Notes Outside Bulk Oxygen Supply Air Gas Outside behind Building 3 No permanent guard posts
B Oxygen Supply System Room C007A Mechanical
(Inside Room C007) Painted pipe inside Room C007A Mechanical Threaded check valve on main line (Inside) Threaded plug on main line (Inside) Service valve inside Room C007A Mechanical needs labeling
Outside EOSC Left of Urgent Care Entrance
B Medical Air Compressor Beacon Medaes Room A021 Service valve needs labeling Dew Point +16.3
B New Vacuum Pump Beacon Medaes Room A021 Source valve needs relabel Threaded cap after source valve Soft soldered exhaust outside
B Old Vacuum Pump Room A025 Relabel source valve Pressure switch has no demand check Pressure switch on source side of source valve Final line gauge has no demand check Galvanized threaded exhaust Multiple gate vales, not labeled Unable to locate exhaust outside
Floor Source Manufacturer Location Notes Outside Bulk Oxygen Supply Air Gas Behind Building 4, By the Loading Dock Pipe needs protection where it enters ground
Bulk pad has only one gate No permanent guard posts
B Oxygen Supply System Pressure switch located in contaminated area (Inside)
Outside EOSC Right of Landing Dock
Penthouse Vacuum Pump Standard Pneumatic
Products PHC-04 Exhaust is soft soldered Exhaust is not 10 feet from windows
Floor Source Manufacturer Location Notes Outside Bulk Oxygen Supply Airgas Outside Pad behind Building 8 Bulk pad has only one gate
No permanent guard posts No 3' clearance around components
B Oxygen Supply System Basement Storage Room Threaded check valve on main piping (Inside)
Outside EOSC Right of Main Entrance
B Emergency Oxygen Supply
System Basement Storage Room Threaded check valve on EOSC piping (Inside)
B Medical Air Compressor Powerex Room A006 No Source Valve Soft soldered piping, main and exhaust Soft copper on pressure switch extension Dew Point +38.8
B Vacuum Pump Ingersoll Rand Room A006 Source valve not 3-piece Exhaust soft soldered Soft copper on pressure switch extension No demand check on pressure switch No final line gauge