36C10G20Q0027-000.docx

DOCX document Posted

Attached to
6515--NX EQ Multifunction Corneal Analyzer
Solicitation number
36C10G20Q0027
Issued by

About this file

36C10G20Q0027 36C10G20Q0027.docx

View the file

Other files

File Type Posted
36C10G20Q0027-0001003.docx DOCX document
36C10G20Q0027-001.xlsx XLSX spreadsheet
36C10G20Q0027-004.docx DOCX document
36C10G20Q0027-005.xlsx XLSX spreadsheet
36C10G20Q0027-002.docx DOCX document
36C10G20Q0027-003.docx DOCX document

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Contract Opportunity Combined Synopsis/Solicitation Notice

CLASSIFICATION CODE

SUBJECT

CONTRACTING OFFICE'S

ZIP-CODE

SOLICITATION NUMBER

RESPONSE DATE (MM-DD-YYYY)

ARCHIVE

DAYS AFTER THE RESPONSE DATE

RECOVERY ACT FUNDS

SET-ASIDE

NAICS CODE

CONTRACTING OFFICE

ADDRESS

POINT OF CONTACT

(POC Information Automatically Filled from User Profile Unless Entered)

DESCRIPTION

See Attachment

AGENCY'S URL

URL DESCRIPTION

AGENCY CONTACT'S EMAIL

ADDRESS

EMAIL DESCRIPTION

ADDRESS

POSTAL CODE

COUNTRY

ADDITIONAL INFORMATION

GENERAL INFORMATION

PLACE OF PERFORMANCE

* = Required Field Contract Opportunity Combined Synopsis/Solicitation Notice NX EQ Multifunction Corneal Analyzer 22408-2697 36C10G20Q0027 04-27-2020 N 339112 U.S. Department of Veterans Affairs OPAL | Strategic Acquisition Center 10300 Spotsylvania Ave | STE 400 Fredericksburg VA 22408-2697 Ninatile Bacote Ninatile.Bacote@va.gov Nationwide https://www.va.gov/OAO/sac.asp

VA | OALC | OPAL | SAC

Ninatile.Bacote@va.gov Ninatile.Bacote@va.gov Solicitation Number:

36C10G20Q0027

Notice Type:

Combined Synopsis/Solicitation

Synopsis:

COMBINED SYNOPSIS/SOLICITATION

This is a combined synopsis/solicitation for brand name or equal commercial items prepared in accordance with the format in Federal Acquisition Regulation (FAR) subpart 12.6, “Streamlined Procedures for Evaluation and Solicitation for Commercial Items,” as supplemented with additional information included in this notice. This announcement constitutes the only solicitation; quotes are being requested and a written solicitation will not be issued.

Solicitation number 36C10G20Q0027 is issued for this combined synopsis/solicitation; this solicitation is being issued as a Request for Quote (RFQ). This combined synopsis/solicitation incorporates provisions and clauses in effect through Federal Acquisition Circular (FAC) 2020-04, dated 15 January 2020.

Requirement Description:

VA has a requirement to purchase Multifunction Corneal Analyzer with Keratometry, Refractometry and Topography to deliver patient care at its VA Medical Centers around the country. Multifunction Corneal Analyzer with Keratometry, Refractometry and Topography is an advanced vision-assessment system that combines topography, autorefraction and keratometry, allowing accurate and reliable analysis of corneal aberration. Some units may also include wave front (a technology that makes custom laser eye surgery possible by precisely measuring and diagramming the aberrations of an optical system), and pupillometry (a measurement of the pupil size and reactivity) analysis.

A single-award Firm-Fixed Price Requirements contract will be awarded for the purchase of brand name or equal equipment in accordance with all terms, conditions, provisions, specifications, and schedule of this solicitation herein. Quotes shall contain the terms for cost/price and technical capabilities of the band name or equal equipment. The Government reserves the right to award without discussions.

The associated North American Industrial Classification System (NAICS) code for this procurement is 339112 – Surgical and Medical Instrument Manufacturing, Product Service Code: 6515 – Medical and Surgical Instruments, Equipment, and Supplies and the associated size standard is 1,000 employees. This procurement action is issued as full and open competition.

This acquisition is for a single-award Firm-Fixed Price Requirements contract in accordance with FAR Part 16.5 for Topcon® Wave-Front Analyzer brand name or equal Multifunction Corneal Analyzer and related equipment and accessories as identified in ATTACHMENT A – NX EQ MULTIFUNCTION CORNEAL ANALYZER – PRICE COST SCHEDULE. All interested companies shall provide a quote for all Contract Line Item Numbers (CLINs) listed in ATTACHMENT A – NX EQ MULTIFUNCTION CORNEAL ANALYZER – PRICE COST SCHEDULE.

The ordering period is for one year with two 12-month option years. Delivery is FOB destination. Firm-Fixed Price Orders will be placed against this contract in writing and will provide the delivery locations, delivery dates and exact quantities.

The FAR provision at 52.212-1, Instructions to Offerors – Commercial, applies to this acquisition including attached addenda to the provision.

The FAR provision at 52.212-2, Evaluation -- Commercial Items, and the specific evaluation criteria as attached addenda also applies to this acquisition The Offeror is to include a completed copy of the provision at 52.212-3, Offeror Representations and Certifications -- Commercial Items, with its offer.

Clause at 52.212-4, Contract Terms and Conditions -- Commercial Items, applies to this acquisition and a statement regarding any addenda to the clause. Addendum is attached.

Clause at 52.212-5, Contract Terms and Conditions Required to Implement Statutes or Executive Orders -- Commercial Items, applies to this acquisition and a statement regarding which, if any, of the additional FAR clauses cited in the clause are applicable to the acquisition. Addendum is attached.

All interested Offerors must be registered in the System for Award Management (SAM) prior to submitting an offer or quotation. You may access SAM at https://sam.gov/SAM/.

Questions regarding this combined synopsis/solicitation are due via email by 11:59PM EST on April 8, 2020 to Contract Specialist, Ninatile Bacote at Ninatile.Bacote@va.gov and Contracting Officer, Timothy Johnson at Timothy.Johnson13@va.gov. No calls will be accepted.

Quotes are due via email by 11:59PM EST on April 27, 2020 to Contract Specialist, Ninatile Bacote at Ninatile.Bacote@va.gov and Contracting Officer, Timothy Johnson at Timothy.Johnson13@va.gov.

Product Description Salient Characteristics Literature Review

1. SCOPE OF WORK:

The Veterans Health Administration (VHA) Non-Expendable Equipment National Program has identified the Multifunction Corneal Analyzer, with Keratometry, Refractometry and Topography product line as a candidate for a VA-wide (otherwise referred to as “national single Requirements contract award.

The Multifunction Corneal Analyzer is an advanced vision-assessment system that combines topography, autorefraction and keratometry, allowing accurate and reliable analysis of corneal aberration. Some units may also include wavefront (a technology that makes custom laser eye surgery possible by precisely measuring and diagramming the aberrations of an optical system), and pupillometry (a measurement of the pupil size and reactivity) analysis.

Corneal Topography provides intuitive maps and numerical data for the corneal surface. The Refractometer provides exceptionally accurate refractions for various pupil diameters including refractions under photopic and mesopic conditions, critical for proper assessment of both refractive surgery patients and common refractive problems.

The Keratometer is a device used for the measurement of cornea anterior surface curvature across a fixed chord length.

The versatility incorporated in one compact unit allows clinicians to obtain broad and precise information about the refractive status of the eye enabling comprehensive analysis and assessment, utilizing state-of-the-art data.

The Contractor must demonstrate the ability to meet all requirements for the solicitation. The objective is to provide Multifunction Corneal Analyzers to be used by clinicians throughout the VA medical centers and facilities. The period of performance is for one 12-month base period with two 12-month option periods.

2. REQUIREMENT

This requirement will be IAW FAR clause 52.211-6, “Brand Name or Equal” which requires the quoter indicate that product being offered as an “equal” product to the TOPCON® Wave-Front Analyzer brand name Multifunction Corneal Analyzer. For each equal product, the offeror must include a description reflecting the salient characteristics (SC) and level of quality that will satisfy the salient physical, functional, or performance characteristics of the equal product(s) specified in the solicitation. The quoter must also clearly identify the item by brand name (if any) and make/model number. Finally, the quoter must include descriptive literature, such as illustrations, drawings, or a clear reference to previously furnished descriptive data or information available to the Contracting Officer, and clearly describe any modifications it plans to make to a product to make it conform to the solicitation requirements.

CONTRACT LINE

ITEM

MANUFACTURER
PART

NUMBER

DESCRIPTION

0001
Topcon
15091
KR-1W, Wave-Front Analyzer.

Five functions in one machine:

1. Auto-Refraction

2. Auto-Keratometry

3. Aberrometry

4. Topography

5. Pupillometry with Printer

The Department of Veterans Affairs (VA) is seeking vendors who can provide TOPCON® Wave-Front Analyzer brand name Multifunction Corneal Analyzer as listed above or equal commodities which meet all the following salient characteristics:

CLIN’s 0001 applies to the following SC’s

SC #
SALIENT CHARACTERISTICS
METHOD OF EVALUATION
SC 1
Unit Must Combine a Minimum of a Topographer, Auto Refractometer and Auto Keratometer
Literature Review
SC 2
Refractometer Pupil Diameter Must Measure a Minimum of 2.0 mm
Literature Review
SC 3
Keratometer Must Measure a Curvature Radius of 5.00 to 10.00 mm
Literature Review
SC4
Keratometer Must Have a Corneal Refraction Range from 67.5 D to 33.75 D
Literature Review
SC 5
Must be Able to Connect to a Printer
Literature Review
SC 6
Must Have a Minimum 8.5-inch LCD Touch Screen
Literature Review
SC 7
Must be Able to Interface with Electronic Medical/Health Records
Literature Review
SC 8
Refractometer must have minimum spherical refractive range of -20 diopters to +22 diopters
Literature Review
SC 9
Refractometer Must Have Minimum Cylinder Refractive Range of 0 D to +/- 10 Diopters
Literature Review

See attached document: ATTACHMENT A - PRICE COST SCHEDULE - NX EQ MULTIFUNCTION CORNEAL ANALYZER.

See attached document: ATTACHMENT B - CONTRACT ADMINISTRATION - NX EQ MULTIFUNCTION CORNEAL ANALYZER.

See attached document: ATTACHMENT C - CONTRACT CLAUSES - NX EQ MULTIFUNCTION CORNEAL ANALYZER.

See attached document: ATTACHMENT D - SOLICITATION PROVISIONS - NX EQ MULTIFUNCTION CORNEAL ANALYZER.

See attached document: ATTACHMENT E - REPORTING TOOL - NX EQ MULTIFUNCTION CORNEAL ANALYZER.

File details come from the government source that posted it. Updated .