PRODUCT DESCRIPTION
For
NON-EXPENDABLE EQUIPMENT:
Medical Graphics Corporation Ultima PF Pulmonary Function System
Prepared By:
Strategic Acquisition Center Department of Veterans Affairs 10300 Spotsylvania Avenue, Suite 400 Fredericksburg, VA 22408
1. SCOPE OF WORK:
The Veterans Health Administration (VHA) Non-Expendable Equipment National Program has identified the Medical Graphics Corporation Ultima PF Brand Name or Equal Pulmonary Function System as a candidate for a VA-wide (otherwise referred to as “national”) single-award Requirements contract.
Pulmonary function analyzers measure the respiratory performance of patients with suspected pulmonary disease. Pulmonary function testing provides objective data about the clinical status of a patient's lungs, such as ventilation, diffusion, and distribution of gases in the lungs. With this information, the physician can determine baseline ventilatory function, along with types and degrees of impairments; effects of subsequent therapy; and, often, a disease's underlying mechanism. Pulmonary function testing is typically performed on patients with known or suspected pulmonary disease, often after screening with a spirometer or in the presence of symptoms.
The Contractor must demonstrate the ability to meet all requirements for the solicitation. The objective is to provide a pulmonary function analyzer to be used by clinicians throughout the VA medical centers and facilities. The period of performance will be for one 12-month base period with two 12-month option periods.
2. REQUIREMENT
This requirement will be IAW FAR clause 52.211-6, “Brand Name or Equal” which requires the vendor to indicate that each product being offered is an “equal” product to the Medical Graphics Corporation Ultima PF Pulmonary Function System and associated equipment listed below in Table A.
For each equal product, the vendor must include a description reflecting the salient characteristics (SC) and level of quality that will satisfy the salient physical, functional, or performance characteristics of the equal product(s) specified in the solicitation. The vendor must also clearly identify the item by brand name (if any) and make/model number. Finally, the vendor must include descriptive literature, such as illustrations and drawings and clearly describe any modifications it plans to make to a product to make it conform to the solicitation requirements.
TABLE A
| Contract Line Item |
| Manufacturer |
| Part Number |
| Description |
| 0001 |
| Medical Graphics Corporation |
| 800820-011 |
| Ultima PF Pulmonary Function System RTD with UPS (Uninterruptable Power Supply) |
| 0002 |
| Medical Graphics Corporation |
| 147612-311 |
| Bronchial Provocation Software (software used for methacholine, mannitol, exercise or cold-air challenges) |
| 0003 |
| Medical Graphics Corporation |
| 147612-312 |
| Multi-User Software License (license for software for devices that are networked or interface) |
| 0004 |
| Medical Graphics Corporation |
| 147612-309 |
| Pulmonary Consult Software (allows for computerized assessment of testing results) |
| 0005 |
| Medical Graphics Corporation |
| 147612-308 |
| Query Database Software License (allows for queries of the patient database) |
| 0006 |
| Medical Graphics Corporation |
| 704001-004 |
| Calibration Syringe, 3L (device used for calibration) |
The Department of Veterans Affairs (VA) is seeking vendors who can provide the Medical Graphics Corporation Ultima PF Pulmonary Function System and associated equipment as listed above or Brand Name or Equal commodities which meet all the following salient characteristics:
CLIN 0001
Ultima PF Pulmonary Function System RTD with UPS (Uninterruptable Power Supply) Part Number 800820-011
| SC # |
| MINIMUM SALIENT CHARACTERISTICS |
| METHOD OF EVALUATION |
| LOCATION IN LITERATURE |
SC 1
| Shall include a mobile medical cart with wheels for room to room portability. |
| Literature Review |
| SC 2 |
| Shall be capable of Spirometry testing to include FVC, SVC, and MVV. |
| Literature Review |
| SC 3 |
| Shall be capable of performing Respiratory Mechanics to include MIP and MEP. |
| Literature Review |
| SC4 |
| Shall be capable of Diffusing Capacity testing. |
| Literature Review |
| SC 5 |
| Shall be capable of Nitrogen Washout testing. |
| Literature Review |
| SC 6 |
| Shall be capable of DLco Single Breath testing. |
| Literature Review |
| SC 7 |
| Shall be capable of calibration. |
| Literature Review |
CLIN 0002
Bronchial Provocation Software Part Number 147612-311
| SC # |
| MINIMUM SALIENT CHARACTERISTICS |
| METHOD OF EVALUATION |
| LOCATION IN LITERATURE |
| Shall be capable of providing bronchial challenge testing. |
| Literature Review |
CLIN 0003
Multi-User Software License Part Number 147612-312
| SC # |
| MINIMUM SALIENT CHARACTERISTICS |
| METHOD OF EVALUATION |
| LOCATION IN LITERATURE |
| Shall enable devices to be networked or interfaced. |
| Literature Review |
CLIN 0004
Pulmonary Consult Software Part Number 147612-309
| SC # |
| MINIMUM SALIENT CHARACTERISTICS |
| METHOD OF EVALUATION |
| LOCATION IN LITERATURE |
| Shall allow computerized assessment and/or interpretation of testing results. |
| Literature Review |
CLIN 0005
Query Database Software License Part Number 147612-308
| SC # |
| MINIMUM SALIENT CHARACTERISTICS |
| METHOD OF EVALUATION |
| LOCATION IN LITERATURE |
| Shall allow for queries of the patient database. |
| Literature Review |
image1.png