Alcon Constellation Vitrectomy Units
This requirement will be IAW FAR clause 52.211-6, “Brand Name or Equal” which requires that the vendor indicate each product that is being offered as an “equal” product. For each equal product, the vendor must include a description reflecting the characteristics and level of quality that will satisfy the salient physical, functional, or performance characteristics of the equal product(s) specified below The vendor must clearly identify the item by brand name (if any) and make/model number. The vendor must include descriptive literature, such as illustrations, drawings, or a clear reference to previously furnished descriptive data or information available to the contracting officer.
VENDORS SUBMITTING AN "OR EQUAL" QUOTE MUST MEET ALL THE MINIMUM SALIENT CHARACTERISTICS LISTED BELOW.
Minimum Salient Characteristics (SC) for the Vitrectomy Units (Literature Review)
| CLIN |
| Brand |
| Part Number |
| Description |
| 0001 |
| Alcon |
| Model |
| CONSTELLATION LXT |
| 0002 |
| Alcon |
| 8065750888 |
| Fragmentation Handpiece |
| 0003 |
| Alcon |
| 8065797005 |
| Constellation SF6 Tank 125GM |
| 0004 |
| Alcon |
| 8065751051 |
| Zeiss/Moeller Fixed Laser Safety Filter |
| 0005 |
| Alcon |
| 8065797105 |
| Constellation C3F8 Tank 125GM |
| 0006 |
| Alcon |
| 8065751502 |
| Leica/Wild Fixed Quad Filter |
| SC # |
| Applicable CLINs |
| SALIENT CHARACTERISTICS |
| METHOD OF EVALUATION |
| SC 1 |
| 0001-0002 |
| Vitrectomy Cut Rate must be a minimum of 100-800 cuts per minute |
| Literature Review |
| SC 2 |
| 0001, 0003, 0005, 0006 |
| Aspiration Operating Range must be 0-600 mmHg |
| Literature Review |
| SC 3 |
| 0001 |
| Operational Modes must include Irrigation, Aspiration, & Vitrectomy |
| Literature Review |
| SC 4 |
| 0001 |
| Pressurized Infusion rate must be a minimum of 1-100 mmHg |
| Literature Review |
| SC 5 |
| 0001 |
| Must have Remote Control |
| Literature Review |