Marco Brand Keratometers or Equal

This requirement will be IAW FAR clause 52.211-6, “Brand Name or Equal” which requires that the vendor indicate each product that is being offered as an “equal” product. For each equal product, the vendor must include a description reflecting the characteristics and level of quality that will satisfy the salient physical, functional, or performance characteristics of the equal product(s) specified below The vendor must clearly identify the item by brand name (if any) and make/model number. The vendor must include descriptive literature, such as illustrations, drawings, or a clear reference to previously furnished descriptive data or information available to the contracting officer.

VENDORS SUBMITTING AN "OR EQUAL" QUOTE MUST MEET ALL THE MINIMUM SALIENT CHARACTERISTICS LISTED BELOW.

Minimum Salient Characteristics (SC) for the Keratometers (Literature Review)

CLINSREF
SC #
SALIENT CHARACTERISIT
METHOD OF EVALUATION
0001-0004
SC 1
Must be Food & Drug Administration Approved
Literature Review
0001-0002
SC 2
Refractive Measurement Range:

Sphere: -30D to +25D (in case VD=12) (Step: 0.12/0.25D) Cylinder: 0 to ± 12D (Step: 0.12/.25D) Axis angle: 0 to 180° (Step: 1°) Literature Review

0001-0002
SC 3
Minimum Measurable Pupil Diameter: 2.0mm
Literature Review
0001-0002
SC 4
Keratometric Measurement Range:

Radius of curvature: 5.0 to 13.0 mm (Step: 0.01 mm) Corneal Power: 25.96 to 67.5D (Corneal refractive n=1.3375) (Step: 0.12/0.25D) Degree of corneal astigmatism: 0 to ±12D (Step: 0.12/0.25D) Axis angle: 1 to 180° (Step: 1°) Literature Review

0001-0002
SC 5
Pupillary Distance Measurement: 30 to 85 mm (1 mm increments)
Literature Review
0001-0002
SC 6
Pupil Size Measurement: 1.0 to 10.0 mm (0.1 mm increments)
Literature Review
0001-0002
SC 7
Observation/Display Method: minimum 5.7-inch color LCD
Literature Review
0001-0002
SC 8
Printer Thermal line printer with auto cutter
Literature Review
0001-0002
SC 9
Power Supply AC 100 to 240 V ±10% 50/60 Hz
Literature Review
0001-0002
SC 10
Minimum Dimensions: 260 (W) mm × 495 (D) mm × 457 (H) mm
Literature Review
0003
SC 11
Warranty-12 months
Literature Review
0004
SC 12
User Guides (2)
Literature Review

File details come from the government source that posted it.