36C10G18Q0159-005.pdf
PDF 89 KB Posted
- Solicitation number
- 36C10G18Q0159
About this file
36C10G18Q0159 S02 Attachment 1 - Salient Characteristics.pdf
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 36C10G18Q0159-001.pdf | ||
| 36C10G18Q0159-000.docx | DOCX document | |
| 36C10G18Q0159-003.docx | DOCX document | |
| 36C10G18Q0159-002.xlsx | XLSX spreadsheet | |
| 36C10G18Q0159-004.xls | XLS spreadsheet |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
36C10G18Q0159 | NX EQ Automated Coverslipper Machine
Salient Characteristics Attachment
Salient Characteristics (SC) Literature Review
Automated Coverslipper Machine
CLIN Brand Product Number Description
0001 Leica 14047839700 Leica CV5030 (1 Basic Instrument)
Leica CV5030 - Robotic coverslipper Basic instrument - 100-250 VAC / 50 60 Hz The basic instrument is ready for use without any further accessories!
0002 Leica 149CVTS5015 Leica CV5030-TS5015 Leica CV5030 / TS5015 Configuration 2
Leica CV5030 - Robotic Coverslipper 100-250 VAC / 50-60 Hz - Configuration connected with transfer station TS5015 for use with Leica ST5010 - Autostainer XL as automated workstation.
0003 Leica 149CVTS5025 Leica CV5030-TS5025 Leica CV5030 / TS5025 Configuration 1
Leica CV5030 - Robotic Coverslipper 100-250 VAC / 50-60 Hz - Configuration connected with transfer station TS5025 for use with Leica ST5020 - Multistainer as automated workstation.
The following SCs pertain to CLINs 0001, 0002, and 0003:
SC 1 High-throughput glass coverslipper (min 300 slides/hour)
SC 2 Compact bench top
SC3 Works with coverslips in various sizes from 22-24 mm x 40-60 mm
SC 4 100-250 VAC / 50-60 Hz
SC 5 Broken coverslips can be detected and discarded automatically
SC 6 Minimum volume of 250-ml glass mounted bottle
SC 7 Allows for adjustment of mounting media flow volume and in accordance with mounted viscosity
SC 8 Must be compatible with commercially available types of mounting media (not brand specific)
SC 9 Wet coverslipping from xylene capability
SC 10 Dispenser nozzle must be able to rest in xylene-filled receptacle
SC 11 Permanent fume extraction via activated carbon filter (integrated fume extraction system)
36C10G18Q0159 | NX EQ Automated Coverslipper Machine
SC 12 Must be compatible with commercially available slide racks (not brand specific)
SC 13 Permanent self monitoring function with acoustic and optical instrument status indications
SC 14 Error codes displayed in the event of malfunctions and other functional errors
--- End ---