This document is a government file related to a federal contract opportunity for medical imaging equipment. It contains a schedule of line items for x-ray systems, ultrasound systems, and radiation therapy systems that are being procured across multiple VA medical facilities. The file includes technical specifications, compliance matrices, and a question and answer section where the government has provided responses to vendor inquiries about the requirements. Key details include the types of equipment required (e.g. radiographic, fluoroscopic, portable x-ray, ultrasound, linear accelerator), specific technical parameters and features, quantities needed at each facility, trade-in information, and responses to vendor requests for changes to certain specifications. The federal contract opportunity is being managed by the Department of Veterans Affairs National Acquisition Center, with a solicitation number of 36A79724Q0011.
View the file
Other files
Show all 24
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
MAR 24 Schedule of Line Items
| HTME VISN | Station | EQ PO | Equipment Type | QTY |
| VISN 19: VA Rocky Mountain Network | 436 BILLINGS, MT | 436B40002 | XR US | 1 |
| VISN 19: VA Rocky Mountain Network | 436 HARRISON, MT | 436B40001 | XR US | 1 |
| VISN 21: VA Sierra Pacific Network | 459 HONOLULU, HI | 459B46013 | XR PORT C ARM | 1 |
| VISN 4: VA Stars & Stripes Healthcare Network | 460 WILMINGTON, DE | 460B48004 | XR US | 2 |
| VISN 22: VA Desert Pacific Healthcare Network | 501 ALBUQUERQUE, NM | 501B40008 | XR US | 1 |
| VISN 17: VA Heart of Texas Health Care Network | 504 AMARILLO, TX | 504B40006 | XR US | 4 |
| VISN 7: VA Southeast Health Care Network | 509 AUGUSTA, GA | 509B42004 | XR US | 1 |
| VISN 5: VA Capitol Health Care Network | 512 BALTIMORE, MD | 512B40405 | XR US | 1 |
| VISN 8: VA Sunshine Healthcare Network | 516 BAY PINES, FL | 516B40021 | XR US | 1 |
| VISN 2: VA Healthcare Network Upstate New York | 526 BRONX, NY | 526B41018 | RT LINAC | 1 |
| VISN 2: VA Healthcare Network Upstate New York | 528 ALBANY, NY | 528B43003 | XR US | 1 |
| VISN 2: VA Healthcare Network Upstate New York | 528 BATH, NY | 528B47008 | XR RF | 1 |
| VISN 2: VA Healthcare Network Upstate New York | 528 CANANDAIGUA, NY | 528B47009 | XR CT | 1 |
| VISN 2: VA Healthcare Network Upstate New York | 528 SYRACUSE, NY | 528B44245 | XR US | 3 |
| VISN 20: VA Northwest Network | 531 BOISE, ID | 531B43001 | XR PORT C ARM | 2 |
| VISN 20: VA Northwest Network | 531 BOISE, ID | 531B43002 | XR PORT RAD | 5 |
| VISN 7: VA Southeast Health Care Network | 534 CHARLESTON, SC | 534B42014 | XR US | 1 |
| VISN 10: VA Healthcare System of Ohio | 541 CLEVELAND, OH | 541B40007 | XR US | 5 |
| VISN 10: VA Healthcare System of Ohio | 541 CLEVELAND, OH | 541B40010 | XR US | 2 |
| VISN 7: VA Southeast Health Care Network | 557 DUBLIN, GA | 557B42004 | XR US | 3 |
| VISN 6: VA Mid-Atlantic Health Care Network | 558 DURHAM, NC | 558B49013 | XR US | 1 |
| VISN 6: VA Mid-Atlantic Health Care Network | 558 DURHAM, NC | 558B49015 | XR US | 1 |
| VISN 6: VA Mid-Atlantic Health Care Network | 558 DURHAM, NC | 558B49017 | XR US | 1 |
| VISN 6: VA Mid-Atlantic Health Care Network | 558 DURHAM, NC | 558B49025 | XR US | 1 |
| VISN 16: South Central VA Health Care Network | 580 HOUSTON, TX | 580B30050 | XR CT | 1 |
| VISN 5: VA Capitol Health Care Network | 581 HUNTINGTON, WV | 581B40159 | XR US | 3 |
| VISN 4: VA Stars & Stripes Healthcare Network | 595 LEBANON, PA | 595B49006 | XR US | 1 |
| VISN 23: VA Midwest Health Care Network | 618 MINNEAPOLIS, MN | 499B49003 | XR US | 1 |
| VISN 23: VA Midwest Health Care Network | 618 MINNEAPOLIS, MN | 499B49004 | XR US | 1 |
| VISN 23: VA Midwest Health Care Network | 618 MINNEAPOLIS, MN | 499B49005 | XR US | 1 |
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44001 | NM SPECT CT | 1 |
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44002 | XR RAD | 1 |
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44003 | XR RAD | 1 |
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44006 | XR PORT C ARM | 1 |
| VISN 19: VA Rocky Mountain Network | 623 MUSKOGGE, OK | 623B34014 | XR US | 1 |
| VISN 9: VA Mid South Healthcare Network | 626 NASHVILLE, TN | 626B44014 | XR US | 3 |
| VISN 9: VA Mid South Healthcare Network | 626 NASHVILLE, TN | 626B44019 | XR PORT C ARM | 2 |
| VISN 16: South Central VA Health Care Network | 629 NEW ORLEANS, LA | 629B30005 | RT LINAC | 1 |
| VISN 16: South Central VA Health Care Network | 629 NEW ORLEANS, LA | 629B30007 | RT LINAC | 1 |
| VISN 22: VA Desert Pacific Healthcare Network | 644 PHOENIX, AZ | 644B48007 | XR US | 2 |
| VISN 4: VA Stars & Stripes Healthcare Network | 646 PITTSBURGH, PA | 646B32011 | XR US | 1 |
| VISN 4: VA Stars & Stripes Healthcare Network | 646 PITTSBURGH, PA | 646B42001 | XR PORT RAD | 1 |
| VISN 4: VA Stars & Stripes Healthcare Network | 646 PITTSBURGH, PA | 646B42002 | XR US | 4 |
| VISN 4: VA Stars & Stripes Healthcare Network | 646 PITTSBURGH, PA | 646B42004 | XR US | 1 |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49005 | XR US | 8 |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49006 | XR US | 1 |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49007 | NM BONE DEN | 1 |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49008 | XR PORT C ARM | 1 |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49009 | XR PORT C ARM | 1 |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49010 | XR PORT C ARM | 3 |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49011 | XR PORT RAD | 1 |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49012 | XR US | 2 |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49013 | XR US | 2 |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49016 | XR US | 3 |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49017 | XR PORT C ARM | 4 |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49018 | XR US | 5 |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49019 | XR US | 2 |
| VISN 21: VA Sierra Pacific Network | 654 RENO, NV | 654B30018 | XR PORT RAD | 6 |
| VISN 10: VA Healthcare System of Ohio | 655 SAGINAW, MI | 655B40002 | XR U ARM | 1 |
| VISN 10: VA Healthcare System of Ohio | 655 SAGINAW, MI | 655B40009 | XR RF | 1 |
| VISN 19: VA Rocky Mountain Network | 660 SALT LAKE CITY, UT | 660B30006 | XR US | 3 |
| VISN 19: VA Rocky Mountain Network | 660 SALT LAKE CITY, UT | 660B30007 | XR US | 1 |
| VISN 20: VA Northwest Network | 663 PUGET SOUND, WA | 663B43000 | XR US | 2 |
| VISN 20: VA Northwest Network | 663 PUGET SOUND, WA | 663B43008 | XR US | 2 |
| VISN 20: VA Northwest Network | 663 PUGET SOUND, WA | 663B43010 | XR US | 1 |
| VISN 17: VA Heart of Texas Health Care Network | 671 SAN ANTONIO, TX | 671B40003 | XR RAD | 1 |
| VISN 8: VA Sunshine Healthcare Network | 673 TAMPA, FL | 673B40037 | XR US | 1 |
| VISN 17: VA Heart of Texas Health Care Network | 674 TEMPLE, TX | 674B30033 | XR US | 2 |
| VISN 8: VA Sunshine Healthcare Network | 675 ORLANDO, FL | 675B47007 | XR US | 1 |
| VISN 8: VA Sunshine Healthcare Network | 675 ORLANDO, FL | 675B47008 | RT LINAC | 1 |
| VISN 22: VA Desert Pacific Healthcare Network | 691 WEST LOS ANGELES, CA | 691B40038 | XR US | 1 |
| VISN 17: VA Heart of Texas Health Care Network | 740 HARLINGEN, TX | 740B40014 | XR US | 1 |
| VISN 17: VA Heart of Texas Health Care Network | 756 EL PASO, TX | 756B45003 | XR US | 2 |
| VISN 17: VA Heart of Texas Health Care Network | 756 EL PASO, TX | 756B45004 | XR MAMMO | 1 |
Q&A
| HTME VISN | Station | Equipment Requisition Number | Equipment Questions | GOVT Equipment Response | |
| VISN 4: VA Stars & Stripes Healthcare Network | 460 WILMINGTON, DE | 460B48004 | Request item 1.m. Diagonal (in) 21 inches be reduced to 19 inches. | ||
| Request item 8.a. and 8.b. Fusion Target Biopsy System Capability of accepting MR Images and Prostate Gland volume be moved to added value or removed as these are not typical procedures based on the description in the matrix. | Updated Tech Specs | ||||
| VISN 4: VA Stars & Stripes Healthcare Network | 460 WILMINGTON, DE | 460B48004 | Are the following advanced features required for TAP Procedures? |
- Fusion-Targeted Biopsy System
- Capable of accepting MRI (both 1.5T and 3.0T) and ultrasound images from other modalities as input and displaying these images on a screen
- Prostate gland volume and edge boundaries computation capabilities
| Deleted | ||||
| VISN 4: VA Stars & Stripes Healthcare Network | 460 WILMINGTON, DE | 460B48004 | Line 95 specifies Sonosite Institute Training. Please remove and revise as this requirement is vendor specific. | Deleted |
| VISN 4: VA Stars & Stripes Healthcare Network | 460 WILMINGTON, DE | 460B48004 | Are analysis packages 6G, H,I,J and K required for TAP procedures? | NO |
| VISN 23: VA Midwest Health Care Network | 618 MINNEAPOLIS, MN | 499B49003 | Can the requirement for "automatic initialization of probes when removed from holder" be deemed not clinically necessary and removed to increase competition? | Specification is clinical required and will remain |
| VISN 23: VA Midwest Health Care Network | 618 MINNEAPOLIS, MN | 499B49003 | Can the minimum monitor size be reduced to 21.5" to increase competition? | We will reduce the monitor size requirement to 21.5 inches. |
| VISN 22: VA Desert Pacific Healthcare Network | 501 ALBUQUERQUE, NM | 501B40008 | Request item 1.a. Minimum screen size or 21 in be reduced to 19 in. |
Request item 1.b. Minimum active size or 21 in be reduced to 19 in.
| Request item 1.d. Minimum Monitor Rotation of 90 degrees be changed to 80 degrees. | Items are clinical required and offered by two or more vendors | ||||
| VISN 17: VA Heart of Texas Health Care Network | 504 AMARILLO, TX | 504B40006 | Lines 81-91: This states that there are 4 systems requested with all of our transdcucers with a quanity of one. We need to clarify what transducers are needed for a quanity of 4 systems. | All the transducers are required for the 4 identical systems | |
| VISN 17: VA Heart of Texas Health Care Network | 504 AMARILLO, TX | 504B40006 | Request item 1.a. Minimum screen size or 21 in be reduced to 19 in. |
Request item 1.b. Minimum active size or 21 in be reduced to 19 in.
Request item 1.d. Minimum Monitor Rotation of 90 degrees be changed to 80 degrees.
Transducers: This system has a significant number of transducers - can you verify that all of the transducers are required? All specifications have been changed to the recommendations.
All transducers are required VISN 7: VA Southeast Health Care Network 509 AUGUSTA, GA 509B42004 Request item 1.a. Minimum screen size or 20 in be reduced to 19 in.
Request item 1.b. Minimum active size or 20 in be reduced to 19 in.
Request item 1.d. Minimum Monitor Rotation of 90 degrees be changed to 80 degrees.
| Request item 1.m. Maximum equipment weight 125 lbs be increased to 140 lbs. | Technical specification updated. | ||||
| VISN 5: VA Capitol Health Care Network | 512 BALTIMORE, MD | 512B40405 | Can the minimum battery life be reduced to 40 minutes to increase competition? | Technical specification updated. | |
| VISN 5: VA Capitol Health Care Network | 512 BALTIMORE, MD | 512B40405 | Can the Depth be increased to 33.8" to increase competition? | Technical specification updated. | |
| VISN 5: VA Capitol Health Care Network | 512 BALTIMORE, MD | 512B40405 | Can the max equipment weight be increased to 177 lbs to increase competition? | Technical specification updated. | |
| VISN 5: VA Capitol Health Care Network | 512 BALTIMORE, MD | 512B40405 | Request 1.a. Minimum screen size or 21 in be reduced to 19 in. |
Request 1.c. Minimum Monitor Rotation of 90 degrees be changed to 80 degrees.
| Request 1.i. Maximum equipment weight 125 lbs be increased to 140 lbs. | Technical specification updated. | ||||
| VISN 8: VA Sunshine Healthcare Network | 516 BAY PINES, FL | 516B40021 | Request 1.a. Minimum screen size or 21 in be reduced to 19 in. | ||
| Request 1.c. Minimum Monitor Rotation of 90 degrees be changed to 80 degrees. | Technical specification updated. | ||||
| VISN 2: VA Healthcare Network Upstate New York | 528 BATH, NY | 528B47008 | Please provide serial number and additional system information for trade-in. | SN#07000094 | |
| VISN 2: VA Healthcare Network Upstate New York | 528 BATH, NY | 528B47008 | Please change table height from floor to 19.7” to allow full and fair competition |
Please verify if obligation 528B47008 is for a fixed RAD Room or an RF room Please verify location of obligation 528B47008 is Bath or Buffalo NY Please change weight requirement from 6.2lbs to 7 lbs or all three detectors to allow full and fair competition (1ac, 1bb, 1cb) Please advise if you prefer a 17x17 as one of your three detectors (9lbs) 15 minutes of uptime (full functionality) would require a full UPS. Does this site expect a full UPS ($$$) for this XR system?
If obligation is in fact for a Fixed Rad room, please remove all RF specifications.
Minimum table tilt is an RF specification please remove if this is a Fixed RAD opportunity -Ability to Tilt 90 degrees in both directions/full tilt is an RF specification please remove if this is a Fixed RAD opportunity -Please remove Imaging tower longitudinal movement requirement if this is a fixed rad system -Please remove Imaging tower lateral movement requirement if this is a fixed rad system -Please remove dual focal spots if this is a fixed rad system -Please remove full articulation if this is a fixed rad system. 1.Technical specifications updated
2.Fixed Rad Room
3.Bath, NY
4.Technical specifications updated
5.Technical specifications updated
6.No full functionality UPS required. Acq station only
7.Removed all RF specs
| VISN 2: VA Healthcare Network Upstate New York | 528 BATH, NY | 528B47008 | would you please provide us with the room dimensions? | Equipment room 15ft 10 inch x 11ft 1 inch. Control room 6 ft 11 inch x 11ft 1 inch |
| VISN 2: VA Healthcare Network Upstate New York | 528 BATH, NY | 528B47008 | Is a rad-fluouro system requested or a DR? Specifications indicate that a RF room is requested, but in the description a diagnostic Radiology system is explicitly called out. | This procurement is for DR. |
| VISN 2: VA Healthcare Network Upstate New York | 528 BATH, NY | 528B47008 | 4. Table Requirements j and k: Which tabletop is requested? Freefloating or motor driven? These two requirements contradict each other. Also, if motor driven, what is the weight capacity desired? | Motor Driven / 700lbs max |
| VISN 2: VA Healthcare Network Upstate New York | 528 CANANDAIGUA, NY | 528B47009 | Please remove offsite clinical applications requirement to align with NAC instructions. |
Please remove gantry mounted injector to allow full and fair competition, or update to other types of injectors beside gantry mounted. offsite clinical applications training removed from requirement
Gantry mounted injector changed to ceiling mounted
| VISN 2: VA Healthcare Network Upstate New York | 528 CANANDAIGUA, NY | 528B47009 | Technical Requirements 1.j. What is the actual maximum system weight allowed? | 676.82 lbs | |
| VISN 2: VA Healthcare Network Upstate New York | 528 CANANDAIGUA, NY | 528B47009 | Training Requirements 1.c. Offsite training is not allowed by the NAC. Please remove or make optional. | offsite clinical applications training removed from requirement | |
| VISN 2: VA Healthcare Network Upstate New York | 528 CANANDAIGUA, NY | 528B47009 | For fair competition, can we please request a change to the “gantry mounted contrast injector” to “pedestal or ceiling mounted contrast injector”? (For internal reference, this is a lock out spec and specific to only 1 vendor) | Gantry mounted injector changed to ceiling mounted | |
| VISN 2: VA Healthcare Network Upstate New York | 528 CANANDAIGUA, NY | 528B47009 | The Specs state “UPS – maintains full functionality of the system for a minimum of minutes during a power outage”, please specify the number of minutes required for this spec. | Specification has been updated | |
| VISN 20: VA Northwest Network | 531 BOISE, ID | 531B43001 | Please provide the Biomedical Engineering Point of Contact. | Ben Sedlacek Donald.sedlacek@va.gov 208-422-1594 | |
| VISN 20: VA Northwest Network | 531 BOISE, ID | 531B43002 | 1.c Tube max kV Range | ||
| Please change to 140 kV to allow competition. | Technical specification updated. | ||||
| VISN 20: VA Northwest Network | 531 BOISE, ID | 531B43002 | 1.f Max SID of 900 inches is a typo. Please change to 80". | Technical specification updated. | |
| VISN 20: VA Northwest Network | 531 BOISE, ID | 531B43002 | 1.g Min SID, please change to 21" to allow competition | Technical specification updated. | |
| VISN 20: VA Northwest Network | 531 BOISE, ID | 531B43002 | 1.h Max tube reach, please change to 45 to allow competition | Technical specification updated. | |
| VISN 20: VA Northwest Network | 531 BOISE, ID | 531B43002 | Please review if 900in is the correct requirement. Seems like a typo | Technical specification updated. | |
| VISN 20: VA Northwest Network | 531 BOISE, ID | 531B43002 | Please review if 20in is the correct requirement. Seems like a typo | Technical specification updated. | |
| VISN 20: VA Northwest Network | 531 BOISE, ID | 531B43002 | Please alleviate maximum kVP to 150kvP to increase competition. | Technical specification updated. | |
| VISN 20: VA Northwest Network | 531 BOISE, ID | 531B43002 | Please verify if there are qty 4 or qty 5 trade ins for this opportunity (4 listed) | There are 4 devices for trade-in | |
| VISN 20: VA Northwest Network | 531 BOISE, ID | 531B43002 | 1c) Please reduce max kVp to 150 to maximize competition |
1f) Please reduce SID to 80inches; 900 inches appears to be a typo
| 1h) Please reduce maximum tube reach to 50 inches to broaded competition | Technical specification updated. | ||||
| VISN 7: VA Southeast Health Care Network | 534 CHARLESTON, SC | 534B42014 | Request 1.a. Minimum screen size or 21 in be reduced to 19 in. | ||
| Request 1.c. Minimum Monitor Rotation of 90 degrees be changed to 80 degrees. | Technical specification updated. | ||||
| VISN 10: VA Healthcare System of Ohio | 541 CLEVELAND, OH | 541B40007 | Request 1.j. Maximum equipment weight 250 ls be increased to 265 lbs. | Technical specification updated. | |
| VISN 10: VA Healthcare System of Ohio | 541 CLEVELAND, OH | 541B40010 | Can the max height be increased to 55" to increase competition? | Technical specification updated. | |
| VISN 10: VA Healthcare System of Ohio | 541 CLEVELAND, OH | 541B40010 | Can the max depth be increased to 38.5" to increase competition? | Technical specification updated. | |
| VISN 10: VA Healthcare System of Ohio | 541 CLEVELAND, OH | 541B40010 | Can the max weight be increased to 290 lbs to increase competition? | Technical specification updated. | |
| VISN 10: VA Healthcare System of Ohio | 541 CLEVELAND, OH | 541B40010 | Item 2.b. Image Fusion Software, please remove or move to added value. This option is normally associated with systems that are confgured for advanced interventional procedures and is not required for routine vascular exams. |
Item 2.g. Continuous Wave Doppler, please remove or move to added value. Continuous Wave Doppler is typically applicable to cardiac exams and is not included as part of a standard vascular package.
| Item 2.j. Tissue Doppler Imaging, please remove or move to added value. Tissue Doppler Imaging is typically applicable to cardiac exams and is not included as part of a standard vascular package. | Technical specification updated. | ||||
| VISN 10: VA Healthcare System of Ohio | 541 CLEVELAND, OH | 541B40010 | Will there be an opportunity to demo our system with the local site? | No. Patient schedule for studies does not afford the time for a demo at the local site. | |
| VISN 7: VA Southeast Health Care Network | 557 DUBLIN, GA | 557B42004 | Can the minimum active (touchscreen) screen size be decreased to 14" to increase competition? | No, there are at least two vendors who can meet this | |
| VISN 7: VA Southeast Health Care Network | 557 DUBLIN, GA | 557B42004 | Can the minimum # of transducer ports be decreased to 5 to increase competition? | No, but this was changed to 6 transducer ports on the updated Tech Specs and Compliance Matrix, as there are at least two vendors who can meet this. | |
| VISN 7: VA Southeast Health Care Network | 557 DUBLIN, GA | 557B42004 | Can the minimum battery life be reduced to 45 minutes to increase competition? | Technical specification updated. | |
| VISN 7: VA Southeast Health Care Network | 557 DUBLIN, GA | 557B42004 | Can the max width be increased to 22" to increase competition? | Technical specification updated. | |
| VISN 7: VA Southeast Health Care Network | 557 DUBLIN, GA | 557B42004 | Can the max depth be increased to 38.5" to increase competition? | Technical specification updated. | |
| VISN 7: VA Southeast Health Care Network | 557 DUBLIN, GA | 557B42004 | Request Item 1.i. Number of Transducer Ports 8 be reduced to 4 Transducer ports. | ||
| Request item 1.k. Minimum battery life 240 min be reduced to 40 min. | No, but this was changed to 6 transducer ports on the updated Tech Specs and Compliance Matrix, as there are at least two vendors who can meet this. |
Yes, this was changed to 40 minutes on the updated Tech Specs and Compliance Matrix.
| VISN 6: VA Mid-Atlantic Health Care Network | 558 DURHAM, NC | 558B49017 | Can the minimum # of gel warmer holders be decreased to 1 to increase competition? | Technical specification updated. | |
| VISN 6: VA Mid-Atlantic Health Care Network | 558 DURHAM, NC | 558B49017 | Request 1.h. Maximum equipment weight 250 ls be increased to 282 lbs. | ||
| Request item 1.i. Minimum Gel Warmers 2 be reduced to 1. | Technical specification updated. | ||||
| VISN 6: VA Mid-Atlantic Health Care Network | 558 DURHAM, NC | 558B49025 | Can the maximum weight be increased to 125 lbs to increase competition? | Technical specification updated. | |
| VISN 6: VA Mid-Atlantic Health Care Network | 558 DURHAM, NC | 558B49025 | Can the minimum # of gel warmer holders be decreased to 1 to increase competition? | Technical specification updated. | |
| VISN 4: VA Stars & Stripes Healthcare Network | 595 LEBANON, PA | 595B49006 | Request 1.a. Minimum screen size or 20 in be reduced to 19 in. | ||
| Request 1.e. Maximum equipment weight of 100lbs be revised to 140lbs.. | Minimum screen size reduced to 18 in. Maximum weight increased to 150 lbs. | ||||
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44001 | Line 22 – Please change to 50cm to allow any vendor to participate as all current SPECT/CT offerings have a 50cm FOV with an optional 70cm FOV. |
Line 44 – Please remove to allow for full and fair competition as only one vendor offers this.
Line 70 – Which 3rd party cardiac package is preferred (Cedars, ECTb, or 4DM)?
Line 91 - Please remove to allow for full and fair competition as only one vendor offers this. Line 22 - Updated tech specs Line 44 - Moved to Added Value section Line 70 - Added preference in tech specs Line 91 - Moved to Added Value Section
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44002 | Please modify from 100 µm pixel pitch detector to allow 140 µm pixel pitch. 100 micron pixel pitch detector aggregates 4 pixels for chest images in background and it is equivalent as 200 µm pixel pitch for chest images that mostly common exams in DR(Binning). It means 100 micron pixel pitch resolution has no clear clinical benefits for chest X-Ray exam. | 2 or more vendors are able to meet this requirement. No change. | |
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44002 | Please reconsider whether or not 1000mA is really required at 100kVp. It means 100kW system is required for this. Please change to 100kVp = 800 mA to allow for full and fair competition. | 2 or more vendors are able to meet this requirement. No change. | |
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44002 | Please modify from 100 µm pixel pitch detector to allow 140 µm pixel pitch. 100 micron pixel pitch detector aggregates 4 pixels for chest images in background and it is equivalent as 200 µm pixel pitch for chest images that mostly common exams in DR(Binning). It means 100 micron pixel pitch resolution has no clear clinical benefits for chest X-Ray exam. | ||
| (3.4lp/mm is reasonable as other requirements) | 2 or more vendors are able to meet this requirement. No change. | ||||
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44002 | Please modify from 100 µm pixel pitch detector to allow 140 µm pixel pitch. 100 micron pixel pitch detector aggregates 4 pixels for chest images in background and it is equivalent as 200 µm pixel pitch for chest images that mostly common exams in DR(Binning). It means 100 micron pixel pitch resolution has not clinically beneficial for chest X-Ray exam. | 2 or more vendors are able to meet this requirement. No change. | |
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44002 | Please change from 800 lbs to 770 lbs to ensure maximum competition. | Changed to 750 lbs static and 700 lbs dynamic | |
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44002 | Motor-driven X-Y table top movement is not available for all DR room system. It is a function for fluoroscopy. Please remove to ensure maximum competition | Requirement removed. | |
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44002 | Please modify to equivalent function of bone suppression(to see soft tissue as most common clinical requirement) in SW to ensure maximum competition. | 2 or more vendors are able to meet this requirement. No change. | |
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44002 | Please remove 'Tomography' to ensure maximum competition. It is not common practice in general DR due to its processing speed and image quality. | Requirement removed. | |
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44002 | For the request of the wireless detector in the wall bucky, can this be 17x17 fixed? |
For the wireless detector in the table, can the spec be changed to either wireless 14x17 or fixed 17x17?
Can you clarify the request is for 3 detectors, 1 in wallstand, 1 in table and 1 free?
If there is a 3rd detector, can the spec be changed to 14x17 wireless?
Can the minimum table weight for dynamic be switched to 660 lbs?
Can the lowest table height from floor be increased to 20.5 inches?
Can you clarify the maximum system weight?
Is digital tomosynthesis required for clinical use or is it an optional feature you would like to evaluate?
Is the clinical need for dual energy for chest imaging? If so, can the specification be Dual energy imaging or equivalent function of bone suppression in software in place of dual energy?
Can you clarify what AI Technology you are requesting for common department errors and inefficiencies with a Live Streaming Camera?
- Facility requests that this detector to be wireless to move the detector from the bucky for exams or move between rooms.
- This detector must be 17x17 wireless for patient and staff safety to allow for flexibilty within the exam.
- Requesting 2 detectors: 1 for wall bucky and 1 for table.
- No 3rd detector.
- Updated to 700 lbs dynamic which is able to be met by 2 or more vendors.
- Updated to 21"
- Requirement removed.
- Requirement removed.
- Dual energy is used for chest imaging. See requirement 5.b in the tech specs for equivalent.
- Application would be for camera to monitor patient and ensure that they stay in the position needed for the imaging once the tech walks away to the control station. Requirement moved to Added Value section.
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44002 | Specific to table and wall detector requirements, would an alternative be a fixed 17x17 detector? If not, would a smaller detector suffice? | No, that is not an acceptable alternative. Detectors must be 17x17 wireless. | |
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44002 | For the free floating detector requested, would an alternative response be a smaller detector, such as a 14x17? | 3rd detector not requested - only bucky and table. | |
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44002 | Are wireless detector sharing licenses to be included in these solicitations? | Yes | |
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44003 | Please modify from 100 µm pixel pitch detector to allow 140 µm pixel pitch. 100 micron pixel pitch detector aggregates 4 pixels for chest images in background and it is equivalent as 200 µm pixel pitch for chest images that mostly common exams in DR(Binning). It means 100 micron pixel pitch resolution has no clear clinical benefits for chest X-Ray exam. | 2 or more vendors are able to meet this requirement. No change. | |
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44003 | Please reconsider whether or not 1000mA is really required at 100kVp. It means 100kW system is required for this. Please change to 100kVp = 800 mA to allow for full and fair competition. | 2 or more vendors are able to meet this requirement. No change. | |
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44003 | Please modify from 100 µm pixel pitch detector to allow 140 µm pixel pitch. 100 micron pixel pitch detector aggregates 4 pixels for chest images in background and it is equivalent as 200 µm pixel pitch for chest images that mostly common exams in DR(Binning). It means 100 micron pixel pitch resolution has no clear clinical benefits for chest X-Ray exam. | ||
| (3.4lp/mm is reasonable as other requirements) | 2 or more vendors are able to meet this requirement. No change. | ||||
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44003 | Please modify from 100 µm pixel pitch detector to allow 140 µm pixel pitch. 100 micron pixel pitch detector aggregates 4 pixels for chest images in background and it is equivalent as 200 µm pixel pitch for chest images that mostly common exams in DR(Binning). It means 100 micron pixel pitch resolution has not clinically beneficial for chest X-Ray exam. | 2 or more vendors are able to meet this requirement. No change. | |
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44003 | Please change from 800 lbs to 770 lbs to ensure maximum competition. | Changed to 750 lbs static and 700 lbs dynamic | |
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44003 | Motor-driven X-Y table top movement is not available for all DR room system. It is a function for fluoroscopy. Please remove to ensure maximum competition | Requirement removed. | |
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44003 | Please modify to equivalent function of bone suppression(to see soft tissue as most common clinical requirement) in SW to ensure maximum competition. | 2 or more vendors are able to meet this requirement. No change. | |
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44003 | Please remove 'Tomography' to ensure maximum competition. It is not common practice in general DR due to its processing speed and image quality. | Requirement removed. | |
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44003 | Specific to table and wall detector requirements, would an alternative be a fixed 17x17 detector? If not, would a smaller detector suffice? | No, that is not an acceptable alternative. Detectors must be 17x17 wireless. | |
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44003 | For the free floating detector requested, would an alternative response be a smaller detector, such as a 14x17? | 3rd detector not requested - only bucky and table. | |
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44003 | Are wireless detector sharing licenses to be included in these solicitations? | Yes | |
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44003 | For the request of the wireless detector in the wall bucky, can this be 17x17 fixed? |
For the wireless detector in the table, can the spec be changed to either wireless 14x17 or fixed 17x17?
Can you clarify the request is for 3 detectors, 1 in wallstand, 1 in table and 1 free?
If there is a 3rd detector, can the spec be changed to 14x17 wireless?
Can the minimum table weight for dynamic be switched to 660 lbs?
Can the lowest table height from floor be increased to 20.5 inches?
Can you clarify the maximum system weight?
Is digital tomosynthesis required for clinical use or is it an optional feature you would like to evaluate?
Is the clinical need for dual energy for chest imaging? If so, can the specification be Dual energy imaging or equivalent function of bone suppression in software in place of dual energy?
Can you clarify what AI Technology you are requesting for common department errors and inefficiencies with a Live Streaming Camera?
- Facility requests that this detector to be wireless to move the detector from the bucky for exams or move between rooms.
- This detector must be 17x17 wireless for patient and staff safety to allow for flexibilty within the exam.
- Requesting 2 detectors: 1 for wall bucky and 1 for table.
- No 3rd detector.
- Updated to 700 lbs dynamic which is able to be met by 2 or more vendors.
- Updated to 21"
- Requirement removed.
- Requirement removed.
- Dual energy is used for chest imaging. See requirement 5.b in the tech specs for equivalent.
- Application would be for camera to monitor patient and ensure that they stay in the position needed for the imaging once the tech walks away to the control station. Requirement moved to Added Value section.
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44006 | The Matrix details that the system will be used in the pain management department. What specific case work and exams will be performed? Will motorized, fluoroscopic sweep images be needed for any of those types? | Floroscopic sweep images will not be needed. Exams to be performed are spine injections. |
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44006 | Is there a specific need for CMOS detector? | CMOS detector not required - Requirement moved to Added Value section |
| VISN 9: VA Mid South Healthcare Network | 621 MOUNTAIN HOME, TN | 621B44006 | Is there a specific detector size required for this care area? Matrix size is included but not specific detector size | Requirement added for Minimum Detector Size of 30x30 cm. |
| VISN 19: VA Rocky Mountain Network | 623 MUSKOGGE, OK | 623B34014 | Can the Depth be increased to 33.8" to increase competition? | Technical specification updated. |
| VISN 19: VA Rocky Mountain Network | 623 MUSKOGGE, OK | 623B34014 | Is the Ophthalmic analysis package clinically necessary? | Technical specification updated. |
| VISN 9: VA Mid South Healthcare Network | 626 NASHVILLE, TN | 626B44014 | Can the minimum active (touchscreen) screen size be decreased to 14" to increase competition? | Yes |
| VISN 9: VA Mid South Healthcare Network | 626 NASHVILLE, TN | 626B44014 | Can the minimum battery life be reduced to 45 minutes to increase competition? | Yes |
| VISN 9: VA Mid South Healthcare Network | 626 NASHVILLE, TN | 626B44014 | Can the max weight be increased to 290 lbs to increase competition? | Yes |
| VISN 9: VA Mid South Healthcare Network | 626 NASHVILLE, TN | 626B44014 | Request item 1.h. Control panel height range from floor 55-70 be revised to 30-36 inches. |
Request item 1.m. Maximum equipment weight of 250 lbs be revised to 255 lbs.
Request item Training Requirement 1.f. remove Travel cost requirement. Yes Yes - increase to 290 lbs. above Yes
| VISN 9: VA Mid South Healthcare Network | 626 NASHVILLE, TN | 626B44014 | After meeting with the local site last month, they determined they would like the mL26-8 transducer included with each system. Are we able to add those to the quote? (Philips high frequency micro linear transducer) | Vendors may add any extra features to their quote that they believe will meet the needs of the facility. |
| VISN 9: VA Mid South Healthcare Network | 626 NASHVILLE, TN | 626B44014 | After meeting with the local site last month, they determined they would like xMATRIX, xPlane, and Live 3D GI software included with each system. Are we able to add those to the quote? | Vendors may add any extra features to their quote that they believe will meet the needs of the facility. |
| VISN 9: VA Mid South Healthcare Network | 626 NASHVILLE, TN | 626B44014 | After meeting with the local site last month, they determined they would like the XL14-3 transducer included with each system. Are we able to add those to the quote? (Philips Linear array transducer that supports 3D/4D and vessel cast imaging; vascular, thyroid, msk applications) | Vendors may add any extra features to their quote that they believe will meet the needs of the facility. |
| VISN 16: South Central VA Health Care Network | 629 NEW ORLEANS, LA | 629B30005 | • | Line 40 3. Image Guidance and Portal dosimetry c. Intrafraction Triggered Planar (MV) |
Question: Can you please provide more clarification on how intrafraction triggered planar (MC) is used? Is this for clinical use or QA?
• Line 45 4. Additional specifications: a. ROEMS interface with CPRS and/or CERNER HER Question: Can you please clarify if this is meant to be ROEMR Radiation Oncology Electronic Medical Record?
• Line Item 96 ☒ Ability for Patient Positioning Devices to provide backup connection to 3rd party system (purchased separately) and allow output of patient treatment images & correction vectors and System configuration, DICOM import/export settings Question: Is this a requirement for immobilization device compatibility or TPS/OIS DICOM import/export features? What do you mean by ‘back up to a 3rd party sys?
| • | Line 92 ☒ Ability for Contouring/Registration System to provide backup connection to 3rd party system (purchased separately) and allow output of Co-registered imaging studies DICOM / DICOM-RT and System configuration, DICOM import configurations, 3rd party scripts and workflows |
| Question: Are you referring to 3rd party sys like MiM or Velosity? Could you please clarify what you mean by ‘backing up to a third-party system’? | Line 40 3: It is triggered planar imaging for intrafraction motion management, often useful in SBRT treatments. |
Line 45 4: ROEMR is the Radiation Oncology Electronic Medical Record system (such as Varian Aria or Elekta Mosaiq).
Line item 92: Patient positioning devices from, for example, VisionRT generate patient treatment images and correction vector data elements. We are requiring these patient specific images and correction vectors including system configurations datasets to be backed up in 3rd party data storage systems in DICOM-DICOM-RT formats.
Line 93: The 3rd party systems can be for example MIM or Velocity like true DICOM-RT archive systems or Commvault like drive backup systems which can store non-DICOM-RT datasets too.
| VISN 16: South Central VA Health Care Network | 629 NEW ORLEANS, LA | 629B30005 | Is online adaptive a requirement? | It is preferred. However, the vendor needs to explain their vision for adaptive workflow. |
| VISN 16: South Central VA Health Care Network | 629 NEW ORLEANS, LA | 629B30007 | • | Line 52. 4. Additional specifications: a. ROEMS interface |
Question: Can you please clarify if this was meant to be ROEMR Radiation Oncology Electronic Medical Record?
• Line 105 Added Value ☒ n. Ability for Patient Positioning Devices to provide backup connection to 3rd party system (purchased separately) and allow output of patient treatment images & correction vectors and System configuration, DICOM import/export settings Question: Is this a requirement for immobilization device compatibility or TPS/OIS DICOM import/export features? Could you please clarify what you mean by ‘backing up to a third-party system’?
Line 52 4: Yes, ROEMR is the Radiation Oncology Electronic Medical Record system (such as Varian Aria or Elekta Mosaiq).
Line 105: Backing up to a third party system means disk/drive backup which creates a redudent copy of the data and can be used to restore the data incase of a disk corruption or other like events. Commvault is one 3rd party system used for these kind of backups.
| VISN 22: VA Desert Pacific Healthcare Network | 644 PHOENIX, AZ | 644B48007 | Request item 1.e. Minimum Battery life of 180 min be revised to 120 min. | Technical specification updated. | |
| VISN 4: VA Stars & Stripes Healthcare Network | 646 PITTSBURGH, PA | 646B32011 | Can the minimum # of transducer holders be reduced to 4 to increase competition? | Yes. This has been amended | |
| VISN 4: VA Stars & Stripes Healthcare Network | 646 PITTSBURGH, PA | 646B32011 | Is the 3D/4D doppler display for echo/cardiac clinically necessary? | Yes this is needed to provide 3D access to imaging without the ned to purchase further software or hardware. | |
| VISN 4: VA Stars & Stripes Healthcare Network | 646 PITTSBURGH, PA | 646B32011 | Is the Fusion-Targeted Biopsy System clinically necessary? | Yes. This is included in transducers. | |
| VISN 4: VA Stars & Stripes Healthcare Network | 646 PITTSBURGH, PA | 646B32011 | Request item 1.j. Maximum equipment weight of 245 lbs be revised to 255 lbs. | ||
| This has been changed to 255 lbs | |||||
| VISN 4: VA Stars & Stripes Healthcare Network | 646 PITTSBURGH, PA | 646B42001 | Please review if 200in is the correct requirement. Seems like a typo | This has been changed to 78 inches | |
| VISN 4: VA Stars & Stripes Healthcare Network | 646 PITTSBURGH, PA | 646B42001 | Please alleviate the reach distance to 800mm to increase competition. | Yes. This has been amended | |
| VISN 4: VA Stars & Stripes Healthcare Network | 646 PITTSBURGH, PA | 646B42001 | Please modify from 100 µm pixel pitch detector to allow 140 µm pixel pitch. 100 micron pixel pitch detector aggregates 4 pixels for chest images in background and it is equivalent as 200 µm pixel pitch for chest images that mostly common exams in DR(Binning). It means 100 micron pixel pitch resolution has no clear clinical benefits for chest X-Ray exam. | ||
| (3.4lp/mm is reasonable as other requirements) | changed to 140 micron pixel. | ||||
| VISN 4: VA Stars & Stripes Healthcare Network | 646 PITTSBURGH, PA | 646B42001 | Please modify from 100 µm pixel pitch detector to allow 140 µm pixel pitch. 100 micron pixel pitch detector aggregates 4 pixels for chest images in background and it is equivalent as 200 µm pixel pitch for chest images that mostly common exams in DR(Binning). It means 100 micron pixel pitch resolution has not clinically beneficial for chest X-Ray exam. | changed to 140 micron pixel. | |
| VISN 4: VA Stars & Stripes Healthcare Network | 646 PITTSBURGH, PA | 646B42001 | Please review if 70 lp/mm is the correct requirement. Seems like a typo | Eliminated requirement. | |
| VISN 4: VA Stars & Stripes Healthcare Network | 646 PITTSBURGH, PA | 646B42001 | 2e. Please change minimum pixel size to 100 to align with 100 requirement listed in 1o |
Please verify if you are looking for qty 2 14x17 detectors or one 14x17 and one 10x12 to ensure detectors fit in bins.
| Please remove ability to apply multiple image processing algorithms both pre and post-acquisition to allow for soft tissue and or bone enhancement to allow full and fair competition. | 2e. Changed to 100 | |||
| VISN 4: VA Stars & Stripes Healthcare Network | 646 PITTSBURGH, PA | 646B42001 | 1b) Please adjust generator range from 0 to 100 mAs |
1d) Please reduce max kVp to 150 to broaden competition 1i) Please reduce max SID to 80 inches as 200inches appears to be a typo 1m) Please confirm a wireless detector is required for this portable solution 1n) Please confirm a wireless detector and reduce lp/mm to 3.6 lp/mm to broaden competition 1o) Please increase pixel size to 140 micron to broaden competition 1w) Please increase maximum system weight to 1250 to broaden competition
| All changed to recommendation. | |||||
| VISN 4: VA Stars & Stripes Healthcare Network | 646 PITTSBURGH, PA | 646B42001 | V. Positioning handles on collimator,not on tube head: please either clarify the below question or eliminate this from a spec | Eliminated | |
| VISN 4: VA Stars & Stripes Healthcare Network | 646 PITTSBURGH, PA | 646B42002 | Request item 4.c. Retractable Keyboard allow for Keyboard on Control panel. | ||
| Request item 8.a. MRI Fusion and b. Prostate Gland volumes be removed as they are not typically used in the Cardiology Department. | 4.C Will remia the same to provide ergonimical convenience for our techniticians. 8.A. Removed and B. Removed | ||||
| VISN 4: VA Stars & Stripes Healthcare Network | 646 PITTSBURGH, PA | 646B42002 | Under Section 8. Advanced Features, Is "Prostate gland volume and edge boundaries computation capabilities" needed? | Removed. | |
| VISN 4: VA Stars & Stripes Healthcare Network | 646 PITTSBURGH, PA | 646B42004 | Request 1.j. Maximum equipment weight 130 lbs be increased to 140 lbs. |
Request item 6.k. Mitral Value Navigator be removed or moved to Added Value as this is not typically used by this department.
| Request item 8.a. MRI Fusion and b. Prostate Gland volumes be removed as they are not typically used in this Department. | 1.J. increased. 6. K. moved to added value. 8.A. moved to added value. | ||||
| VISN 4: VA Stars & Stripes Healthcare Network | 646 PITTSBURGH, PA | 646B42004 | Line 13: This states ICU Sonosite LX has reached standard end of life and will need to be replaced. The LX is not end of life. This should state X-Porte as it is end of life. Needs clarification. | Changed to X-Porte | |
| VISN 4: VA Stars & Stripes Healthcare Network | 646 PITTSBURGH, PA | 646B42004 | Lines 74 & 76: This states C5-1 Curved transducer. This is listed twice. This should be a Phased transducer. Needs clarification. | Changed requirement. | |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49005 | Can the minimum battery life be reduced to 45 minutes to increase competition? | Yes, these will be stationary and plugged in to the wall. | |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49005 | Can the max weight be increased to 290 lbs to increase competition? | Technical specification updated. | |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49005 | Request 1.h. Maximum equipment weight 250 ls be increased to 282 lbs. | Technical specification updated. | |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49005 | Is this for shared service or to include TCI (Transcranial imaging)? | No | |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49005 | Is a battery required for quick boot-up or is standby and hibernate also an option? Standby and hibernate do not require a battery. | Standby and hibernate do not require a battery, but also renders the unit inoperable until plugged back in. | |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49005 | Can the Max weight be adjusted from 250 lbs. to 275lbs | Technical specification updated. | |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49005 | Is Freehand 3D a requirement? If so, for what application? | Yes, this allows whole organ imaging and the internal organs to be viewed in 3 planes, not just as a 2D image. | |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49005 | Is Navigation and Fusion a requirement or do you just need to know if it is compatible? If it is a requirement, do you need it on all 8 systems? Historically it is only purchased for 1 system if at all. | Navigation and Fusion is used for prostate biopsies. It would be needed on minimum 1 unit | |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49005 | Is DTI required? This is a Cardiac feature. | No - The acronym used, DTI, is believed to actually be TDI, tissue doppler imaging. | |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49005 | Please specify if you want shear wave elastography for Liver applications, small parts (Linear) applications or both. | Yes, on all units | |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49005 | Is ECG required for radiology applications? | Not for Ultrasound, only for Cardiac which we do not do in Radiology | |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49005 | 7. Transducers |
- D. Sector Array
- E. PW/CW Doppler
These are used in Cardiac Scanning and not generally in a General Imaging/Radiology environment. Will these machines be used in “Shared Service” capacity between Radiology Ultrasound and Cardiology?
Or will the Sector Array be used just for Transcranial Doppler?
| No, these will not be shared with Cardiology and we will not perform transcranial doppler. | ||||
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49006 | Can the minimum battery life be reduced to 40 mins to increase competition? | Technical specification updated. |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49006 | Can the max depth be increased to 38" to increase competition? | Technical specification updated. |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49007 | 1h: Please Change spec to 24" |
1l: Length dimension appears to be a typo - Change spec to 113" (requested dimension wouldn't fit a compact table, please confirm) 1m: Change height dimension to 53" Minimum distance between table and c-arm is 24 in.
The maximum system dimensions (HxWxL) are 56 in. x 52 in. x 113 in.
The maximum operator console dimensions (HxWxD) are 53 in. x 26.5 in. x 24 in.
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49008 | Please provide the Biomedical Engineering Point of Contact. | Joshua Brady |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49008 | Is a large screen required for procedures? | Yes |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49011 | Please clarify "trickle charge method". | Yes, we have to have this. It keeps the batteries charged for the whole unit and increases battery life |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49011 | Why do the positioning handles need to be on the collimator vs the tube head? | Handles will need to be on the tube head vs the collimator unless available on both |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49011 | 1d) Please reduce max mAs to 320 to broaden competition |
1t) Please increase system weight to 1300lbs to broaden competition 1t) Please reduce dimension to 55" High x 23" Wide x 47" Long to broaden competition 2b) Please reduce to industry standard 10"x12" detector; 11"x13" seems to be proprietary or for a grid cover for a 10"x12" detector The generator range of 01. - 500 mAs is needed.
The maximum system weight is 1300 lbs.
The maximum system dimensions in parked position, HxWxD is 55 in. x 23 in. x 47 in.
A 14 in. x 17 in. detector is bare minimum required, a 10 in. x 12 in. is an added bonus, but body imaging can not be completed on a 10 in. x 12 in. detector.
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49012 | Can the max depth be increased to 38" to increase competition? | Technical specification updated. |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49012 | Is 3D needed for vascular access? | Yes, it is needed for vascular lab. |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49012 | Pulse Inversion Harmonics – Is harmonics sufficient or does it have to be PIH? | PIH for vascular lab |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49012 | Is TDI needed for vascular access? – this is a cardiology feature. | This is needed for vascular lab. |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49012 | Is Shear wave needed for vascular access? |
If so, please specify application for shear wave i.e. Liver/small parts or both
| Shear wave is needed for vascular lab. | ||||
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49012 | Is ECG needed for vascular access? | Yes, it is needed for vascular lab. |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49012 | Questions regarding 652B49012: |
| A. | Equipment Overview |
| - | The Central Virginia Healthcare System is requesting two (2) ultrasounds for the vascular access team at the Fredericksburg Healthcare Center |
6. Analysis packages
7. Transducers
The overview states these machines are to be used in a Vascular Access capacity, but the analysis package and transducers suggest a machine that is used in a Vascular Lab.
| These are for vascular lab. | |||||
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49013 | Can you verify item 1.e. Minimum titling of 720 degrees is correct? | The minimum rotation is 180 degrees. | |
| VISN 6: VA Mid-Atlantic Health Care Network | 652 RICHMOND, VA | 652B49013 | Please clarify monitor rotation and tilt of 720 degrees | The minimum tilt is 100 degrees. |