0001_QA_Schedule of Line Items_Revised.xlsx

XLSX spreadsheet 23 KB Posted

Attached to
6525--FY22 Women's Health Consolidation
Solicitation number
36A79722Q0003
Issued by

View the file

Other files

Show all 20

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Schedule of Line Items

HTME VISNVISN #StationRequisition Number in ActionEquipment TypeQuantityBenchmark (BM) or Technical Specification (TS)
VISN 1: VA New England Healthcare System1405 WHITE RIVER, VT405B18012NM BONE DEN1TS
VISN 22: VA Desert Pacific Healthcare Network22501 ALBUQUERQUE, NM501B10052XR MAMMO1BM
VISN 7: VA Southeast Health Care Network7509 AUGUSTA, GA509B12014XR MAMMO1BM
VISN 7: VA Southeast Health Care Network7509 AUGUSTA, GA509B12016NM BONE DEN1BM
VISN 5: VA Capitol Health Care Network5512 BALTIMORE, MD512B11009XR MAMMO1TS
VISN 8: VA Sunshine Healthcare Network8516 ST PETERSBURG, F516B10016XR MAMMO1TS
VISN 8: VA Sunshine Healthcare Network8516 ST PETERSBURG, F516B10020XR MAMMO1TS
VISN 5: VA Capitol Health Care Network5517 BECKLEY, WV517B11001NM BONE DEN1TS
VISN 7: VA Southeast Health Care Network7521 BIRMINGHAM, AL521B12017XR MAMMO1BM
VISN 1: VA New England Healthcare System1523 BOSTON, MA523B11035NM BONE DEN1TS
VISN 2: VA Healthcare Network Upstate New York2526 BRONX, NY526B10007XR MAMMO1BM
VISN 2: VA Healthcare Network Upstate New York2528 BUFFALO, NY528B10171XR MAMMO1TS
VISN 2: VA Healthcare Network Upstate New York2528 ALBANY, NY528B13039XR MAMMO1BM
VISN 12: VA Great Lakes Health Care System12537 CHICAGO, IL537B18013NM BONE DEN1TS
VISN 5: VA Capitol Health Care Network5540 CLARKSBURG, WV540B11003NM BONE DEN1TS
VISN 7: VA Southeast Health Care Network7544 COLUMBIA, SC544B12068NM BONE DEN1BM
VISN 7: VA Southeast Health Care Network7544 COLUMBIA, SC544B12069NM BONE DEN1BM
VISN 8: VA Sunshine Healthcare Network8546 MIAMI, FL546B10041XR MAMMO1TS
VISN 8: VA Sunshine Healthcare Network8546 MIAMI, FL546B10042XR MAMMO1TS
VISN 8: VA Sunshine Healthcare Network8548 WPB, FL548B05040XR MAMMO1TS
VISN 8: VA Sunshine Healthcare Network8548 WPB, FL548B15019NM BONE DEN1TS
VISN 17: VA Heart of Texas Health Care Network17549 DALLAS, TX549B10594NM BONE DEN1TS
VISN 6: VA Mid-Atlantic Health Care Network6558 DURHMAN, NC558B10361NM BONE DEN1TS
VISN 6: VA Mid-Atlantic Health Care Network6558 DURHAM, NC558B10362XR MOBILE MAMMO1TS
VISN 6: VA Mid-Atlantic Health Care Network6565 FAYETTEVILLE, NC565B10079XR MAMMO1TS
VISN 6: VA Mid-Atlantic Health Care Network6565 FAYETTEVILLE, NC565B10080NM BONE DEN2TS
VISN 8: VA Sunshine Healthcare Network8573 JACKSONVILLE, FL573B15068XR MAMMO1TS
VISN 8: VA Sunshine Healthcare Network8573 LAKE CITY, FL573B15075NM BONE DEN1TS
VISN 19: VA Rocky Mountain Network19575 GR JUNCTION, CO575B16024NM BONE DEN1TS
VISN 5: VA Capitol Health Care Network5581 HUNTINGTON, WV581B11246NM BONE DEN1TS
VISN 15: VA Heartland Network15589 TOPEKA, KS589B12024NM BONE DEN1TS
VISN 15: VA Heartland Network15589 COLUMBIA, MO589B13015XR MAMMO1TS
VISN 15: VA Heartland Network15589 COLUMBIA, MO589B13016NM BONE DEN1TS
VISN 21: VA Sierra Pacific Network21593 LAS VEGAS, NV593B10063XR MAMMO2TS
VISN 4: VA Stars & Stripes Healthcare Network4595 LEBANON, PA595B19026NM BONE DEN1TS
VISN 22: VA Desert Pacific Healthcare Network22605 LOMA LINDA, CA605B10023XR MAMMO1BM
VISN 22: VA Desert Pacific Healthcare Network22605 LOMA LINDA, CA605B10024XR MAMMO1BM
VISN 21: VA Sierra Pacific Network21612 MATHER, CA612B18012XR MAMMO1TS
VISN 21: VA Sierra Pacific Network21612 MATHER, CA612B18013XR MAMMO1TS
VISN 22: VA Desert Pacific Healthcare Network22644 PHOENIX, AZ644B18154XR MAMMO1BM
VISN 22: VA Desert Pacific Healthcare Network22644 PHOENIX, AZ644B18165NM BONE DEN1TS
VISN 4: VA Stars & Stripes Healthcare Network4646 PITTSBURGH, PA646B12013NM BONE DEN1TS
VISN 1: VA New England Healthcare System1650 PROVIDENCE, RI650B11028NM BONE DEN1TS
VISN 15: VA Heartland Network15657 ST LOUIS, MO657B10039NM BONE DEN1TS
VISN 15: VA Heartland Network15657 MARION, IL657B12014NM BONE DEN1TS
VISN 15: VA Heartland Network15657 POPLAR BLUFF, MO657B13018NM BONE DEN1TS
VISN 6: VA Mid-Atlantic Health Care Network6659 SALISBURY, NC659B10008XR MAMMO1TS
VISN 19: VA Rocky Mountain Network19660 SALT LAKE CY, UT660B10016NM BONE DEN1TS
VISN 8: VA Sunshine Healthcare Network8673 TAMPA, FL673B12045NM BONE DEN1BM
VISN 8: VA Sunshine Healthcare Network8673 TAMPA, FL673B12046XR MAMMO1BM
VISN 8: VA Sunshine Healthcare Network8673 TAMPA, FL673B12047NM BONE DEN1TS
VISN 17: VA Heart of Texas Health Care Network17674 TEMPLE, TX674B10156XR MAMMO1TS
VISN 17: VA Heart of Texas Health Care Network17674 TEMPLE, TX674B10158XR MAMMO1TS
VISN 5: VA Capitol Health Care Network5688 WASHINGTON, DC688B10019XR MAMMO1TS
VISN 1: VA New England Healthcare System1689 WEST HAVEN, CT689B11009NM BONE DEN1TS
VISN 1: VA New England Healthcare System1689 WEST HAVEN, CT689B11011NM BONE DEN1TS
VISN 10: VA Healthcare System of Ohio10757 COLUMBUS, OH757B19021NM BONE DEN1TS

Vendor Questions_Answers

#STATIONPO#VENDOR QUESTIONGOVT RESPONSE
W01405 WHITE RIVER, VT405B180121c.Please change to 62 seconds for 15.8x12cm scan window using standard mode to allow fair competition.

1d. Please change to 24 seconds for 14.2 x 10cm window 1f. Please change to 350lbs maximum 1H. Please change minimum to 17.9" 1c. Change accepted - specification has been updated 1d. Change accepted - specification has been updated 1f. Adjusted to 400 lbs to allow for additional competition - Clinically required for patient population 1h. adjusted to 20 in to allow for additional competition - clinically required for patient population W02 512 BALTIMORE, MD 512B11009 What is your current vacuum assisted biopsy device used in breast imaging?

Do you need a biopsy chair quoted What is your current PACS and version number?

What Change Healthcare software version are you on and do you have the mammo module? The Baltimore VA Medical Center does not currently have a vacuum assisted biopsy device used in breast imaging A biopsy chair has been added to the Added Value section rather than a requirement because not all biopsy systems require a biopsy chair. A biopsy chair should be quoted if it is needed in the quoted system’s workflow The Medical Center is using Change Healthcare Radiology Solution v12.4 with the mammography module

W03516 ST PETERSBURG, F516B100162.e. Please change to 19 inches to allow full and fair competition
Please provide Software version of Phillips Intelespace?2e. Updated tech specs

Philips Intellispace 4.4.544.00

W04516 ST PETERSBURG, F516B100161.Should a needle localization kit be included?
2.Is a breast biopsy vacuum system required?
3.Pleae clarify if a magnification stand and paddles are required?1. Yes

2. Yes

3. Yes W05 516 ST PETERSBURG, F 516B10020 2. e. Please change to 19 inches to allow full and fair competition What is your current vacuum assisted biopsy device used in breast imaging?

Do you require 2D and 3D CAD?

What is your Software version of Phillips Intelespace? 2e. Updated tech specs No vacuum assisted biopsy device at this location.

Yes, 2D and 3D CAD is required Philips Intellispace 4.4.544.00

W06516 ST PETERSBURG, FL516B100201.Should a needle localization kit be included?
2.Is a breast biopsy vacuum system required?
3.Pleae clarify if a magnification stand and paddles are required?1. No

2. No

3. No

W07517 BECKLEY, WV517B1100123f. Please change to 350lb maximum table load.
27j. Please adjust heigh to 113 x 52 x 49 inches.23f. specification has been adjusted

27f. specification has been adjusted for height, but length remains as specified.

W08 523 BOSTON, MA 523B11035 1a. Please raise minimum to 339 seconds.

3c. Please raise minimum to 62 seconds.

1e. Please consider changing specification to 350lb . 1a. not changed- clinically necessary 3c. is “Project Management”, believe vendor meant 1c : not changed - clinically necessary 1e: changed to 400lbs.

W09 526 BRONX, NY 526B10007 What is your current vacuum assisted biopsy device used in breast imaging?

What is your current PACS and version number?

What Change Healthcare software version are you on and do you have the mammo module?

The siemens unit has the paddles with grid and perforations. There are but according to techs, no vacuum suction assist device for biopsy.

Current PACS is Philips Ver4.4 , soon to be AGFA Enterprise Imaging version 8.1.4.170

N/A W10 528 BUFFALO, NY 528B10171 4.a States biopsy paddle, is the request for needle loc paddle - or just wire localization?

What is your current version number of your PACS?

Would you like 2D Cad and 3D Artifical Intellegence? Needle localization paddle Current version is Philips IntelliSpace 4.4, moving to Agfa Enterprise Imaging for Radiology and Cardiology version 8.1.4.170 in FY22 Yes, please include both W11 537 CHICAGO, IL 537B18013 18a. Please update Minimum scan time at 1% CV for whole body to 339 seconds.

1c. Please update maximum patient load to 350lbs 1d. Please change minimum distance between table and carm to 17.9 inches to allow fair competition. 18a. No change to technical specification 1c. No change to technical specification 1d. Technical specification updated W12 540 CLARKSBURG, WV 540B11003 23f. Please change to 350lb maximum.

25h. PleaseLower minimum to 17.9 in.

28k. Please adjust heigh to 113 x 52 x 49 inches. 23f. specification has been adjusted 25h specification states minimum monitor size, if a larger monitor size is provided it will meet the minimum requirement – specification will not change 27f. specification has been adjusted for height, but length remains as specified.

W13 546 MIAMI, FL 546B10041 2. e. Please change to 19 inches to allow full and fair competition

2. f. This specification typically refers to available RAM. Are you referencing storage? If you are asking about hard disk space please change to 32 GB of RAM to allow full competition 1.k. Please change to 930 lbs to allow full and fair competition.

What is your current vaccum assisted biopsy device used in breast imaging?

What is your current PACS and software version number?

CAD is in the proposal would that include 3D CAD? 2e. Change to 19 inches 2f. change to 32 GB 1k. change to 930 lbs Atec Sapphire Model 100 Biopsy PACS: Philips Intellispace ; Software version 4.4 Include 3D CAD (Yes)

W14546 MIAMI, FL546B100411.Should a needle localization kit be included?
2.Is a breast biopsy vacuum system required?
3.Pleae clarify if a magnification stand and paddles are required?1. Yes

2. Yes

3. Yes W15 546 MIAMI, FL 546B10042 2. e. Please change to 19 inches to allow full and fair competition

2. f. This specification typically refers to available RAM. Are you referencing storage? If you are asking about hard disk space please change to 32 GB of RAM to allow full competition.

1.k. Please change to 930 lbs to allow full and fair competition.

What is your current vaccum assisted biopsy device used in breast imaging?

What is your current PACS and software version number?

CAD is in the proposal would that include 3D CAD? 2e. Change to 19 inches 2f. change to 32 GB 1k. change to 930 lbs Atec Sapphire Model 100 Biopsy PACS: Philips Intellispace ; Software version 4.4 Include 3D CAD (Yes)

W16546 MIAMI, FL546B100421.Should a needle localization kit be included?
2.Is a breast biopsy vacuum system required?
3.Pleae clarify if a magnification stand and paddles are required?1. Yes

2. Yes

3. Yes W17 548 WPB, FL 548B05040 1. h. Please change to 66cm to allow full and fair competition, 70cm is a vendor specific requirement.

1. k. Please change to 930 lbs to allow full and fair competition

2. e. Please change to 19cm to allow full and fair competition What is your current vacuum assisted biopsy device used in breast imaging?

What is your current Software version of Phillips PACS? 1.h - specification has been updated 1.k - specification has been updated 2.e. - specification has been updated Atec 4.4.544.0

W18548 WPB, FL548B050401.Should a needle localization kit be included?
2.Is a breast biopsy vacuum system required?
3.Pleae clarify if a magnification stand and paddles are required?1. Yes

2. Yes

3. Yes

W19548 WPB, FL548B150191. f. Please change this requirement to 350 lbs to allow for full and fair competition
1. j. Please change to 113x52x49 to allow full and fair competition1.f. - specification has been updated

1.j. - specification has been updated W20 549 DALLAS, TX 549B10594 1a. Please change minimum scan time at 1% CV for whole body to 339 sec for 153.1x66cm scan window.

1f. Please change exposure time to 339seconds 1g. Please change entrance dose to .047mGy.

1h. Please change to entrance dose to .047mGy 1i. Please make single energy femur optional 1j. Please change exposure time to 183 seconds and entrance dose to .329mGy.

1k. Please update scan region to 8.7x27.4in for IVA, 78x26 for WB 1p. Please clarify dimensions. Please update to 113L,52W,49H inches 2b. Please make Apex productivity tools optional as this is a Hologic specific software. 1a. No, this is needed for clinical studies and more than two vendors can meet specifications.

1f. No, this is needed for clinical studies and more than two vendors can meet specifications.

1g. No, this is needed for clinical studies and more than two vendors can meet specifications.

1h. No, this is needed for clinical studies and more than two vendors can meet specifications.

1i. No, this is needed for clinical studies and more than two vendors can meet specifications.

1j. No, this is needed for clinical studies and more than two vendors can meet specifications.

1k. No, this is needed for clinical studies and more than two vendors can meet specifications.

1p. Updated specifications to 50H 120L 55W 2b. Updated specifications to optional

W21558 DURHMAN, NC558B103611. f. Please change this requirement to 350 lbs to allow for full and fair competitionChanged this requirement to 350 lbs to allow for full and fair competition.
W22558 DURHAM, NC558B10362What is your current vacuum assisted biopsy device?Hologic ATEC
W23558 DURHAM, NC558B10362Should we submit the quote with just the mammo unit or should the quote consist of both the mammo and coach (trailer).Changed this requirement. The quote should not contain a trailer (coach).
W24565 FAYETTEVILLE, NC565B10079What is your current vacuum assisted biopsy device?

In regards to CAD is 2D & 3D CAD are desired?

What Change Software versiondo you currently have and are you on mammo module? Hologic ATEC YES …. 2D & 3D desired MK PACS = v12.4.1 no Mammo Module

W25565 FAYETTEVILLE, NC565B100801. f. Please change this requirement to 350 lbs to allow for full and fair competitionChanged this requirement to 350 lbs to allow for full and fair competition.
W26573 JACKSONVILLE, FL573B150682. e. Please change to 19 inches to allow full and fair competition

2. f. This specification typically refers to available RAM. Are you referencing storage? If you are asking about hard disk space please change to 32 GB of RAM to allow full competition What is your current vacuum assisted biopsy device?

Are both 2D & 3D CAD required or wanted?

What is your current Phillps PACS sofware version? 2.e: Changed to 19in 2.f.: Changed Bx device: Celero Yes, both 2D and 3D CAD are required PACS: Intellispace 4.4.544.0

W27573 LAKE CITY, FL573B150751. f. Please change this requirement to 350 lbs to allow for full and fair competition1f: Changed
W28575 GR JUNCTION, CO575B160241a. please update minimum scan time at 1% CV for whole body to 339 seconds to allow fair competition.

1c. Please update minimum scan time at 1% CV for lateral spine to 62 seconds to allow fair competition.

1d. Please allow minimum scan time at 1% CV for forearms to 24 seconds.

1h. Please update minimum distance between table and carm to 17.9 inches to allow fair competion.

1k. Please clairfy dimensions required for this spec and consider optimal dimensions of 113x52x49" 1l. Please consider updating maximum system weight to 778lbs to allow fair competition. 1a. Updated to 340 seconds to allow for fair competition 1c. Updated to 65 seconds to allow for fair competition 1d. Updated to 25 seconds to allow for fair competition 1h. Updated to 20 inches to allow for fair competition 1k. Updated to 100x55x60 inches which are the max dimensions for the room it will be in 1l. Updated to 800lbs to allow for fair competition W29 581 HUNTINGTON, WV 581B11246 23f. Please change to 350lb maximum. (Not sure why we are asking them to change this when we meet the spec.?)

25h. Lower minimum to 17.9 in.

28k. Please adjust heigh to 113 x 52 x 49 inches. 23f. specification has been adjusted 25h specification states minimum monitor size, if a larger monitor size is provided it will meet the minimum requirement – specification will not change 27f. specification has been adjusted for height, but length remains as specified.

W30 589 TOPEKA, KS 589B12024 1a. Please change minimum scan time at 1% CV for whole body to 339 seconds.

1c. Please change minimum scan time at 1% CV for lateral spine to 62 seconds 1d. Please change minimum scan time at 1% CV for forearms to 24 seconds.

1h. Please change minimum distance between table and carm to 17.9 inches to allow fair competition.

1k. Please specify dimensions needed and consider 113x52x49 inches.

1l. Please consider updating maximum system weight to 778lbs to allow fair competition. 1a. Technical specification will not be changed.

1c. Technical specification will not be changed.

1d. Technical specification will not be changed.

1h. Technical specification will not be changed.

1k. There is no technical specification for this questions 1l. Technical specification will not be changed.

W31 589 COLUMBIA, MO 589B13015 3. c. What vacuum assisted biopsy system are you using?

3. c. Do you require a biopsy chair to be quoted?

3.k. In regards to CAD is 2D & 3D CAD are desired?

5. f. Would you please specify the Philips PACS software version?" 3. c. Technical specification updated. No biopsy chair

3. k. 2D and 3D

5. f. Philips Intellispace v4.4 W32 593 LAS VEGAS, NV 593B10063 1. k. Please make weight less then 930 lbs for fair competition

2. e. Please change monitor sie to 19" to allow for fair competition Do you require a wire localization package for both gantries?

In regards to CAD, are you looking for 2D CAD and 3D Artificial Intelligence?

What is your PACS software version? Does it include a mamo Module, and if so what is the mamo module.

Do you require full biopsy solution with hardware on both gantries being quoted? 1) 1.k. Maximum weight changed to 930 lbs.

2) 2.e. Minimum review workstation monitor size changed to 19”

3) Yes

4) Yes

5) SecureView FCCURE

6) At least one room

W33595 LEBANON, PA595B190261e. Please change to 350lbs to allow for fair competition.1e. Technical specification updated to 350 lbs
W34605 LOMA LINDA, CA605B10023What is you is your current reading solution and version for interpreting mammography studies?

Will you be performing biopsies on site?

AGFA IMPAX Version 6.6.1 with BARCO Monitor Model MDMC-12133. License # 0000133731-1 Yes W35 605 LOMA LINDA, CA 605B10024 What is you is your current reading solution and version for interpreting mammography studies?

Will you be performing biopsies on site?

AGFA IMPAX Version 6.6.1 with BARCO Monitor Model MDMC-12133. License # 0000133731-1 Yes W36 612 MATHER, CA 612B18012 1. j. Please make weight less then 930 lbs for fair competition.

3.u is vendor specfic can it state vacuum assited device to allow fair competition?

Since biopsy is being quoted does the site need a biopsy chair quoted as well?

3.n and 3.p and 4.h are vendor specific specifications, please consider a generic specification or move to Value Add/optional.

What is your current PACS software version?

In regards to CAD, are you looking for 2D CAD and 3D Artificial Intelligence?

What is your PACS software version?

1) 1.j. Maximum weight is set to not exceed 930 lbs.

2) 3.u. Changed

3) Biopsy chair does not need to be quoted and that specification has been taken out.

4) 3.n and 4.h moved to value added.

5) 3.p has been revised to not limit to one vendor’s specific specifications.

6) Yes

7) Current PACS software is Philips IntelliSpace v 4.4.554.0. Mammo PACS software is MRS7.

W37 612 MATHER, CA 612B18013 1. j. Please make weight less then 930 lbs for fair competition.

3.u is vendor specfic can it state vacuum assited device to allow fair competition?

Since biopsy is being quoted does the site need a biopsy chair quoted as well?

3.n and 3.p and 4.h are vendor specific specifications, please consider a generic specification or move to Value Add/optional.

What is your current PACS software version?

In regards to CAD, are you looking for 2D CAD and 3D Artificial Intelligence?

What is your PACS software version? 1) 1.j. Maximum weight is set to not exceed 930 lbs.

2) 3.u. Changed

3) Biopsy chair does not need to be quoted and that specification has been taken out.

4) 3.n and 4.h moved to value added.

5) 3.p has been revised to not limit to one vendor’s specific specifications.

6) Yes

7) Current PACS software is Philips IntelliSpace v 4.4.554.0. Mammo PACS software is MRS7.

W38 644 PHOENIX, AZ 644B18154 Can you provide serial number for trade in unit?

Can you provide current PACS vendor and software level?

Will you require needle locs as part of the requirement? SN: 81006143576 EE: 87260 Hologic SecureXChange, AGFA IMPAXs 6.7, Medicom Yes

W39644 PHOENIX, AZ644B181651f. Please change to 350lb maximum.Technical specification document has been updated.
W40646 PITTSBURGH, PA646B1201324f. Please change to 350lb maximum.24f. No change to technical specification due to bariatric population
W41650 PROVIDENCE, RI650B110281f. Please change to 350lbs to allow for fair competition.
1i. Please change to 113x52x49 inches1f. Change accepted - specification has been updated

1i. Change accepted - specification has been updated W42 659 SALISBURY, NC 659B10008 What is your current vacuum assisted biopsy device?

In regards to CAD is 2D & 3D CAD are desired?

What Change Software versiondo you currently have and are you on mammo module?

Hologic ATEC YES …. 2D & 3D desired MK PACS = v12.4.1 no Mammo Module

W43660 SALT LAKE CY, UT660B100161f. Please update minimum patient table load capacity to 350lbs to allow fair competition.
1h. Please update minimum distance between table and carm to 17.9 inches to allow fair competition.1.Technical specification updated to 350 lbs
2.Technical specification updated to 17 inches
W44673 TAMPA, FL673B12046In regards to CAD, would you please specify if 2D & 3D CAD are desired?Both are required
W45673 TAMPA, FL673B120471. f. Please change this requirement to 350 lbs to allow for full and fair competition

1. h. Please change to 17.9 inches to allow full and fair competition

1. k. Please change to 10.5'x11' to allow full and fair competition 1.f. - specification has been updated 1.h. - specification has been updated 1.k. - specification has been updated W46 673 TAMPA, FL 673B12047 1. f. Please change this requirement to 350 lbs to allow for full and fair competition

1. h. Please change to 17.9 inches to allow full and fair competition

1. k. Please change to 10.5'x11' to allow full and fair competition 1.f. - specification has been updated 1.h. - specification has been updated 1.k. - specification has been updated W47 674 TEMPLE, TX 674B10156 3. c. What vacuum assisted biopsy system are you using?

3. k. In regards to CAD is 2D & 3D CAD are desired?

5. f. Would you please specify the McKesson software version? On mammo module?" 3.c. Not currently performing biopsies 3.k. Yes 2D and 3D 5.f. 12.4.1 W48 674 TEMPLE, TX 674B10158 3. c. What vacuum assisted biopsy system are you using?

3. k. In regards to CAD is 2D & 3D CAD are desired?

5. f. Would you please specify the McKesson software version? On mammo module?" 3.c. Not currently performing biopsies 3.k. Yes 2D and 3D 5.f. 12.4.1 W49 688 WASHINGTON, DC 688B10019 1a. Please change Generator Voltage Range to 22-49kV to allow fair competition.

What is your current vacuum assisted biopsy device used in breast imaging?

What software version are you on with Change Healthcare and do you have the mammo module? 1a. Specification will not change – the range of 22-49 meets the minimum requirement We have not selected a vacuum assisted biopsy device at this time.

The current version of Change Healthcare PACS is 12.4.1 and yes we do have the Mammo module.

W50689 WEST HAVEN, CT689B110091f. Please change to 350lbs to allow for fair competition.1f. Change accepted - specification has been updated
W51689 WEST HAVEN, CT689B110111f. Please change to 350lbs to allow for fair competition.1f. Change accepted - specification has been updated
W52757 COLUMBUS, OH757B1902121d. Please update minimum patient table load capacity to 350lbs1.d No change to Technical Specification

File details come from the government source that posted it. Updated .