Schedule of Line Items
| HTME VISN | Requisition Number in Action | Equipment Type | Station | Consolidation Period | Benchmark (BM) or Technical Specification (TS) |
| VISN 10: VA Healthcare System of Ohio | 541B13025 | XR US | 541 CLEVELAND, OH | FEB 2021 | BM |
| VISN 10: VA Healthcare System of Ohio | 541B13026 | XR US | 541 CLEVELAND, OH | FEB 2021 | BM |
| VISN 10: VA Healthcare System of Ohio | 552B00002 | XR US | 552 DAYTON, OH | FEB 2021 | BM |
| VISN 10: VA Healthcare System of Ohio | 552B00005 | NM PET CT | 552 DAYTON, OH | FEB 2021 | BM |
| VISN 12: VA Great Lakes Health Care System | 556B13010 | XR MAMMO | 556 N. CHICAGO, IL | FEB 2021 | BM |
| VISN 12: VA Great Lakes Health Care System | 607B10005 | XR US | 607 MADISON, WI | FEB 2021 | BM |
| VISN 15: VA Heartland Network | 589B10025 | XR US | 589 KANSAS CITY, MO | FEB 2021 | BM |
| VISN 15: VA Heartland Network | 657B10032 | XR US | 657 ST LOUIS, MO | FEB 2021 | BM |
| VISN 15: VA Heartland Network | 657B10033 | XR US | 657 ST LOUIS, MO | FEB 2021 | BM |
| VISN 15: VA Heartland Network | 657B10034 | XR US | 657 ST LOUIS, MO | FEB 2021 | BM |
| VISN 17: VA Heart of Texas Health Care Network | 504B07037 | XR US | 504 AMARILLO, TX | FEB 2021 | TS |
| VISN 17: VA Heart of Texas Health Care Network | 674B00187 | XR RF | 674 TEMPLE, TX | FEB 2021 | TS |
| VISN 17: VA Heart of Texas Health Care Network | 674B00189 | XR RAD | 674 TEMPLE, TX | FEB 2021 | TS |
| VISN 17: VA Heart of Texas Health Care Network | 674B90190 | XR RF | 674 TEMPLE, TX | FEB 2021 | TS |
| VISN 17: VA Heart of Texas Health Care Network | 740B10003 | XR MAMMO | 740 HARLINGEN, TX | FEB 2021 | TS |
| VISN 17: VA Heart of Texas Health Care Network | 740B10004 | XR RF | 740 HARLINGEN, TX | FEB 2021 | TS |
| VISN 17: VA Heart of Texas Health Care Network | 740B10005 | NM BONE DEN | 740 HARLINGEN, TX | FEB 2021 | TS |
| VISN 17: VA Heart of Texas Health Care Network | 740B10006 | XR RAD | 740 HARLINGEN, TX | FEB 2021 | TS |
| VISN 17: VA Heart of Texas Health Care Network | 740B10007 | XR CT | 740 HARLINGEN, TX | FEB 2021 | TS |
| VISN 2: VA Healthcare Network Upstate New York | 528B07145 | XR RF | 528 BATH, NY | FEB 2021 | BM |
| VISN 2: VA Healthcare Network Upstate New York | 561B00009 | XR US | 561 EAST ORANGE, NJ | FEB 2021 | BM |
| VISN 21: VA Sierra Pacific Network | 640B14002 | XR US | 640 PALO ALTO, CA | FEB 2021 | TS |
| VISN 21: VA Sierra Pacific Network | 640B14003 | XR US | 640 PALO ALTO, CA | FEB 2021 | TS |
| VISN 21: VA Sierra Pacific Network | 640B14004 | XR US | 640 PALO ALTO, CA | FEB 2021 | TS |
| VISN 21: VA Sierra Pacific Network | 654B19023 | XR PORT C-ARM | 654 RENO, NV | FEB 2021 | BM |
| VISN 21: VA Sierra Pacific Network | 662B10004 | XR US | 662 SAN FRANCISCO, C | FEB 2021 | TS |
| VISN 21: VA Sierra Pacific Network | 662B10005 | XR PORT C-ARM | 662 SAN FRANCISCO, C | FEB 2021 | TS |
| VISN 23: VA Midwest Health Care Network | 568B10014 | XR RAD | 568 FT MEADE, SD | FEB 2021 | TS |
| VISN 23: VA Midwest Health Care Network | 636B19000 | XR RAD | 636 DES MOINES, IA | FEB 2021 | TS |
| VISN 4: VA Stars & Stripes Healthcare Network | 460B08007 | XR US | WILMINGTON, DE | FEB 2021 | BM |
| VISN 4: VA Stars & Stripes Healthcare Network | 460B08008 | XR US | WILMINGTON, DE | FEB 2021 | BM |
| VISN 4: VA Stars & Stripes Healthcare Network | 460B08009 | XR RAD | WILMINGTON, DE | FEB 2021 | BM |
| VISN 4: VA Stars & Stripes Healthcare Network | 693B10004 | XR MAMMO | 693 WILKES-BARRE, PA | FEB 2021 | TS |
| VISN 6: VA Mid-Atlantic Health Care Network | 652B00022 | XR PORT C-ARM | 652 RICHMOND, VA | FEB 2021 | TS |
| VISN 6: VA Mid-Atlantic Health Care Network | 652B00023 | XR PORT C-ARM | 652 RICHMOND, VA | FEB 2021 | TS |
| VISN 6: VA Mid-Atlantic Health Care Network | 652B00027 | XR PORT RAD | 652 RICHMOND, VA | FEB 2021 | TS |
| VISN 7: VA Southeast Health Care Network | 509B12002 | XR MAMMO | 509 AUGUSTA, GA | FEB 2021 | BM |
| VISN 7: VA Southeast Health Care Network | 544B12021 | XR US | 544 COLUMBIA, SC | FEB 2021 | BM |
Q&A
| VISN | Requisition Number | QUESTIONS | ANSWERS |
| VISN 17 | 504B07037 | We believe the semi-robotic arm specification is a reference to the Artemis 3D Semi-Robotic Prostate Fusion Biopsy system from InnoMedicus. We contend that the Fusion Targeted Biopsy System is a separate product from the Ultrasound and should not be bundled with it. | 9g. Moved to “Added Value” section. |
| VISN 17 | 504B07037 | 1. Request 1.g. Minimum control panel rotation 25 degrees be removed |
2. Request 1.i. Minimum number of transducer ports from 5 be revised to 4
3. Request 2.c. Is Navigation software a requirement? If not recommend removal or move to 10. Added Value.
4. Request 2.d. Is Image Fusion Software a requirement? If not recommend removal or move to 10. Added Value.
5. Request 5.j. Is shear wave and strain-based elastography a requirement? If not recommend removal or move to 10. Added Value.
6. Request 8.b. Recommend linear frequency range 18-5 be revised to 15-5 MHz
7. Request 8.c. Recommend linear frequency range 12-2 be revised to 12-3 MHz
8. Request 8.d. Recommend curved array EC frequency range 10-2 be revised to 10-3 MHz.
9. Request 9.a. Is capability of accepting MRI images a requirement? If not recommend removal or move to 10. Added Value.
10. Request 9.g. Is semi-robotic arm, physically-attached to biopsy system a requirement? If not recommend removal or move to 10. Added Value. 1g. 25 degrees was removed.
1i. 4 ports acceptable.
2c. Moved to “ Added Value” section 2d. Moved to “ Added Value” section 5j. Moved to “ Added Value” section 8b. 15-5 MHz is acceptable.
8c. 12-3 MHz is acceptable.
8d. 10-3 MHz is acceptable.
9a. Moved to “ Added Value” section 9g. Moved to “ Added Value” section
| VISN 10 | 541B13025 | The quote is labeled as quantity 4. However, line #1 says quantity 2 “VIVID E95 4D-Ultra edition V204” Catalog # H8010EA. Line #2 says quantity 2 “VIVID E95 4D-Ultra edition V204 ALT”, with the different catalog number H8010EC. Is the intent to purchase 4 systems, two of each model? Or is the requirement for 2 systems, with a primary and alternate model to select from? Please clarify what ALT means here, and how it changes the second two systems. | Yes, the intent is to purchase 4 systems, two of each model. |
| VISN 12 | 556B13010 | 1. What are the system serial #’s |
2. Does the station want 3D CAD?
3. Does the station want HD technology?
4. Does the station want a workstation or does the station read from PACS?
5. Does the station want breast density software? 1. 847688JLMAM1A/ 847688JLMAM2 2.Not Required
3. N/A
4. Yes
5. Yes
| VISN 12 | 556B13010 | Please clarify if you are requiring just 2D CAD or 3D CAD to be quoted as well? | |
| Please clarify the PBX TRADE HOLX – Trade In Manufacturer and Trade In Serial number? | 2D CAD- yes 3D Not required. |
Trade in Manufacturer: GE and Serial number 847688IDI VISN 02 561B00009 A. Which Department is this system intended to be used in?
B. What is the clinical use for the system?
C. Please provide transducers and serial numbers that will accompany the trade-in system A. Radiology Service B. Interventional Radiology
C. 10L-1959651YM9
M12L-99840BYM9
3.5SC-40270PD1
M7C-935201YM4
E8C-1462WX8
VISN 23 568B10014 Technical Specifications Matrix, Section 2, question n Minimum table height from floor (in) 21"
| Please change to 22" to allow for more competition. | This is a clinical requirement and is offered by two or more vendors. | ||
| VISN 23 | 568B10014 | 2j - Which detector does this specification apply to? 17"x17" or 10"x12" | |
| 5a - Would the site consider an alternate quote for a single shot long length imaging detector in place of stitching software? | 2j. Both detectors |
5a. Alternate offers are always evaluated.
| VISN 23 | 568B10014 | 4. Workstation Requirements | ||
| d. LCD Touchscreen acquisition workstation - Please removed the word "touch" from the spec. | This is a clinical requirement and is offered by two or more vendors. | |||
| VISN 15 | 589B10025 | Trade information Logiq E needed: Serial number, probes that will be traded in and confirm model | There is no trade in, this request is for an additional US. | |
| VISN 23 | 636B19000 | Technical Specifications Matrix, Section 2, question n |
Minimum table height from floor (in) 21"
| Please change to 22" to allow for more competition. | This is a clinical requirement and is offered by two or more vendors. | ||
| VISN 23 | 636B19000 | 2j - Which detector does this specification apply to? 17"x17" or 10"x12" | |
| 5a - Would the site consider an alternate quote for a single shot long length imaging detector in place of stitching software? | 2j. Both detectors |
5a. Alternate offers are always evaluated.
| VISN 23 | 636B19000 | 4. Workstation Requirements | ||
| d. LCD Touchscreen acquisition workstation - Please removed the word "touch" from the spec. | This is a clinical requirement and is offered by two or more vendors. | |||
| VISN 21 | 640B14002 | 1. | Request 1.h. Control Panel Height be revised from 36 to 48 inches to 30 to 36 inches. | |
| 2. | Request 1.k. Minimum battery life 30 min be moved to 9. Added Value. | |||
| 3. | Request 1.m. Maximum equipment weight of 230 lbs be increased to 265 lbs. | |||
| 4. | Request 5.j. Ability to enter standby mode or sleep mode be moved to 9. Added Value. | 1h. - This is a clinical requirement and is offered by two or more vendors. |
1k. - This is a clinical requirement and is offered by two or more vendors.
1m. - This is a clinical requirement and is offered by two or more vendors.
5j. - This is a clinical requirement and is offered by two or more vendors.
| VISN 21 | 640B14003 | 1. | Request 1.h. Control Panel Height be revised from 36 to 48 inches to 30 to 36 inches. | |
| 2. | Request 1.k. Minimum battery life 30 min be moved to 9. Added Value. | |||
| 3. | Request 1.m. Maximum equipment weight of 230 lbs be increased to 265 lbs. | |||
| 4. | Request 5.j. Ability to enter standby mode or sleep mode be moved to 9. Added Value. | 1h. - This is a clinical requirement and is offered by two or more vendors. |
1k. - This is a clinical requirement and is offered by two or more vendors.
1m. - This is a clinical requirement and is offered by two or more vendors.
5j. - This is a clinical requirement and is offered by two or more vendors.
| VISN 21 | 640B14004 | 1. | Request 1.h. Control Panel Height be revised from 36 to 48 inches to 30 to 36 inches. | |
| 2. | Request 1.k. Minimum battery life 30 min be moved to 9. Added Value. | |||
| 3. | Request 1.m. Maximum equipment weight of 230 lbs be increased to 265 lbs. | |||
| 4. | Request 5.j. Ability to enter standby mode or sleep mode be moved to 9. Added Value. | 1h. - This is a clinical requirement and is offered by two or more vendors. |
1k. - This is a clinical requirement and is offered by two or more vendors.
1m. - This is a clinical requirement and is offered by two or more vendors.
5j. - This is a clinical requirement and is offered by two or more vendors.
VISN 06 652B00027 1a - To ensure maximum competition, please change to 32kW 1d - To ensure maximum competition, please change to 0.1 - 320. There are no portable x-ray exams that should ever require 500 mAs given the efficiency (DQE and TQF) of today's detectors.
1f - To ensure maximum competition, please change to 130cm 1j - Please remove specification; given most, if not all, vendors have detector charging mechanisms either on the cart, within the detector bins, or via external chargers, battery life without recharge is not a necessary spec. Additionally, site has included spec 2g, which requires detector battery charging on the unit. Further, as all batteries degrade over time, the accuracy of this specification cannot be verified, nor are kVp and mAs consistent throughout all exposures, making the spec obsolete.
1l - To ensure maximum competition, please change to -90/+90. Only one vendor features -90/+180.
1m - Please clarify or remove spec; is this final image? Preview image?
1n - To ensure maximum competition, please change to 19" or less.
2n - To ensure maximum competition, please remove this requirement; only one vendor features this software. 1a - Changed specification to "minimum 30KW" 1d - Changed specification to "minimum 0.1-320 mAs."
1f - Changed specification to "at least 100cm" 1j - Removed from technical specifications.
1l - Changed specification to -90 to +90 1m - Removed from technical specifications.
1n - Changed specification to 19" or greater.
2n - Moved specification to the added value section.
| VISN 06 | 652B00027 | 1. Unit Physical Specifications a. High Frequency Power Generator 40kw. | |
| Question: This is an unduly restrictive requirement to allow open compeitition, please change the minimum generator to 30KW. | Changed specification to "minimum 30KW" | ||
| VISN 06 | 652B00027 | 1. Unit Physical Specifications b. generator 40-150 kvp range | |
| Question: To allow for open competition please change the minimum kv range to 50-125 kv | Changed specification to at least "50 - 125." | ||
| VISN 06 | 652B00027 | 1. Unit Physical Specifications c. tube maximum 150 kVp | |
| Question: To allow for open competition please change the maximum kVp to 125 | Changed specification to "at least 125". | ||
| VISN 06 | 652B00027 | 1. Unit Physical Specifications d. Generator 00.1-500 mAs range | |
| Question: To allow for open competition please the minimum requirement to .1~400 mAs | Changed specification to "Minimum range 0.1 - 320 mAs". | ||
| VISN 06 | 652B00027 | Technical Requirements |
1. Unit physical specifications
a. High frequency generator power [kW]: 40 Question: Please adjust this requirement to 30 kW to allow for fair competition.
f. Maximum tube reach (cm): 135.5 Question: This requirement is restrictive and specific to one vendor. Please adjust to 100 cm to allow for fair competition.
j. Minimum detector battery life without recharge [exposure/hour at kVp and mAs]: 220exp/hr at 80kv & 2Mas Question: This requirement is restrictive and specific to one vendor in how they determine battery life. Please remove this requirement or change to a more generic specification.
2. Additional specifications
e. Collapsible column Question: Please remove this requirement or make optional to allow for fair competition.
f. LED positioning light Question: Please remove “LED” from this requirement to allow for fair competition.
k. Display monitor at tube head Question: Please remove this requirement or make optional to allow for fair competition.
n. Tube and detector angle alignment software Question: Restrictive. Please remove this requirement or make optional to allow for fair competition.
Section F. Trade-In Question: Please calrify if there is a trade-in? If so, please provide the following information to determine an accurate trade-in value: make, model, quantity, number of detectors include their sizes, and year of Installation.
1a. Changed specification to minimum 30 kW.
1f. Changed specification to "at least 100cm".
1j. Removed from technical specifications.
2e. Moved specification to the added value section.
2f. Removed from technical specifications.
2k. Removed from technical specifications.
2n. Moved specification to the added value section.
F. Yes, we will trade in three Shimadzu Mobiledart Evolution portables. Each will come with a 14 by 7 detector. The portables were installed in 2013. Please refer to page 4 of the updated technical specifications document for more information.
VISN 21 654B19023 1) Is the system operational?
2) What size is the ii detector?
3) What software and hardware packages are included in the system?
4) When was it manufactured?
5) What monitors are on the trolly? 1. Yes.
2. Nine (9) inches.
3. Standard OEC Ortho package.
4. The acquisition date is 05/22/2009.
5. Standard OEC touchscreen monitors.
VISN 17 674B00187 1a - To ensure maximum competition, please change to 19" 1d - Please remove spec or clarify why variable speed is necessary or desired?
1s - Please define "articulation" with respect to the table 1a. Has been updated to 19" 1d. This is a clinical requirement and is offered by two or more vendors.
2s.(2r) Has been defined in updated specs VISN 17 674B00187 1.d.) – Please adjust this requirement to +/-45 degrees to allow for fair competition?
2.g.) – Please make the tilting functionality optionl to allow for fair competition.
2.r.) – Please change spec to allow for multiple options (ie, step, lowering table, etc) to allow for fair competition. 1d. This is a clinical requirement and is offered by two or more vendors.
2g. Moved to "Added Value" Section 2r. (2q) This is a clinical requirement and is offered by two or more vendors.
| VISN 17 | 674B00189 | 5a - Would the site consider an alternate quote to include a single shot long length imaging detector in place of stitching software be of interest to the site? | 5a. Alternate offers are always evaluated. |
| VISN 17 | 674B00189 | Can you please adjust minimum touchscreen monitor from 20" to 19" to allow for open & fair competition | 1a. Has been updated to 19" |
| VISN 17 | 674B90190 | 1a - To ensure maximum competition, please change to 19" |
1d - Please remove spec or clarify why variable speed is necessary or desired?
1s - Please define "articulation" with respect to the table 1a. Has been updated to 19" 1d. This is a clinical requirement and is offered by two or more vendors.
2s.(2r) Has been defined in updated specs VISN 17 674B90190 1.d.) – Please adjust this requirement to +/-45 degrees to allow for fair competition?
2.g.) – Please make the tilting functionality optionl to allow for fair competition.
2.r.) – Please change spec to allow for multiple options (ie, step, lowering table, etc) to allow for fair competition. 1d. This is a clinical requirement and is offered by two or more vendors.
2g. Moved to "Added Value" Section 2r. (2q) This is a clinical requirement and is offered by two or more vendors.
| VISN 04 | 693B10004 | No quantity is listed in the compliance matrix. Please clarify quantity requested? | As noted in the technical specification document, the requirement is for one system | |
| VISN 04 | 693B10004 | 2. Workstation Requirements: Question: Is a dedicated radiologist workstation for reading required? | ||
| 3. Additional specifications n. Computer-aided detection (CAD): Question: Is this requirement 2D CAD only or 2D + 3D CAD? | 2. The review workstation referenced in the technical specification is considered the reading workstation. |
3. 2D CAD required, 2D + 3D Preferred VISN 17 740B10003 Please confirm what vacuum-assisted biopsy (VAD) device you currently use today?
Please confirm if a biopsy chair needed?
1.h.) Please update SID to a range of 60-70 cm to allow for fair competition.
1. j.) Please update Maximum system weight to 926 lbs to allow for fair competition.
3.m.) Please confirm if just 2D CAD only is needed, or are you requiring 3D CAD as well?
5.e.) Please confirm the current version of the McKesson PACS and the version of the Mammo module that operates on the PACS? No VAD is currently being utilized.Added to "Added Value" section if this items is on vendors IDIQ contract please quote as optional No Biopsy chair is currently being utilized. Added to "Added Value" section if this items is on vendors IDIQ contract please quote as optional 1h. This is a clinical requirement and is offered by two or more vendors.
1j. Requirement has been removed 3m. 3D CAD added to "Added Value" section, 2D added to minimum requirement 5e. additional information added to UPDATED tech spec.
| VISN 17 | 740B10003 | No quantity is listed in the compliance matrix. Please clarify quantity requested? | Quantity requested is 1 |
| VISN 17 | 740B10004 | Please provide room drawings, or a VA POC to provide them, for the new Corpus Christi. | After selection of a vendor for this consolidation, the Valley Coastal Bend Healthcare System will provide all necessary drawings for the space allocated for this system. |
| VISN 17 | 740B10004 | Question for Line 18 and Line 19 : The spec is asking for a wall detector size of 14x17 and a table detector size of 17x17. Should these be swapped? |
Question Line 24 : Clarification on the patient table load of 650 pounds-ls this correct? 1a. Changed to Fixed Table Detector size - 14x17 1b. Changed to wireless detector size - 14x17 1g. Patient table load capacity changed to 550lbs VISN 17 740B10004 1c - Please confirm specification of 1lp/mm or update 1g - To ensure maximum competition, please change to 550 lbs 1v - Please reduce to 19", as this is industry standard 1y - Please provide rationale for why weight limit is required 2j - Please remove spec or provide further information for why variable speed is required 1c. Requirement has been modified to 3.38lp/mm 1(f)g. Requirement has been modified.
1(u)v. Requirement has been updated.
1y. Requirement has been removed 2j. It’s 2i. This requirement has been removed In addition:
1p. Modified Minimum acquisition matrix size to 2880X2880 2c. Fluoroscopic tube Configuration: Overhead VISN 17 740B10004 1. f.) Please adjust to 80 - 120 in. to allow for fair competition.
2.dd.) Please make the tilting functionality optional to allow for fair competition.
3.b.) Please make the Bone suppression software optional to allow for fair competition.
1f. Requirement has been removed 2dd. Requirement has been removed 3b This is a clinical requirement and is offered by two or more vendors.
VISN 17 740B10005 1. a. Scan time depends on scan size and scan mode. Please change to 339 to allow for fair competition.
1. c. Scan time depends on scan size and scan mode. Please change to 62 sec to allow for fair competition.
1. d. Scan time depends on scan size. Please change to 24 sec to allow for fair competition.
1. i. please change to 778 lbs to allow for fair competition 1a. Requirement has been modified to 400s 1c. Requirement has been modified to 65s 1d. Requirement has been modified to 25s 1.i. (1j.) Requirement has been removed
| VISN 17 | 740B10006 | Please provide room drawings, or a VA POC to provide them, for the new Corpus Christi. | After selection of a vendor for this consolidation, the Valley Coastal Bend Healthcare System will provide all necessary drawings for the space allocated for this system. |
| VISN 17 | 740B10006 | Line 23 Need clarification on Minimum patient table load capacity | The requirement has been modified to 600lbs |
| VISN 17 | 740B10006 | The Compliance Matrix has identified the following requirements for detectors: |
a. wall stand digital detector
b. table digital detector.
| Please advise the site's preference for fixed or wireless detectors for a. and b. | e and f: One fixed table detector(14x17) and one wireless detector (14x17) | ||
| VISN 17 | 740B10006 | 1m - Please increase to 20" to ensure maximum competition |
1p - Please change to 140 to ensure maximum competition. Only 1 vendor features 100.
1q - Please clarify what is meant by spec. Is this preview image time?
1t - Please clarify spec. Is this inches, centimeters, millimeters?
4a - Please remove requirement for a minimum of 2 acquisition workstations. There is no need for 2 acquisition workstations.
4d - Please remove requirement for a minimum of 2 processing workstations. There should never be a need for 2.
5b - Would the site consider an alternate quote that replaces the stitching package with a single shot long length imaging detector? 2m. Requirement has been modified.
2p. Requirement has been modified.
2q. Nominal changed to Minimum 2t. Requirement is in “cm”, added to Tech Spec 4a Modified to 1 acquisition workstation 4d Modified to 1 processing workstation 5b Alternate offers are always evaluated.
| VISN 17 | 740B10006 | Can you please adjust maximim pixel pitch from 100 microns to 150 microns, to allow for open & fair competition | 2p. Requirement has been modified. |
| VISN 17 | 740B10007 | Please provide room drawings, or a VA POC to provide them, for the new Corpus Christi. | After selection of a vendor for this consolidation, the Valley Coastal Bend Healthcare System will provide all necessary drawings for the space allocated for this system. |
| VISN 17 | 740B10007 | 1a. 128 Acquired or reconstructed slices? |
3c. In 4g, they require integration into Tera Recon post processing platform. So, do the want vendors to quote their OEM software and workstations?
4a. Do they want real live fluoroscopy or just biopsy mode (i.e. step & shoot)?
4g. Integration into Tera Recon – Again if they are using Tera as their post processing platform do they need the OEM to quote their post processing software with workstations?
4i. – Please elaborate on clinical need? Only one vendor offers variable helical pitch due to its unproven clinical value.
Do they want a Bayer injector included? With or without Bayer’s contrast informatics? 1a. Requirement has been modified to 128 reconstructed slices 3c. Requirement changed to 1 processing workstation 4a. Biopsy mode is requested 4g. moved to section 5 connectivity requirements 4i. Requirement has been removed 2i. updated to state Bayer constrast injector and 4f. moved to section 5 connectivity requirements
| VISN 17 | 740B10007 | 1.a) Please clarify, is the system to provide 128 slices acquired per rotation or can the system reconstruct to 128 slices per rotation? | |
| 1.c.) To allow for open competition please change the minimum requirement to 70cm? | 1.a) Modified to 128 reconstructed slices |
1.c) Modified to 70cm