Vendor Q_A May 2020 Consolidation - FINAL.xlsx

XLSX spreadsheet 28 KB Posted

Attached to
6525--MAY FY20 CONSOLIDATION Federal contract opportunity
Solicitation number
36A79720Q0018
Issued by
Department of Veterans Affairs National Acquisition Center

View the file

Other files for this federal contract opportunity

Other files attached to 6525--MAY FY20 CONSOLIDATION, newest first.
File Type Posted
Amendment 0002 - Continuation Page.pdf PDF
662B00017 - 674B00244.zip ZIP file
436B00014 - 648B03001.zip ZIP file
36A79720Q0018 0002.docx DOCX document
Updated RFQ Instructions to Offerors.pdf PDF
36A79720Q0018 0001.docx DOCX document
528B00078 - 570B00026.zip ZIP file
Amendment 0001 - Continuation Page.pdf PDF
May 2020 Consolidation Schedule of Line Items.xlsx XLSX spreadsheet
Memorandum for Turnkey.pdf PDF
436B00014 - 504B07014.zip ZIP file
Memorandum for Training.pdf PDF
654B09018 - 662B00024.zip ZIP file
612B09000 - 649B07012.zip ZIP file
RFQ Instructions to Offerors.pdf PDF
May 2020 Consolidation Schedule.pdf PDF
675B07001 - 740B00008.zip ZIP file
546B07001 - 603B03029.zip ZIP file
528B00078 - 541B03022.zip ZIP file
541B03024 - 541B93031.zip ZIP file
Vendor Workup Spreadsheet - 36A79720Q0018.xlsx XLSX spreadsheet
Vendor Folder.zip ZIP file
508B02012 - 509B03007.zip ZIP file
VA Directive 6550 - New Appendix A - published Jun. 2019.pdf PDF
663B90037.zip ZIP file
36A79720Q0018_1.docx DOCX document
671B00015 - 674B00244.zip ZIP file
Show all 27

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

MAY 2020 Schedule of Line Items

HTME VISNObligation NumberEquipment TypeStationConsolidation PeriodBenchmark (BM) or Technical Specification (TS)
VISN 19: VA Rocky Mountain Network436B00014XR RAD436 MISSOULA, MTMAY 2020TS
VISN 17: VA Heart of Texas Health Care Network504B07002XR PORT C-ARM504 AMARILLO, TXMAY 2020TS
VISN 17: VA Heart of Texas Health Care Network504B07004XR PORT C-ARM504 AMARILLO, TXMAY 2020TS
VISN 17: VA Heart of Texas Health Care Network504B07010XR US504 AMARILLO, TXMAY 2020TS
VISN 17: VA Heart of Texas Health Care Network504B07012XR PORT RAD504 AMARILLO, TXMAY 2020TS
VISN 17: VA Heart of Texas Health Care Network504B07014XR US504 AMARILLO, TXMAY 2020TS
VISN 7: VA Southeast Health Care Network508B02012XR US508 DECATUR, GAMAY 2020BM
VISN 7: VA Southeast Health Care Network508B02017XR US508 DECATUR, GAMAY 2020BM
VISN 7: VA Southeast Health Care Network509B02004XR PORT C-ARM509 AUGUSTA, GAMAY 2020TS
VISN 7: VA Southeast Health Care Network509B03004XR US509 AUGUSTA, GAMAY 2020BM
VISN 7: VA Southeast Health Care Network509B03005XR PORT RAD509 AUGUSTA, GAMAY 2020BM
VISN 7: VA Southeast Health Care Network509B03007XR MAMMO509 AUGUSTA, GAMAY 2020BM
VISN 2: VA Healthcare Network Upstate New York528B00078NM BONE DEN528 BUFFALO, NYMAY 2020BM
VISN 7: VA Southeast Health Care Network534B02016XR US534 CHARLESTON, SCMAY 2020BM
VISN 7: VA Southeast Health Care Network534B02017NM BONE DEN534 CHARLESTON, SCMAY 2020BM
VISN 7: VA Southeast Health Care Network534B02018XR US534 CHARLESTON, SCMAY 2020BM
VISN 7: VA Southeast Health Care Network534B02019NM BONE DEN534 CHARLESTON, SCMAY 2020BM
VISN 10: VA Healthcare System of Ohio541B03011XR RAD541 CLEVELAND, OHMAY 2020BM
VISN 10: VA Healthcare System of Ohio541B03013XR RAD541 CLEVELAND, OHMAY 2020BM
VISN 10: VA Healthcare System of Ohio541B03016XR RAD541 CLEVELAND, OHMAY 2020BM
VISN 10: VA Healthcare System of Ohio541B03019XR PORT RAD541 CLEVELAND, OHMAY 2020BM
VISN 10: VA Healthcare System of Ohio541B03022XR US541 CLEVELAND, OHMAY 2020BM
VISN 10: VA Healthcare System of Ohio541B03024XR RAD541 CLEVELAND, OHMAY 2020BM
VISN 10: VA Healthcare System of Ohio541B03026XR RF541 CLEVELAND, OHMAY 2020BM
VISN 10: VA Healthcare System of Ohio541B03030XR US541 CLEVELAND, OHMAY 2020BM
VISN 10: VA Healthcare System of Ohio541B03031XR US541 CLEVELAND, OHMAY 2020BM
VISN 10: VA Healthcare System of Ohio541B03032XR US541 CLEVELAND, OHMAY 2020BM
VISN 10: VA Healthcare System of Ohio541B03033XR US541 CLEVELAND, OHMAY 2020BM
VISN 10: VA Healthcare System of Ohio541B03035XR US541 CLEVELAND, OHMAY 2020BM
VISN 10: VA Healthcare System of Ohio541B03037XR US541 CLEVELAND, OHMAY 2020BM
VISN 10: VA Healthcare System of Ohio541B93029XR CT541 CLEVELAND, OHMAY 2020BM
VISN 10: VA Healthcare System of Ohio541B93030XR CT541 CLEVELAND, OHMAY 2020BM
VISN 10: VA Healthcare System of Ohio541B93031XR CT541 CLEVELAND, OHMAY 2020BM
VISN 8: VA Sunshine Healthcare Network546B07001XR HYBRID546 MIAMI, FLMAY 2020BM
VISN 10: VA Healthcare System of Ohio553B00035XR CT553 DETROIT, MIMAY 2020BM
VISN 6: VA Mid-Atlantic Health Care Network558B00056XR US558 DURHAM, NCMAY 2020BM
VISN 21: VA Sierra Pacific Network570B00026XR CATH LAB570 FRESNO, CAMAY 2020BM
VISN 12: VA Great Lakes Health Care System578B98075RT LINAC578 HINES, ILMAY 2020BM
VISN 12: VA Great Lakes Health Care System578B98076XR US578 HINES, ILMAY 2020BM
VISN 4: VA Stars & Stripes Healthcare Network595B09000XR US595 LEBANON, PAMAY 2020TS
VISN 9: VA Mid South Healthcare Network603B03028XR RF603 LOUISVILLE, KYMAY 2020BM
VISN 9: VA Mid South Healthcare Network603B03029XR RF603 LOUISVILLE, KYMAY 2020BM
VISN 21: VA Sierra Pacific Network612B09000XR RAD612 REDDING, CAMAY 2020TS
VISN 23: VA Midwest Health Care Network618B99010XR US618 Minneapolis, MNMAY 2020BM
VISN 23: VA Midwest Health Care Network618B99011XR US618 Minneapolis, MNMAY 2020BM
VISN 7: VA Southeast Health Care Network619B92002XR CT619 MONTGOMERY, ALMAY 2020TS
VISN 9: VA Mid South Healthcare Network626B09001XR US626 NASHVILLE, TNMAY 2020BM
VISN 21: VA Sierra Pacific Network640B04003XR US640 PALO ALTO, CAMAY 2020BM
VISN 4: VA Stars & Stripes Healthcare Network642B90029XR US642 PHILADELPHIA, PAMAY 2020BM
VISN 22: VA Southwest Health Care Network644B08040XR US644 PHOENIX, AZMAY 2020BM
VISN 20: VA Northwest Network648B03001XR CATH LAB648 VANCOUVER, WAMAY 2020TS
VISN 22: VA Southwest Health Care Network649B07012NM BONE DEN649 PRESCOTT, AZMAY 2020BM
VISN 21: VA Sierra Pacific Network654B09018XR US654 RENO, NVMAY 2020BM
VISN 21: VA Sierra Pacific Network662B00016XR PORT C-ARM662 SAN FRAN, CAMAY 2020TS
VISN 21: VA Sierra Pacific Network662B00017XR US662 SAN FRAN, CAMAY 2020TS
VISN 21: VA Sierra Pacific Network662B00020XR CT662 SAN FRAN, CAMAY 2020TS
VISN 21: VA Sierra Pacific Network662B00024XR RAD662 SAN FRAN, CAMAY 2020TS
VISN 20: VA Northwest Network663B90037RT LINAC663 SEATTLE, WAMAY 2020BM
VISN 17: VA Heart of Texas Health Care Network671B00015XR US671 SAN ANTONIO, TXMAY 2020TS
VISN 17: VA Heart of Texas Health Care Network671B00022XR US671 SAN ANTONIO, TXMAY 2020TS
VISN 17: VA Heart of Texas Health Care Network671B00052XR US671 SAN ANTONIO, TXMAY 2020TS
VISN 17: VA Heart of Texas Health Care Network674B00184XR US674 TEMPLE, TXMAY 2020TS
VISN 17: VA Heart of Texas Health Care Network674B00185XR US674 TEMPLE, TXMAY 2020TS
VISN 17: VA Heart of Texas Health Care Network674B00244XR PORT RAD674 TEMPLE, TXMAY 2020TS
VISN 8: VA Sunshine Healthcare Network675B07001XR US675 ORLANDO, FLMAY 2020BM
VISN 22: VA Southwest Health Care Network678B09014XR PORT C-ARM678 TUCSON, AZMAY 2020BM
VISN 22: VA Southwest Health Care Network678B09015XR PORT C-ARM678 TUCSON, AZMAY 2020BM
VISN 22: VA Desert Pacific Healthcare Network691B90020XR US691 LOS ANGELES, CAMAY 2020TS
VISN 17: VA Heart of Texas Health Care Network740B00008XR C-ARM740 HARLINGEN, TXMAY 2020TS

Q&As

Requisition NumberQUESTIONSANSWERS
570B00026Please provide the Excel version of the generic technical specification.There was a compliance matrix posted on 8/12/2020 and an updated one will be posted with this amendment
528B00078VAMC Buffalo, NY for a NM Bone Den appears to be a Technical Spec however it does not have a Compliance Matrix. Should there be one?There was a compliance matrix posted on 8/12/2020 and an updated one will be posted with this amendment
General2.If there are additional VA sites that need a vault assessment during this quiet period, what is the process we follow?
a.Do we work with the NAC to schedule those appointments?
b.Can the NAC direct us to the individual sites biomedical team for access to drawings and facilitate that process?2. Site that have LINACs in the current consolidation solication should NOT be connected to conduct vault assessments.

If there are VA sites that need a vault assessment that are NOT part of the current consolidation/solication they can be contacted directly. The VISN Biomedical Engineers can assist with this.

VISN 12 - VA Hines, ILCan we get built drawings for their existing larger vault space? Can the NAC approve an onsite visit to assess that vault and the equipment delivery path for siting our system?Once the vendor selection is annouced the selected vendor can meet with the VA staff on site.
509B03007Is this an upgrade or a net new system?A net new system.
757B09002Please confirm that this was removed from the consolidation.This was removed
757B09002Please confirm that the following requirements have been removed from the consolidation. They were included in the pre-solicitation.This was removed
C-ARMSFor the C-Arms there seems to be different naming conventions. Typically they are the same for Mini C-Arm, such as, ‘XR PORT C-ARM’, all of ours are Mini C-Arms. Which naming convention should we use?Please use XR PORT C-ARM for Mini or full size C-arms
TurnkeyWe would like to confirm that the NAC is accepting construction statement of work proposals (turnkey) after the vendor of choice is awarded and does not need to be submitted as part of this consolidation for obligation # 663B90037 VAMC Puget Sound, WA. Please note, Varian has submitted construction services proposals previously and would need to verify the scope of work before submitting final pricing for construction services.Confirmed. NAC is NOT accepting turnkey quotes during the consolidation/solicitation only once the selected vendor is awarded.
528B00078Buffalo BMD – 1.c. Minimum distance between table and C-arm 24 inches
Question: To allow for full and open competition, please change this minimum specification to 17.7 inches.The minimum distance has been updated to 17.7 inches.
595B09000XR US VAMC Lebanon
Request 2.c. Navigation Software be removed or made optional as this is not typically used in the Pain Clinic.Removed from specification
595B09000Benchmark 1.k.: In what clinical situation will this ultrasound be used away from electrical socket? If it will always be used in a department with electrical connections, we request that this benchmark be removed to allow for competition.Moved to added value section.
508B02018Please confirm that this was removed from the consolidation.Confirmed.
508B02018Please confirm that the following requirements have been removed from the consolidation. They were included in the pre-solicitation.Confirmed
509B03005Please advise if a Keypad or an RFID unit is preferred for security access to the system.Keypad
619B92002VA Montgomery CT, we are asking for additional information to help us value the trade-in:

• Model name

• Model Year

• Serial #

• Generator size

• Year of detector

• Manufacturer of Detector

• Detector serial # Tube count (In exposures or seconds) and gantry count

• Original Year of Manufacture

• Model Size of Detector

• Is the system fully functional?

• Is the system currently in use?

• Overall System Condition (1-10 with 1 being Not Functional and 10 being Like New)

• How is the System Serviced? (OEM Full-Service, Third Party, Time & Material, etc.)

• System Location (i.e. Ground Floor, Basement, Upper Floor, etc.)

• Removal Path (Doors, Elevators, etc.)

In normal times we would have obtained this information through site visits before the quiet period, but that hasn’t been an option due to Covid-19.

Your help is much appreciated. Aquilion 64

JDA0892055

72 KW

Analogic Corp., MA, USA Not Available Tube: 266142 Slices Gantry: 912103 Slices 2017 (Tube)

0.5 mm X 64 Row Yes Yes OEM Full Service Ground Floor 619B92002 Montgomery CT – Section 2.g, 3.e, and 3.f.

Question: The technical spec requires Full UPS and then additional UPS’s for the workstations. Normally you would have one or the other. A Full UPS should cover the entire system; however, the facility needs to have the additional power and facility space to install a large full system UPS. Most customers have a partial system UPS to cover the workstations. Please clarify what is required.

619B92002 Montgomery CT – Question: The CT specification is very simple, many VA hospitals require metal artifact reduction, should this be a requirement for Montgomery? Full UPS and additional UPS's for workstations are required as stated in the specifications.

Metal artifact reduction was added to the specifications

546B07001ami Hybrid – The benchmark identifies a CORE Mobile, VOCOREMB120. This looks like a service contract on a third-party item. Please clarify this requirement.What Siemens did is itemized what include with the installation of the ARTIS icono floor. This is not a mobile unit.
603B03028a.Are there any trade’s associated with these obligations numbers? If so we need make model tube(usage)
b.For both obligations numbers do they want an the elevating table or the doctors prefer a remote room?a. There is no trade-in with the order. It will go in room A-223. b. We want an elevating table
603B03028Louisville RF – Ascend Tabletop Lateral Cassette Holder is included for a wireless detector and then later in the benchmark, it includes a DRX-1 System without detector. The specification also includes a Military Grade FIPS Data in Transit wireless Encryption to detector. It appears the benchmarks doesn’t require a wireless detector, is that correct? How do you plan to use the lateral cassette holder without a wireless detector?At minimum, one (1) wireless detector should be included so that the room is capable of wireless functionality
603B03028Louisville The Digital Subtraction Angiography (DSA) package includes a monitor. The benchmark doesn’t include a cart or a place for the second monitor on the suspension. Where will the monitor be located? Should a cart be added to the proposal?The intention is for both monitors to be ceiling mounted.
603B03029a.Are there any trade’s associated with these obligations numbers? If so we need make model tube(usage)
b.For both obligations numbers do they want an the elevating table or the doctors prefer a remote room?A. Yes, there is a trade-in..This is for room A-218. We will trade-in a 2008 Philips EasyDiagnost Eleva RF system. This system performs 10 fluoro studies a week. B. The site would like an elevating table
603B03029Louisville RF – Ascend Tabletop Lateral Cassette Holder is included for a wireless detector and then later in the benchmark, it includes a DRX-1 System without detector. The specification also includes a Military Grade FIPS Data in Transit wireless Encryption to detector. It appears the benchmarks doesn’t require a wireless detector, is that correct? How do you plan to use the lateral cassette holder without a wireless detector?The Ascend Tabletop Lateral Cassette Holder can be removed from 603-B03029. At minimum, one (1) wireless detector should be included so that the room is capable of wireless functionality.
603B03029Louisville The Digital Subtraction Angiography (DSA) package includes a monitor. The benchmark doesn’t include a cart or a place for the second monitor on the suspension. Where will the monitor be located? Should a cart be added to the proposal?The intention is for both monitors to be ceiling mounted.
541B03011Please provide the location of where this will be usedLorain CBOC
541B03013Please provide the location of where this will be usedWade Park CBOC
541B03016Please provide the location of where this will be usedParma CBOC
541B03022Please provide the location of where this will be usedMedical Service Wade Park CBOC
541B03024What is the model of the GE Rad Room?At minimum, one (1) wireless detector should be included so that the room is capable of wireless functionality.
541B03024Please provide the location of where this will be used4. Yes, have the vendors quote a post-processing workstation.
541B03026What is the model and serial # of the GE Rad Room?Precision RXI S/N: 216VAPXI
541B03026Please provide the location of where this will be used5. Yes, include it if not on the scanner software.
541B03030Please provide the location of where this will be usedRadiology Ultrasound
541B03031Please provide the location of where this will be usedMICU
541B03032Please provide the location of where this will be usedPulmonary
541B03033Please provide the location of where this will be usedMassillon CBOC
541B03035Please provide the location of where this will be usedAkron CBOC Cardiology
541B03037Please provide the location of where this will be usedAkron CBOC Cardiology
541B93029Is a server required for this Requisition?Yes Physical part of the order
541B93031Is a server required for this Requisition?Yes Physical part of the order
553B00035This redacted Siemens benchmark includes this part number and description: 14447460 CT Replacement Confidence 20
Conversion to Siemens SOMATOM Confidence 20. What does conversion mean? Does that mean they included a trade-in, however there are no details tied to a trade-in in the benchmark. If there is a trade-in, please provide specifics tied to that CT.Requisition to be Cancelled
553B00035Is a UPS required for this Requisition?Requisition to be Cancelled
578B98076The benchmark says they want 16 days of applications educational support. Is this a possible typo? Wondering if they wanted an additional 16 hours?16 hours
504B07010XR US VAMC Amarillo
1.Request 1.d. Minimum monitor rotation [degrees] changed from 180 degrees to 170 degrees.
2.Request 1.f Monitor height range from floor [in] be revised from 30” to 50” - 65”.1. Tech specs have been changed to 170 degrees. 2. Tech specs have been changed to 50-65 degrees.
504B07012a.Will both portables be used in the same department? Which department? Will one charger suffice in the quote submission for two units or would you like dedicated chargers for both?

b. Two detectors are requested in the RFP. Is that two per unit, totaling 4 detectors? Or is that two total, one for each portable?

c. If two detectors per portable are requested, the minimum size requirement listed in the RFP is 14x17. For the second detector we more often see a selection of a 10x12. Would that be your preferred size for the second plate or do you want two 14x17 plates?

d. Several vendors offer a “virtual interface” to connect to the hospital EMR and RIS. Will this software be required?

e. What are the different shifts that you need covered during applications go live? A. Units will be used in Radiology. Two chargers needed.

b. 1 detector per unit, size 14x17

c. 1 detector per unit, size 14x17

d. yes; All vendors must provide any VA approved interface to connect to the VA network.

e. 7a – 3:30p; 10a-6:30p 504B07012 1a – Please reduce kW to Minimum of 32kW. There are currently no images that require a 40kW generator, regardless of patient size.

1b – 65 is not a range. Please re-state as a range (ex. 50 – 125).

1c – 500 is not a range. Please provide the required range (ex. 0.2 – 320).

1d – Is this 300 degrees total in rotation from the axis (+150/-150) or +300/-300 degrees of rotation?

1g – Please confirm the requirement is for quantity 2 14”x17” detectors total (1 detector for each unit) and not 2 per unit.

5b – Please explain what information is being requested in “Version/platform long-range plan”. 1a. Accepted, updated.

1b. 50kV – 100kV is acceptable, updated 1c. .2mAs – 500mAs is acceptable, updated 1d. +150/-150 degrees is acceptable, updated 1g. 1 detector per unit.

5b. Version/platform must not be near end of life support, updated.

504B07012Contained within the list of requirements for XR Portable Radiographic system for VISN 17 Amarillo VAMC is the following technical requirement,a.) High frequency generator power is noted as 40kW.
As such, can you please clarify if the use of a 40KW generator is meant to represent a minimum specification? If so, we would respectfully request that this specification be amended to reflect a 32KW generator or above which is in line with industry standards. Reason being, if this most unusual specification is not changed it will result in a lock out spec that will eliminate FMSU and other vendors from being able to respond and compete for this digital portable system. For example, our GO Plus portable system contains a 32KW generator and we have a very large installed base of our GO Plus digital portable systems here in the U.S. and abroad and have never experienced any clinical issues whatsoever concerning the ability to generate the necessary power for any standard and advanced imaging applications with our portable systemTech specs have been updated to 32kW.
504B07012Amarillo Portable Rad – Section 1.a. High frequency Generator Power 40 kW.
Question: If the mA and kV exceed the technical requirements indicated would a 30kW generator be acceptable?Tech specs have been updated to 32kW.
504B07012Amarillo Portable Rad – Section 2.h. Collapsible Column
Question: A Collapsible Column is an option offered by some vendors. Would a secondary monitor on the tower of the portable be acceptable in lieu of a collapsible column and display on tube head? The secondary monitor is much larger than a tube head display which is ideal for physician review and image manipulation.Collapsible column is preferred.
504B07012Amarillo Portable Rad – Section 2.aa. Display monitor at tube head
Question: In order to allow for full competition, please change this spec to read Display Monitor…please remove the Tube Head language.Tech specs have been updated to remove language.
504B07012Amarillo Portable Rad – Section 2 dd. Tube and detector angle alignment software
Question: This is a restrictive requirement only offered by one vendor, please remove this requirement.This has been removed from the requirement.
504B07012Please advise if a Keypad or an RFID unit is preferred for security access to the system.Keypad.
504B07014XR US VAMC Amarillo
1.Request 1.d. Minimum monitor rotation [degrees] changed from 180 degrees to 170 degrees.
2.Request 1.f Monitor height range from floor [in] be revised from 30” to 50” - 65”.
3.Request the 1.j. Minimum Battery Life be removed.
4.Request 2.c. Navigation Software be removed or made optional as most VA sites do not use this in Cardiology.
5.Request 5.g. Shear Wave and Strain-based Elastography be removed or optional as Cardiology Dept do not use Elastography.
6.Request 8.d. 3D Transducer Frequency Range be revised from 5 - 1 to 5 - 1.5.1. Acceptable, updated.

2. Acceptable, updated.

3. 30 minutes is preferable; will be an “exceeds” quality.

4. Moved to optional.

5. Removed.

6. Acceptable, updated.

671B00015XR US VAMC South Texas
1.Request 1.a. Minimum monitor rotation [degrees] changed from 180 degrees to 170 degrees.
2.Request 1.e. Minimum battery life 240 min be revised to 30 min.
3.Request 8.a. Vascular linear Frequency Range from 3 – 11 MHz be revised to 3 to 9 MHz.
4.Request 8.c. Abdominal curvilinear Frequency Range 1 – 8 MHZ be revised to 1 – 6 MHz.
5.Request 8.d. MSK Hockey Stick Frequency Range 10 – 17 MHz be revised to 10 – 15 MHz1. Minimum monitor rotation has been changed to 170 degrees

2. Minimum battery life has been changed to 30 min.

3. Vascular linear probe frequency has been changed to 3 - 9 MHz

4. Abdominal curvilinear probe frequency has been changed to 1 - 6 MHz

5. MSK Hockey Stick probe frequency has been changed to 10 - 15 MHz

671B00022XR US South Texas
1.Request 1.a. Minimum monitor rotation [degrees] changed from 180 degrees to 170 degrees.
2.Request 1.e. Minimum battery life 240 min be revised to 30 min.
3.Request 8.a. Vascular linear Frequency Range from 3 – 11 MHz be revised to 3 to 9 MHz.
4.Request 8.c. Abdominal curvilinear Frequency Range 1 – 8 MHZ be revised to 1 – 6 MHz.
5.Request 8.d. MSK Hockey Stick Frequency Range 10 – 17 MHz be revised to 10 – 15 MHz1. Minimum monitor rotation has been changed to 170 degrees

2. Minimum battery life has been changed to 30 min.

3. Vascular linear probe frequency has been changed to 3 - 9 MHz

4. Abdominal curvilinear probe frequency has been changed to 1 - 6 MHz

5. MSK Hockey Stick probe frequency has been changed to 10 - 15 MHz

671B00052XR US South Texas
1.Request 1.a. Minimum monitor rotation [degrees] changed from 180 degrees to 170 degrees.
2.Minimum equipment dimensions: 14 in. Recommended change to 14.25in
3.Request 2.c. Navigation Software be removed or made optional as this is not typically used in the Cardiology Dept.
4.Request 2.d. Image Fusion Software be removed or made optional as most VA sites do not use this in Cardiology.
5.Request 5.i. Shear Wave and Strain-based Elastography be removed or optional as Cardiology Dept do not use Elastography.
6.Request 8.a. Sector Frequency Range from 1 – 5 MHz be revised to 1.5 to 4.6 MHz.
7.Request 8.b. Non-Imaging Doppler Frequency Range 2.5 – 12 MHZ be revised to 2 MHz.
8.Request 8.c. Transesophageal Frequency Range 2 – 7 MHz be revised to 3 – 7 MHz
9.Request 8.d. Pediatric TEE Frequency Range 3.2 – 7.4 MHz be revised to 4 – 7.4 MHz
10.Request 8.e. Linear Frequency Range 3 – 12 MHz be revised to 3 – 10 MHz1. This does not apply to B00052. The specification has been changed to 170 degrees on B00022 and B00015.

2. Minimum equipment dimensions has been changed to 14.25 in.

3. Navigation software is needed as it provides an opportunity to use the combination of echo and x-ray together in the cath lab to provide an internal picture of the surrounding tissues around the valves before making decisions for surgery.

4. Image Fusion software is needed as it provides an opportunity to use the combination of echo and x-ray together in the cath lab to provide an internal picture of the surrounding tissues around the valves before making decisions for surgery.

5. Shear Wave and Strain-based elastography is needed, since strain is used for patients undergoing chemotherapy. When the heart is affected by the chemo agents, strain provides specific measurements of affected areas of the heart muscle that is affected.

6. Sector probe frequency has been changed to 1.5 - 4.6 MHz

7. Non-Imaging Doppler probe frequency has been changed to 2 MHz

8. TEE probe frequency has been changed to 3 - 7 MHz

9. Pediatric TEE probe frequency has been changed to 4 - 7.4 MHz

10. Linear probe frequency has been changed to 3 - 10 MHz

674B00184XR US Central Texas
1.Request 2.c. Navigation Software be removed or made optional as this is not typically used in the Anesthesia Dept.
2.Request 7.a. 25 mm Linear Array Frequency Range from 13 – 6 MHz be revised to Linear Array 13 to 9 MHz.
3.Request 7.b. 50 mm Linear Array Frequency Range 15 – 6 MHZ be revised to Linear Array 12 – 7 MHz.1.Accepted, updated.
2.Accepted, updated.
3.Accepted, updated.
674B00185XR US Central Texas
1.Request 7.a. 25 mm Linear Array Frequency Range from 13 – 6 MHz be revised to Linear Array 13 to 9 MHz.
2.Request 7.b. 50 mm Linear Array Frequency Range 15 – 6 MHZ be revised to Linear Array 12 – 7 MHz.
3.Request 7.c. 35 mm Curvilinear Array Frequency Range 8 - 3 MHZ be revised to Curvilinear Array 3 -5- MHz.
4.Request 9. Fusion-Targeted Biopsy System be removed or made optional as this is not typically used in the Anesthesia Dept.1.Accepted, updated.
2.Accepted, updated.
3.Accepted, updated.
4.Accepted, updated.
674B002441k – Please remove maximum tube reach. Having a maximum tube reach doesn’t make sense. Having longer tube reach cannot ever hurt. If the site wants a 47” tube reach, then the site can simply extend the tube to 47”. Having a longer tube reach does not negatively impact the user.
1ii – Please remove the requirement or increase to 1200 lbs. There is no technical or facility-specific reason for limiting the weight to 1000 lbs. Minimum requirements for elevator weight capacity is 1800 - 2100 lbs, depending on the size of the facility. The spec is a direct pull from the Shimadzu spec sheet.1.Accepted.
2.Accepted, updated – but less weight is more desirable.
674B00244Central Texas Section 1.a. High frequency Generator Power 42 kW.
Question: If the mA and kV exceed the technical requirements indicated would a 30kW generator be acceptable?Accepted, updated.
674B00244Central Texas Section 1.g. Minimum tube rotation from axis in both direction 360 degrees
Question: In order to allow for full and open competition please change this requirement to 270 degrees.Accepted, updated.
674B00244Please advise if a Keypad or an RFID unit is preferred for security access to the system.Keypad, no RFID unit.
436B000142d – Please change Maximum radiographic mA exposure @100kVp from 630 to 800. 800 is the industry standard for high efficiency generators; this appears to be a vendor-specific offering. There is no reason why a facility would want anything less than what the maximum allowable is (800), as the mA can be reduced to whatever is desired at various kVp.

2k – There is only one known vendor who offers a 100 micron 17”x17” detector, which is unduly limiting on competition. Please re-state the requirement to provide for a range (150um to 100um). As stated, the requirement is that only 100um or below will be accepted, as resolution is measured as the lower the um the more resolution.

4c – Please reduce the Minimum Acquisition Workstation requirement size to 250GB. Given all vendors utilize a first-in-first-out method of clearing HD space, 1TB HD size is unnecessary from a technical and expense perspective, particularly at a CBOC location that is not a major medical center. Further, if a vendor offers 1TB, they can get credit for exceeding the minimum.

Trade-in – Please confirm there is no trade-in 2d: Tech specs have been updated to 800 2K: for image quality we would like 100 micron spec 4c: agree change to 320 GB min New space no trade in

436B00014436 MISSOULA, MT, Section 4.Workstation Requirements Section d. LCD touchscreen acquisition workstation monitor
Question: To allow for open competition please remove the word “touchscreen” from this minimum requirement.Multiple vendors can provide the touchscreen requirement which is required for clinical workflow and quality.
436B00014Line 32 - Minimum spatial resolution of 5 lp/mm

Line 33 - Maximum pixel size of 100 μm

The above are lock-out specs. Please consider revising.for image quality we would like 100 micron spec and 5lp/mm
541B03019Please provide the location of where this will be usedWomen’s health Bldg
648B03001the Station identified by the NAC was Vancouver but the generic benchmark indicates Portland. Can you get us clarification so we don’t create the quote for the wrong site?The correct location is Portland, this is where the EP lab is located.
648B030011.b. PA gantry RAO projection rotation range [deg] 120
Question to allow for full and open competition please change this minimum spec to 117 deg

1.c. PA gantry LAO projection rotation range [deg] 120 Question to allow for full and open competition please change this minimum spec to 105 deg

1.d. PA gantry minimum rotation rate [deg/s] 25 Question to allow for full and open competition please change this minimum spec to 20 deg

1.i. Minimum table height from floor [in] 30.5 Question to allow for full and open competition please change this minimum spec to 30.7 inches

1.k. Table lateral motion range [cm] 36 Question to allow for full and open competition please change this minimum spec to 28cm

1.l. Table longitudinal motion range [cm] 120 Question to allow for full and open competition please change this minimum spec to 170cm

1.m. Minimum patient table load capacity [lbs.] 550 Question to allow for full and open competition please change this minimum spec to 450 lbs.

1.p. Radiographic kV range [kV] 40-125 Question to allow for full and open competition please change this minimum spec to 50-125 kV

1.u. Minimum workstation monitor size [in] 27” Question to allow for full and open competition please change this minimum spec to 19”

1.v. Minimum workstation storage 150 GB Question to allow for full and open competition please change this minimum spec to 1.28TB

2.o. Ceiling mounted equipment rack/boom at foot of Bed Question: The required boom is not part of the imaging system, this should be moved to the extended installation, please remove from this specification.

2. p. Ceiling mounted anesthesia boom available at head of bed.

Question: The required boom is not part of the imaging system, this should be moved to the extended installation, please remove from this specification. 1.b. Revised in spec.

1.c. The LAO rotation range should not be decreased to 105 degrees, as that would significantly limit imaging during procedures. I could be changed to 117 to match RAO project, but not 105.

1.d. Revised in spec.

1.i. Revised in spec.

1.k. Revised in spec.

1.l. Revised in spec.

1.m. The weight capacity cannot be changed, there are too many patients in the 550 lb. range who may require procedures and a 450 capacity would be prohibitive to providing care to all patients.

1.p. Revised in spec.

1.u. Revised in spec.

1.v. Revised in spec.

2.o Revised in spec.

2.p Revised in spec.

570B00026Fresno Cath Lab Section 1.h. h. Table lateral motion range [in] 14
In order to allow for full and open competition, please change this spec to minimum of 11 inches. With a robotic system the movement of the gantry offsets need for extreme lateral motion.Specification was changed to allow for full and open competition. New Table lateral motion range is [in] 11
570B00026Fresno Cath Lab Section 1.m. Radiographic kV range [kV] 40-125
In order to allow for full and open competition, please change this to 50-125kVpSpecification was changed to allow for full and open competition. New Radiographic kV range was changed to 50-125kVp
570B00026Fresno Cath Lab Section 1.o. Fluoroscopic kV range [kV] 40-125 kVp
In order to allow for full and open competition, please change this to 60-120 kVpSpecification was changed to allow for full and open competition. New Fluoroscopic kV range was changed to 60-125kVp
612B090002c – Please define the range (ex. 40-150kV), as 150 kV is not a range.
3u – Please define or further explain “Full articulation”. DR tables generally do not articulate; this is a D/RF function and believe the requirement should be removed.Requirement 2.c. has been changed to 40 – 150kV. Requirement 3.u. “Full articulation” has been removed.
612B09000Redding, 3.t. Ability to lower tabletop height to at least 18.5” from the floor in the horizontal position at stated maximum weight capacity.
Question: This is a redundant requirement to 2.q. Please remove this requirement, or change this one to 20” from the floor so it is the same as 2.q.Requirement 3.t. “. Ability to lower tabletop height to at least 18.5”” has been removed.
612B09000TABLE HEIGHT LOWERS TO FLOOR -Contradicting specs for table height

Question 2.q. is 20" Question 3.t. is 18.5"

Please consider changing to a typical table height of 22" to allow all vendors to compete.Requirement 2.q. “TABLE HEIGHT LOWERS TO FLOOR“ has been changed to 22”.
662B00017What are the procedures being performed with the 3D FreeHand?We’re not clear what 3D FreeHand is, and we don’t currently perform any procedures that we think require this option. Clinical staff are requesting a standard machine for transrectal ultrasound and other ultrasound procedures.
662B00017XR US VAMC San Francisco
1.Request 1.e. Maximum equipment dimension for Depth be increased from 33” to 34.5”
2.Request Item 2.b. Navigation Software be removed.
3.Request Item 2.c. Image Fusion Software be removed.1. Dimension increase is fine. 2. Software can be removed. 3. Software can be removed.
662B000201.Typically 128 slice count is aligned with a CT scanner that has 4 cm of coverage and a 16 cm (160mm) detector has a slice count of over 500 slices. Do you want 4 cm or 16 cm of coverage?
2.Do you want Dual Energy / Spectral imaging capabilities?
3.The bid asks to include Full UPS and later asks for Console UPS for computers. If the intent is going to get a Full UPS, there is no reason to quote/purchase the Console UPS. Do they want a Full or Console UPS.
4.Do they have the ability to post process Cardiac Studies or do they want they vendors to quote a Post Processing Workstation – like Vital.
5.Do they want Calcium Scoring to be included with their Cardiac CTA.1. 16 cm 2. Dual energy is not required 3. Full UPS 4. Yes, post-processing station is required. 5. Yes, include if not on the scanner software.
662B00024San Francisco RAD, Section 4.Workstation Requirements Section d. LCD touchscreen acquisition workstation monitor
Question: To allow for open competition please remove the word “touchscreen” from this minimum requirement.Touchscreen moved to Added Value section
662B00024San Francisco RAD Section 1a., b., c.- 1.a Fixed Wall Digital Detector, 1.b. Fixed Digital Detector, 1.c. Wireless digital detector
Question: Please clarify When the wall stand and/or table detectors can be used as fixed or wireless do you still require a third wireless detector?1.b Fixed Digital Detector specification has been removed