This document is a Combined Synopsis/Solicitation for a commercial item - a sterilizer unit (autoclave) for the Insectary at the San Joaquin Valley Agricultural Sciences Center in Parlier, California. The solicitation is a Request for Quotation (RFQ) with a required delivery date of November 27, 2024. The product specifications call for a 20" x 20" x 38" chamber, single door sterilizer that is freestanding. This is a Total Small Business Set-Aside under NAICS code 339113 with an 800 employee size standard. Offers are due by 12:00 pm Central Time on August 21, 2024 and should be submitted to the email address provided. Questions regarding the solicitation must be submitted in writing by 12:00 pm Central Time on August 19, 2024. The Federal Agency issuing this solicitation is the Department of Agriculture, Agricultural Research Service, Pacific West Area.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 12905B24Q0253.pdf | ||
| 24Q0253 - Combined Syn-SOL - Description Notice - Supply.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Brand Name or Equal
The Commodity Protection and Quality research unit needs a sterilization unit (autoclave) for the Insectary. Through market research we identified Consolidated Sterilizer System that meets our requirements. Use the specifications below to provide a brand name or equal ICP-MS System to continue our research at the San Joaquin Valley Agricultural Sciences Center in Parlier, California.
Specifications for Insectary Sterilization Unit
1. Features CSS 3AV: 20” X 20” x 38” chamber single door sterilizer
• The system should be freestanding and occupy a footprint of >20”and <27”
• X1 Controls: Top Mount 7" Color Touchscreen, up to 50 programmable cycles SteriNET Dataport - USB Password Protection (up to 50 users & passwords) and Lockable Cycle Parameters
• Thermal Printer Controls: 110 Volts, 60 Hz (Standard)
• Service access on left side and limited to 6” rear access
• ProGEN Model Integral Electrically Heated Steam Generator - 25 KW
** Generator Voltage & Phase TBD - Please provide at time of order **
• Automatic Generator Blowdown With Motorized Ball Valve
• Hi-Vacuum System with Water Ejector (MINIMUM 45 psi dynamic water required)
Brass, Bronze, and Copper Piping
• EZ-Glide Vertical Sliding Door with Non-Stick Gasket
• Cabinet Mounting with Stainless Steel Front and Side Panels (Standard)
• WaterEco® Water Conservation & Monitoring System
• Automatic Jacket Blowdown for Liquids Cycle
• Bowie Dick Test and Vacuum Leak Test Whisper-Quiet On-Board Air Compressor
(Supplied if house air is not available) Emergency Stop Button
• 2 shelves – 1 stationary
• Service access on left side and limited to 6” rear access
• Emergency Stop Button
2. Standard Features
• EcoCalendar & Auto Idle Shut-Off capability for energy efficient control of steam
• ASME (American Society of Mechanical Engineers) U-1 Code Stamping
• 1/2" Thermocouple Port UL, cUL Listed
• End-of-Cycle Vacuum Drying
• Fully Opening Front Service Panel
• CSS Starter Kit including C3 cleaner and scrubbing pads, a pair of autoclave gloves, printer paper and printer ribbon (if applicable), and electronic manuals
3. Installation
• The system should be delivered with all necessary supplies and accessories required for the installation and start-up. The installation and training must be included.
• Old system should be de-installed and removed.
4. Delivery
USDA ARS
9611 S. Riverbend
Parlier, CA 93648
By 11/27/2024