This document contains technical specifications for the manufacture of ConFlat vacuum chambers in two sizes: 8 inches and 10 inches. All components must be fabricated from 316LN stainless steel flanges and 316L or 316LN stainless steel tubing, with an electropolished finish suitable for ultra-high vacuum (UHV) service. Welds must be cleaned, and all parts must undergo UHV service cleaning prior to delivery. Dimensional tolerances are set at +/- 0.060 inches with 1-degree angular tolerance on welded dimensions unless otherwise specified in referenced documentation. Government approval of vendor-supplied drawings is mandatory before manufacturing commences.

Each chamber must meet strict performance and quality standards for acceptance. All deliverables must pass leak-rate testing at less than 10^-9 mbar·L/s at NIST facilities. For both the 8-inch and 10-inch ConFlat chamber variants, dimensions, flanges, and welded mounting points must precisely match attached CAD files, with all tolerances conforming to ranges specified in referenced PDF files. These specifications ensure compliance with ultra-high vacuum system requirements for precision vacuum chamber construction.

View the file

Other files

File Type Posted
26-00736 trap_chamber.pdf PDF
Sources Sought Notice.pdf PDF
26-00736 Requirements and Specifications.pdf PDF
26-00736 chamber_tolerances.pdf PDF
26-00736 cryostat_chamber.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Technical requirements:

• Flanges must be ConFlat and made of 316LN stainless steel

• Tubing must be made of 316L or 316LN stainless steel

• Electropolished finish for UHV service

• Welds must be cleaned

• Parts must be cleaned for UHV service

• +/-.060", and 1 degree angular tolerance on all welded dimensions, unless otherwise noted

• Must pass leak check at < 10^-9 mbar * L / s at NIST for acceptance

• Government must approve drawings from vendor prior to manufacture

• 8” ConFlat chamber requirements:

o Dimensions, flanges, and welded mounting points must match attached CAD file(s) o Tolerances must be within stated tolerance ranges in attached pdf file

• 10” ConFlat chamber requirements:

o Dimensions, flanges, and welded mounting points must match attached CAD file(s) o Tolerances must be within stated tolerance ranges in attached pdf file