This document is a Salient Characteristics specification for High Power Direct Current Electronic Loads for the Naval Surface Warfare Center (NSWC) Crane. The specification details comprehensive technical requirements for electronic load units with a power range of 0-10kW, voltage range of 0-600V, and current range of 0-700A. Key requirements include remote and manual control capabilities via Ethernet and front panel, adjustable protection features for overvoltage, overcurrent, and overpower scenarios, and the ability to be paralleled with up to 4 identical units. Additional specifications mandate rack-mountable design (19" server rack, no taller than 7U), air cooling, weight under 160 lbs, and precise output accuracy parameters for voltage, current, and power, with automatic cooling fans and temperature protection.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 25Q0690_SOL_0001.pdf | ||
| 25Q0690_SOL.zip | ZIP file |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Salient Characteristics – High Power Direct Current Electronic Loads
• Power Range of Oto l0kW
• Voltage Range of 0-G00V
• Current Range of 0-700A
• Resistance Range of .67 to 13330
• Ability to output maximum current at a voltage down to 14V
• DC Output Terminals isolated from ground by at least 900 VDC
• Input voltage of lPH 100-240VAC 60Hz
• Ability to conduct constant current, constant voltage, constant current, and constant power discharges
• Must be able of remote control via SCPI commands over Ethernet. At least two Ethernet ports must be included.
• Must be able to support manual control via front panel controls
• Front panel must display output voltage, current, and power.
• Unit must have ability for adjustable overvoltage protection, overcurrent protection, overpower protection up to at least 600V, 700A, and l0kW respectively.
• Unit shall have over temperature protection which shuts off DC output if unit exceeds its safe temperature operating parameters
• Unit must mount within a standard 19" server rack. At least two Rack Mount Kits must be included.
• Unit must be no taller than 7U
• Unit must be capable of being paralleled in master-slave arrangement with up to at least 4 identical units, increasing the power and current capability
• Unit shall be capable of remote sense compensation
• Slew rate shall be l0mA/µs to 20A/µs with a resolution of 5mA/µs.
• Constant voltage output accuracy shall be less than or equal to .025% of set point plus .15V.
• Constant current output accuracy shall be less than or equal to .05% of the set point plus .35A.
• Temperature coefficient shall be less than or equal to 30ppm/!!C.
• Constant power output accuracy shall be less than or equal to .2% of the set point plus 20W.
• In the event of facility power failure, the unit shall disable terminal output.
• The unit shall not consume more than 500VA.
• The unit shall have cooling fans which automatically activate as needed to keep internal temperatures within operational range.
• The unit shall weigh no more than 1601bs.
• The unit shall dissipate all energy into heat, no returning any energy to the grid/facility.
• The unit shall be air cooled. No liquid cooling permitted.