SOW Implant Console South.docx
DOCX document 29 KB Posted
- Solicitation number
- 36C25924Q0335
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 36C25924Q0335 0001.docx | DOCX document | |
| 36C25924Q0335.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Oklahoma City VA Medical Center Oklahoma City, OK 73104 scope of work and GENERAL Specifications FOR equipment PURCHASE
1. Background The Dental Service Department is a vital service within the Oklahoma City VAMC to provide dentistry for our veterans. The new activation of the South OKC Dental clinic will require an implant console.
Project Scope The Oklahoma City VA Medical Center is looking to purchase consoles, foot pedals, motors, and handpieces for the new South OKC Dental Clinic. This is new equipment for a new clinic. Brand name or equal is acceptable see below minimum requirements
2. Specifications The Oklahoma City VAMC Implant system equipment specifications are designed to outline all the Oklahoma City VAMC Dental Service and Biomedical Engineering requirements. These requirements will be provided to the Point of Contract (POC) before final approval. The system must be compatible with ZimVie implants. W&H is what is currently being used with a high level of success. The Implant system equipment specifications are provided below:
· 6.2 Ncm for all bone qualities and 90 ml/min coolant to prevent necrosis during osteotomy.
· Controlled insertion of all types of implants with a torque of up to 80 Ncm.
· User-friendly menu.
· Glass touchscreen surface for ease of cleanability.
· Main Voltage: 120V.
· Maximum torque at motor: 6.3Ncm
· Motor speed range in the rated voltage range: 200-40,000 min-1.
· Coolant flow volume at 100%: at least 90 ml/min
· Dimensions in mm (height x width x depth): 100 x 262 x 291
· Weight in kg: 3.5
· Foot control: wired S-N2 full-function foot control.
· Ability to use straight and contra-angle handpieces.
· Electric motors must be disinfectable and sterilizable.
The C&A requirements do not apply, and a Security Accreditation Package is not required.
3. Delivery Location:
Department of Veteran Affairs Oklahoma City VA Heathcare System ATTN: South Dental Clinic 50 NE 42nd Street Oklahoma City, OK 73105
File details come from the government source that posted it. Updated .