2_Item list and special notes Miltenyi 2161661.xlsx
XLSX spreadsheet 11 KB Posted
- Solicitation number
- RFQ-NIAID-2161661
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 5_TC_FAR PART 52.212-5 DEC 2022.docx | DOCX document | |
| 5_TC_HHSAR 352.222-70 Contrctr Coop in EEO Dec 2015.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Sheet1
| Equipment: | Special Notes: |
| MACSQuant Tyto S/N: 160-002-053 | Equipment is housed in a BSL-3 facility and as such equipment is routinely decontaminated and is decontaminated after spills, splashes, or other potential contamination and before repair, maintenance, or removal from the laboratory. In general, BSL-3 laboratory areas are decontaminated by exposure to paraformaldehyde followed by ammonium bicarbonate neutralization. |
| Maintenance and Service Requirements: | |
| 2-day Priority On-Site Response | |
| All Labor | |
| All travel | |
| All OEM Parts | The decontamination process may shorten the normal lifespan of equipment and accessories. Extra-time and parts are expected, please quote accordingly. |
| Reagent replacement upon HW failures | |
| Vendor coordinates PM on behalf of the lab | |
| Remote technical support 24 hours a day X 5 days a week (M-F) | |
| Only OEM parts are accepted for all repairs and servcies | |
| Software and Hardware coverage updates | |
| Two (2) PM includes: |
- Performance assurance checks
- Functionality and safety checks
- Safety inspections in accordance with all applicable regulations
- Cleaning of operational contaminations
- Cleaning and greasing of mechanical parts
- Examination of all connecting tubes
- Calibration of instrument components
- Optical alignment
- Performance testing with beads
- Testing of components according to Miltenyi SOP BGL-IS-SOP-000026 Routine inspection and replacement of wearing parts