2_Item list and special notes Miltenyi 2161661.xlsx

XLSX spreadsheet 11 KB Posted

Attached to
Maintenance/Service Agreement for Miltenyi Biotec Equipment
Solicitation number
RFQ-NIAID-2161661
Issued by

View the file

Other files

File Type Posted
5_TC_FAR PART 52.212-5 DEC 2022.docx DOCX document
5_TC_HHSAR 352.222-70 Contrctr Coop in EEO Dec 2015.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Sheet1

Equipment:Special Notes:
MACSQuant Tyto S/N: 160-002-053Equipment is housed in a BSL-3 facility and as such equipment is routinely decontaminated and is decontaminated after spills, splashes, or other potential contamination and before repair, maintenance, or removal from the laboratory. In general, BSL-3 laboratory areas are decontaminated by exposure to paraformaldehyde followed by ammonium bicarbonate neutralization.
Maintenance and Service Requirements:
2-day Priority On-Site Response
All Labor
All travel
All OEM PartsThe decontamination process may shorten the normal lifespan of equipment and accessories. Extra-time and parts are expected, please quote accordingly.
Reagent replacement upon HW failures
Vendor coordinates PM on behalf of the lab
Remote technical support 24 hours a day X 5 days a week (M-F)
Only OEM parts are accepted for all repairs and servcies
Software and Hardware coverage updates
Two (2) PM includes:

- Performance assurance checks

- Functionality and safety checks

- Safety inspections in accordance with all applicable regulations

- Cleaning of operational contaminations

- Cleaning and greasing of mechanical parts

- Examination of all connecting tubes

- Calibration of instrument components

- Optical alignment

- Performance testing with beads

- Testing of components according to Miltenyi SOP BGL-IS-SOP-000026 Routine inspection and replacement of wearing parts