This document is a Request for Quote (RFQ) for the procurement of 4 Portable Open-Path Tunable Diode Lasers to measure ammonia and methane concentrations. The solicitation, issued by the USDA Agricultural Research Service (ARS) Coastal Plains Research Center in Florence, SC, seeks 2 instruments to measure ammonia and 2 to measure methane. The technical requirements include the ability to measure concentrations within specific ranges, operate over a path length of 5-500 meters, and have remote diagnostics and data transmission capabilities. The Government will award a Firm Fixed Price contract to the responsible offeror whose quote conforms to the RFQ and specifications using Lowest Price Technically Acceptable source selection. Past performance will be evaluated on an "acceptable" or "unacceptable" basis. Quotes are due by April 12, 2024 at 4:00 PM CST.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Commercial Items Clauses FAC 2020-01 eff. 11-12-19 under $250k 1117271.docx | DOCX document |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Statement of Requirements Project: Upgrades to Existing Open-Path (OP) Ammonia (NH3) and Methane (CH4) Tunable Diode Lasers (TDL)
GENERAL INFORMATION
1.0 Scope of Work: Delivery of 2 OP-TDLs for NH3 and 2 OP-TDLs for CH4 2.0 Background: These instruments replace existing instruments that are obsolete.
CONTRACTOR REQUIREMENTS
3.0 Technical Requirements/Tasks:
Minimum Requirements
1) The OP TDL instrument must be able to measure either NH3 or CH4.
2) OP CH4. TDL must be able to measure CH4.concentration at a minimum of 2 ppm-m; Sensitivity: 0.5 ppm-m; Full Scale: 8,500 ppm-m.
3) OP NH3. TDL must be able to measure NH3 concentration at a minimum of 8 ppm-m; Sensitivity: 2 ppm-m; Full Scale: 6,500 ppm-m.
4) OP TDL path length (separation path) of operation should be between 5 to 500 m.
5) OP TDL must be able to store data for months at a time.
6) OP TDL must be able to have remote diagnostics and data transmission via wireless communications
7) OP TDL must be able to compensate for pressure and temperature from internal pressure and temperature sensors.
8) OP TDL must include Onboard Software (no computer required)
9) OP TDL must have little to no temperature drift from -40 to 50C.
10) OP TDL must have a touchscreen interface with graphical data displays, including spectral waveforms.
11) Instrument operates for weeks with little to no intervention.
4.0 Government Furnished: Not applicable.
5.0 Deliverables/Schedule: Open path instrument to measure NH3 or CH4., sighting scope, alignment mount for instrument on tripod, power supply and cable for instrument, minimum 3-year warranty on vendors manufacture parts, and technical support including remote diagnostic. Package must be delivered to USDA-ARS, Coastal Plains Soil, Water & Plant Research Center, 2611 West Lucas Street, Florence, SC 29501.
6.0 Evaluation Factors: Lowest Price Technically Acceptable