This is a Technical Specifications/Salient Characteristics document for federal contract PR No. 1301376106 outlining requirements for ashing and high temperature furnaces for processing scientific samples.

Two specific furnace models are required. The LHT 03/17 D is a high temperature furnace featuring molybdenum disilicide heating elements with a maximum temperature of 1650°C, an interior volume of 4 liters, maximum exterior dimensions of 450mm depth by 412mm width, and operates on 208V single phase power. The LVT 5/11 is an ashing furnace with a maximum temperature of 1100°C, an interior volume of 5 liters, maximum exterior dimensions of 415mm depth by 385mm width, and operates on 120V power. The ashing furnace includes heating from two sides with ceramic heating plates and integral heating elements, a lift door with the hot surface facing away from the operator for safety, air exchange of more than 6 times per minute, and good temperature uniformity achieved through preheating of incoming air.

View the file

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Ashing and High Temp Furnaces PR No.: 1301376106 Specifications / Salient Characteristics

TECHNICAL SPECIFICATIONS/ SALIENT CHARACTERISTICS

Description of Requirement

Ashing furnace and high temperature furnace for processing scientific samples.

Primary Capabilities/ Salient Characteristics/Technical Specifications:

LHT 03/17 D

High temperature furnace with the following specifications:

molybdenum disilicide heating elements max temperature 1650 C maximum exterior depth 450mm maximum exterior width 412mm interior volume 4 liter operates on 208V single phase

LVT 5/11

Ashing furnace with the following specifications:

maximum temperature 1100C.

Heating from two sides.

Ceramic heating plates with integral heating element.

lift door with hot surface facing away from the operator.

Air exchange of more than 6 times per minute.

Good temperature uniformity due to preheating of incoming air.

interior volume 5 liter maximum exterior depth 415mm maximum exterior width 385mm operates on 120V