Question and Answer

1. Regarding the Appendix A, Kit #11- Eight (8) New Vortex Mixer, page 55. Can INL clarify if the refrigerated cooling characteristic is correct? Since the vortex mixers do not count with refrigerated cooling.

Does not belong to this kit. Please disregard.

2. Regarding the Appendix A, Kit #23- Forty-Eight (48) New UPS device, page 60. Can

INL clarify if the UPS capacity is 15 KVAS or 1.5 KVAS?

15KVA

3. Regarding the Appendix A, Kit #24- Twenty-Two (22) New 0.5-10 UL Micropipettes, page 60. Can INL accept a Single Channel Pipette 1-10 µL?

No, 0.05 range is needed.

4. Regarding Appendix A, Kit #27- Twenty-Two (22) New Multichannel Micropipette 1-

20UL, Biological safety cabinet, page 61. Can INL confirm if the name of the kit

"Multichannel Micropipette 1-20ul, Biological safety cabinet" is correct? As it infers it includes a "Biological safety cabinet ". If the name is correct, this kit will include two items: the micropipette and the safety cabinet.

Only pippets. No biological safety cabinet is needed.

5. Regarding Appendix A, Kit #27- Twenty-Two (22) New Multichannel Micropipette 1-

20UL, Biological safety cabinet, page 61. Can INL accept an 8-channels Pipette 1-10 µL?

1-10UL is not acceptable.

6. Kit # 5, The customer required 5-20 plug type or 5-15 plug type ( this last recommended to México)

5-15 plug

7. For Kit # 9, Do you know the tube’s length needed for the installation of this equipment?

Each laboratory (site) might have different layout. This information might be provided after award as part of pre-installation requirements.

8. For Kit #12 and #13 The equipment that we’ll offer is on 120 V, doesn´t reach 240 V, is this ok for your customer?

120-220V is enough

9. For Kit # 21, What Readability does your customer needs? 0.001mg or 0.01mg

0.01mg

10. For Kit # 27 the multichannel micropipette requires 8 or 12 channels?

Both are acceptable

11. For Kit # 25 and 26, Can we offer kits that include all the ranges, or do you prefer single presentations?

All the ranges are ok

12. Kit # 23, Could you be more specific, what equipment UPSs are required for the

Appendix A?

15KVA

220V capacity

13. For kit # 28, is this for inside (lab) or for outside work?

Indoor work analysis