This document contains the Salient Characteristics for a Hand-Held Fluorometer required under a federal contract opportunity.

The solicitation, number 140G0124Q0269, is for the delivery of a Hand-Held Fluorometer to the U.S. Geological Idaho Water Science Center within 30 days of contract award. The fluorometer must meet specific technical requirements, including the ability to measure green algae, blue-green algae, and diatoms, cable-free operation, an internal GPS, data logging capacity, and an IP68 rating. The anticipated award date is 08/31/2024, and the acquisition is a total small business set-aside under NAICS code 334516 with a size standard of 1000 employees. Vendors must be registered in SAM.gov to be eligible for award. Quotes are due by 08/15/2024, and questions must be submitted to the Contracting Officer by 08/12/2024.

View the file

Other files

File Type Posted
Sol_140G0124Q0269.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Salient Characteristics

S24135

Hand-Held Fluorometer

SALIENT CHARACTERISTICS:

1 ea. FluoroProbe:

1. Shall consist of a field portable, hand-held, fluorometer for real-time measurement of benthic algal concentrations.

2. Shall be capable of measuring green algae, blue-green algae, and diatoms.

3. Shall be capable of cable-free operation.

4. Shall be pre-calibrated and have a measurement range of at least 0-10 μg chl-a/cm2.

5. Shall include an internal GPS for location recording.

6. Shall include an internal display with soft keys.

7. Shall include internal data logging with a capacity of storing at least 2,000 datasets..

8. Shall not require sampling or preparation and be capable of performing a measurement within 20 seconds or less.

9. Shall have a USB communication interface.

10. Shall have an Ingress Protection (IP) rating of IP68, or better.

11. Shall be capable of operating over a temperature range from at least 0 to 400 C.

12. Shall include internal rechargeable batteries.

13. Shall be no larger than 2.5” in diameter x 20” long and weigh no more than 3 lbs.

14. Shall include a power supply, carrying case, manual, and compatible PC operational software.

15. The software shall support various functions including, but not limited to data retrieval and management, displaying of measurement results, graphical displaying of data, and exporting of data.

1 ea. Shoulder Carry Bag:

16. Shall be compatible with the probe described above.

17. Shall consist of a nylon (shoulder style) carry bag that is suitable for transporting the probe and its various accessories.