This Request for Quote (RFQ) No. 1333ND25QNB640405 seeks a specialized Incubator Shaker Growth System with specific technical requirements for scientific research. The requested instrument is a double-stack incubator capable of supporting mammalian cell cultures, with precise environmental controls including temperature (ambient to 60°C, ±0.3°C uniformity), CO2 gassing (0.2-20% with 0.1% accuracy), and humidity (60-90% Rh with 1% accuracy).
The system must feature a 25 mm throw platform for shaking suspended cell cultures, support flask formats from 1ml-5L, and accommodate various clamp and holder configurations with a weight capacity of 20 kg. The required consumables include 3 adhesion mats, an incubator stand, 3 microwell plate holders (ranging from 96 to 6 deep well plates), a water carboy for humidity control, and optional set-up support. The shake speed range is specified as 30-350 rpm, with a slideout platform design for versatile research applications.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 1333ND25QNB640405-RFQ.pdf | ||
| 1333ND25QNB640405 RFQ.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
RFQ NO. 1333ND25QNB640405 MINIMUM SPECIFICATIONS
Contract Line Item Number (CLIN) 0001 shall meet the following minimum requirements:
The Contractor shall provide one Incubator Shaker Growth System with CO2 gassing and humidity control that meets the following minimum specifications with required consumables:
a. The Incubator shall be a double stack incubator.
b. Each of the the incubator in the double stack shall have separate controls.
c. The instrument shall be able to support growth of mammalian cell cultures containing suspended cells.
d. The instrument shall provide tight temperature control from ambient to 60°C with uniformity of
0.3°C in the entire chamber.
e. The instrument shall provide tight CO2 range control from 0.2-20% with an accuracy of 0.1%.
f. The instrument shall provide tight humidity control from 60-90% Rh with accuracy of 1%.
g. The instrument shall be equipped with a 25 mm throw platform for shaking of suspended cell cultures.
h. The instrument shall be able to support various flask formats from 1ml-5L cultures.
i. The instrument shall provide a platform with slideout operation compatible with various clamps and holders for sticky pads and holders and hold up to 20 kg of weight.
j. The instrument shall be capable of shake speed at 25 mm throw orbit from 30 rpm to 350 rpm.
Required Consumables and Set-Up Support:
a. Adhesion mats - 3
b. Stand for incubator
c. Microwell plate holders from 96 deep well to 6 deep well plates - 3
d. Water carboy for humidity control.
e. Set- up support, if needed.