Final SOR Blackbody Calibration Source.pdf

PDF 83 KB Posted

Attached to
BLACKBODY CALIBRATION SOURCE Federal contract opportunity
Solicitation number
1333ND24QNB640596
Issued by
Department of Commerce National Institute of Standards and Technology

About this file

This document is a Statement of Requirements (SOR) for a Blackbody Calibration Source and the associated Federal Contract Opportunity.

The SOR outlines the requirements for a portable blackbody calibration source to be used by the Material Measurement Laboratory at the National Institute of Standards and Technology (NIST) for elevated-temperature mechanical testing experiments. The source must have a temperature range of room temperature to 450°C, an aperture size of 75mm or larger, an effective emissivity of 0.95 or higher, 115V AC input power, and weigh 10kg or less. It must include a certificate of calibration traceable to NIST and a one-year warranty.

The related Federal Contract Opportunity is a combined synopsis/solicitation for the Blackbody Calibration Source, with a NAICS code of 334516 - Analytical Laboratory Instrument Manufacturing. Quotes are due by 10:00am ET on August 29, 2024, and must include a technical description, pricing, and completed provisions/clauses. The Government intends to award a firm-fixed-price purchase order to the responsible contractor submitting the lowest-priced, technically acceptable quote that meets the SOR requirements.

View the file

Other files for this federal contract opportunity

Other files attached to BLACKBODY CALIBRATION SOURCE, newest first.
File Type Posted
Provisions and Clauses Final.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

STATEMENT OF REQUIREMENTS

TITLE: Blackbody Calibration Source

LAB REQUESTING SERVICE: Material Measurement Laboratory, Materials Science and Engineering Division

I. BACKGROUND INFORMATION

The Mechanical Performance Group’s (MPG’s) mission is to develop and advance measurement techniques to study the mechanical behavior of structural materials of interest to US Industry. An essential component of this mission is to understand the behavior of structural materials under elevated temperature conditions to predict material response for advanced manufacturing applications, including additive manufacturing. In response to this measurement need, the MPG developed several elevated-temperature mechanical testing experiments to perform the needed measurements. Elevated temperatures are achieved in some cases using direct-current electrical heating with temperature measurement and control provided through the use of thermal radiation detectors (e.g., pyrometers or thermal imaging devices). Such detectors require calibration using sources of thermal radiation with well-defined, stable temperatures and emissivities, such as blackbody calibration sources. A blackbody calibration source provides thermal radiation with an emissivity close to 1.0 (definition of a blackbody radiator) over a range of controllable temperatures.

II. SPECIFICATIONS

The Contractor shall provide one (1) portable blackbody calibration source with the following requirements:

Blackbody Calibration Source Requirements

• Temperature range: Room Temperature to 450 °C

• Aperture size: 75 mm diameter or larger

• An effective emissivity of 0.95 or higher over the entire temperature range, using the values specified in the technical manual supplied with the instrument (see Deliverables below).

• 115 V AC input power

• Weight: 10 kg or less

• Certificate of calibration traceable to NIST

• One set of technical manuals, either printed or electronic

Warranty:

The Contractor shall provide a minimum one-year warranty for the equipment. The warranty shall cover all parts, labor, and travel and commence upon successful acceptance by the NIST

Subject Matter Expert (SME).

III. DELIVERY

Delivery of the instrument shall be accomplished at NIST, Gaithersburg, MD, during NIST's normal business hours, Monday through Friday, 8:30 a.m. to 5:00 p.m., with the exception of

Federal holidays.

Delivery shall be made to:

NIST

100 Bureau Drive

Gaithersburg, MD 20899-8553

Bldg 223/A121

IV. DELIVERABLES

Description Quantity Recipient Due Date

Portable Blackbody

Calibration Source

One (1) SME 60 calendar days after the award of this requirement.

Operations, Maintenance, and

Technical Manual for the

Instrument (Either printed or

Hard Copy)

One (1) SME 60 calendar days after the award of this requirement

V. ACCEPTANCE

The NIST SME will test the deliverable for compliance with specifications within five (5) calendar days of receipt and provide any comments to the contractor.

File details come from the government source that posted it. Updated .