AQM_REVISED_confocal_Raman_specifications_with-3-lasers_V2.docx

DOCX document 20 KB Posted

Attached to
Confocal Raman Microscope Federal contract opportunity
Solicitation number
12-5680-19-Q-0008
Issued by
Department of Agriculture Forest Service Forest Products Laboratory

About this file

Required Specifications & Best Value Criteria

View the file

Other files for this federal contract opportunity

Other files attached to Confocal Raman Microscope, newest first.
File Type Posted
Confocal_Raman_Microscope_amendment.docx DOCX document
Award_Criteria_Confocal.doc DOC document
confocal_best_value.docx DOCX document

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Required specifications

A new Confocal Raman Microscope system that meets or exceeds the following specifications:

MINIMUM REQUIRED SPECIFICATIONS

(offers made must meet all of the minimum technical specifications below)

TECHNICAL

· Capability of confocal Raman imaging of wood cell walls (in 30 microns or lower thickness sections) with lateral resolution of less than 1 micron and axial resolution of 2 microns at 633 nm laser excitation.

· For 532 and 633 nm excitations, with the lowest g/mm grating, spectral resolution better than 3.9 cm‐1 and for the highest g/mm grating, spectral resolution better than 1.1 cm‐1.

· For the 785 nm excitation, with the lowest g/mm grating, spectral resolution 2.7 cm‐1 and for the highest g/mm grating, spectral resolution 0.7 cm‐1.

· Spectral measurement capability up to 100 cm‐1 from the laser line.

· High through-put confocal spectrometer for the three laser excitations (532, 633, and 785 nm) and gratings.

· The microscope must be equipped with a high definition color camera to visualize the sample under white light illumination and the laser spot simultaneously.

· Optical microscope directly coupled to the spectrometer with turret and at least three microscope objectives – 10X, 50X, and 100X. Adapter rings applicable to the objectives shall be included. Neutral density filter is required for modulating laser power.

· Warranty that includes all parts and labor to be one year from the date of installation.

OPERATING

· Computer system with software for controlling Raman microscope and for data acquisition and processing.

· Multiple Perpetual Licensing minimum requirements

· One to run the instrument

· One for data analysis at user’s computer.

· Capability of the confocal Raman for future upgrade to a Raman/AFM/TERS system.

· Computer controlled grating turret and confocal hole.

TRAINING and ASSISTANCE

· At least one complete set of operating manuals shall be provided at delivery

· On-site training and any necessary calibration required

· User troubleshooting and virtual assistance available for the duration of the warranty

BEST VALUE TECHNICAL SPECIFICATIONS CRITERIA

(in order of importance)

· Piezo-driven scanning platform – mapping stage with at least 100 microns movement in each X and Y directions, and 20 microns in Z directions with 1 nm resolution.

· 532 (power 50 mW), 633 (power 17 mW), and 785 (power 100 mW) nm laser excitation sources.

· Integrated confocal imaging spectrometer with three gratings Preference for 300, 600, and 1200 set or 600, 900, and 1800 g/mm set.

· Efficient (quantum efficiency at least 30% from 500 nm to 800 nm) thermoelectrically cooled (Peltier cooling to approx. – 65 degree C at room temperature) CCD detector (1024 x 256 pixels) with near-IR optimized anti-reflection coating: the detector shall be optimized for high-end confocal Raman imaging.

· Systems that can provide evidence that when cellulose fiber is under tensile strain, the spectrometer shows a shift to lower wavenumber in the cellulose Raman band seen at 1095 cm-1 (in unstrained sample) will receive a higher technical rating.

a. Resulting data analysis, reports or signed statements from stated company officers may all be submit for this requirement.

File details come from the government source that posted it.