1 - Bill of Material for RFQ-NIAID-2155094.xlsx

XLSX spreadsheet 13 KB Posted

Attached to
Brand Name or Equal Leica HistoCore PEGASUS and accessories
Solicitation number
RFQ-NIAID-2155094
Issued by

View the file

Other files

File Type Posted
FAR 52.212-5 DEC 2022.docx DOCX document
HHSAR_352.22-70.pdf PDF
HHSAR 352.232-71 Electronic Submission of Payment Requests (2-2-2022).docx DOCX document

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Sheet1

RFQ-NIAID-2155094
Brand Name or equal
Leica HistoCore Pegasus
https://www.leicabiosystems.com/us/histology-equipment/tissue-processors/histocore-pegasus-iscan/
#DescriptionCatalogQty
1HistoCore PEGASUS140488580051 each
2HistoCore I-Scan kit140488609461 each
3Power cord USA - 20 AMP140411496131 each
4SHIPPING1 each
FOB Shipping destination: Rockville, MD 20852
Warranty Information must be included on quote
Country of Origin must be included for each line item on quote
Equal products must include a brochure that demonstrate in detail being an equal product
Quotes will be evaluated per FAR Part 25
Quote must list items in the exact order as this bill of material
Specifications:
Dual Retort Tissue Processor
Reagent Management System
Automated Reporting
Dimensions must not exceed (W x D x H): 804 x 715 x 1533 mm
Number of reagent containers: 17 reagent bottles, 1 condensate bottle
Bottle volume: 3.88 L
Number of paraffin baths: 4
Capacity: 3.9 L per paraffin bath
Melting time: 7 hours an ambient temperature of 23°C-25°C
Temperature settings range: Paraffin Melting Point setting: 50°C-65°C
Standby temperature setting: Paraffin Melting Point +1°C and max. 71°C
Cassettes Capacity: 2x200 cassettes
Temperature (paraffin): Paraffin temperature selectable from 58°C to 85°C
Temperature (processing reagents): Ambient temperature or 35°C-65°C
Temperature (cleaning reagents): 35°C -85°C
Filling vacuum: -60 kPa
Draining pressure: 45 kPa
Two protocols to run in parallel
Must be compatible with current Leica ASP6025 S tissue processor:
https://www.leicabiosystems.com/us/histology-equipment/tissue-processors/asp6025-s/
52.211-6 Brand Name or Equal.

Brand Name or Equal (Aug 1999)

(a) If an item in this solicitation is identified as "brand name or equal," the purchase description reflects the characteristics and level of quality that will satisfy the Government’s needs. The salient physical, functional, or performance characteristics that "equal" products must meet are specified in the solicitation.

(b) To be considered for award, offers of "equal" products, including "equal" products of the brand name manufacturer, must-

(1) Meet the salient physical, functional, or performance characteristic specified in this solicitation;

(2) Clearly identify the item by-

(i) Brand name, if any; and

(ii) Make or model number;

(3) Include descriptive literature such as illustrations, drawings, or a clear reference to previously furnished descriptive data or information available to the Contracting Officer; and

(4) Clearly describe any modifications the offeror plans to make in a product to make it conform to the solicitation requirements. Mark any descriptive material to clearly show the modifications.

(c) The Contracting Officer will evaluate "equal" products on the basis of information furnished by the offeror or identified in the offer and reasonably available to the Contracting Officer. The Contracting Officer is not responsible for locating or obtaining any information not identified in the offer.

(d) Unless the offeror clearly indicates in its offer that the product being offered is an "equal" product, the offeror shall provide the brand name product referenced in the solicitation.

(End of provision) https://www.leicabiosystems.com/us/histology-equipment/tissue-processors/histocore-pegasus-iscan/https://www.leicabiosystems.com/us/histology-equipment/tissue-processors/asp6025-s/

File details come from the government source that posted it. Updated .