This document is a request for information from the Naval Research Laboratory seeking information on an electrical network analyzer system. The RFI specifies requirements for a millimeter-wave electrical network analyzer system with frequency ranges from 10MHz to 43GHz and extension bands to 75-110GHz and 140-220GHz, along with detailed technical requirements for dynamic range, noise floor, power output, trace noise, reflection and transmission tracking, and temperature stability. Responses are requested by August 24, 2020 for planning purposes, with the objective of identifying potential sources capable of meeting the needs for this system. The anticipated follow-on action is to award a firm fixed price contract for delivery within 6 to 18 weeks of order receipt.
View the file
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
PR# 72-7203-20
SALIENT CHARACTERISTICS
The Naval Research Laboratory (NRL) has a requirement for the following pieces of equipment. The specifications below shall be met in order to be a compatible electrical network analyzer systems. For responses to the RFI, please provide technical backup that the product offered meets all required specifications and is fully compatible electrical network analyzer systems bits.
Millimeter-wave Electrical Network Analyzer System (MENAS)
-Frequency requirements, base unit: 10MHz – 43 GHz; extension bands to 75-110 GHz, and 140-220 GHz
-System dynamic range at 20 GHz, 10 Hz IFBW: 125 dB
-Noise floor at 20 GHz, 10 Hz IFBW: -114 dBm
-Power output at 20 GHz, adjustable; up to or greater than: -14 dBm
-Trace noise, at 10 KHz IFBW Mag (dBrms)/ Phase (degrms): 0.01/0.07
-Test set: 4 ports, with reference mixer switch
(The 4 port capability is only up to 43 GHz, not extension modules)
-Reflection tracking magnitude (10-20 GHz): +/- 0.01
-Reflection tracking phase (10-20 GHz): +/- 0.05
-Transmission tracking magnitude (10-20 GHz): +/- 0.05
-Transmission tracking phase (10-20 GHz): +/- 0.4
-Temperature stability (20 GHz):
Magnitude: 0.017 dB/deg C
Phase: 0.45 deg/deg C
-Standard wall power, 110-120 VAC, 60 Hz operation
Supplied noise figure measurement feature
Spectrum analysis software for multichannel analysis at millimeter-wave frequencies to support frequencies to 120 GHz, minimum.
System shall be capable of making standard scattering parameter measurements, as well as active measurement applications that include swept-frequency gain compression, two-tone IMD, pulse, noise figure measurements for amplifiers and frequency converters
(The Noise Figure, Pulse, IMD, Gain Compression measurements only pertain to the mainframe unit, up to 43 GHz.)
PR# 72-7203-20
SALIENT CHARACTERISTICS
Expansion capability to employ automatic calibration modules
System shall be contained as a desktop or tabletop configuration.