STATEMENT OF WORK

TITLE: Keysight Equipment Calibration

LAB REQUESTING SERVICE: Communications Technology Laboratory, Public Safety Communications Research (671.01)

I. BACKGROUND INFORMATION

PSCR laboratories provide research, development, testing, and evaluation to foster nationwide public safety communications interoperability. In order to be able to accurately test, monitor, and measure LMR and LTE signals within NIST standards, calibrated periodically to manufacturer standards. In order to meet these standards, PSCR must maintain a manufacturer-approved calibration with uncertainties and guardbanding for all existing equipment manufactured by Keysight.

II. SCOPE OF WORK

- Three-year (once per year) manufacturer-approved calibration, uncertainties, and guardbanding service of the government-owned equipment:

• Keysight N9030B Signal Analyzer – Serial# MY57140856 o Required Functional Tests Internal Alignment Effective DANL Option NFE Frequency Response Unpreselected Preamp On o Required Performance Tests with Stated Measurement Uncertainty Frequency Reference Accuracy Power Bandwidth Accuracy Resolution Bandwidth Switching Uncertainty Residual Responses Noise Density Displayed Average Noise Level

• Preamp On

• Preamp Off

• Low Noise Path

Frequency Readout Accuracy Frequency Span Accuracy Count Accuracy IF Frequency Response at 1.825 GHz, 6.0 GHz, 11.0 GHz, 15.35 GHz,

21.80 GHz

• Span 10 MHz

• Span 40 MHz

• Span 85 MHz

Spurious Responses

• First Order Spurs

• Higher Order Spurs

• Image/Multiple/Feedthru Spurs, Standard IF Path

• Image/Multiple/Feedthru Spurs, 85 MHz IF Path

Gain Compression Third Order Intermodulation Distortion Second Harmonic Distortion

• Standard Path

• Low Noise Path

Absolute Amplitude Accuracy

• Preamp Off/Electronic Attenuator Off

• Preamp On

• Electronic Attenuator On

Input Attenuation Switching Uncertainty Display Scale Fidelity Phase Noise Frequency Response 300 kHz to 3.6 GHz Preamp Off Frequency Response 300 kHz to 3.6 GHz AC Coupled Frequency Response 300 kHz to 3.6 GHz Option EA3 Frequency Response 300 kHz to 3.6 GHz Preamp On Frequency Response Above 3.6 GHz Preamp Off Frequency Response Above 3.6 GHz Preamp On Frequency Response Below 300 kHz

• Keysight PNA-X N5249B Network Analyzer Serial# MY57181586 o Required Tests/Measurements with Stated Measurement Uncertainty IF Frequency Response Total Absolute Amplitude Accuracy High Frequency Total Absolute Amplitude Accuracy High Frequency Preamp On Source Power Absolute Amplitude Accuracy Dynamic Accuracy

• Keysight DSOX3034T Oscilloscope Serial# SG56310019

DC Vertical Gain Accuracy DC Dual Cursor Accuracy Bandwidth Time Base Accuracy Internal Trigger Sensitivity External Trigger Sensitivity

• Keysight MSOX6004A Oscilloscope Serial #SG58030024

DC Vertical Gain Accuracy DC Dual Cursor Accuracy Bandwidth

Time Base Accuracy Internal Trigger Sensitivity External Trigger Sensitivity

• Keysight MSOX2014A Oscilloscope Serial #MY58100689 o Required Tests/Measurements with Stated Measurement Uncertainty

DC Vertical Gain Accuracy DC Dual Cursor Accuracy Bandwidth Time Base Accuracy Internal Trigger Sensitivity External Trigger Sensitivity

• Keysight N4001A Noise Source, Serial# SG44420153 o Required Tests/Measurements

Reflection

ENR

Write Calibration Data to SNS

- Calibration reports and measurement reports for all equipment listed above

III. SPECIFICATIONS

See II. Scope of Work.

IV. PERIOD OF PERFORMANCE

Service shall be completed within one week of receipt of equipment by contractor. The government shall schedule the service dates with the contractor within 90 days of contract.

V. PLACE OF PERFORMANCE

All work shall be completed at the contractor facility. Equipment to be upgraded/calibrated will be picked up at the NIST Boulder campus (Colorado) by a courier and shipped to contractor facility at expense of contractor. Contractor will return equipment to NIST Boulder campus at expense of contractor.

VI. DELIVERABLES

Description Quantity Due Date Certificate of Calibration 1 per piece of equipment Completion of service

Measurement Results 1 per piece Completion of service

Description Quantity Due Date of equipment

VII. PERFORMANCE REQUIREMENT SUMMARY

The Contractor shall demonstrate that all performance standards in this statement of work have been met.