This solicitation notice is for an MRI site preparation project at the Oklahoma City VA Medical Center. The scope of work involves demolishing and constructing a revised MRI Suite 1 and Control Room along with adjoining areas. The project period of performance is 300 calendar days from the notice to proceed date. The estimated value is between $1 million to $2 million. This solicitation will be set aside for Service-Disabled Veteran Owned Small Businesses. The work must comply with all Oklahoma VA infection control requirements and MRI Suite 2 must remain operational for staff and patients throughout construction. The full solicitation will be posted on or after October 21, 2020.
View the file
Other files for this federal contract opportunity
Show all 38
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
OKLAHOMA CITY VA HEALTH CARE CENTER
JULY 24, 2020
OKLAHOMA CITY, OKLAHOMA
PROJECT NO. 635-CSI-105
Architectural
Sheet Index:
FourFront Design
517 7th Street
Rapid City, South Dakota 57701
Architectural
G-000 COVER SHEET
G-001 PHASING PLAN
A-100 FLOOR PLANS
A-101 SCHEDULES AND DETAILS
Mechanical
Electrical
E000 ELECTRICAL INFORMATION SHEET
E101 FLOOR PLANS - ELECTRICAL
E102 FLOOR PLANS - ELECTRICAL
M000 MECHANICAL INFORMATION SHEET
M001 MECHANICAL DEMOLITION PLANS
MRI SITE PREPARATION
M101 MECHANICAL PLANS
MRI EQUIPMENT DRAWINGS
A-101 ARCHITECTURAL EQUIPMENT PLAN
A-102 SAFETY AND SERVICE CLEARANCE PLAN
A-501 DETAILS
PROJECT
LOCATION
M102 FLOOR PLAN - PLUMBING
M401 QUENCH VENT DETAILS
M501 MECHANICAL SCHEDULES AND DETAILS
E201 ELECTRICAL DETAILS AND SCHEDULES
(THESE DRAWINGS ARE INCLUDED IN THE CONSTRUCTION
DOCUMENT FOR THE GENERAL CONTRACTORS INFORMATION
AND COORDINATION. SEE THE PHASING PLAN NOTES
INCLUDED ON THE PHASING PLAN)
A-502 DETAILS
S-101 STRUCTURAL FLOOR PLAN
E-101 ELECTRICAL RACEWAY PLAN
E-102 ELECTRICAL DIMENSION PLAN
E-501 DETAIL
M-101 MECHANICAL PLAN
M-501 DETAIL
ATTENTION!
All contractors will be required to comply with the VA provided "MRI
PERSONNEL SCREENING FORM" for each individual working on site, in any capacity. This form will be completed, signed and dated by each individual with a copy turned over to the Resident Engineer, and a copy kept on site with the project Superintendent. Individuals not submitting this form will not be allowed on the site. A copy of this form is included in the
General Requirements section of the specifications. The project
Superintendent will be responsible for verifying daily, all individuals working, or visiting the site are compliant.
ALL CONTRACTORS SHALL BE FAMILIAR WITH THE FOLLOWING
TRAINING VIDEOS:
https://www.youtube.com/watch?v=IZrfDGI1mUO&feature=youtu.be https://www.youtube.com/watch?v=vn5H-6BbesQ&feature=youtu.be https://www.youtube.com/watch?v=C721AzAoNAO&feature=youtu.be
Office of
Construction and Facilities
Management
Project Number
Building Number
Drawing Number
Dwg.
DrawnCheckedDate
Location
Project TitleDrawing Title
Approved: FMS Project Manager/COR
ARCHITECT/ENGINEERS:CONSULTANTS:
DateRevisions:
VA FORM 08-6231
o n e e i g h t h i n c h o n e f o o t o n e q u a r t e r i n c h o n e f o o t t h r e e e i g h t h s i n c h o n e f o o t o n e h a l f i n c h o n e f o o t t h r e e q u a r t e r s i n c h o n e f o o t o n e i n c h o n e f o o t o n e a n d o n e h a l f i n c h e s o n e f o o t t h r e e i n c h e s o n e f o o t
F
E
D
C
B
A
1 2 3 4 5 6 7 8 9
F
E
D
C
B
A
1 2 3 4 5 6 7 8 9
Department of
Veterans Affairs of
100% CONSTRUCTION
07/24/2020
OKLAHOMA CITY, OKLAHOMA
OWNER
635-CSI-105
OKLAHOMA CITY VA HEALTH
CARE SYSTEM MRI SITE PREP
SCHEMMER
1044 NORTH 115TH STREET SUITE 300
OMAHA, NE 68154
517 Seventh Street
Rapid City, SD 57701
(605) 342-9470
(605) 348-0571 (fax)
DOCUMENTS
OKLAHOMA CITY OK. 2ND FLOOR.1
G-000
PROJECT LOCATION PLAN. VA MEDICAL CENTER
AutoCAD SHX Text
FOURFRONTFRONTT
AutoCAD SHX Text
DESIGN INC.C.
AutoCAD SHX Text
CHASE
AutoCAD SHX Text
CHASE
AutoCAD SHX Text
CHASE
AutoCAD SHX Text
CHASE
AutoCAD SHX Text
CHASE
AutoCAD SHX Text
CHASE
AutoCAD SHX Text
(BLOCKED)
AutoCAD SHX Text Cover Sheet and Sheet Index
AutoCAD SHX Text G-000
AutoCAD SHX Text
FOURFRONTFRONTT
AutoCAD SHX Text
DESIGN INC.C.
CONTROL
MRI 1
EQUIPMENT
TOILET
MRI 2
CONTROL
MRI 1
EQUIPMENT
TOILET
MRI 2
CONTROL
MRI 1
EQUIPMENT
TOILET
MRI 2
Phase 1 Plan 1/4"=1'-0"
Phase 2 Plan 1/4"=1'-0"
Phase 3 Plan 1/4"=1'-0"G-001
G-001
G-001
GYP. BOARD PARTITION, EXTENDING UP
TO DECK ABOVE, 1HR. CONSTRUCTION.
NO ACCESS THRU THIS PARTITION.
GYP. BOARD PARTITION FOR INFECTION
CONTROL ACCESS TO WORK AREAS.
GYP. BOARD PARTITION FOR INFECTION
CONTROL ACCESS TO WORK AREAS.
SEAL EXISTING DOOR WITH
PLASTIC SHEET AS REQUIRED
GYP. BOARD PARTITION, EXTENDING UP
TO DECK ABOVE, 1HR. CONSTRUCTION.
PROVIDE ACCESS THRU THIS PARTITION
TO WORK AREA.
PHASE 1
PHASE 2
PHASE 3
GYP. BOARD PARTITION FOR INFECTION
CONTROL ACCESS TO WORK AREAS.
Office of
Construction and Facilities
Management
Project Number
Building Number
Drawing Number
Dwg.
DrawnCheckedDate
Location
Project TitleDrawing Title
Approved: FMS Project Manager/COR
ARCHITECT/ENGINEERS:CONSULTANTS:
DateRevisions:
VA FORM 08-6231
o n e e i g h t h i n c h o n e f o o t o n e q u a r t e r i n c h o n e f o o t t h r e e e i g h t h s i n c h o n e f o o t o n e h a l f i n c h o n e f o o t t h r e e q u a r t e r s i n c h o n e f o o t o n e i n c h o n e f o o t o n e a n d o n e h a l f i n c h e s o n e f o o t t h r e e i n c h e s o n e f o o t
F
E
D
C
B
A
1 2 3 4 5 6 7 8 9
F
E
D
C
B
A
1 2 3 4 5 6 7 8 9
Department of
Veterans Affairs of
100% CONSTRUCTION
07/24/2020
OKLAHOMA CITY, OKLAHOMA
OWNER
635-CSI-105
OKLAHOMA CITY VA HEALTH
CARE SYSTEM MRI SITE PREP
SCHEMMER
1044 NORTH 115TH STREET SUITE 300
OMAHA, NE 68154
517 Seventh Street
Rapid City, SD 57701
(605) 342-9470
(605) 348-0571 (fax)
DOCUMENTS
AutoCAD SHX Text Phasing and Temp Walls Floor Plan
AutoCAD SHX Text G-001
AutoCAD SHX Text
FOURFRONTFRONTT
MRI 1
EQUIPMENT
CHASE
TOILET
MRI 2
D4
COORIDOR
D3
D2
D1 D7
D8
D9
D5
D6
D8
D7
D11
D10
D1 As needed for new equipment
D13
D12D7
D14 D9
D7
Partial
Partial
D17
D16
D14 (Ramp)
D15
D18
REMOVE & REINSTALL EXTERIOR WALL
AS REQUIRED FOR EQUIPMENT INSTALLATION
D4 D17
Remove exg. wall as shown for enlarged EQUIPMENT Room. Salvage exg. CT for reinstallation. See Floor Plan.
D2
D1
DEMOLITION KEY NOTES:
Remove exg. floor and wall tile and ACT ceiling and grid as required for enlarged EQUIPMENT Room. Modify toilet ACT ceiling tile and grid as required for new toilet space. See Floor Plan.
D3 Relocate exg. door, frame and hardware as located on floor plan.
D4 Remove exg. wall as required for relocated door and new temporary door. See Floor Plan.
D5 Remove and reinstall new aluminum window A assembly. Remove and reinstall new RF shielding window A by shielding contractor.
D6 Remove exg. equipment enclosure, including raised floor. Existing MRI equipment will be de-energized, shut down and removed by others.
D7 Remove exg. ACT ceiling and grid.
D8 Remove exg. floor tile. Patch floor for new finish.
D9 Remove exg. Vinyl Base.
Moveable furnishings, equipment and MRI affects will be removed by owner in coordination with general contractors schedule. During construction of temporary construction walls provide temporary tenting of remaining MRI affects.
D11
D10
Remove exg. casework.
All demolition will be coordinated with Phasing Plan and staff / Patient access to MRI rooms.
DEMOLITION NOTES:
1.
D12 Remove exg. shielding door and reinstall new door by shielding system installer.
D13 Remove exg. gyp. bd. and framing to expose exg. shielding system. Shielding system to be removed and replaced by shielding Installer in its entirety.
D14 Remove exg. flooring to expose shielding system. Shielding system to be removed in its entirety by shielding contractor.
D15 Remove existing bench structure in its entirety.
D16 Exg. Clg. Hung Med. Gas to be relocated. See Mechanical.
D17 Modify Exg. Handrail as required for new opening.
D18 Provide demolition as required for installation of new floor sink. See Mechanical.
CHASE
CONTROL
MRI 1
EQUIPMENT
TOILET
COORIDOR
101A 102A 104A 104B
103A
105A
2'-0"
WALL TYPE 2
Match exg. wall structure and finish. Install 3 1/2" sound batt.
A-101
A-101
A-101
V I F
18'-9"
V.I.F.
18'-7
A-101
A-101
A-101
New Ramp by G.C.
See B on A-100
New Med Gas Outlet 5' to center and center on wall, See Mechanical
A-100 B
Note:
All equipment provided by owner, Coordinate installation with provider.
V I F
V I F
8'-0
V.I.F.
1'-6
V.I.F.
V I F
V I F
6'-0"
A
UP
9'-7" V.I.F.
B
A-101
Restructure handrails as needed for door 101A relocation, see detail 6/A-100
Exg. MRI Operations Desk
INTREGAL BASE.
SEE DETAIL 8/A-101
TYPICAL
See Phasing Plan sheet G-001.
GENERAL NOTES:
1.
Relocate exg. sprinkler heads per original location.2.
CHASE
CONTROL
MRI 1
MRI 2
COORIDOR
TOILET
EQUIPMENT
Reinstall remove ACT at new wall to finish as required.
New RF Modular Floor System
1/4" Magnetic Shielding where shown, in other areas 1/4" Cement Board.
Exg Concrete Floor, prep as required for new floor system.
GC. provide floor prep/filler at RF panel joints
Welded Seam Sht Vinyl Floor, Typ.
New RF Modular Floor System
Prep floor to recieve new ramp.
Exg Concrete Floor, prep as required for new flooring.
New Welded Sheet Vinyl Flooring.
Jamb beyond
Lean concrete fill.
Provide 1:24 slope.
18"
101A 102A
Relocated Door 101A
Shorten exg handrail as required to accommodate door relocation.
Provide wall blocking if required.
Modify Exg VB as required to accommodate door relocation.
Perimeter of door
101A Demolition
New handrail provide blocking as needed.
Remove and reinstall new VB this run.
Remove and install new wall fabric
Demo Plan 1/4"=1'-0"A-100
Office of
Construction and Facilities
Management
Project Number
Building Number
Drawing Number
Dwg.
DrawnCheckedDate
Location
Project TitleDrawing Title
Approved: FMS Project Manager/COR
ARCHITECT/ENGINEERS:CONSULTANTS:
DateRevisions:
VA FORM 08-6231
o n e e i g h t h i n c h o n e f o o t o n e q u a r t e r i n c h o n e f o o t t h r e e e i g h t h s i n c h o n e f o o t o n e h a l f i n c h o n e f o o t t h r e e q u a r t e r s i n c h o n e f o o t o n e i n c h o n e f o o t o n e a n d o n e h a l f i n c h e s o n e f o o t t h r e e i n c h e s o n e f o o t
F
E
D
C
B
A
1 2 3 4 5 6 7 8 9
F
E
D
C
B
A
1 2 3 4 5 6 7 8 9
Department of
Veterans Affairs of
100% CONSTRUCTION
07/24/2020
OKLAHOMA CITY, OKLAHOMA
OWNER
635-CSI-105
OKLAHOMA CITY VA HEALTH
CARE SYSTEM MRI SITE PREP
SCHEMMER
1044 NORTH 115TH STREET SUITE 300
OMAHA, NE 68154
517 Seventh Street
Rapid City, SD 57701
(605) 342-9470
(605) 348-0571 (fax)
DOCUMENTS
Floor Plan 1/4"=1'-0"A-100
Reflected Ceiling Plan 1/4"=1'-0"A-100
Typ RF Modular Flooring Detail 3/4"=1'-0"A-100
4 Door Threshold/Ramp Section 3/4"=1'-0"A-100
Partial Wall Elevation 1/4"=1'-0"A-100
AutoCAD SHX Text
NORTH
AutoCAD SHX Text Demo Plan, Floor Plan and Reflected Ceiling Plan
AutoCAD SHX Text A-100
AutoCAD SHX Text
FOURFRONTFRONTT
5/8" TYPE 'X' GYP. BD.
2x4 @ 16"O.C.
TILE (REUSED)
5/8" TYPE 'X' GYP. BD.
3/4" RCI CHANNEL
3 1/2" SOUND BATT.
EQUIPMENT
TOILET
FRAME
101A 102A 103A 104A 104B 105A
EXG 3'-0"x7'-0" EXG 3'-0"x7'-0" NEW 3'-10"x7'-0" EXG 2'-0"x7'-0" EXG 3'-0"x7'-0" EXG 3'-10"x7'-0"
DOOR SCHEDULE
DOOR
HM
HM
HM
HM
HM
HM
PFWD
PFWD
PFWD
PFWD
PFWD
PFWD
RELOCATE DOOR FRAME. DOOR AND HARDWARE.
PROTECT IN PLACE
DEMO / SUPPLY AND INSTALL BY SHIELDING SYSTEM INSTALLER.
PROTECT IN PLACE
PROTECT IN PLACE
PROTECT IN PLACE
DOOR LEGEND:
HM
PFWD
HOLLOW METAL
PRE-FINISHED WOOD DOOR
30"V.I.F.
SUPPORT PNL.
BEYOND
NEW DOOR & HARDWARE TO MRI PER
SHIELDING DESIGN. G.C. PREP NEW
OPENING AS NEEDED FOR
NEW RF FLOOR/RAMP,
PAINT FRAME
NEW RAMP AND FLOORING
BY G.C., SEE B/A-100
EXISTING WINDOW, PAINT
FRAME PER SPECIFICATIONS
SOLID SURFACE TOP
AND BACKSPLASH.
NEW VB, TYP.
NEW GYP. BD.
PAINT
COORD. AVAILABLE
WIDTH WITH MRI
EQUIPMENT
INSTALLATION
30"EQ.EQ.EQ.30"
SUPPORT PNL.
TYP.
SOLID SURFACE TOP
AND BACKSPLASH
3'-0
NEW V.B.
NEW GYP. BD.
PAINT
30"
3'-0
EXG. WINDOW FRAME,
PAINT AS SPECIFIED
EXG. DOOR FRAME,
PAINT AS SPECIFIED
NEW V.B.EXISTING MRI OPERATION
DESK TO REMAIN
COORD. AVAILABLE WIDTH
WITH MRI EQUIPMENT
INSTALLATION
MRI
2x4 WOOD FRAMING @ 16"O.C.
5/8" TYPE 'X' GYP. BD.
VB
NEW RF SHIELDING SYSTEM
EXISTING WALL
3 1/2" SOUND BATT.
INSTALL 5/8" GYP. BD. @ EQUIP. ROOM WALL
6'-0"
SOLID SURFACE COUNTERTOP
AND BACKSPLASH
3'-0
PROVIDE POWER UC AND
ABOVE COUNTER, COOR.
WITH ELECTRICAL
1" CENTER PARTITION
(2) 1" ADJUSTABLE
SHELVES
5'-0"
SOLID SURFACE COUNTERTOP
AND BACKSPLASH
3'-0
1" CENTER PARTITION
(2) 1" ADJUSTABLE
SHELVES
FINISHED END
PANELS, TYP.
NEW V.B.
PROVIDE POWER UC AND
ABOVE COUNTER, COOR.
WITH ELECTRICAL
NOTE:
TYP. BOTH CABINETS, ALL METALIC
HARDWARE TO BE HD, STAINLESS
STEEL HARDWARE.
FINISHED END
PANELS, TYP.
NEW V.B.
NOTE:
CABINETS TYP. 24" DEEP.
ALL METALIC HARDWARE TO
BE BRASS.
NOTE:
EXTEND ALL POWER OUTLET DEVICE BOX
THRU BACK OF BOTH CABINETS.
1/4
7'-0
17'-101 2" M.O.
(F
IE
LD
V
ER
IF
Y)
(F
IE
LD
V
ER
IF
Y)
(FIELD VERIFY)
WINDOW ELEVATIONS
B Remove and reinstall new RF shielding window.
Surface mounted window shade, provided and installed by V.A.
A-101
A Remove and reinstall new aluminum window assembly.
1/4
7'-0
10'-83
8" FRAMED OPENING
(F
IE
LD
V
ER
IF
Y)
(F
IE
LD
V
ER
IF
Y)
(FIELD VERIFY)
Surface mounted window shade, provided and installed by V.A.
A-101
1" Spandrel Panels 1" Insulated Glazing
Glazing as designed by RF Shielding
A B
SOLID SURFACE MID-SILL. PROVIDE SEALANT AT PERIMETER
SHIELDING WINDOW BY RF DESIGN
SEALANT AT PERIMETER
SOLID SURFACE SILL AND APRON
RF SHIELDING SYSTEM
5/8" TYPE X GYP. BD.
TYP.
2x4 STUD WALL, FULL BATT INSULATION
NOTE:
PROVIDE GYP. BD. RETURN AT JAMBS
AND HEAD CONDITIONS.
SHIELDING SYSTEM AND SHIELDING WINDOW
CONNECTION BY RF DESIGN.
EXISTING STRUCTURE
SEALANT AT PERIMETER
SHIM AS REQUIRED
NEW ALUMINUM WINDOW FRAME AND
IG PANEL. SET IN ORIGINAL LOCATION
SPACE
REPAIR IN PLACE THE INTERSITIAL JAMB
AND HEAD GYP. BD. AND FINISHES
4"
PRE-FINISHED METAL
CAP
FILLER
SHEET VINYL FLOOR
& BASE
WALL TYPE 1
1 1/2"=1'-0"
A-101 ELEVATION
1/4"=1'-0"
A-101
ELEVATION
1/4"=1'-0"
A-101
ELEVATION
1/4"=1'-0"
A-101
Office of
Construction and Facilities
Management
Project Number
Building Number
Drawing Number
Dwg.
DrawnCheckedDate
Location
Project TitleDrawing Title
Approved: FMS Project Manager/COR
ARCHITECT/ENGINEERS:CONSULTANTS:
DateRevisions:
VA FORM 08-6231
o n e e i g h t h i n c h o n e f o o t o n e q u a r t e r i n c h o n e f o o t t h r e e e i g h t h s i n c h o n e f o o t o n e h a l f i n c h o n e f o o t t h r e e q u a r t e r s i n c h o n e f o o t o n e i n c h o n e f o o t o n e a n d o n e h a l f i n c h e s o n e f o o t t h r e e i n c h e s o n e f o o t
F
E
D
C
B
A
1 2 3 4 5 6 7 8 9
F
E
D
C
B
A
1 2 3 4 5 6 7 8 9
Department of
Veterans Affairs of
100% CONSTRUCTION
07/24/2020
OKLAHOMA CITY, OKLAHOMA
OWNER
635-CSI-105
OKLAHOMA CITY VA HEALTH
CARE SYSTEM MRI SITE PREP
SCHEMMER
1044 NORTH 115TH STREET SUITE 300
OMAHA, NE 68154
517 Seventh Street
Rapid City, SD 57701
(605) 342-9470
(605) 348-0571 (fax)
DOCUMENTS
WALL TYPE 3
1 1/2"=1'-0"
1A A-101
ELEVATION
1/4"=1'-0"
A-101
ELEVATION
1/4"=1'-0"
A-101
EXTERIOR WINDOW DETAIL
3/4"=1'-0"
A-101
SHEET VINYL & BASE DETAIL
NONE
A-101
AutoCAD SHX Text Details and Schedules
AutoCAD SHX Text A-101
AutoCAD SHX Text
FOURFRONTFRONTT
A. EXISTING CONDITIONS ARE BASED ON DRAWINGS PROVIDED BY
THE OWNER AND LIMITED FIELD VERIFICATION. THE CONTRACTOR
SHALL FIELD VERIFY ALL EXISTING CONDITIONS AND DIMENSIONS
AND NOTIFY THE ARCHITECT/ENGINEER OF ANY DISCREPANCIES
OR VARIATIONS FROM THOSE SHOWN ON THE DRAWINGS. THE
CONTRACTOR SHALL ADJUST TO ACTUAL FIELD CONDITIONS AT NO
ADDITIONAL EXPENSE TO THE OWNER.
B. ALL CONTRACTORS ARE RESPONSIBLE FOR REVIEWING ENTIRE
SET OF DOCUMENTS TO DETERMINE THEIR FULL SCOPE OF WORK.
CONTRACTOR SHALL NOT BE ALLOWED EXTRA COSTS DUE TO
FAILURE TO REVIEW ENTIRE SET OF DOCUMENTS.
C. CONTRACTOR SHALL TAKE ALL MEASUREMENTS FOR WORK AND
BE RESPONSIBLE FOR SAME. CONTRACTOR SHALL ADJUST FOR
ACTUAL FIELD CONDITIONS AT NO ADDITIONAL EXPENSE TO
OWNER. COORDINATE THE WORK AND SHOP DRAWINGS WITH ALL
OTHER TRADES AFFECTED.
D. SCREENED HALF-TONE LINES INDICATE EXISTING CONSTRUCTION
TO REMAIN. FULL-TONE LINES INDICATED NEW OR RELOCATED
CONSTRUCTION.
E. CONTRACTOR SHALL COORDINATE THE ROUTING OF ALL
DUCTWORK WITH STRUCTURE, CEILING, LIGHTING, AND PIPING.
F. MAINTAIN MANUFACTURER'S REQUIRED CLEARANCE AROUND ALL
EQUIPMENT TO ALLOW FOR MAINTENANCE ACCESS.
G. ALL PIPING IN CONCEALED LOCATIONS SHALL BE ROUTED AT
RIGHT ANGLES AND PARALLEL TO BUILDING WALLS UNLESS
OTHERWISE INDICATED.
H. DO NOT ROUTE DUCTWORK OR PIPING ABOVE ELECTRICAL
PANELS. MAINTAIN ALL CODE REQUIRED CLEARANCES.
I. IN THE EVENT THE CONTRACTOR ENCOUNTERS ON THE SITE,
MATERIAL REASONABLY BELIEVED TO BE ASBESTOS, LEAD-BASED
PAINT, OR ANY OTHER HAZARDOUS MATERIAL WHICH HAS NOT
BEEN RENDERED HARMLESS, THE CONTRACTOR SHALL
IMMEDIATELY STOP WORK IN THE AREA AFFECTED AND REPORT
THE CONDITION TO THE OWNER AND ARCHITECT IN WRITING. THE
WORK IN THE AFFECTED AREA SHALL NOT THEREAFTER BE
RESUMED EXCEPT BY WRITTEN AGREEMENT OF THE OWNER AND
CONTRACTOR.
J. ANY UTILITY OR SYSTEM DISRUPTIONS OR OUTAGES DURING
CONSTRUCTION SHALL BE COORDINATED WITH THE OWNER AND
GENERAL CONTRACTOR.
A. THE DEMOLITION PLANS WERE PRODUCED BY DUPLICATING THE ORIGINAL
DRAWINGS PROVIDED BY THE OWNER AND LIMITED FIELD VERIFICATION.
DIMENSIONS AND EXISTING CONDITIONS ARE APPROXIMATE. CONTRACTOR
SHALL FIELD VERIFY EXISTING CONDITIONS. ITEMS WHICH ARE INTENDED
TO BE REMOVED, RELOCATED, OR SALVAGED ARE SHOWN AS SOLID, FULL-
TONE LINES. ALL OTHER ITEMS ARE INTENDED TO REMAIN IN PLACE AND
ARE SHOWN AS HALF-TONE.
B. DEMOLITION SHALL INCLUDE, BUT IS NOT LIMITED TO, THE ITEMS LISTED
BELOW OR HERE-IN. THE CONTRACTOR SHALL COORDINATE ALL
REQUIRED RENOVATION AND NEW CONSTRUCTION WITH THE EXISTING
BUILDING TO IDENTIFY THE TOTAL EXTENT OF THE DEMOLITION REQUIRED
AND AS LISTED BELOW.
C. THE REMOVAL OF HVAC AND PLUMBING EQUIPMENT FROM FINISHED AREAS
AND THROUGH WEATHER BARRIERS SHALL BE COORDINATED WITH
ARCHITECTURAL PLANS AND DETAILS AS WELL AS NOTES ON THE
MECHANICAL DRAWING TO ENSURE PROPER PATCHING AND REPAIR IS
MADE TO THE BUILDING.
D. THE CONTRACTOR SHALL BE RESPONSIBLE FOR THE DEMOLITION AND
REMOVAL OF ALL EXISTING COMPONENTS, MATERIALS, EQUIPMENT, AND
APPURTENANCES AS REQUIRED TO BUILD, ERECT, INSTALL, OR
ACCOMMODATE ALL NEW CONSTRUCTION WITH THE OWNER HAVING FIRST
RIGHT OF REFUSAL ON ALL ITEMS.
E. ITEMS NOTED TO BE REMOVED AND SALVAGED OR REINSTALLED SHALL BE
CAREFULLY REMOVED BY THE CONTRACTOR WITHOUT DAMAGE AND
STORED OR REINSTALLED ON THE SITE AS DIRECTED. REMOVED OR
SALVAGED ITEMS SHALL REMAIN THE PROPERTY OF THE OWNER.
F. COORDINATE THE DISCONNECTION OF ELECTRICAL MOTORS AND
CONTROLS WITH ELECTRICAL CONTRACTOR PRIOR TO MECHANICAL
REMOVALS.
G. IN THE EVENT THE CONTRACTOR ENCOUNTERS ON THE SITE, MATERIAL
REASONABLY BELIEVED TO BE ASBESTOS, LEAD-BASED PAINT, OR ANY
OTHER HAZARDOUS MATERIAL WHICH HAS NOT BEEN RENDERED
HARMLESS, THE CONTRACTOR SHALL IMMEDIATELY STOP WORK IN THE
AREA AFFECTED AND REPORT THE CONDITION TO THE OWNER AND
ARCHITECT IN WRITING. THE WORK IN THE AFFECTED AREA SHALL NOT
THEREAFTER BE RESUMED EXCEPT BY WRITTEN AGREEMENT OF THE
OWNER AND ARCHITECT.
H. ANY UTILITY OR SYSTEMS DISRUPTION OR OUTAGES DURING DEMOLITION
SHALL BE COORDINATED WITH THE OWNER AND GENERAL CONTRACTOR.
FOURFRONT DESIGN, INC.
517 7TH STREET
RAPID CITY, SOUTH DAKOTA
57701
PH: (605) 342-9470
FAX: (605) 342-2377
WWW.FOURFRONTDESIGN.COMFOURFRONT
D E S I G N I N C.
Office of Construction and Facilities Management
Project Number
Building Number
Drawing Number
Dwg.
DrawnCheckedDate
Location
Project TitleDrawing Title
Approved: Project Director
ARCHITECTCONSULTANTS:
DateRevisions:
VA FORM 08-6231
o n e e ig h th in c h o n e fo o t o n e q u a rt e r in c h o n e fo o t th re e e ig h th s in c h o n e fo o t o n e h a lf in c h o n e fo o t th re e q u a rt e rs in c h o n e fo o t o n e in c h o n e fo o t o n e a n d o n e h a lf in c h e s o n e fo o t th re e in c h e s o n e fo o t
F
E
D
C
B
A
1 2 3 4 5 6 7 8 9
F
E
D
C
B
A
1 2 3 4 5 6 7 8 9
Department of Veterans Affairsof
635-CSI-105
SCHEMMER
1044 NORTH 115TH STREET, SUITE 3300
OMAHA, NEBRASKA 68154
/2
/2
:1
:4
A
M
07/22/20
MECHANICAL INFO SHEET
M 00
OKLAHOMA CITY VA HEALTH
CARE SYSTEM MRI SITE PREP
AuthorChecker
OKLAHOMA CITY, OKLAHOMA
OWNER
07868.001
TRUE
NORTH
PLAN
NORTH
NORTH
NORTH
ARROW
SYMBOL LEGEND
MATCHLINE
AREA A
AREA B
EXISTING
COLUMN
NEW
COLUMN
X
X
X
XX-XX
SIM
X
XX-XX
XXXX
DX
X
SECTION/
DETAIL
PLAN
DETAIL
REFERENCE
NOTE
REFERENCE
DEMOLITION
NOTE
EQUIPMENT
NUMBER
ROOM / SPACE
NUMBER
LETTER INDICATES
NEW COLUMN LINE
LETTER INDICATES
EXISTING COLUMN LINE
SECTION/DETAIL NUMBER
SHEET NUMBER
INDICATES DIRECTION
OF VIEW
DETAIL NUMBER
SHEET NUMBER
REFERENCE
DESIGNATION
DEMOLITION
DESIGNATION
EQUIPMENT
NUMBER
REFERENCE LEGEND
NAME
ELEVATION
LEVEL
TITLE
MARKER
VIEW NAME
SCALE: 1/8" = 1'-0"
DETAIL NUMBER
LEVEL CALLOUT
DESIGNATION
ELEVATION
DESIGNATION
GRAPHIC AREA FOR
DRAWING
TITLE
SCALE OF THE DRAWING
RECIRCULATING DOMESTIC
HOT WATER (RDHW)
DOMESTIC COLD WATER (DCW)
DOMESTIC HOT WATER (DHW)
PIPE GUIDE
HOSE BIBB
THERMOMETER
PIPE ANCHOR
HB
STEAM TRAP
PH
WH
YH
SIGHT GLASS-REFRIGERANT
POST HYDRANT
YARD HYDRANT
WALL HYDRANT
PRESSURE GAUGE
CA
PN
CONTROL AIR SUPPLY
NEW TO EXISTING CONNECTION
PNEUMATIC TUBE
FIRE WATER (BURIED)
SPRINKLER DRAIN
FIRE SPRINKLER MAIN
TAMPER SWITCH
FLOW SWITCH
FIRE DEPARTMENT CONNECTION
MISCELLANEOUS SYMBOLS
FLOOR SINK
AREA DRAIN
FLOOR DRAIN
GRADE CLEANOUT
PLUMBING VENT THRU ROOF
CLEANOUT / WALL CLEANOUT
MANHOLE
TRENCH DRAIN
OVERFLOW ROOF DRAIN
RUNNING TRAP
OXYGEN OUTLET
PIPE CAP
ELBOW UP
AIR OUTLET
TEE
ELBOW DOWN
VACUUM OUTLET
RUNNING TRAP WITH CLEANOUTS
ROOF DRAIN
UNION
TEE DOWN
STRAINER
DIELECTRIC UNION
THERMOMETER WELL
CONCENTRIC REDUCER
EXPANSION JOINT
FLEXIBLE PIPE CONNECTION
ECCENTRIC REDUCER
AIR VENT (MANUAL UNLESS
NOTED)
PITCH PIPE DOWN IN DIR OF
ARROW
TEE UP
SM
FW
D
TS
FS
LPC
HPS
MPC
MPS
LPS
HVAC PIPING
GHWR
GHWS
HWR
HWS
PC
HPC
ES
BBD BOILER BLOW DOWN
EXHAUST STEAM
PUMPED CONDENSATE
HIGH PRESSURE CONDENSATE
HIGH PRESSURE STEAM
MEDIUM PRESSURE CONDENSATE
MEDIUM PRESSURE STEAM
LOW PRESSURE CONDENSATE
LOW PRESSURE STEAM
GLYCOL HOT WATER RETURN
GLYCOL HOT WATER SUPPLY
HOT WATER RETURN
HOT WATER SUPPLY
_"FS-_
_"AD-_
_"FD-_
CO
VTR
GCO
_"OD-_
MH
X
A
V
_"RD-_
RD
SAN
SAN
PLUMBING PIPING
ACID VENT
TW
AV
AW
W
DS
LS
SCW
RTW
SHW
V SANITARY VENT
BURIED EXTERIOR WATER
RECIRCULATING TEMPERED
WATER
TEMPERED WATER
DISTILLED WATER
LAWN SPRINKLER
SOFTENED HOT WATER
SOFTENED COLD WATER
ACID WASTE
SUBSOIL OR FOUNDATION DRAIN
STORM DRAIN (BELOW FLOOR)
STORM DRAIN (ABOVE FLOOR)
SANITARY DRAIN (BELOW FLOOR)
SANITARY DRAIN (ABOVE FLOOR)
OXYGEN
GASOLINE
VACUUM
OX
GS
G
VAC
RO
A
FOF
FOG
LPG
FOV
FOR
FOS
FUEL OIL RETURN
FUEL OIL SUPPLY
FUEL OIL GAGE
FUEL OIL FILL
FUEL OIL VENT
REVERSE OSMOSIS PROCESSED
WATER
COMPRESSED AIR
NATURAL GAS
LIQUEFIED PETROLEUM GAS
DEIONIZED WATERDI
PIPING SPECIALTIES AND FITTINGS
WCO
FIRE PROTECTION
XXX
PLUMBING
FIXTURES PLUMBING FIXTURE
DESIGNATION
X"PIPE
SIZE
PIPE SIZE (DIAMETER)
X"/X"
DUCT
SIZE
WIDTH / HEIGHT
REF. DEMO NOTES ( ) XX
CWS
GCWS
CWR
GCS
CR
CS
RL
RD
RS
GCWR
GLYCOL CONDENSER WATER
SUPPLY
GLYCOL CHILLED WATER RETURN
GLYCOL CHILLED WATER SUPPLY
REFRIGERANT LIQUID LINE
CHILLED WATER RETURN
CHILLED WATER SUPPLY
CONDENSER WATER RETURN
CONDENSER WATER SUPPLY
REFRIGERANT DISCHARGE
REFRIGERANT SUCTION LINE
GHCS
GHCR
CD
HCS
HCR
GCR
HOT/CHILLED WATER SUPPLY
HOT/CHILLED WATER RETURN
GLYCOL CONDENSER WATER
RETURN
GLYCOL HOT/CHILLED WATER
RETURN
GLYCOL HOT/CHILLED WATER
SUPPLY
CONDENSATE DRAIN
REF. NOTES ( )ABBREVIATIONS
ROOM NAME
150 SF
ROOM NAME
DESIGNATION
ROOM NUMBER
DESIGNATION
ROOM AREA
DESIGNATION
TSTAT THERMOSTAT
TWU TROUGH WALL AIR CONDITIONER
UH UNIT HEATER
UR URINAL
US UTILITY SINK
V VENT
VAC VACUUM
VAV VARIABLE AIR VOLUME
VD VOLUME DAMPER
VEL VELOCITY
VFD VARIABLE FREQUENCY DRIVE
VTR VENT THRU ROOF
WB WET BULB
WC WATER CLOSETS
WCO WALL CLEANOUT
WF WASH FOUNTAINS
WFU WATER FIXTURE UNITS
WH WALL HYDRANT
WPD WATER PRESSURE DROP
WS WATER SOFTENER
WSHP WATER SOURCE HEAT PUMP
WWHP WATER TO WATER HEAT PUMP
YH YARD HYDRANT
GPH GALLONS PER HOUR
GRV GRAVITY RELIEF VENTILATOR
GPM GALLONS PER MINUTE
GV GAS VENT
GWH GAS WATER HEATER
H HYDRANTS
HB HOSE BIBB
HC HEATING COIL
HCR HOT/COLD WATER RETURN
HCS HOT/COLD WATER SUPPLY
HG GAS: HIGH-PRESSURE
HP HEAT PUMP (AIR SOURCE)
HP HORSEPOWER
HPC HIGH PRESSURE CONDENSTATE
HPS HIGH PRESURE STEAM
HRC HEAT RECOVERY COIL
HU HUMIDIFIER
HWR HOT WATER RETURN (HEATING)
HWS HOT WATER SUPPLY (HEATING)
HX HEAT EXCHANGER
IH INTAKE HOOD
IRH INFRARED RADIANT HEATER
KDH KITCHEN DISHWASHING HOOD
KH KITCHEN HOOD
KW KILOWATT
KWH KILOWATT HOURS
L LAVATORY
LAT LEAVING AIR TEMPERATURE
LH LAB HOOD
LPC LOW PRESSURE CONDENSATE
LPG LIQUIFIED PETROLEUM GAS
LPS LOW PRESSURE STEAM
LRA LOCKED ROTOR AMPS
LS LAUNDRY SINK
LSS LAWN SPRINKLER DESIGN
LV LOUVER
LWT LEAVING WATER TEMPERATURE
MAU MAKE-UP AIR UNIT
MAX MAXIMUM
MB MACHINE BOLT
MBH 1000 BTUH
MCA MINIMUM CIRCUIT AMPS
MD MOTORIZED DAMPER
MG GAS: MEDIUM-PRESSURE
MIN MINIMUM
MOCP MAXIMUM OVERCURRENT
PROTECTION
MPC MEDIUM PRESSURE
CONDENSATE
MPS MEDIUM PRESSURE STEAM
MS MOP SINK
N NITROGEN
NC NORMALLY CLOSED
NG NATURAL GAS
NO NORMALLY OPEN
NPCW NONPOTABLE COLD WATER
NPHW NONPOTABLE HOT WATER
NPHWR NONPOTABLE HOT WATER
RETURN
OA OUTSIDE AIR
OD OVERFLOW ROOF DRAIN
OX OXYGEN
P PUMP
PC POINT OF CURVATURE
PHC PREHEAT COIL
PIV POST INDICATING VALVE
PRV PRESSURE REGULATING VALVE
PS PLASTER SINKS
PSF POUNDS PER SQUARE FOOT
PSI POUNDS PER SQUARE INCH
PSIA POUNDS PER SQUARE INCH,
ABSOLUTE
PSIG POUNDS PER SQUARE INCH,
GAUGE
PTAC PACKAGED TERMINAL AIR
CONDITIONER
RA RETURN AIR
RCP RADIANT CEILING PANELS
RD ROOF AREA DRAINS
RDL REFRIGERANT DISCHARGE LINE
RF RETURN FAN
RH RELATIVE HUMIDITY
RHC REHEAT COIL
RL REFRIGERANT LIQUID LINE
RLA RELIEF AIR
RLF RELIEF FAN
RO REVERSE OSMOSIS WATER
RPM REVOLUTIONS PER MINUTE
RS REFRIGERANT SUCTION LINE
RTU ROOF TOP UNIT
RV RELIEF VALVE
S SINKS
SA SUPPLY AIR
SAN SANITARY SEWER
SB SITZ BATH
SD SMOKE DAMPER
SF SUPPLY FAN
SFU SANITARY FIXTURE UNITS
SH SHOWERS
SHB SHAMPOO BOWLS
SP STATIC PRESSURE
SS SERVICE SINK
SSCU DUSTLESS SPLIT SYSTEM
CONDENSING UNIT
SSFC DUCTLESS SPLIT SYSTEM
SSS SURGEON SCRUB SINKS
ST STORAGE TANK
SUP SUMP PUMP
SW STORM WATER
TD TRENCH DRAIN
TMV THERMOSTATIC MIXING VALVES
TONS TONS OF REFRIGERATION
TSP TOTAL STATIC PRESSURE
APD AIR PRESSURE DROP
A COMPRESSED AIR
AFCV AUTO FLOW CONTROL VALVE
AC AIR COMPRESSOR
AD AIR DOOR
AW ACID WASTE
AV ACID VENT
AS AIR SEPARATOR
AW AREA DRAIN
ARD AIR PRESSURE DROP
AMP AMPERE
AHU AIR HANDLING UNIT
BW BEDPAN WASHERS
BV BALL VALVE
BT BATHTUBS
BLV BALANCING VALVE
BLKG BLOCKING
BHP BREAK HORSEPOWER
BFV BUTTERFLY VALVE
BF BACKFLOW PREVENTER
BD BIDET
BC BLOWER COIL
B BOILER
CWS CHILLED WATER SUPPLY
CWR CHILLED WATER RETURN
CU AIR COOLED CONDENSING UNIT
CT COOLING TOWER
CSS CLINICAL SERVICE SINKS
CS CONDENSER WATER SUPPLY
CRAC COMPUTER ROOM AIR
CONDITIONER
CR CONDENSER WATER RETURN
CMU COMBINATION UNITS
CI CAST IRON
CHV CHECK VALVE
CHIU CABINET HEATER
CFM CUBIC FEET PER MINUTE
CF CEILING FANS
CD CONDENSATE DRAIN
CC COOLING COIL
C CHILLER
D DIFFUSER
DB DRY-BULB TEMPERATURE
DCW DOMESTIC COLD WATER
DEG DEGREE
DF DRINKING FOUNTAIN
DGCO DOUBLE GRADE CLEANOUT
DH EMERGENCY DRENCH HOSES
DHW DOMESTIC HOT WATER
DHWR DOMESTIC HOT WATER RECIRC
DI DEIONIZED WATER
DOA DEDICATED OUTDOOR AIR UNIT
DP DOMESTIC PUMP
DS DISTILLED WATER
DSA DUCT SOUND ATTENUATORS
DSCW DOMESTIC SOF COLD WATER
DSHW DOMESTIC SOFT HOT WATER
DSHWR DOMESTIC SOFT HOT WATER
RETURN
DST DOMESTIC STORAGE TANK
DTW DOMESTIC TEMPERED WATER
DTWR DOMESTIC TEMPERED WATER
RETURN
DXT DOMESTIC EXPANSION TANK
EA EACH
EAT ENTERING AIR TEMPERATURE
EDH ELECTRIC DUCT HEATER
EEW EMERGENCY EYE WASH
EEWS EMERGENCY EYE
WASH/SHOWERS
EF EXHAUST FAN
EFF EFFICIENCY
EFW EMERGENCY FACE WASH
EH EXHAUST HOOD
ERV ENERGY RECOVERY
VENTILATOR
ES EMERGENCY SHOWER
ESP EXTERNAL STATIC PRESSURE
EWC ELECTRIC WATER COOLER
EWH ELECTRIC WATER HEATER
EWT ENTERING WATER
TEMPERATURE
EX EXPANSION TANK
F FIRE PROTECTION WATER
SUPPLY
F FURNACE
FCO FOOT CANDLE
FCT FLUID COOLER TOWERS
FCU FAN COIL UNIT
FD FIRE DAMPER
FLA FULL LOAD AMPS
FOR FUEL OIL RETURN
FOS FUEL OIL SUPPLY
FOV FUEL OIL VENT
FPM FEET PER MINUTE
FPS FEET PER SECOND
FR FLOOR SINKS
FSD FIRE/SMOKE DAMPER
FTR FINNED TUBE RADIATION
G GAS: LOW-PRESSURE
GCR GLYCOL CONDENSER WATER
RETURN
GCS GLYCOL CONDENSER WATER
SUPPLY
GCWR GLYCOL CHILLED WATER
RETURN
GCWS GLYCOL CHILLED WATER
SUPPLY
GD GARBAGE DISPOSAL
GD GRAVITY DAMPER
GDH GAS DUCT HEATER
GHCR GLYCOL HOT/COLD WATER
RETURN
GHCS GLYCOL HOT/COLD WATER
SUPPLY
GHWR GLYCOL HOT WATER RETURN
GHWS GLYCOL HOT WATER SUPPLY
GI GREASE INTERCEPTOR
GIV GRAVITY INTAKE VENTILATOR
GPD GALLONS PER DAY
18 Ø
18x10
18/10
ROUND DUCT
VD
FLAT OVAL DUCT-1ST FIG IS
SIDE SHOWN
FLEXIBLE DUCTWORK CONNECTION
RECT DUCT-1ST FIG IS SIDE SHOWN
ACOUSTICAL DUCT LINER
VOLUME DAMPER
MANUAL BALANCING VALVE
CONTROL VALVE (3 WAY)
CONTROL VALVE (2 WAY)
HOSE END DRAIN VALVE
PRESSURE REDUCING VALVE
AUTO FLOW CONTROL VALVE
VALVE
CHECK VALVE
HVAC DUCTWORK AND SPECIALTIES
EXTRACTOR
SIZE/TYPE
RETURN OR EXHAUST DUCT DOWN
RETURN OR EXHAUST REGISTER
OR GRILLE (VERTICAL MOUNTING)
SUPPLY REGISTER OR GRILLE
ROUND DUCT UP / DOUBLE LINE PIPE
ROUND DUCT DOWN / DOUBLE
LINE PIPE
RETURN OR EXHAUST DUCT UP
(R OR G) (VERTICAL MOUNTING)
TURNING VANES
SUPPLY AND FRESH AIR DUCT DOWN
SUPPLY AND FRESH AIR DUCT
SIZE/TYPE
SIZE/TYPE
FLOW
FLOW
SIZE/TYPE
FLOW
FLOW
SMOKE DAMPER - FIRE SEPARATION
FIRE DAMPER - FIRE SEPARATION
MOTORIZED DAMPER-OPPOSED BLADE
MOTORIZED DAMPER-PARALLEL BLADE
ROUND CEILING SUPPLY DIFFUSER
(HORIZONTAL MOUNTING)
GRAVITY DAMPER
(OR) SAFETY VALVE-STEAM
TEMP & PRESS RELIEF
THERMOSTATIC MIXING VALVE
SIZE/TYPE
FLOW
HUMIDISTAT
THERMOSTAT
FLEXIBLE DUCT
S
H
T
TEMPERATURE SENSOR
FIRE/SMOKE DAMPER - FIRE
SEPARATION
RECTANGULAR RETURN REG OR
GRILLE (HORIZONTAL MOUNTING)
RECTANGULAR SUPPLY DIFF, REG
OR GRILLE (HORIZONTAL
MOUNTING)
VALVES
M
M SIZE/TYPE
FLOW
RECTANGULAR EXHAUST REG OR
GRILLE (HORIZONTAL MOUNTING)
M
VOLUME DAMPER
M
F
S
FS
DEMOLITION GENERAL NOTES GENERAL NOTES
SEE INDIVIDUAL SHEET FOR REFERENCE DEMO NOTES SEE INDIVIDUAL SHEET FOR REFERENCE NOTES
MECHANICAL SHEET LIST
M 00 MECHANICAL INFO SHEET
M001 MECHANICAL DEMOLITION PLANS
M101 MECHANICAL PLANS
M102 FLOOR PLAN - PLUMBING
M401 QUENCH VENT DETAILS
M501 MECHANICAL SCHEDULES & DETAILS
T
MRI 1
MRI 2
EQUIPMENT
TOILET
CORRIDOR
VTR (E)
SAN
V
V
V
V S
A N
4" SAN DN
VAC
O
(E) VAV(E) VAV
(E) VAV
3 3
2 2
(E) AHU
(E
C
(E
C U
N
FOURFRONT DESIGN, INC.
517 7TH STREET
RAPID CITY, SOUTH DAKOTA
57701
PH: (605) 342-9470
FAX: (605) 342-2377
WWW.FOURFRONTDESIGN.COMFOURFRONT
D E S I G N I N C.
Office of Construction and Facilities Management
Project Number
Building Number
Drawing Number
Dwg.
DrawnCheckedDate
Location
Project TitleDrawing Title
Approved: Project Director
ARCHITECTCONSULTANTS:
DateRevisions:
VA FORM 08-6231
o n e e ig h th in c h o n e fo o t o n e q u a rt e r in c h o n e fo o t th re e e ig h th s in c h o n e fo o t o n e h a lf in c h o n e fo o t th re e q u a rt e rs in c h o n e fo o t o n e in c h o n e fo o t o n e a n d o n e h a lf in c h e s o n e fo o t th re e in c h e s o n e fo o t
F
E
D
C
B
A
1 2 3 4 5 6 7 8 9
F
E
D
C
B
A
1 2 3 4 5 6 7 8 9
Department of Veterans Affairsof
635-CSI-105
M R
I
KEY PLAN
NOT TO SCALE
SCHEMMER
1044 NORTH 115TH STREET, SUITE 3300
OMAHA, NEBRASKA 68154
ROOF
/2
/2
:2
:0
P
M
07/22/20
MECHANICAL DEMOLITION PLANS
M001
OKLAHOMA CITY VA HEALTH
CARE SYSTEM MRI SITE PREP
AJPJEK
OKLAHOMA CITY, OKLAHOMA
OWNER
07868.001
REF. DEMO NOTES ( ):X
SCALE: 1/4" = 1'-0"
SECOND FLOOR PLAN - MECHANICAL & PLUMBING DEMOLITION
SCALE: 1/8" = 1'-0"
THIRD FLOOR ROOF PLAN - MECHANICAL DEMOLITION
1 REMOVE ANY GRILLES OR DIFFUSERS SERVING EQUIPMENT 102.
DEMO ASSOCIATED DUCTWORK BACK TO MAIN AND CAP. REPAIR
INSULATION TO MATCH ADJACENT CONDITION WITH PROPER VAPOR
BARRIER.
2 EXISTING EQUIPMENT TO REMAIN.
3 EXISTING VAV BOX AND ASSOCIATED DUCTWORK, ACCESSORIES,
GRILLES, AND DIFFUSERS ARE TO REMAIN.
4 DEMO EXISTING FAN COIL SYSTEM. FAN COIL IS LOCATED ON THIRD
FLOOR. REMOVE ALL ASSOCIATED DUCTWORK, DIFFUSERS, GRILLES,
AND DUCT ACCESSORIES. REMOVE ASSOCIATED PIPING AND
SPECIALTIES BACK TO MAIN AND CAP. SEAL ALL FLOOR
PENETRATIONS.
5 EXISTING SUPPLY GRILLE TO BE RELOCATED TO NEW CEILING GRID IN
TOILET 101. SEE M101 FOR RELOCATION.
6 DEMO MEDICAL GAS OUTLETS AND ASSOCIATED PIPING BACK TO
ISOLATION VALVE AND PREPARE FOR CONNECTION TO NEW PIPING.
SEE PLANS ON M101.
7 CONTRACTOR TO COORDINATE WITH THE OWNER ON THE REMOVAL
OF ANY ASSOCIATED MECHANICAL COMPONENTS SERVING THE MRI
MACHINE BEING REMOVED. DISCONNECT ANY ASSOCIATED PIPING
AND REMOVE BACK TO MAINS AND CAP. REPAIR INSULATION TO MATCH
ADJACENT CONDITION WITH PROPER VAPOR BARRIER. CONTRACTOR
TO SEAL ALL WALL OR FLOOR OPENINGS TO MATCH ADJACENT
CONDITIONS.
8 CONTRACTOR TO REMOVE EXISTING CRYOGEN VENT SYSTEM AND
ANY ASSOCIATED ACCESSORIES. CONTRACTOR TO SEAL ALL WALL
OR FLOOR OPENINGS TO MATCH ADJACENT CONDITIONS.
CONTRACTOR TO FIELD VERIFY EXACT LOCATION.
NORTH
CONTROL
MRI 1
MRI 2
EQUIPMENT
TOILET
CORRIDOR
SS-1
QUENCH
VENT
10x10 R-1
10 Ø S-1
30010 Ø SA
10x8 RA
VAV-104
6 Ø SA
3/4" HWR
3/4" HWS
RS/RLRS/RL
3/4" CD
2" CWS
2" CWR
8 Ø
(E
V A
V
(E
V A
V
(E) VAV
SSCU-1
RS/RLRS/RL
1 2" CWS
2" CWR
M R
I C
H
IL
L E
R
(E
C
(E
C U
(E) AHU
EQUIPMENT
V
V V
SAN
S A
N
V
O
VAC
141414
2" DCW
2"
3"
2"
V
SAN
VTR (E)
FS-1
TOILET
MRI 1
CONTROL
MRI 2
CORRIDOR
106CORRIDOR
4" SAN DN
N
FOURFRONT DESIGN, INC.
517 7TH STREET
RAPID CITY, SOUTH DAKOTA
57701
PH: (605) 342-9470
FAX: (605) 342-2377
WWW.FOURFRONTDESIGN.COMFOURFRONT
D E S I G N I N C.
Office of Construction and Facilities Management
Project Number
Building Number
Drawing Number
Dwg.
DrawnCheckedDate
Location
Project TitleDrawing Title
Approved: Project Director
ARCHITECTCONSULTANTS:
DateRevisions:
VA FORM 08-6231
o n e e ig h th in c h o n e fo o t o n e q u a rt e r in c h o n e fo o t th re e e ig h th s in c h o n e fo o t o n e h a lf in c h o n e fo o t th re e q u a rt e rs in c h o n e fo o t o n e in c h o n e fo o t o n e a n d o n e h a lf in c h e s o n e fo o t th re e in c h e s o n e fo o t
F
E
D
C
B
A
1 2 3 4 5 6 7 8 9
F
E
D
C
B
A
1 2 3 4 5 6 7 8 9
Department of Veterans Affairsof
635-CSI-105
M R
I
KEY PLAN
NOT TO SCALE
SCHEMMER
1044 NORTH 115TH STREET, SUITE 3300
OMAHA, NEBRASKA 68154
ROOF
/2
/2
:0
:3
P
M
07/22/20
MECHANICAL PLANS
M101
OKLAHOMA CITY VA HEALTH
CARE SYSTEM MRI SITE PREP
AJPJEK
OKLAHOMA CITY, OKLAHOMA
OWNER
07868.001
REF. NOTES ( ):X
16 EXTEND EXISTING OXYGEN AND VACUUM PIPING TO WALL OUTLET.
FIELD VERIFY EXACT LOCATION OF PIPING MAINS AND VALVES. IF
EXISTING ISOLATION VALVE IS NOT ABOVE THE CEILING OF THE
ADJACENT CORRIDOR, CONTRACTOR SHALL RELOCATE EXISTING
VALVE IN AN ACCESSIBLE SPACE IN THE CEILING OF THE ADJACENT
CORRIDOR. COORDINATE PIPE ROUTING WITH MRI SHIELDING.
17 CONTRACTOR TO PROVIDE KOOLANT KOOLERS MANUAL
SWITCH-OVER PANEL. INSTALL A 50 MICRON STRAINER OR FILTER
BETWEEN DOMESTIC WATER SUPPLY AND MANUAL SWITCH-OVER
PANEL.
18 CONDENSATE PIPING TO DISCHARGE TO FLOOR SINK.
19 RS/RL PIPING BETWEEN SS-1 AND SSCU-1 TO BE ROUTED THROUGH
THE THIRD FLOOR MECHANICAL SPACE. PROVIDE FIRE CAULKING
AROUND ANY PENETRATIONS THROUGH RATED WALLS AND FLOORS.
20 PROVIDE FIRE CAULKING AROUND ANY PIPE PENETRATIONS
THROUGH RATED WALLS AND FLOORS. (TYP.)
21 CONNECT 3" SANITARY LINE TO EXISTING 4" VERTICAL SANITARY MAIN
LOCATED IN CHASE. CONTRACTOR TO FIELD VERIFY EXACT LOCATION
OF SANITARY MAIN IN CHASE.
22 CONTRACTOR IS TO INSTALL QUENCH LINE PER MANUFACTURER
GUIDELINES. SEE DETAILS ON M401 FOR CONSTRUCTION TYPES,
MATERIALS, AND GUIDELINES. QUENCH LINE IS TO CONSIST OF THE
FOLLOWING COMPONENTS: (1) VIBRATION DECOUPLER, 25 FEET OF
SMOOTH STRAIGHT RUN, (4) 90 DEGREE SMOOTH ELBOW, (2) 45
DEGREE SMOOTH ELBOWS, AND MANUFACTURER APPROVED
HORIZONTAL OUTLET. ALL COMPONENTS ARE TO BE 8" DIAMETER
SIZE. THE DESIGN SHOWN ON PLANS ONLY REQUIRES (2) 90 DEGREE
ELBOWS, THE ADDITIONAL (2) 90 DEGREE ELBOWS AND (2) 45 DEGREE
ELBOWS ARE INCLUDED TO ALLOW CONTRACTOR TO ROUTE QUENCH
VENT AROUND EXISTING EQUIPMENT OR STRUCTURES IF NECESSARY,
CONTRACTOR IS TO REQUEST APPROVAL IF ANY ADDITIONAL
STRAIGHT RUN OR ADDITIONAL ELBOWS ARE REQUIRED. SEE DETAILS
ON M401 FOR ADDITIONAL INFORMATION.
23 QUENCH VENT TO PENETRATE FLOOR ABOVE MRI ROOM AND BE
ROUTED TO EXTERIOR THROUGH EXISTING EXHAUST LOUVER.
DASHED LINE INDICATES PORTION OF QUENCH VENT THAT IS TO BE
ROUTED IN THE THIRD FLOOR MECHANICAL SPACE.
24 PROVIDE OPTION 3 OUTLET TYPE WITH A DEFLECTOR PLATE FOR
QUENCH VENT. SEE OPTION 3 DETAIL ON SHEET M401. MAINTAIN
REQUIRED CLEARANCES AS SHOWN ON SHEET M401.
1 SIZE AND INSTALL REFRIGERANT PIPING PER MANUFACTURER'S
REQUIREMENTS AND THE SPECIFICATIONS. ALL PIPE PENETRATIONS
SHALL BE DONE VIA A CORE DRILL TO BE NEAT AND CLEAN.
2 EXISTING MECHANICAL EQUIPMENT ON ROOF TO REMAIN.
3 CONDENSATE PIPING TO DISCHARGE TO FLOOR SINK. CONDENSATE
PIPING TO BE DRAWN-TEMPER COPPER TUBING, TYPE M AND IS TO
COMPLY WITH ASTM B 88. PROVIDE 3/4" FLEXIBLE ELASTOMERIC
INSULATION. INSULATION IS TO COMPLY WITH ASTM C534.
4 2" CHILLED WATER SUPPLY AND RETURN TO CONNCET TO KOOLANT
KOOLERS MANUAL SWITCH-OVER PANEL. CONTRACTOR TO PROVIDE
KOOLANT KOOLERS MANUAL SWITCH-OVER PANEL.
5 INSTALL CHILLER ON MANUFACTURER RECOMMENDED ROOF CURB.
CURB IS TO BE ANCHORED TO STRUCTURE.
6 EXISTING VAV BOX IS TO REMAIN.
7 INSTALL CONDENSING UNIT ON 12" ROOF CURB. CURB IS TO BE
ANCHORED TO BUILDING STRUCTURE.
8 CONNECT 6" Ø SA DUCT TO EXISTING SUPPLY AIR MAIN LOCATED IN
CORRIDOR.
9 CONNECT 10X8 RA DUCT TO EXISTING RETURN AIR MAIN LOCATED IN
CORRIDOR.
10 CONNECT 3/4" HWS/R TO MAIN. CONTRACTOR TO FIELD VERIFY EXACT
LOCATION.
11 RELOCATE EXISTING DIFFUSER TO NEW CEILING GRID IN TOILET 101.
CONNECT TO EXISTING DUCTWORK AND MATCH EXISTING DUCT SIZE.
CONTRACTOR TO FIELD VERIFY EXISTING DUCT'S EXACT LOCATION.
12 ROUTE CHILLED WATER SUPPLY/RETURN TO CHILLER THROUGH THE
THIRD FLOOR MECHANICAL SPACE. PROVIDE FIRE CAULKING AROUND
ANY PENETRATIONS THROUGH RATED WALLS AND FLOORS.
13 ROUTE CHILLED WATER SUPPLY/RETURN TO MRI EQUIPMENT
THROUGH THE THIRD FLOOR MECHANICAL SPACE AND DOWN
THROUGH THE EQUIPMENT ROOM CEILING. PROVIDE FIRE CAULKING
AROUND ANY PENETRATIONS THROUGH RATED WALLS AND FLOORS.
14 FURNISH AND INSTALL NEW MEDICAL GAS WALL OUTLETS. PROVIDE
(1) OXYGEN OUTLET, (1) MEDICAL VACUUM OUTLET, AND (1) SLIDE
BRACKET. CONTRACTOR TO MATCH CONNECTION TYPE OF OUTLETS
BEING REMOVED. NEW MEDICAL GAS WALL OUTLETS SHALL BE
NON-FERROUS CONSTRUCTION FOR INSTALLATION WITHIN AN MRI
SHIELD. SEE ARCHITECTURAL PLANS FOR MOUNTING HEIGHT AND
PLACEMENT.
15 SEE CONTINUATION ON M102.
NORTH
SCALE: 1/4" = 1'-0"
SECOND FLOOR PLAN - MECHANICAL
SCALE: 1/8" = 1'-0"
THIRD FLOOR ROOF PLAN - MECHANICAL
NORTH
SCALE: 1/4" = 1'-0"
SECOND FLOOR PLAN - PLUMBING
STAIR 6
C H
A S
E
ELEVATOR
P-11
STAIR 7
CHASE
C H
A S
E
S T
A
IR
DN
CHASE
ELEVATOR
P-8
ELEVATOR
P-7
ELEVATOR
P-9
ELEVATOR
P-10
CHASE
STAIR 8
PATIENT
HOLDING
PHLEBOTOMY
DRESSING
ROOM
RADIOLOG
Y
WAITING
SPS
STORAG
E
BONE
DENSITOMETR
Y
EMS
STORAG
E
TELE/DATA
ELECTRICAL EQUIPMENT
PATIENT
HOLDING
E Q
U
IP
M E
N T
EQUIPMENT
CONTROL
WOR
K
AREA
SCRUB
ROOM
H.A.C.
IR STORAGE
FILES
SPS
STORAG
E
TOILET
OFFICE
PHYSICIA
N
OFFICEOFFICEOFFICE
OFFICE
PHYSICIA
N
STAFF
RADIOLOGY P.A.C.U.
TOILET
TRASH LINEN
PATIENT
WAITIN
G
HOLDING ROOM
TOILET
TOILET
80 SLICE CT WORKROOM
64 SLICE CT
TOILETTOILET
TOILET TOILET
FLUOROSCOPY
CONTROL
X-RAY X-RAY
READING ROOM
EXPEDITE
DESKELECTRIC
PROCESSING ALCOVEH.A.C.
OFFICE
X-RAY
FLUOROSCOPY
CONTROL
X-RAYX-RAY
READING ROOM
PATIENT CHECK-IN
ULTRASOUND
TOILET
TOILET
ULTRASOUND
ULTRASOUND
TOILET
TOILET
MALE
TOILET
FEMALE
S T
O R
A G
E
80 SLICE CT
CONTROL
MRI
EQUIPMENT
INTERVENTIONA
L
RADIOLOGY
INTERVENTIONA
L
RADIOLOGY
SOILED
UTILITY
PATIENT
HOLDING
RECEPTION/CHECK IN
TOILET TOILET
RADIOLOGY
CONFERENC
E
DRESSING
ROOM
DRESSING
ROOM
DRESSING
ROOM
DRESSING
ROOM
OFFICE
RN
OFFICE
TECHNICIA
N
OFFICE
AO
OFFICE
CHIEFADPAC
S
OFFICE
COMPUTE
R
STORAGE
TECHNICIA
N
OFFICE
TELE/DATA
ELECT.
COMPUTER
ROOM
TOILET
MALE
TOILET
FEMALE
LOCKER
ROOM
BREAK
ROOM
OFFICE
IV ROOM
M101
2"
2"
DCW
DCW 1
GENERAL NOTES:
1. THIS SHEET IS TO BE USED FOR REFERENCE TO THE TIE-IN LOCATION OF
THE EXISTING DOMESTIC WATER MAIN. THE CONTRACTOR IS TO FIELD
VERIFY EXISTING CONDITIONS AND MAKE ADJUSTMENTS ACCORDINGLY.
2. PIPE IS TO BE ROUTED IN THE CEILING ABOVE CORRIDORS TO THE
INDICATED TIE-IN POINT. CONTRACTOR TO REQUEST APPROVAL IF
ALTERNATE PIPE ROUTING IS REQUIRED.
3. CONTRACTOR TO COORDINATE INSTALLATION WITH FACILITY STAFF.
4. CONTRACTOR TO REQUEST OWNER APPROVAL OF THE TIE-IN METHOD
USED TO CONNECT INTO THE EXISTING PIPING AS WELL AS ANY REQUIRED
DOWNTIME TO THE EXISTING SYSTEM.
N
FOURFRONT DESIGN, INC.
517 7TH STREET
RAPID CITY, SOUTH DAKOTA
57701
PH: (605) 342-9470
FAX: (605) 342-2377
WWW.FOURFRONTDESIGN.COMFOURFRONT
D E S I G N I N C.
Office of Construction and Facilities Management
Project Number
Building Number
Drawing Number
Dwg.
DrawnCheckedDate
Location
Project TitleDrawing Title
Approved: Project Director
ARCHITECTCONSULTANTS:
DateRevisions:
VA FORM 08-6231
o n e e ig h th in c h o n e fo o t o n e q u a rt e r in c h o n e fo o t th re e e ig h th s in c h o n e fo o t o n e h a lf in c h o n e fo o t th re e q u a rt e rs in c h o n e fo o t o n e in c h o n e fo o t o n e a n d o n e h a lf in c h e s o n e fo o t th re e in c h e s o n e fo o t
F
E
D
C
B
A
1 2 3 4 5 6 7 8 9
F
E
D
C
B
A
1 2 3 4 5 6 7 8 9
Department of Veterans Affairsof
635-CSI-105
KEY PLAN
NOT TO SCALE
SCHEMMER
1044 NORTH 115TH STREET, SUITE 3300
OMAHA, NEBRASKA 68154
AREA F
/2
/2
:1
:4
A
M
07/22/20
FLOOR PLAN - PLUMBING
M102
OKLAHOMA CITY VA HEALTH
CARE SYSTEM MRI SITE PREP
AuthorChecker
OKLAHOMA CITY, OKLAHOMA
OWNER
07868.001
REF. NOTES ( ):X
1 CONNECT 2" DCW TO EXISTING 5" DCW LOCATED IN CHASE. CONTRACTOR
TO VERIFY EXACT LOCATION OF 5" DCW MAIN LOCATED IN CHASE.
SCALE: 3/32" = 1'-0"
SECOND FLOOR PLAN - AREA K - PLUMBING
120°
Deflector plate
FOURFRONT DESIGN, INC.
517 7TH STREET
RAPID CITY, SOUTH DAKOTA
57701
PH: (605) 342-9470
FAX: (605) 342-2377
WWW.FOURFRONTDESIGN.COMFOURFRONT
D E S I G N I N C.
Office of Construction and Facilities Management
Project Number
Building Number
Drawing Number
Dwg.
DrawnCheckedDate
Location
Project TitleDrawing Title
Approved: Project Director
ARCHITECTCONSULTANTS:
DateRevisions:
VA FORM 08-6231
o n e e ig h th in c h o n e fo o t o n e q u a rt e r in c h o n e fo o t th re e e ig h th s in c h o n e fo o t o n e h a lf in c h o n e fo o t th re e q u a rt e rs in c h o n e fo o t o n e in c h o n e fo o t o n e a n d o n e h a lf in c h e s o n e fo o t th re e in c h e s o n e fo o t
F
E
D
C
B
A
1 2 3 4 5 6 7 8 9
F
E
D
C
B
A
1 2 3 4 5 6 7 8 9
Department of
Veterans Affairsof
635-CSI-105
SCHEMMER
1044 NORTH 115TH STREET, SUITE 3300
OMAHA, NEBRASKA 68154
/2
/2
:1
:4
A
M
07/22/20
QUENCH VENT DETAILS
M401
OKLAHOMA CITY VA HEALTH
CARE SYSTEM MRI SITE PREP
AuthorChecker
OKLAHOMA CITY, OKLAHOMA
OWNER
07868.001
SCALE: NOT TO SCALE
INTERIOR PIPE SLEEVE DETAILS
CONCRETE OR
MASONRY PARTITION.
GROUT, SEE
SPECIFICATIONS.
1/4" CLEAR AROUND
PIPE OR INSULATION.
CAULKING, SEE
SPECIFICATIONS.
SLEEVE, SEE
SPECIFICATIONS.
NOTE:
1. CAULK FULL PERIMETER OF PIPE AT BOTH INSIDE & OUTSIDE OF WALL.
1/4" CLEAR AROUND
PIPE OR INSULATION
RISER CLAMP
CONCRETE
FLOOR
CAULKING, SEE
SPECIFICATIONS.
NOTE:
1. CAULK FULL PERIMETER OF PIPE AT BOTH TOP AND BOTTOM OF
FLOOR.
DETAIL OF SLEEVE FOR PIPING
PENETRATION OF FLOOR
DETAIL OF SLEEVE FOR PIPING
PENETRATION OF VERTICAL PARTITIONS
SLEEVE, SEE
SPECIFICATIONS.
SLEEVE HEIGHT
SHALL BE 2" HIGH IN
WET LOCATIONS, AT
FLOOR LEVEL IN ALL
OTHER LOCATIONS
RISER
CLAMP LEG
TOTAL
HEIGHT
OF TRAP
TO FLOOR DRAIN
TOTAL HEIGHT OF TRAP:
X + H + (1 1/2 x PIPE DIAMETER)
H = AT LEAST 1 INCH PLUS
CASING STATIC PRESSURE
X = 1/2 "H"
COIL CONDENSATE
DRAIN PAN
SLOPE DOWN
X
H
CLEAN OUT
HOUSEKEEPING PAD,
HEIGHT IS BASED ON
TRAP HEIGHT
SEP GPA EPC
SIEMENS SUPPLIED AND
INSTALLED CABINETS
THE PIPING FROM THE CHILLED WATER SUPPLY
TO THE SEP CABINET MUST MEET THE CHILLER
MANUFACTURER'S GUIDELINES FOR ELBOWS
AND LENGTH SEE MANUFACTURER FOR
MINIMUM REQUIRMENTS.
22 FOOT MAX
HOSE LENGTH
9'-10" MAX HOSE LENGTH
1-1/4" MALE THREAD
PROVIDED BY SIEMENS FOR 60 kW SEP
FULL PORT SHUT OFF VALVES
FOR SERVICING SEP CABINET
1" MINIMUM SYSTEM BYPASS WITH BOILER
DRAIN AND FULL PORT BALL VALVE
PROVIDE FULL PORT SHUT OFF VALVES
INSTALLED ON CHILLER SUPPLY AND
RETURN LINES AT CHILLER
DIMPLEX CHILLER
PRESSURE GAUGE WITH RANGE FROM 40
TO 110 PSI (LOCATED NEAR SEP)
THERMOMETER WITH RANGE FROM 30°F
TO 80°F (LOCATED NEAR SEP)
BALL VALVE
BOILER DRAIN
FILTER - PROVIDED BY DIMPLEX,
INSTALLED BY CONTRACTOR
ALL PIPING AND PLUMBING FIXTURES SHALL BE FURNISHED, INSTALLED, PRESSURE TESTED AND CHARGED BY THE MECHANICAL CONTRACTOR PRIOR TO THE DELIVERY AND INSTALLATION OF THE
SIEMENS SPPLIED AND INSTALLED EQUIPMENT UNLESS SPECIFIED OTHERWISE.
AT THE HIGHEST POINT OF THE WATER SUPPLY PIPE FROM THE CIHLLER AN AUTOMATIC
DEAERATION DEVICE (AIR VENT) WITH BALL VALVE MUST BE INSTALLED BY THE MECHANICAL
CONTRACTOR.
SYSTEM MUST BE PROVEN TO BE FREE OF LEAKAGE.
THE MECHANICAL ENGINEER OF RECORD SHALL BE ULTIMATELY RESPONSIBLE FOR THE SITE SPECIFIC DESIGN AND SPECIFICATION OF THE MECHANICAL AND PIPING SYSTEMS AS SHOWN AND SHALL BE IN ACCORDANCE WITH ALL APPLICABLE LOCAL, STATE, AND NATIONAL CODES. ALL WORK SHALL BE PERFORMED BY THE MECHANICAL CONTRACTOR AND SHALL BE SUBJECT TO COMPLIANCE WITH ALL
APPLICABLE LOCAL, STATE, AND NATIONAL CODES.
THE SUPPLY AND RETURN PIPES FROM THE CHILLED WATER SUPPLY TO THE SEP MUST BE LABELED TO
SHOW FLOW DIRECTION AND CONTENT (WATER/GLYCOL).
NOTES:
LIQUID AND SUCTION
LINES DOWN TO INDOOR
UNIT
INSULATE REFRIGERANT LINES
CONTINUOUSLY THROUGH
PENETRATION. SEAL
PENETRATION WEATHER TIGHT.
ROOF
CONDSENSING UNIT
METAL CURB CAP
FASTEN CONDENSING UNIT
SECURELY TO CURB CAP
PREFRABRIACTED ROOF
CURB
REHEAT COIL
CONTROL
VALVE
UNION
(TYP.)
STRAINER
VALVE
(TYP.)
AUTOFLOW
VALVE
HWS
HWR
MANUAL AIR VENT
HOSE END DRAIN WITH CAP
NOTE: SEE PLANS FOR PIPE SIZE, MINIMUM 3/4".
SHEET METAL 90° ELBOW
DIFFUSER NECK
TRANSITION
(AS REQUIRED)
CEILING
SUPPLY DUCT-
SEE PLANS FOR
SIZES
SUPPLY DIFFUSER
DUCT INSULATION
FLEX DUCT
INSULATE BACKPAN
FOURFRONT DESIGN, INC.
517 7TH STREET
RAPID CITY, SOUTH DAKOTA
57701
PH: (605) 342-9470
FAX: (605) 342-2377
WWW.FOURFRONTDESIGN.COMFOURFRONT
D E S I G N I N C.
Office of Construction and Facilities Management
Project Number
Building Number
Drawing Number
Dwg.
DrawnCheckedDate
Location
Project TitleDrawing Title
Approved: Project Director
ARCHITECTCONSULTANTS:
DateRevisions:
VA FORM 08-6231
o n e e ig h th in c h o n e fo o t o n e q u a rt e r in c h o n e fo o t th re e e ig h th s in c h o n e fo o t o n e h a lf in c h o n e fo o t th re e q u a rt e rs in c h o n e fo o t o n e in c h o n e fo o t o n e a n d o n e h a lf in c h e s o n e fo o t th re e in c h e s o n e fo o t
F
E
D
C
B
A
1 2 3 4 5 6 7 8 9
F
E
D
C
B
A
1 2 3 4 5 6 7 8 9
Department of Veterans Affairsof
635-CSI-105
SCHEMMER
1044 NORTH 115TH STREET, SUITE 3300
OMAHA, NEBRASKA 68154
/2
/2
:1
:4
A
M
07/22/20
MECHANICAL SCHEDULES & DETAILS
M501
OKLAHOMA CITY VA HEALTH
CARE SYSTEM MRI SITE PREP
AJPJEK
OKLAHOMA CITY, OKLAHOMA
OWNER
07868.001
2. 24"x24" FACE FRAME WITH 1"x1"x1" EGG CRATE OPENING AS SHOWN ON THE PLANS.
NOTES: 1. PROVIDE WITH STRAIGHTENING GRID.
R-1 RETURN AIR LAY-IN 24x24 SEE PLANS WHITE 0.1 25 EGC10 50F 80 2
S-1 SUPPLY AIR LAY-IN 24x24 SEE PLANS WHITE 0.1 25 PLQ OMNI SPD 1
MARK SERVES TYPE
FACE SIZE
(IN.) NECK SIZE (IN.) FINISH
MAX PD (IN
WG)
MAX NC AT
PD SHOWN
KRUEGER
MODEL
NUMBER
TITUS MODEL
NUMBER
PRICE
MODEL
NUMBER REMARKS
DIFFUSER, REGISTER AND GRILLE SCHEDULE
2. PROVIDE UNIT WITH ADVANCED WIND BAFFLE KIT FOR OPERATION DOWN TO -20 °F.
NOTES: 1. PROVIDE REFRIGERANT LINESETS SIZED AND INSTALLED PER MANUFACTURER REQUIREMENTS. PROVIDE LONG LINE LENGTH ACCESSORIES AS REQUIRED.
SS-1 SEE PLANS 705 80 67 24.0 208 1 60 MITSUBISHI PKA-A24KA7 SSCU-1 WALL 21.4 95 R410A MITSUBISHI PUY-A24NHA7 19 26 208 1 60 1,2
DB WB
AIRSIDE DATA EAT (°F)
MIN CAP
(MBH) VOLTS PH HZ MCA MOCP VOLTS PH HZLOCATION CFM
COOLING COIL DATA ELECTRICAL DATA
MANUFACTURER AND
MODEL NUMBER MARK LOCATION
COOLING
EFFICIENCY
(SEER)
AMB
TEMP (°F)
REFER
TYPE
MANUFACTURER AND
MODEL NUMBER
ELECTRICAL DATA
MARK
INDOOR UNIT DATA OUTDOOR UNIT DATA
REMARKS
SYSTEM AIR CONDITIONING SCHEDULE
2. CONTRACTOR IS RESPONSIBLE FOR DETERMINING LEFT OR RIGHT HAND CONFIGURATION.
NOTES: 1. PROVIDE WITH 1/2" FIBER-FREE LINER AND CONTROL ENCLOSURE.
VAV-104 CONTROL 104 300 90 6 0.12 1 16 18 90 55 100 4.4 1 0.06 0.5 180 160 0.14 KRUEGER LMHS 1,2
MAX MIN MIN INLET CFM EAT (°F) LAT (°F) CAP (MBH) NO ROWS MAX PD GPM EWT (°F) LWT (°F)
PRESS DROP
FT WATMARK SERVES
CFM
INLET SIZE
(IN)
OP PRESS (IN WG)
MAX RAD NC @ TOTAL
INLET PRESS SHOWN
REHEAT COIL DATA WATER DATA
MANUFACTURER AND
MODEL NUMBER REMARKS
AIR TERMINAL UNIT SCHEDULE
SCALE: NOT TO SCALE
INTERIOR PIPE SLEEVE DETAILS
SCALE: NOT TO SCALE
CONDENSATE DRAIN TRAP NEGATIVE STATIC PRESSURE
SCALE: NOT TO SCALE
PIPING SCHEMATIC FOR CHILLED WATER - DIMPLEX CHILLER
SCALE: NOT TO SCALE
CONDENSING UNIT SUPPORT/ PIPING THROUGH ROOF DETAIL
NOTES: 1. PROVIDE WITH ALUMINUM STRAINER AND NICKEL BRONZE RIM AND GRATE WITH 2-1/2" SQU CENTER HOLE.
FS-1 FLOOR SINK --- --- 3" 2" --- --- --- JAY R. SMITH 3100 --- --- --- 1
HOT
WATER
COLD
WATER WASTE VENT OTHER FIXTURE FAUCET/FLUSH VALVE ACCESSORY #1 ACCESSORY #2 REMARKSMARK DESCRIPTION
FIXTURE CONNECTIONS
MOUNTING HEIGHT FLOW RATE
MANUFACTURER AND MODEL NUMBER
PLUMBING FIXTURE SCHEDULE
SCALE: NOT TO SCALE
REHEAT COIL PIPING DETAIL
SCALE: NOT TO SCALE
SUPPLY DIFFUSER/ DUCT CONNECTION DETAIL
A. SEE ARCHITECTURAL SHEETS FOR EXACT AREAS AND WALLS AFFECTED BY THE DEMOLITION AND
REMODEL.
B. DEMOLITION SHOWN IS AS REQUIRED FOR THE COMPLETE PROJECT. THE CONTRACTOR SHALL COORDINATE THE DEMOLITION EFFORT PER THE PHASING PLANS. COORDINATE DEMOLITION WITH THE WORK OF OTHER TRADES. PROVIDE TEMPORARY POWER AND LIGHTS AS REQUIRED TO ALLOW
THE WORK OF OTHER TRADES TO PROCEED OR AS REQUIRED TO ALLOW THE OWNER TO OCCUPY
THE SPACE.
C. ALL OF THE EQUIPMENT AND DEVICES SHOWN ON THIS DRAWING ARE EXISTING. THE LOCATIONS OF
THE EQUIPMENT AND DEVICES WERE OBTAINED FROM EXISTING DRAWING AND FIELD
INVESTIGATION. THE LOCATIONS OF THE EXISTING EQUIPMENT AND DEVICES ARE SHOWN FOR THE
CONVENIENCE OF THE CONTRACTOR. ACCURACY OF THE INFORMATION SHOWN IS NOT
GUARANTEED. THE CONTRACTOR IS RESPONSIBLE FOR THE VERIFICATIONS OF ALL EXISTING
CONDITIONS PRIOR TO SUBMITTING BIDS. SITE VISITS PRIOR TO SUBMISSION OF BIDS NEED TO BE
FULLY COORDINATED WITH THE OWNER.
D. CIRCUIT INFORMATION FOR ALL AFFECTED CIRCUITS SHALL BE VERIFIED PRIOR TO COMMENCING WORK. ACCURACY OF THE INFORMATION SHOWN AND INCLUDED ON THE PANEL SCHEDULES IS NOT
GUARANTEED.
E. LUMINAIRES AND DEVICES INDICATED TO BE REMOVED, WHICH ARE LOCATED ON CEILINGS OR WALLS, SHALL HAVE LUMINAIRES, DEVICE AND ALL ASSOCIATED CIRCUIT WIRING AND CONDUIT REMOVED BACK TO NEAREST JUNCTION BOX, OR DEVICE NOT INDICATED TO BE REMOVED. ACTUAL
CIRCUITING MUST BE DETERMINED IN THE FIELD. THE CONTRACTOR SHALL PROVIDE ALL
NECESSARY WIRE, CONDUIT, DEVICES AND CONNECTIONS NECESSARY TO ENSURE CIRCUIT
CONTINUITY OF THE REMAINING DEVICES.
F. THE OWNER HAS THE RIGHT TO RETAIN ALL SALVAGEABLE MATERIAL. ANY MATERIAL THE OWNER
CHOOSES NOT TO ACCEPT SHALL BE REMOVED FROM THE SITE AND PROPERLY DISPOSED OF BY
THE CONTRACTOR. CONTRACTOR IS RESPONSIBLE FOR THE REMOVAL OF ANY PCB BALLAST OR
HAZARDOUS LAMPS.
G. ELECTRICAL CONTRACTOR SHALL BE RESPONSIBLE FOR THE LABELING OF SPARE OR ABANDONED
CIRCUITS AND/OR CIRCUIT BREAKERS IN ALL PANELS AFFECTED BY THE DEMOLITION. SPARE
CIRCUITS AND/OR CIRCUIT BREAKERS MAY BE REUSED FOR THE CIRCUITING OF THE NEW OR
REMODELED AREA AS SPACE AND PANEL LOADING WILL ALLOW.
H. THE CONTRACTOR SHALL BE HELD FULLY RESPONSIBLE FOR THE PROPER RESTORATION OF ALL EXISTING SURFACES REQUIRING PATCHING, PLASTERING, PAINTING AND/OR OTHER REPAIR DUE TO
THE DEMOLITION WORK DONE BY THE ELECTRICAL CONTRACTOR, UNDER THE TERMS OF THIS
SPECIFICATION. CLOSE AND SEAL ALL OPENINGS, REPAIR ALL SURFACES, ETC., AS REQUIRED.
I. EXISTING FIRE ALARM DEVICES, EMERGENCY LIGHTS AND EXIT SIGNS SHALL BE COMPLETE AND OPERABLE AT ALL TIMES. PROVIDE TEMPORARY DEVICES IN AREAS BEING WORKED ON TO ALLOW
FOR A WORKABLE SYSTEM DURING CONSTRUCTION.
J. PROVIDE MINIMUM [2] DAYS ADVANCE NOTICE OF OUTAGES. COORDINATE WITH OWNER FOR ALL ANTICIPATED OUTAGES. OWNER HAS FINAL APPROVAL ON TIME AND EXTENT OF POWER OUTAGE.
K. PROTECT EXISTING FURNITURE, CARPET AND OTHER APPURTENANCES FROM DAMAGE DURING THE
COURSE OF CONSTRUCTION. REPLACE DAMAGED ITEMS FREE OF COST TO THE OWNER.
L. ITEMS WHICH ARE INTENDED TO BE REMOVED, RELOCATED, OR SALVAGED ARE SHOWN AS FULL- TONE LINES ON DEMO PLANS. ALL OTHER ITEMS ARE INTENDED TO REMAIN IN PLACE AND ARE
SHOWN AS HALF-TONE.
A. DRAWINGS DO NOT SHOW TOTAL NUMBER OF CONDUCTORS FOR CIRCUITS REQUIRED. PROVIDE
THE NUMBER OF CONDUCTORS NEEDED TO COMPLETE INSTALLATION OF SPECIFIC EQUIPMENT
FURNISHED. A GREEN GROUND CONDUCTOR, SIZED PER NEC, SHALL ALSO BE INCLUDED IN EACH
CIRCUIT. ADJUST CONDUCTOR SIZES AS NEEDED TO LIMIT VOLTAGE DROP TO LEVELS
RECOMMENDED BY NEC. THE CONTRACTOR SHALL ALSO COORDINATE INSPECTIONS…