Addendum and Acknowledgement 2.doc

DOC document 2 MB Posted

Attached to
Plasma Enhanced Chemical Vapor Deposition (PECVD) System
Solicitation number
5-B176-Q-00304-00
Issued by

About this file

This is Addendum #2 to Request for Quotation (RFQ) 5-B176-Q-00304-00 issued by Argonne National Laboratory's Procurement Department on February 25, 2025, for a Plasma Enhanced Chemical Vapor Deposition (PECVD) System.

The addendum provides answers to four technical questions regarding system specifications: the required chamber base pressure must be < 2x10^-6 Torr; a pressurized reactor design with an exterior chamber at lower pressure than the process chamber is required; the hot wall PECVD reactor chamber wall temperature must be 300 degrees C; and confirmation that this is not a replacement for an existing system at CNM. The original RFQ seeks domestic manufactured items meeting Buy American Act requirements, with proposals due by February 28, 2025 at 5:00pm CST. Evaluation will be conducted using Lowest Price Technically Acceptable criteria.

View the file

Other files

File Type Posted
Addendum and Acknowledgement Doc.doc DOC document
ANL-70A Reps and Certs Short-04-FEB-2025.PDF PDF
ANL-70 Request for Quotation.doc DOC document
Appendix D8C-12-DEC-2023.PDF PDF
Appendix A-71COM-13-FEB-2023.PDF PDF
PECVD SOW 2.13.25.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

MOLLY MURTAUGH

Procurement Specialist 3 Office of the Chief Financial Officer

Procurement Department

Argonne National Laboratory

9700 South Cass Avenue, Bldg. 201

Lemont, IL 60439

(630)252-3094 mmurtaugh@anl.gov

02/25/2025

TO ALL OFFERORS:

Subject:

Request for 5-B176-Q-00304-00 No.

Addendum No. 002 Attached herewith for your information and use in preparing your proposal for subject project is one (1) copy of Addendum No.002.

Kindly acknowledge receipt of this Addendum by signing and returning one (1) copy of this letter and including this signed document with the completed Request for Quotation by the offer deadline.

Sincerely, MOLLY MURTAUGH

Procurement Specialist 3 Procurement Department

Receipt Acknowledged:

Company:

By:

cc:

File

CF

Addendum No.

TO

REQUEST FOR 5-B176-Q-00304-00

FOR

Argonne National Laboratory

BLDG. 201 OCF

9700 S. Cass Ave.

Lemont, IL 60439

02/25/2025 This addendum is being issued to incorporate the addition of questions and responses as follows:

1. What is the required chamber base pressure prior to deposition?

< 2x10^-6 Torr.

2. If the reaction chamber can reach a base pressure of 10 E-06 Torr pressure and the loadlock chamber can also reach a base pressure of 10E-06 Torr and the net direction of gas flow is from the reaction chamber to the loadlock chamber, will the pressurized reactor design with an exterior chamber at a much lower pressure than the process chamber still be required to prevent contamination from entering the deposition zone?

YES, "the pressurized reactor design with an exterior chamber at a much lower pressure than the process chamber" is required.

3. What is the required chamber wall temperature of the hot wall PECVD reactor?

300 degree C.

4. Is the system to replace an existing system with the same specifications at CNM?

No.

U. S. Department of Energy laboratory managed by UChicago Argonne, LLC