PD-AFLCMC-EZPT-17-01
23 November 2016
DISTRIBUTION STATEMENT A. Approved for public release; distribution is unlimited.
PRODUCT DESCRIPTION
COMPUTED RADIOGRAPHY IMAGING SYSTEM
1.0 SCOPE. This Product Description (PD) provides the minimum requirements for a Computed Radiography (CR) digital radiographic imaging system suitable for Nondestructive Inspection of a variety of metallic and non-metallic products used in military aircraft maintenance environments, which may be located at fixed military bases or deployed to austere forward operating locations. These imaging systems shall directly digitize and process radiographic images recorded on both hard and flexible Imaging Plates (IPs). The method of performing radiography shall be by using a storage phosphor system on both hard and flexible IPs. This PD provides for production evaluation as described herein.
2.0 CLASSIFICATION. The system described in this PD is classified as a complete CR system, employing state of the art photostimulable phosphor IPs as the collection media for radiographic data in the same manner as chemically-processed x-ray film used for the purpose of nondestructive inspection/testing/evaluation, except that these IPs shall be capable of being erased and re-used. An automated, programmable method of extracting the data and subsequently erasing the IP; and a method for displaying and storing the images shall be provided.
2.1 This PD contains system specifications necessary to meet Department of Defense (DoD) requirements for radiographic detection of foreign objects, water entrapment, and alignment (i.e., general purpose use) at 2.5 line pairs per millimeter (lp/mm), fatigue crack detection at 6.0 lp/mm, and welder certification at 8.0 lp/mm, in both the scanning and transverse directions using the process control devices described in paragraph 5.6.
2.2 The CR system shall be a table-top, portable CR system with Integral Eraser capable of processing industry standard flexible IPs from 5” x 7” to 14” x 17”, as well as custom sized flexible IPs as small as 2” x 4” and Original Equipment Manufacturer (OEM)-specific hard cassettes. Metric equivalent IP sizes are acceptable.
2.3 Contract Line Item Numbers associated with this PD are:
2.3.1 CLIN 0001AA – Production Evaluation Article
2.3.2 CLIN 0001AB – Production Article (Training Environment)
2.3.3 CLIN 0001AC – Production Article
2.3.4 CLIN 0002 – Data per DD1423
2.3.5 CLIN 1001 – Production Article – Option 1
Beneficial comments, recommendations, additions, deletions, clarifications, etc. and any other data which may improve this document should be sent to AFLCMC/EZPT, 5259 Warehouse Road, Bldg 2211; Room 1, Tinker AFB OK 73145-3307 or email to aflcmc-ezpt-ndio@us.af.mil.
AMSC N/A FSC 6635
3.0 APPLICABLE DOCUMENTS.
The following document(s) form a part of this document to the extent specified herein. Unless otherwise specified, the issues of documents, which are DoD adopted are those listed in the issue of the DoDISS cited in the solicitation. Unless otherwise specified, the issues of documents not listed in the DoDISS are the issues of the documents cited in the solicitation.
DEPARTMENT OF DEFENSE
MIL-PRF-28800F – Test Equipment for use with Electrical and Electronic Equipment, General Specification For, dated 24 June 1996
MIL-STD-810G – Test Method Standard, Change 1, dated 15 April 2014
DI-TMSS-80527C, Commercial Off-the-Shelf Manuals and Associated Supplemental Data, dated 12 August 2011
DI-MISC-80557, Commercial Support Documentation, dated 1 April 1988
All available at http://quicksearch.dla.mil/
DEPARTMENT OF THE AIR FORCE
TECHNICAL ORDER 33B-1-2, NONDESTRUCTIVE INSPECTION – General Procedures and Process Controls; Work Package 106 01 – COMPUTED RADIOGRAPHY PROCESS CONTROL, dated 1 April 2014. This document is currently DISTRIBUTION STATEMENT C: Distribution authorized to U.S. Government agencies and their contractors. A redacted copy is provided as Appendix 1.
AMERICAN SOCIETY FOR TESTING AND MATERIALS (ASTM)
E 746, Standard Practice for Determining Relative Image Quality Response of Industrial Radiographic Imaging Systems E 747, Standard Practice for Design, Manufacture and Material Grouping Classification of Wire Image Quality Indicators (IQI) Used for Radiology E 1742, Standard Practice for Radiographic Examination E 2007, Standard Guide for Computed Radiography E 2033, Standard Practice for Computed Radiology E 2339, Standard Practice for Digital Imaging and Communication in Nondestructive Evaluation
(DICONDE)
E 2445, Standard Practice for Performance Evaluation and Long-Term Stability of Computed Radiography Systems E 2446, Standard Practice for Manufacturing Characterization of Computed Radiography Systems E 2738, Standard Practice for Digital Imaging and Communication Nondestructive Evaluation (DICONDE) for Computed Radiography (CR) Test Methods
(Application for copies should be addressed to ASTM, 100 Barr Harbor Drive, West Conshohocken, PA 19428.)
SOCIETY OF MOTION PICTURE AND TELEVISION ENGINEERS
RP-133, Medical Diagnostic Imaging Test Pattern, see https://www.smpte.org/
4.0 SYSTEM. The CR system described in this document shall be a turnkey system and shall include the following:
4.1 Flexible phosphors IPs.
4.2 Flexible exposure cassettes, and if required, rigid cassettes or adapters for processing.
4.3 Hard cassette IPs.
4.4 Scanner (Image Reader/Erasure Unit/Processing Cassettes).
4.5 System Controller/Workstation.
4.6 Display Monitor.
4.7 Software for image acquisition, image viewing, database, data archiving, data sharing, and printing.
4.8 Hard-shell, appropriately padded, shipping containers/carrying cases suitable for storing and shipping the unit with accessories for all system components and accessories.
4.9 Process control devices.
4.10 Large format printer.
5.0 REQUIREMENTS AND SALIENT CHARACTERISTICS. The CR system shall have the following requirements and salient characteristics as a minimum unless specified in the solicitation:
5.1 The CR System. It shall be capable of producing radiographic images equal to or exceeding the general radiographic requirements of the American Standards for Testing and Materials specification E 1742, Standard Practice for Radiographic Examination. The CR System shall conform to the following requirements:
5.1.1 Environment:
5.1.1.1 The CR system shall be capable of operating in an industrial environment. Suitable mechanisms shall be incorporated to protect hardware, including IPs, from damage due to airborne particulates. All proposals shall describe these protective measures and a plan for accomplishing MIL-STD-810G, change 1, paragraph 510.6, testing by an independent laboratory within 60 days of contract award. Prior testing accomplished to a previous revision of MIL- STD-810 is acceptable, but shall be identified in the proposal.
5.1.1.2 The CR system shall be capable of being used in an area with ambient lighting using hard cassettes or subdued lighting with flexible IPs.
5.1.1.3 The CR system shall be capable to perform to the requirements of this PD throughout temperatures from 50 to 110 degrees Fahrenheit at 95% relative humidity, and from 0 to 7,000 feet above sea level. All proposals shall include a plan for accomplishing MIL-PRF-28800F, paragraph 3.8.2, testing by an independent laboratory within 60 days of contract award. Prior testing accomplished to a previous revision of MIL-PRF-28800 is acceptable, but shall be identified in the proposal.
5.1.1.4 The CR system shall be capable of being stored from 0 to 150 degrees Fahrenheit. All proposals shall include a plan for accomplishing MIL-PRF-28800F, paragraph 3.8.2.1, testing by an independent laboratory within 60 days of contract award. Prior testing accomplished to a previous revision of MIL-PRF-28800 is acceptable, but shall be identified in the proposal.
5.1.1.5 The CR system shall be capable of operation when supplied with 100-250 Volt Alternating Current – 50 or 60 Hertz (100-250 VAC – 50/60 Hz) automatically selected. All proposals shall include a plan for accomplishing MIL-PRF-28800F, paragraph 3.5.1, testing by an independent laboratory within 60 days of contract award. Prior testing accomplished to a previous revision of MIL-PRF-28800 is acceptable, but shall be identified in the proposal.
5.1.2 Performance:
5.1.2.1 The CR system shall demonstrate system performance through successful completion of a production evaluation as described in section 5.9.
5.1.2.2 The CR system shall be capable of scanning and erasing a minimum of 15 each 14” x 17” IPs per hour at 50 micron sampling resolution.
5.1.3 Size and Weight. The complete CR system in shipping cases shall not exceed a volume of 125 cubic feet, with no dimension larger than five (5) feet. The complete CR system in shipping cases shall weigh no more than 500 pounds.
5.2 Hardware. The CR system shall contain the following sub-components:
5.2.1 Imaging Plates, and Hard and Flexible Exposure Cassettes. The types and generation of included IPs shall be identified in the deliverable CR Operational and Technical Manual and Illustrated Parts Breakdown. All flexible IPs shall have a minimum bend radius of 3” or smaller without damage. The following IPs and cassettes shall be included:
5.2.1.1 General Purpose IPs: The following general purpose IPs shall be included: Two (2) flexible 14” x 17” IPs, two (2) flexible 8” x 10” IPs, and two (2) flexible 5” x7” IPs, each with flexible exposure cassettes. Two (2) hard 14” x 17” cassettes with IPs shall be included. One (1) hard cassette with an IP with an area between 5” x 7” and 8” x 10.
5.2.1.2 High Resolution IPs: The following high resolution IPs shall be included: Two (2) flexible 14” x 17” IPs, two (2) flexible 8” x 10” IPs, and two (2) flexible 5” x7” IPs, each with flexible exposure cassettes. Two (2) hard cassettes with 14” x 17” IPs shall be included. One (1) hard cassette with an IP with an area between 5” x 7” and 8” x 10. For the Production Evaluation Article only (CLIN 0001AA), two (2) flexible 2.75” x 8” IPs, each with flexible exposure cassettes, as shown in Appendix 2.
5.2.1.3 Ultra High Resolution IPs: The following ultra high resolution IPs shall be included:
Two (2) flexible 14” x 17” IPs, two (2) flexible 8” x 10” IPs, and two (2) flexible 5” x7” IPs, each with flexible exposure cassettes. One (1) hard cassette with 14” x 17” IP. One (1) hard cassette with an IP with an area between 5” x 7” and 8” x 10”. (Note these are often referred to as “blue” IPs.)
5.2.1.4 Manufacturer’s recommended cleaning system to include fluid, wiping cloths, etc.
necessary to clean one hundred (QTY 100) 14x17 inch IPs shall be included.
5.2.2 Scanner (Image Reader/Erasure Unit): Drum scanners are specifically excluded from consideration due to USAF history of operational damage to Imaging Plates. Drum scanners require the imaging plate to be placed active side down on a half-cylinder and then mechanically moved over the laser-scanning window. Any particulates (e.g., dust, sand, etc.) on the drum cylinder or imaging plate will cause scratches to the imaging plate and cause an unacceptable decrease in the operational life of the imaging plate and increase in life-cycle costs.
5.2.2.1 The scanner shall operate with current USAF inventory of commercially-available IPs, including prior-generation IPs, specifically including NSNs 6635-01-584-0226 (CAGE 1HHA0, VIRTUAL MEDIA INTEGRATION, P/N IP0507HR), 6635-01-584-0227 (P/N CAGE 1HHA0, IP1012HR), 6635-01-584-0233 (CAGE 1HHA0, P/N IP1417HR), 6635-01-606-4999 (CAGE
4QX22, CARESTREAM HEALTH, P/N 8360521) and 6635-01-656-1291 (CAGE 4QX22, P/N 1008564).
5.2.2.2 The scanner shall be capable of automatically reading and erasing the IP in a single operation.
5.2.2.3 The scanner shall read the IP at a 14 Bit minimum gray scale depth.
5.2.2.4 The scanner shall feature integral diagnostics and communication of error conditions with the workstation.
5.2.2.5 The scanner shall be capable of acquiring data from the IP at 50 and 100 micron sampling. Finer sampling may be necessary to achieve welder certification and crack detection requirements.
5.2.2.6 The scanner shall be capable of handling “cut” IPs (i.e. IPs of non-rectangular, unique shapes). It is acceptable to use a special or custom adapter to scan “cut” IPs and a minimum of two adapters shall be included. Appendix 2 provides a drawing of the “cut” IP to be included and used in the production evaluation.
5.2.2.7 Cassettes required for processing shall be included as required. If the processing cassette is not integral to the scanner, a minimum of two (2) processing cassettes shall be provided to accommodate the IP sizes listed in 5.2.1. Any special tools required to insert/extract IP in cassette or hard cassette in scanner shall be included.
5.2.2.8 The scanner shall include an erase function which may be operated independent from the scan/erase cycle.
5.2.3 System Controller/Workstation:
5.2.3.1 The workstation shall include a CPU, keyboard, mouse, and monochrome display monitor. The workstation shall also contain a separate connection for a VGA projector for training or color display monitor for file maintenance; this projector or monitor is not a contract deliverable. Hardware type software keys or dongles shall not be required for workstation functionality. As delivered, there shall not be any user level password. Passwords shall be used for Supervisor (intermediate) and Super User (higher level) access; however, all passwords shall be provided to the cognizant USAF Engineer at AFLCMC-EZPT-NDIO@us.af.mil at time of shipment. Passwords shall not be shipped with the unit.
5.2.3.2 The workstation shall serve as the Controlling Programming device and driver system for the scanner (Image Reader/Eraser Unit) and shall serve as a viewing device to view, manipulate, interpret, and store images.
5.2.3.3 The CPU shall include:
5.2.3.3.1 Microsoft Windows Operating System, version 7 or later.
5.2.3.3.2 Internal dual-format DVD/CD writer where content can be viewed on other DVD-ROM drives, DVD players and DVD recorders.
5.2.3.3.3 Minimum of four (4) USB ports available for transferring data to/from external devices.
5.2.3.3.4 Two (2) internal hard drives in a RAID 1 (mirrored) configuration, each with a minimum of two (2) terabyte (2 TB) storage capacity. Separate hard drive partitions for system files, system recovery, and image storage may be used.
5.2.4 Display Monitor. The CR system shall contain one (1) each 21” (533mm) minimum diagonal measurement, flat panel, five (5) mega pixel (minimum resolution) monochrome display monitor for viewing the computed radiograph. Minimum resolution shall be 2560 by 2048 pixels.
5.3 Software. All software updates made by the manufacturer during the duration of the contract shall be included at no additional charge. All proposals shall include detailed recommended procedures for maintaining configuration control of software.
5.3.1 Image acquisition software.
5.3.1.1 Shall provide a means to record pertinent data related to the image including:
5.3.1.1.1 Shall have the ability to record and organize file data per ASTM E2339 and E2738, (DICONDE). The proposal shall include a Conformance Statement that accurately describes DICONDE implementation. Any deviations from DICONDE standards shall be identified in the proposal.
5.3.1.1.2 Shall have the ability to record technique information including but not limited to kV, mA, time, source-to-detector distance, and general notes.
5.3.1.1.3 Shall automatically record DICONDE data including but not limited to scanner settings, date/time, and operator.
5.3.1.1.4 This metadata shall be automatically recalled with the image during subsequent reviews.
5.3.1.2 Shall activate the start of the Read/Erase cycle.
5.3.2 Image viewing software. The CR system image analysis software shall be capable of leveling the image in no more than two (2) operations. All image manipulations and calculations shall be conducted with at least 14-bit grayscale pixel depth. Image viewing software shall provide the following tools as a minimum:
5.3.2.1 Window/Level. Ability to manually adjust the Window (contrast) and Level (brightness).
5.3.2.2 Pan. Ability to manually pan the image.
5.3.2.3 Magnification:
5.3.2.3.1 Ability to manually adjust the magnification (zoom) up to a minimum of 400%, including the ability to display the image at 1:1 pixel mapping.
5.3.2.3.2 Ability to display a thumbnail view indicating the displayed portion of the image relative to the entire image when using digital magnification.
5.3.2.3.3 Ability to display the digital magnification (zoom) level.
5.3.2.3.4 Ability to magnify an area of interest with a “floating” magnifier tool to 300% magnification.
5.3.2.4 Gray Value Readout. Ability to measure and display the gray scale value (pixel value) of individual pixels.
5.3.2.5 Measurement:
5.3.2.5.1 Ability to obtain calibrated measurements in linear, area, and angle, with user selectable metric or English format, with the ability to compensate for any geometric enlargement.
5.3.2.5.2 Ability to generate a region of interest (ROI). This tool shall display the size of the ROI (e.g. x pixels by y pixels), and as a minimum display the statistical mean, standard deviation, and signal-to-noise ratio of the pixel values in the region of interest.
5.3.2.5.3 Ability to generate a line profile capable of displaying the pixel values along a user defined line as a graph/histogram.
5.3.2.6 Image Manipulation and Enhancements:
5.3.2.6.1 Ability to initially display images without image enhancements and then allow the user to apply filters, edge enhancements or contrast enhancements as desired.
5.3.2.6.2 Ability to display the image in either negative or positive polarity.
5.3.2.6.3 Ability to change image display orientation, including the ability to mirror and rotate the image at 90 degree intervals.
5.3.2.6.4 Ability to include inspector generated annotations on the image including but not limited to text, circle/box outlines, and arrows.
5.3.2.6.5 Ability to display any selected DICONDE data on the image.
5.3.2.6.6 Enable the user to select image processing filters for inspection.
5.3.2.6.7 Ability to display the image menu filters applied to the image.
5.3.3 Display monitor software:
5.3.3.1 Visual test pattern. Monitor software shall include the SMPTE RP133 test pattern, and the ability to display the test pattern on the high resolution monochrome display monitor.
5.3.3.2 Self-diagnostic capability. Monitor software shall include self-diagnostic software, including but not limited to, display brightness checks on the high resolution monochrome display monitor.
5.3.4 Printing software:
5.3.4.1 The CR system shall have the ability to print to the included large-format printer described in paragraph 5.7, with both positive and negative formats, of various sizes and up to at least 14” x 17”.
5.3.4.2 The CR system shall have the ability to print from typical office laser printers, such as the Hewlett-Packard Laserjet series; as well as other specialty DICONDE or DICOM (Digital Imaging and Communications in Medicine) compatible printers offered in commercial CR systems at no additional cost to the government.
5.4 Database, data archiving and data sharing. The CR System’s Database and Archiving program shall be designed and able to meet the database and archiving maintenance requirements of aircraft and other weapons system of the DoD.
5.4.1 Filing structure. The CR system shall be able to logically search image inventory for image selection and retrieval. Data fields will be provided at or before the time of the Government’s Production Decision.
5.4.2 File format:
5.4.2.1 DICONDE. Image files shall be in a DICONDE compliant file format per ASTM E2339 and ASTM E2738. Any deviations shall be identified in the proposal.
5.4.2.2 DICONDE Headers. The following data shall be stored in the DICONDE Headers:
• CR system manufacturer and model
• Viewing software and version number
• Imaging plate serial number or other unique identifier
• Pixel pitch (in microns)
• Unit Name (Nomenclature)
• Aircraft identification
• Component identification
• Technique parameters (kV, mA, time, distance)
• Other notes as required
• Date of Inspection
• Inspecting Activity
5.4.2.3 Other file formats. Image files shall be able to be filed and exported in JPEG, BITMAP, and TIFF.
5.4.3 Data Saving/Archival:
5.4.3.1 Software shall retain the original (raw) data. The original data shall be capable of being re-processed and re-displayed using the same or another process algorithm without having to take another radiographic exposure.
5.4.3.2 The full bit depth of the image, as read from the IP, shall be retained in the saved image.
5.4.3.3 Digital Annotation shall not permanently alter the nature of the image or hinder interpretation of an area within the image.
5.4.4 Data Sharing:
5.4.4.1 The CR system shall be capable of saving and retrieving data directly to/from the system’s hard drive, as well as CDs, DVDs, and external hard drives.
5.4.4.2 The CR system shall be capable of exporting images to other DICONDE Compliant workstations.
5.4.4.3 Images exported to other DICONDE network shall be able to be viewed, manipulated, interpreted, and stored to meet global mission demands.
5.4.4.4 Users shall be able to transmit DICONDE images through DICONDE network without loss of image resolution when viewed at comparable display screens.
5.4.4.5 The CR system software shall have the ability to allow third-party viewing of images, without the requirement of having a system workstation. No special, proprietary software or workstation shall be required to operate the viewing software.
5.4.4.5.1 Viewing software shall include the minimum tools: window/level, magnification, and pan.
5.4.4.5.2 Viewing software shall be created on a CD/DVD or USB interfaced mass storage device.
5.4.4.6 The CR system software shall have the ability to transmit images through the network or posting on the web.
5.4.4.7 A DICONDE compliant image of the line pair gauges of the Process Control Phantom described in paragraph 5.6 shall be transmitted to the AF NDI Office at AFLCMC-EZPT- NDIO@us.af.mil at time of proposal submission to confirm compliance. File size is limited to 10MB. Alternatively, the proposal can include a website URL to download the image file, or a CD-ROM may be sent to AFLCMC/EZPT-NDIO, 5259 Warehouse Road, Bldg 2211; Room 1, Tinker AFB OK 73145. Notional data shall be stored under the DICONDE headers.
5.4.5 Diagnostics: The software shall enable the user to perform system diagnostics.
5.4.6 Passwords: No passwords shall be required to scan, manipulate, retrieve, or store an image. A Supervisor password shall be utilized to allow file maintenance, image erasure, adjusting system pre-sets (e.g. IP size, PMT and resolution settings) and system recovery. A Super-user password shall be utilized for system maintenance. No passwords shall be shipped with the unit, but will be transmitted to the AF NDI Office at AFLCMC-EZPT-NDIO@us.af.mil at time of shipment to manage software configuration.
5.4.7 System Recovery Software. A system recovery disk(s) shall be included to return the system to an as delivered software configuration. A recovery partition on the hard drive may be used.
5.5 Shipping Containers: The CR system shall include hard-shell carrying cases and be suitable for storing and shipping the unit with accessories. Carrying cases shall be appropriately padded to ensure potential shipping damage to the complete systems would be minimized. All proposals shall include a plan for accomplishing a 46 centimeter (18”) drop in the hard cases conducted IAW the procedure in MIL-PRF-28800F, paragraph 4.5.5.4.2, by an independent laboratory within 60 days of contract award. Prior testing accomplished to a previous revision of MIL- PRF-28800 is acceptable, but shall be identified in the proposal. Cases shall have wheels and handles. (Note - it is desirable for cases be of similar size to facilitate stacking & palletizing for shipment.) Shipping cases shall not be included in CLIN 0001AB – Production Article (Training Environment).
5.6 Process Control Devices. The CR System shall include a Process Control Phantom. NSN
6635-01-555-4372 (CAGE 6KCS9, ASES LLC, DBA FIELD AEROSPACE, CP 405-219-3420,
http://www.fieldaero.com, P/N A0295004002) is the initial configuration and drawings are available upon request. It is available commercially at CAGE 3DF34, NDT SUPPLY, CP 913- 685-0675, http://ndtsupply.com/kowotest-usaf-cr-phantom-panel.html, P/N RT-KT-CR-Phan- USAF. The Phantom in NAVAIR Manual 17-15-530, CAGE 6KCS9, P/N A0295224003, is acceptable. The CAGE 81346, AMERICAN SOCIETY FOR TESTING AND MATERIALS (ASTM), https://www.astm.org/, P/N E2445 Phantom may be acceptable if approved by the Contracting Officer to resolve excessive schedule delays. The CR System shall also include ASTM E 746 aluminum Equivalent Penetrameter Sensitivity (EPS) plaques and 0.75” thick aluminum absorber plate; the USAF approved supplier is CAGE 29884, RAY CHECK MANUFACTURING, CP 559-292-0115, http://www.ray-check.com/radiographic.html, P/Ns RIQR2024 and RIOI-AP. Other suppliers require approval by the Contracting Officer. The CR System shall contain the following Image Quality Indicators: a full ASTM E-1742 plaque-type penetrameter set meeting the requirements of NSN 6635-00-185-1983 (Air Force Master Set with AL, CU, FE, MG, SS, TI, CAGE 81346, P/N E1742), and CAGE 81346, P/N E-747 wire penetrameter sets A-C, in aluminum, copper, and stainless steel, and A-B in Inconel and titanium.
5.7 Large Format Printer. A large-format printer, capable of replicating a 14” x 17” radiograph images on 14” x 17” paper shall be included. The printer shall have a minimum spatial resolution of 300 pixels per inch. Throughput shall be a minimum of fifteen (15) copies per hour of 14” x 17” size. Fifty (50) sheets of 14” x 17” paper shall be included. Large format printer shall not be included in CLIN 0001AB – Production Article (Training Environment).
5.8 Production Article (Training Environment) - CLIN 0001AB. Production Article (Training Environment) shall include a workstation with monitor, as described above. Imaging Plates and Cassettes per paragraph 5.2.1, Scanner per paragraph 5.2.2, Shipping Containers per paragraph 5.5, Process Control Devices per paragraph 5.6, and Large Format Printer per paragraph 5.7 are not required. A recovery software disk shall be included to return the workstation to its original software configuration.
5.9 Production Evaluation. The CR system (i.e. scanner, workstation, IPs, display monitor) shall pass the production evaluation specified in the AFMC Form 260 included in the Solicitation using a Spellman/Lorad LPX-160 X-ray Apparatus Set (1.5mm focal spot), and the following:
5.9.1 Successful completion of the CR Systems for USAF Airframe Crack Detection Applications and Development of System Performance Requirements, Report AFRL/RXS 12- 002 (redacted copy provided as Appendix 3) and the following:
5.9.1.1 Not achieving 2.5 lp/mm or greater spatial resolution in both directions using general purpose IPs is cause for rejection.
5.9.1.2 Not achieving 6.0 lp/mm or greater spatial resolution in both directions using high resolution IPs is cause for rejection.
5.9.1.3 Not achieving 1.28% or better Equivalent Penetrameter Sensitivity (EPS) with maximum exposure of 5mA and 100 seconds (500 mA-sec) at 65kV and 48” Source-to-Detector distance using high resolution IPs is cause for rejection.
5.9.1.4 Not achieving 8.0 lp/mm or greater spatial resolution in both directions using ultra high resolution (blue) IPs is cause for rejection.
5.9.1.5 Not achieving 1.28% or better EPS, with maximum exposure of 5mA and 180 seconds (900 mA-sec) at 75kV and 48” Source-to-Detector distance using ultra high resolution (blue) IPs is cause for rejection.
5.9.1.6 Not achieving 6.0 lp/mm or greater spatial resolution in both directions using the high resolution IP shown in Appendix 2 is cause for rejection.
5.9.2 The scanner shall be tested 10 times each without jamming using the following types and sizes of IPs. Any jamming or damage to the IP, or any necessary special or custom adapter, is cause for rejection.
5.9.2.1 The “cut” high-resolution IP shown in Appendix 2.
5.9.2.2 The 14x17 flexible general purpose IP.
5.9.2.3 The hard cassette with 14x17 general purpose IP.
5.9.2.4 The 8x10 flexible high resolution IP.
5.9.2.5 The hard cassette with 14x17 high resolution IP.
5.9.2.6 The 5x7 flexible ultra-high resolution IP.
5.9.2.7 The hard cassette with 14 x 17 ultra-high resolution IP.
5.9.3 The CR scanner erasure function shall be tested by using a 14” x 17” high-resolution IP (50 micron equivalent). The erasure procedures outlined in ASTM E2445 Section 8.3.1 (Paragraphs 8.3.1.1, 8.3.1.2, and 8.3.1.3) and Section 9.2.11 (Paragraph 9.2.11.1) shall be followed with the following conditions. The test shall be performed at a kV value of 160 kV. Using an IP that has been erased, an exposure shall be taken at 160 kV. This shall be accomplished using an absorber plate that covers approximately one half of the imaging plate and results in 5-10% maximum achievable mean linear PV in the region covered by the absorber plate. In the free beam area, the exposure shall result in 80-90% of the maximum achievable mean linear PV. Erase the IP. Scan the IP after erasure. Document the orientation of the IP during exposure and processing on the technique. Ensure the same orientation is maintained for subsequent tests. Erasure shall be calculated by the PV difference method. Erasure shall be calculated from the image taken as a result of scanning the erased IP. Determine the PV difference between the mean linear PV of the free exposed area and the mean linear PV of the area covered by the absorber. The ROI for each area should cover as large a region as possible within each area. The PV difference divided by the maximum achievable PV shall be ≤ 2%. Perform this test a total of 10 times. Upon completion of the test, any residual image with a PV difference divided by the maximum achievable PV above the allowed 2% shall be re-evaluated by repeating the tests with another IP.
Any residual image with a PV difference divided by the maximum achievable PV above the allowed 2% on the second IP is cause for rejection.
5.9.4 Successful completion of TO 1B-52H-36, Wing Trailing Edge Flaptrack Inspection.
Failure to visualize the 2-2T hole in the image quality indicator is cause for rejection.
5.9.5 Successful completion of TO 1C-130A-36, Aft Fuselage Side Structure from FS 770 to 968 (Tube Attach Fittings). Failure to visualize the 2-2T hole in the image quality indicator is cause for rejection.
5.9.6 Successful completion of TO 1F-15A-36, Tip closure rib (P/N 68A115105, without straps) for cracks, L/R first ramp rib support (P/N 68A327122), and L/R horizontal stabilator inboard leading edge rib (P/N 68A210130) for cracks. Failure to visualize the 2-2T hole in the image quality indicator is cause for rejection.
5.9.7 Successful radiographic inspection of a government owned welder certification tube butt joint weld specimen that contains porosity weld defects using existing T.O. 00-25-252 procedures. Failure to visualize defects, equivalent to film radiograph provided by the government, is cause for rejection.
6.0 DOCUMENTATION. See DD Form 1423, Contract Data Requirements List.
6.1 One (1) paper copy of the commercial operation and maintenance instructions, including calibrations procedures and adjustments, as applicable, shall be shipped with each system. An electronic copy shall be loaded on the workstation. All documentation shall be releasable for third-party repairs. A draft copy shall be provided with the production evaluation article for engineering approval. Two (2) paper and two (2) electronic copies of the final version shall be sent to the USAF Equipment Specialist and NDIO Cognizant Engineer. The CR System’s Operation and Technical Manual shall as a minimum include:
6.1.1 CR Systems Technical Operation and Maintenance.
6.1.2 CR Systems Illustrated Parts Breakdown including compatible IPs available commercially.
6.1.3 Service and Calibration Maintenance Schedule Requirements.
6.1.4 A listing of all available accessory items & services, warranty information, and Vendor/Contractor’s Point of Contact.
6.1.5 See DI-TMSS-80527A and DI-MISC-80557 for additional information.
6.2 Radiographic images of the serialized process control devices described in paragraph 5.6, which are actually delivered with each system, shall be included on the workstation to document the manufacture’s quality control processes. The image shall contain serial numbers of the process control devices, the workstation, scanner, and date formed using metallic lead (Pb) letters & numbers, or hand-scribed on metallic lead foil.
6.3 A system recovery disk(s) shall be included to return the CR system to an as delivered software configuration. A recovery partition on the hard drive may be used. All passwords shall be sent to the NDIO Cognizant Engineer at AFLCMC-EZPT-NDIO@us.af.mil at time of shipment to manage software configuration. Passwords shall NOT be shipped with the unit.
6.4 One (1) Digital Video Disk (DVD) providing hardware and software familiarization shall be included with each system. One (1) DVD providing step-by-step instructions to accomplish process control procedures provided in TO 33B-1-2 using the process control devices described in paragraph 5.6 shall be included with each system. Two additional copies of each DVD shall be sent to the Contracting Officer for delivery to the Equipment Specialist and Cognizant Engineer.
6.5 A DICONDE compliant image of the line pair gauges of the Process Control Phantom described in paragraph 5.6 shall be transmitted to the AF NDI Office at AFLCMC-EZPT- NDIO@us.af.mil at time of proposal submission to confirm compliance. File size is limited to 10MB. Alternatively, a CDROM may be sent to AFLCMC/EZPT-NDIO, 5259 Warehouse Road, Bldg 2211; Room 1, Tinker AFB OK 73145. Notional data shall be stored under the DICONDE headers.
7.0 WARRANTY. Each CR system shall be warranted from defects in materials and workmanship for a period of 12 months from the date of acceptance.
8.0 NOTES. This section contains information of a general or explanatory nature which may be helpful, but is not mandatory.
8.1 National Stock Numbers (NSNs). The generic NSN for this equipment is 6635-01-523-5766.
The Contracting Officer will provide a new actual NSN for equipment procured to this Purchase Description. Deliveries shall not be shipped using NSN 6635-01-553-4539 or 6635-01-573- 1461.
8.2 Ordering data. The contract or order should specify the following:
a. Title, number, and date of this description.
b. Product conformance provisions.
c. The solicitation will provide for any required independent-laboratory production testing, or government testing, as appropriate.
d. Packaging requirements deviations from paragraph 5.5.
8.3 Key Words.
Computed Radiography Digital Radiography Imaging Plate Photostimulable Phosphor DICONDE (Digital Imaging and Communication Nondestructive Evaluation)
8.4 Changes from previous issue. Marginal notations are not used in this revision to identify changes with respect to the previous issue due to the extent of the changes.
9.0 APPENDICES:
9.1 Appendix 1, Technical Order 33B-1-2, WP 106 01, Redacted
9.2 Appendix 2, Custom Imaging Plate for Production Evaluation
9.3 Appendix 3, CR Systems for USAF Airframe Crack Detection Applications and Development of System Performance Requirements, Report AFRL/RXS 12-002, Redacted