Attachment 0010 - ANUSM-677 Calibration Parameters and Equipment.pdf

PDF 122 KB Posted

Attached to
FINAL SOLICITATION FOR SPECTRUM ANALYZER W31P4Q-22-R-0034
Solicitation number
W31P4Q-22-R-0034
Issued by

About this file

This document contains calibration parameters and equipment specifications for a spectrum analyzer procurement. The spectrum analyzer must accurately measure input signals from 9 kHz to 26.5 GHz to support various Army systems. Key performance specifications are provided for features such as frequency accuracy, resolution bandwidth accuracy, input attenuator accuracy, reference level accuracy, residual FM, noise sidebands at various offsets, frequency readout accuracy, marker frequency count accuracy, frequency span accuracy, and frequency response. Displayed average noise level specifications are also outlined at different frequency ranges. The spectrum analyzer will be fielded to the Army under the nomenclature Analyzer, Spectrum: CM-523/U in support of avionic, missile, ordnance, and signal equipment testing requirements.

View the file

Other files

Show all 20

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Table 1. Calibration Description.

Test instrument parameters Performance specifications

10 MHz Reference Accuracy Model E4407B

All other models

Frequency: 10 MHz Stability/Settability: 5 X 10-7

Frequency: 10 MHz Stability/Settability: 1 X 10-8

Resolution Bandwidth Accuracy and Selectivity

Range: 1 Hz to 300 Hz Accuracy: 10% Range: 1 kHz to 3 MHz Accuracy: 15% Range: 5 MHz Accuracy: 30% Selectivity: 60 dB/3 dB bandwidth ratio: 15:1 Range: 3 kHz to 3 MHz RBW Accuracy: ±0.3 dB Range: 5 MHz RBW Accuracy: ±0.6 dB Range: 10 Hz to 300 Hz RBW Accuracy: ±0.3 dB Range: 1 Hz to 3 Hz RBW Accuracy: ±0.3 dB

Input Attenuator Accuracy Range: 0 to 60 dB, referenced to 10 dB input attenuation Accuracy: 0 to 5 dB, ±0.3 dB; 10 dB, reference; 15 dB to 60 dB, ±(0.1 dB + 0.01 X attenuator setting)

Reference Level Accuracy Model E4407B

All other models

Range: -149.9 dBm to maximum mixer level + attenuator setting Accuracy: -10 dBm to > -60 dBm, ± 0.3 dB

-60 dBm to > -85 dBm, ± 0.5 dB

Accuracy: -10 dBm to > -60 dBm, ± 0.3 dB -60 dBm to > -85 dBm, ± 0.5 dB -85 dBm to > -90 dBm, ± 0.7 dB

Residual FM Model E4407B

Options 1DR and 1D5

Accuracy: 150 Hz X N1(100 ms)

2 Hz X N p-p in 20 ms

Calibration Description. TI parameters and performance specifications which pertain to this calibration are listed in table 1.

Table 1. Calibration Description - Continued.

Noise Sidebands

Models E4407B, E4407B-E57, E4407B-H57 (S/N prefixes

MY4510, SG4510, US4510)

All models (S/N prefixes

< MY4510, SG4510, US4510)

Models AN/USM-677, E4407B-H76, E4407B1D51DR (S/N prefixes

MY4510, SG4510, US4510)

Noise Sidebands � Wide Offsets Models E4407B, E4407B-E57, E4407B-H57

Option 120

Non option 120

Range: 1.0 GHz, 0 dBm 10, 20, 30 and 100 kHz offset Accuracy: -90 dBc @ 10 kHz offset

-104 dBc @ 20 kHz offset -110 dBc @ 30 kHz offset -118 dBc @ 100 kHz offset

-90 dBc @ 10 kHz offset -100 dBc @ 20 kHz offset -106 dBc @ 30 kHz offset -118 dBc @ 100 kHz offset

-98 dBc @ 10 kHz offset -100 dBc @ 20 kHz offset -106 dBc @ 30 kHz offset -118 dBc @ 100 kHz offset

Range: 1.0 GHz, 0 dBm 100 kHz, 1 MHz, 5 MHz, and 10 MHz offset Accuracy: -119 dBc/Hz @ 100 kHz offset

-133 dBc/Hz @ 1 MHz offset -135 dBc/Hz @ 5 MHz offset -137 dBc/Hz @ 10 MHz offset

Accuracy: -119 dBc/Hz @ 100 kHz offset -125 dBc/Hz @ 1 MHz offset -127 dBc/Hz @ 5 MHz offset -131 dBc/Hz @ 10 MHz offset

Frequency Readout Accuracy Frequency range: 9 kHz to 26.5 GHz2 Accuracy: ±[(Frequency indication X frequency error) + 0.5% of span + (Span ÷ sweep points � 1)) + 15% of RBW + 10 Hz + 1 Hz X N2]

Marker Frequency Count Accuracy

Frequency range: 9 kHz to 26.5 GHz2 Accuracy: ±[ (marker frequency x reference frequency error )

+ counter resolution] Frequency Span Range: 100 Hz to 26.5 GHz2

Accuracy: ±[0.5% X span = 2 X span /(sweep points � 1)] Frequency Response (10 dB input attenuation)

Model E4407B

All other models

Frequency range: 9 kHz to 3 GHz Accuracy 0.46 dB Frequency range: 3 GHz to 6.7 GHz Accuracy 1.5 dB Frequency range: 6.7 GHz to 26.5 GHz2

Accuracy: 2.0 dB

Frequency range: 10 kHz to 26.5 GHz2 Accuracy: 2.0 dB

Table 1. Calibration Description � Continued.

Displayed Average Noise Level Frequency Range (dBm)

Model E4407B

All other models

10 MHz to 2 GHz 2 GHz to 6 GHz 6 GHz to 12 GHz 12 GHz to 22 GHz2 22 GHz to 26.5 GHz2

10 MHz to 2 GHz 2 GHz to 6 GHz 6 GHz to 12 GHz 12 GHz to 22 GHz2 22 GHz to 26.5 GHz2

-116 -112 -111 -107 -106

-135 -131 -130 -126 -125

Residual Responses (no signal at input, 0 dB input attenuation)

Range: 150 kHz to 6.7 GHz Accuracy: <-90 dBm

Absolute Amplitude Accuracy Measured at 50 MHz Reference level: -20 dBm Accuracy: 0.34 dB

SECTION II

EQUIPMENT REQUIREMENTS

4. Equipment Required. Table 2 identifies the specific equipment to be used in this calibration procedure. This equipment is issued with Secondary Transfer Calibration Standards Sets, AN/GSM-287 and AN/GSM-440.

Table 2. Minimum Specifications of Equipment Required.

Common name Minimum use specifications

Manufacturer and model (part number)

ATTENUATOR (FIXED) 10 dB:

Frequency range: 50 MHz to 18 GHz Accuracy: ±0.3 dB

Weinschel, Model 9918-10 dB (9918-( ) dB)

FREQUENCY COUNTER Frequency range: 10 MHz Resolution: 0.001 Hz

Spectracom, Model CNT-

90\OCXO\RA (CNT-90\OCXO\RA)

MEASURING RECEIVER Frequency range: 300 MHz Accuracy: ±300 Hz dBm range: -9.7 to -10.3 Accuracy: ±0.075 dBm

Measuring receiver system N5531S consisting of: Spectrum Analyzer, Agilent Model E4440A (E4440A), Power meter, Agilent Model E4419B (E4419B), or Power meter, Agilent Model N1911A (N1911A) and Sensor module, Agilent Model N5532A opt. 518 (518)

MULTIMETER Range: 1000 V dc Accuracy: ±2 V dc

Fluke, Model 8846A (8846A/G)

Table 2. Minimum Specifications of Equipment Required � Continued.

POWER SPLITTER Frequency range: 10 MHz to 18 GHz Weinschel, Model 1870A (7916839) RF REFERENCE SOURCE Frequency range: 10 kHz to 1 GHz

Amplitude range: -80 to +12 dBm Accuracy: ±0.02 dB per 10 dB step Range: 1.0 GHz, 0 dBm 10 kHz-100 kHz offset Accuracy: -118 dBc

Fluke, Model 9640A-LPN

(9640A-LPN)

SYNTHESIZED SIGNAL

GENERATOR

Frequency range: 10 MHz to 18 GHz Power range: 0 to -15 dBm Accuracy: ±1.525 dBm

Anritsu, Model MG3694C/2B/5/6/28B

(MG3694C/2B/5/6/28B)

TIME/FREQUENCY

WORKSTATION

Frequency: 10 MHz Accuracy: 0.0025 ppm

Symmetricom Inc., Model 1510-602-

142-AR (1510-602-142-AR)