ITEM #6:

GilAir Plus (or equivalent) Air Sampling Pump Kit with 5 pumps

Specifications

Performance

Firmware

Current Version: 2.5.4 (6/20/17)

Flow Range

Constant flow 20 - 5,000 cc/min. without external adaptors

Constant pressure 1 - 5,000 cc/min without external adaptors

Flow Modes

Constant Flow or High and Low Flow Constant Pressure

Accuracy

Flow Display ± 5% or 3 cc/min of set flow, whichever is larger

Constant Flow Control ± 5% or 3 cc/min of set flow, whichever is larger

Constant Pressure Control ± 10% of back-pressure

Back Pressure Capability

(8-hour runtime) 5000 cc/min. up to 12” H20 back pressure

4000 cc/min. up to 20” H20 back pressure

3000 cc/min. up to 30” H20 back pressure

2000 cc/min. up to 30” H20 back pressure

1000 cc/min. up to 35” H20 back pressure

450 - 1000 cc/min. up to 40” H20 back pressure

20 - 449 cc/min. up to 40” H20 back pressure

Run Time 8 hour minimum

Flow Fault if flow changes exceed 5% within back pressure specifications, fault notification appears. If fault exceeds 30 seconds, pump shuts down. Selectable automatic fault recovery allows the pump to attempt restart every 3 minutes for up to 10 times or to hold until manual intervention

Environmental

Temperature Ranges

Operating 32°F to 113°F (0°C to 45°C)

Storage -4°F to 113°F (-20°C to 45°C)

Charging 41°F to 104°F (5°C to 40°C)

Humidity Ranges

Operating 5-95 %RH, non-condensing

Storage 5-98 %RH, non-condensing

GENERAL

Display Flow rate, sample time, and sample volume in actual conditions

Controls

Indicators Green and Red LED’s

Dimensions 4.3 x 2.4 x 2.4 inches (10.9 x 6.1 x 6.1 centimeters)

Weight 20.5 ounce (581 grams)

Warranty 2 years from original shipment, 5 years for pump keypad, and 1 year on rechargeable NiMH battery pack

ELECTRICAL

Power Options Removable rechargeable nickel-metal hydride (NiMH) battery pack, optional alkaline battery pack, or DC power

Battery Level Indicator On-screen icon displays battery level

Interface Connectors All pump interface is carried out via the Dock or keypad

Dock Functions Battery charging, computer interface, and auto-calibration

Charging Time under 3.5 hours

APPROVALS

EN 1232 Compliant Ingress Protection: IP54

Intrinsic Safety

US/Canada NRTL (FM) Class I, Div 1 Groups A, B, C, D, E, F, and G;

Class II, Div 1 Zone 0; Class III, T4

ATEX/IECEx

II 1 G

Ex ia IIC T4

CE

Applicable Directives ATEX 94/9/EC

EMC 2004/108/EC (RFI) LVD 2006/95/EC

RoHS Compliant

File details come from the government source that posted it.