ITEM #6:
GilAir Plus (or equivalent) Air Sampling Pump Kit with 5 pumps
Specifications
Performance
Firmware
Current Version: 2.5.4 (6/20/17)
Flow Range
Constant flow 20 - 5,000 cc/min. without external adaptors
Constant pressure 1 - 5,000 cc/min without external adaptors
Flow Modes
Constant Flow or High and Low Flow Constant Pressure
Accuracy
Flow Display ± 5% or 3 cc/min of set flow, whichever is larger
Constant Flow Control ± 5% or 3 cc/min of set flow, whichever is larger
Constant Pressure Control ± 10% of back-pressure
Back Pressure Capability
(8-hour runtime) 5000 cc/min. up to 12” H20 back pressure
4000 cc/min. up to 20” H20 back pressure
3000 cc/min. up to 30” H20 back pressure
2000 cc/min. up to 30” H20 back pressure
1000 cc/min. up to 35” H20 back pressure
450 - 1000 cc/min. up to 40” H20 back pressure
20 - 449 cc/min. up to 40” H20 back pressure
Run Time 8 hour minimum
Flow Fault if flow changes exceed 5% within back pressure specifications, fault notification appears. If fault exceeds 30 seconds, pump shuts down. Selectable automatic fault recovery allows the pump to attempt restart every 3 minutes for up to 10 times or to hold until manual intervention
Environmental
Temperature Ranges
Operating 32°F to 113°F (0°C to 45°C)
Storage -4°F to 113°F (-20°C to 45°C)
Charging 41°F to 104°F (5°C to 40°C)
Humidity Ranges
Operating 5-95 %RH, non-condensing
Storage 5-98 %RH, non-condensing
GENERAL
Display Flow rate, sample time, and sample volume in actual conditions
Controls
Indicators Green and Red LED’s
Dimensions 4.3 x 2.4 x 2.4 inches (10.9 x 6.1 x 6.1 centimeters)
Weight 20.5 ounce (581 grams)
Warranty 2 years from original shipment, 5 years for pump keypad, and 1 year on rechargeable NiMH battery pack
ELECTRICAL
Power Options Removable rechargeable nickel-metal hydride (NiMH) battery pack, optional alkaline battery pack, or DC power
Battery Level Indicator On-screen icon displays battery level
Interface Connectors All pump interface is carried out via the Dock or keypad
Dock Functions Battery charging, computer interface, and auto-calibration
Charging Time under 3.5 hours
APPROVALS
EN 1232 Compliant Ingress Protection: IP54
Intrinsic Safety
US/Canada NRTL (FM) Class I, Div 1 Groups A, B, C, D, E, F, and G;
Class II, Div 1 Zone 0; Class III, T4
ATEX/IECEx
II 1 G
Ex ia IIC T4
CE
Applicable Directives ATEX 94/9/EC
EMC 2004/108/EC (RFI) LVD 2006/95/EC
RoHS Compliant