ITEM #3:
Quest (or equivalent model) SoundPro Sound Level Meter DL-1 1/1 & 1/3 Octave
Datalogging RTA & Calibrator Meter
General:
Display Languages: Chinese, Czech, English, French, German, Italian, Japanese, Korean, Norwegian, Polish, Portuguese, Russian, Spanish, Swedish, and Turkish
User Interface: 10 push buttons and 4 soft keys, menu driven
Display Type: Transflective 128 X 64 Dot Matrix LCD with additional fiber optic backlighting
Conformance to Standards:
EMC Requirements: EN/IEC 61326-1(2005) Group 1, Class B Emissions/ Industrial
Location Immunity. CFR:47 (2008) Part 15 - Meets FCC Class B Emissions
Performance EN/IEC 61672-1(2002), ANSI S1.4 (R2006), Requirements: ANSI S1.43(R2007), EN/IEC 61260 (2001), ANSI S1.11 (R2009), (also meets requirements of former standards IEC 60651 and 60804) IEC60268-16 (2003) with Speech Intelligibility option
Safety Requirements: IEC61010-1 (2010)
Certifications: CE Mark, WEEE, RoHS
Measurements:
Parameters: SPL, LMax, LMin, LPk(peak), LEQ/LAVG, SEL, LN (selectable L1 to L99), TWA, Taktm, Taktmx, Dose, PDose, Exposure (Pa2H/Pa2S), LDN, CNEL, PTWA, L C-A
Ranges: 120 dB+ (A-weighted) total dynamic measurement range over 8 individual ranges of 90 dB (A-weighted) each (with filters - 80 dB ranges); overall measurement range 0 dB to 140 dB
Peak Range: Up to 143 dB using standard BK4936 microphone; higher with optional microphones and preamps
Frequency Weighting: A, C, Z and F (Flat)
Response Time: Fast, Slow, IEC Impulse
Exchange Rates: 3, 4, 5, and 6 dB
Criterion Level: 40 to 100 dB
Upper Limit Time Logging: 10 to 140 dB selectable
Run Modes: Level triggered run/pause, clock/date triggered power on and run for programmed duration, external logic input run/pause, and keypad initiated run/pause for programmed duration.
Measurement References: SPL: 114 dB Frequency: 1 kHz
Direction: 0 degrees using free-field response microphone
Octave and Third Octave Filters (optional) (base-10 bands, as recommended by
IEC61260 [2001])
Full Octave Filters: 11 bands with center frequencies from 16 Hz to 16 kHz)
Third Octave Filters: 33 bands with center frequencies from 12.5 Hz to 20 kHz
Calibration:
History: Complete calibration history with post study verification logged with calibration history
Specifications: All Specifications Subject to Change.
Logging and Storage
Logging: DL Models only. LMax, LMin, LPk(peak), LN, LEQ/LAVG may be logged at 11 selectable intervals from one second to 60 minutes to the included SD (secure digital) memory card. Use 3M™ Detection Management Software DMS to interpret data files
Summary Data: All session/study data is stored to the SD card.
Summary data may be interpreted with 3M™ Detection Management Software DMS, or exported to spreadsheet or XML file with an available utility
Memory: Accepts 32 MB to 32 GB SD memory cards.
Card included with all models and stores multiple summary sessions/studies and for setup storage (Contact factory for preferred SD card manufacturers)
Special Functions
Back Erase: Selectable 1 to 20 seconds removal of measurement data (data removed by back erasing and retained in session file)
Security: 4 digit code protection for Runs and Setups available Optional Acoustic: Noise
Criterion (NC) Curves, Preferred Noise Spectral Curves: Criterion (PNC) Curves, Room
Criterion (RC) Curves, Balanced Noise Criterion (NCB) Curves, Noise Rating (NR)
Curves, Audiometric Room Curves (per ANSI S3.1, per OSHA Hearing Conservation
Amendment, and per ISO Hearing Screen for Audiology Booths)
Optional Speech Firmware can be installed in the SoundPro
Intelligibility Function: series to allow the testing and evaluation of intelligibility of human speech through public address (PA), fire alarm and mass notification systems (MNS), the STI-PA method in accordance with IEC 60268-16 and NFPA 72 National Fire Alarm
Code. Results are in STI or CIS. On meter post-processing available
Optional Reverberation Used to measure decay time or acoustic decay
Time (RT-60): performance of a room or closed space
Ports and Connections
Power Jack: External power supply 9-16 VDC
AC/DC Output: 3.5 mm stereo (tip-AC, Ring-DC, Ring2-G
Electrical Characteristics
Batteries: 4 disposable AA alkaline cells, typically >10 hours continuous use without backlight (SLM only without filters activated); optional nickel metal hydride (NiMH) cells, typically 10+ hours (SLM only)
External DC Power Input: 100 – 240 VAC, 47-63 Hz transformed to 9 VDC
Standard Microphones: Class/Type 1 Precision – BK4936; Class/Type 2
General Purpose – QE7052; other optional types and sizes available from ¼" to 1" prepolarized or standard condenser types Microphone Polarization: Selectable 0 volts or 200 volts (Class/Type 1 models only) Microphone Sensitivity: Selectable nominal values in decibels relevant to 1 Volt/Pa Meter Input Impedance: 20 kΩ in series with 11 μF capacitance, with 100 pF capacitance to ground
Remote Cable: Will drive up to 15 meters of cable with negligible signal loss
Preamplifier: Removable preamp directly accepts ½" (0.52" or 13.2 mm) microphone;
other sizes require adapter Preamplifier Input Greater than 1 GΩ; less than 2pF
Impedance:
Physical Characteristics
Size: 3.1"W X 11.1"H X 1.6" thick (with preamp & microphone); 7.9 cm X 28.2 cm X 4.1 cm
Weight: 0.54 kg or 1.2 lbs. (including batteries)
Housing: Stainless fiber filled ABS polycarbonate with additional internal EM/RFI shielding
Tripod Mount: Standard photographic mount on rear accepts
¼"- 20 screw threads
Environmental
Temperature: Operating -10ºC to +50ºC (<± 0.5 dB effect); Storage -25ºC to +70ºC
Humidity: 10% to 90% RH, non-condensing
External Fields: Electric – 10 V/meter, 1 kHz modulated, 30 MHz – 1
GHz, <55 dBC; magnetic - 80 A/m, 50/60 Hz, no significant effect