ITEM #5:

Gilibrator (or equivalent model) Standard Flow Calibrator Kit

SPECIFICATIONS

General Specifications

Base Unit Construction: Painted aluminum sheet metal

Unit Orientation: Base: Horizontal, on flat surface

Controls: “ON” Button, “OFF” Button, “DELETE/RESET” Button

Indicators Display (LCD)

Charge Indicator (LED)

Display Data Flow (4-digit)

Average (4-digit)

Sample# (2-digit)

Display Ranges Flow: 0–9999

Average: 0–9999

Sample#: 0–99

Display Messages: Low Bat, Wait, Error 1, Error 2

Viewing Angle: 60°

Performance Specifications

Operating Temperature: 5° to 35°C (41° to 95°F)

Storage Temperature: 0° to 50°C (32° to 122°F)

Operating Humidity: 0–85 %RH, non-condensing

Storage Humidity: 0–100 %RH, non-condensing

Accuracy: Better than 1% (Wet Cell)

Electrical Specifications

Power Source (DC): Internal Battery Pack

Battery Type: Rechargeable

Battery Charge Time: 14 hours

Expected Battery Life: over 300 charge/recharge cycles

Input Voltage: 5.5–7.5 Vdc

Input Current: 50–60 mA

Transmission Link Requirements: RS-232

Interface Connectors: Charger Jack (2.1 mm barrel jack)

Connecting Jack – Sensor Block (DB-9)

Printer Jack (DB-25)