W81XWH20Q0053 Amendment 06.docx
DOCX document 32 KB Posted
- Attached to
- WRAIR Autoclaves RFQ
- Solicitation number
- W81XWH20Q0053
- Issued by
-
Department of the Army Medical Command
About this file
This document contains a request for quote for autoclaves. The Walter Reed Army Institute of Research is soliciting quotes from qualified vendors to provide twelve replacement autoclaves located across various rooms, as well as one autoclave and associated equipment for a dirty cage wash room. The autoclaves must meet specified dimensions and technical requirements including sterilization chambers sizes, touchscreen interfaces, programmable cycles, materials of construction and utilities requirements. Vendors are to provide load carts and other accessories as defined. The quote response deadline is not stated. The NAICS code is 339113 and the soliciting agency is the Department of the Army Medical Command.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Autoclaves Questions and Answers 3.docx | DOCX document | |
| W81XWH20Q0053 Amendment 05.docx | DOCX document | |
| W81XWH20Q0053 Amendment 04.docx | DOCX document | |
| Autoclaves Questions and Answers 2.docx | DOCX document | |
| Polimers (floor mat).jpeg | JPEG file | |
| Autoclaves Questions and Answers Final.docx | DOCX document | |
| 503-G fp.pdf | ||
| 503-3 fp.pdf | ||
| Floor type.pdf | ||
| W81XWH20Q0053 Amendment 03.docx | DOCX document | |
| 503-2 fp.pdf | ||
| GE03 Room Diagram.jpg | JPG image | |
| 503-1 fp.pdf | ||
| W81XWH20Q0053 Amendment 01.docx | DOCX document | |
| Solicitation W81XWH20Q0053.docx | DOCX document |
Show all 15
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
W81XWH20Q0053
SECTION SF 30 BLOCK 14 CONTINUATION PAGE
SUMMARY OF CHANGES
SECTION B - SUPPLIES OR SERVICES AND PRICES
CLIN 0003
The CLIN extended description has changed from:
Sterilizer 49"x64"x100" Single Door; Recessed Wall Room GE01
To:
Sterilizer 49"x64"x 85" to 100" Single Door; Recessed WallRoom GE01
SECTION C - DESCRIPTIONS AND SPECIFICATIONS
The following have been modified:
MECS
Minimal Essential Characteristics
Autoclaves (LOG)
The 12 Autoclaves are located in rooms 1N53, 2N10, 2N45, 2S21, 2W117, 3E13, 3W47, 3N57, 3N81, 3W24, 3W74 and GW58 (Autoclave located at GW58 is a Single Door, recessed in Modular Wall). Vendor shall replace existing autoclave with new autoclave with the following features:
· Shall not exceed more than 6” outside the footprint of the existing autoclave, footprint dimensions: 44” W x 76” H x 72” D
· Sterilization chamber shall be approximately 26” W x 36” H x 48” L
· Single door shall be either right hand swing or sliding
· Approximately 10” (7” is acceptable) touch screen operator interface panel
· Shall be insulated on all sides, ends and top to reduce hot surfaces and heat loss
· Shall have audible and visual alarms
· Vendor shall provide a steam filter
· Parts shall be non-proprietary
· Shall provide a vacuum source that reduces water consumption (Water circulation system that can decrease water usage by more than 90%)
· Shall have a Thermal Printer provided
· Shall provide a pressure relief valve to ensure rated pressure within the vessel is not exceeded
· Shall provide a signal when door seal is energized
· Shall provide an emergency stop button
· Shall add cooling water to ensure discharge temperature is below 140oF
· Shall provide up to 25 programmable cycles to include:
· Prevacuum cycles
· Gravity cycles
· Liquid cycles
· Leak Test cycle
· Shall allow cycle parameters to be selected and programmed by the operator
· Shall allow for operator to name cycles
· Shall provide a minimum of four custom configurable authorization levels
· Shall provide maintenance staff the ability to monitor the life cycle of parts
· Shall allow for visual monitoring of cycle phases
· Shall have two loading levels
· Shall have a minimum 316L SS chamber and 304 SS jacket
· Shall have cabinet enclosure panels
· Cabinet shall be enclosed
· Shall have a Minimum Available utilities Autoclave 1:
· 120V, 1Ph, 30A
· House steam 80 PSIG dynamic pressure
· Floor Drain
· Cold Water
Autoclave (VSP) The 1 Autoclave for VSP is located in room GE01.
Dirty Cage Wash Autoclave Features located in Room GE01:
· Autoclave shall allow four load carts to be rolled into chamber via a ramp
· Vendor shall provide 6 load carts that shall be no less than 48”W x 24”D x 60”H
· Load carts shall be 3-sided carts with a removable or hinged door
· Autoclave (excluding sliding door) shall not exceed a knockdown footprint of 70”W x 84”H x 120”L
· Autoclave shall include a single load/unload door
· Autoclave shall be placed along the left wall (see room diagram below). No mechanicals or need to access the side next to the wall shall be required
· Contractor shall install a new modular wall to separate the GE01 mechanical space from room GE03, dirty cage wash
· Sterilization chamber shall be a minimum of 48” W x 62” H x 85” to 100” L max
· Shall be insulated on all sides to reduce hot surfaces and heat loss
· Vendor shall provide a ramp not to exceed 5’W x 10’L with no more than a 10” rise to the chamber entrance
· Shall have touch screen operator interface panel on the load and unload side
· Parts shall be non-proprietary
· Vendor shall provide a steam filter
· Shall provide audible and visual alarms
· Shall provide a vacuum source that reduces water consumption
· Printer shall be provided
· Shall provide an emergency stop button
· Shall add cooling water to ensure discharge temperature is below 140oF
· Shall provide up to 15 programmable cycles to include:
· Prevacuum cycles
· Gravity cycles
· Liquid cycles
· Leak Test cycle
· Shall allow cycle parameters to be selected and programmed by the operator
· Shall allow for operator to name cycles
· Shall provide a minimum of three custom configurable authorization levels
· Shall provide maintenance staff the ability to monitor the life cycle of parts
· Shall allow for visual monitoring of cycle phases
· Shall have a minimum 316L SS chamber and 304 SS jacket
· Shall have the following minimum available utilities:
· 115V, 1Ph, 15A
· 120 VAC , 60HZ, single phase
· 1” NPT House steam 80 PSIG dynamic pressure
· 80 PSI compressed air
· 1 1/2 “ Floor Drain
· 1” NPT Cold Water 40-60 PSI, 20 GPM
· Enclosure dimensions (Room GE01): 133”W x 109” H x 144”D
(End of Summary of Changes) image1.wmf