Autoclaves Questions and Answers 2.docx
DOCX document 17 KB Posted
- Attached to
- WRAIR Autoclaves RFQ
- Solicitation number
- W81XWH20Q0053
- Issued by
-
Department of the Army Medical Command
About this file
This document contains questions and answers regarding autoclaves to be provided under Solicitation Number W81XWH20Q0053. The solicitation requests quotes for autoclaves according to specifications for the Walter Reed Army Institute of Research. Acceptable autoclaves must have 316L stainless steel chambers and 304L stainless steel jackets. Carbon steel jackets are not acceptable as a substitute. The document clarifies that one room will contain a single phase, 120 volt sterilizer and vacuum pump requiring 3 phase power. All other bulk sterilizers will include this 3 phase vacuum pump.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Autoclaves Questions and Answers 3.docx | DOCX document | |
| W81XWH20Q0053 Amendment 06.docx | DOCX document | |
| W81XWH20Q0053 Amendment 05.docx | DOCX document | |
| W81XWH20Q0053 Amendment 04.docx | DOCX document | |
| Polimers (floor mat).jpeg | JPEG file | |
| Autoclaves Questions and Answers Final.docx | DOCX document | |
| 503-G fp.pdf | ||
| 503-3 fp.pdf | ||
| Floor type.pdf | ||
| W81XWH20Q0053 Amendment 03.docx | DOCX document | |
| 503-2 fp.pdf | ||
| GE03 Room Diagram.jpg | JPG image | |
| 503-1 fp.pdf | ||
| W81XWH20Q0053 Amendment 01.docx | DOCX document | |
| Solicitation W81XWH20Q0053.docx | DOCX document |
Show all 15
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Questions and Answers
| # |
| Question |
| Answer |
| 1 |
| (#49 on the previous Q&A) The specification calls for the 12 autoclaves (LOG) is for 316L SS chamber and 304L SS jackets. If the vendor offers the same warranty and performance on a 316L SS chamber with carbon steel jackets, is that acceptable? Cost savings would be seen across 12 sterilizers if carbon steel jackets are allowed. |
| (Revised Answer) No, carbon steel jackets are not acceptable. |
| 2 |
| We noticed that the vacuum pump for the bulk autoclave (Room GE 01) has a 3 phase power tag on it but in the questions/answers (answer 34) it indicated that there is no 3-phase power, can you please confirm the power phase being supplied to this room? |
The sterilizer that is there is a 120v single phase unit. The box on the left. The vacuum pump for the sterilizer is 3 phase power. The box on the right. All bulk sterilizers will have this vacuum pump.