Exhibit B - Mandatory EquipmentV2.xlsx

XLSX spreadsheet 14 KB Posted

Attached to
IFB 27-190 – Operating Room Surgical Tables
Solicitation number
2138970
Issued by
Kings County, New York

About this file

This is Exhibit B (Mandatory Requirements and Detailed Specifications) from an Invitation for Bid (IFB 27-190) issued by the State University of New York at Downstate Health Sciences University (SUNY-DHSU) for the procurement and installation of operating room surgical tables. The project requires the replacement of twelve (12) obsolete Skytron operating tables with new universal powered operating tables. The scope of work includes safe removal and disposal of existing equipment, delivery and installation of new systems, assembly, functional testing, calibration, commissioning, and coordination with hospital staff to minimize operational disruption. Bidders must submit evidence of comparable experience, provide three hospital references with similar scope operations, demonstrate a minimum five-year history in the operating room surgical table business, and commit to a two-week trial period prior to final clinical decision. The contract term includes a two-year warranty period commencing from the installation date, with preventive maintenance, parts, labor, and software/hardware upgrades included during the warranty period.

Offerors must provide forty (40) hours of clinical in-service training covering both day and night shifts with basic and advanced user training, three technical training sessions for biomedical engineering staff, and certification of achievement upon completion. The solution must be hosted entirely within the United States with 24/7/365 system support and willingness to participate in the Remi Equipment Maintenance Management Program. All operating tables must meet specific technical requirements including a minimum patient capacity of 1,000 pounds in normal surgical position and 800 pounds at full powered articulation, radiolucent carbon-fiber modular tabletop, powered articulation capabilities, handheld wireless controllers, FDA clearance, and compliance with IEC 60601-1, IEC 60601-2, AAMI, and NFPA 99 standards. Installation must be performed by factory-trained, certified technicians, and the offeror must provide as-built drawings, operation manuals, warranty documents, service schedules, and source materials for ongoing education.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Sheet1

IFB 27-190 - Operating Room Surgical Tables - Exhibit B - Mandatory Requirements and Detailed Specifications
SUNY Downstate Health Siences University ( "SUNY-DHSU")
#PriorityRequirement/Specification TextIndicate Whether You Satisfy This Requirement (Yes/No)Indicate what page of your written proposal a detailed response to the requirement/specification can be found.
1MandatoryOfferors must submit evidence of its experience providing Operating Room Surgical Tables to institutions comparable in size and scope to SUNY-DHSU.
2MandatoryOfferors must submit the names and contact information of at least three (3) references from customers that are hospitals for which it has provided OR Surgical Tables and services of similar scale, purpose and scope of SUNY-DHSU.
3MandatoryOfferor has been in business providing OR Surgical Tables for no less than 5 years.
4MandatoryOfferor must provide peer-reviewed publications.
5MandatoryOfferor must be willing to provide two weeks trial period for at least one system prior to providing final clinical decision.
6MandatoryOfferor must have the ability to provide OR Surgical Tables , all maintenance and updates for the operation of the software system and all attendant services thereto, including, inter alia, implementation, set-up, configuration and training
7MandatoryOfferor must provide a detailed project plan, which includes a point-by-point written description of the proposed solution, including any explanatory drawings or screen shots, user acceptance testing scripts, and training workbooks.
8MandatoryOfferor must provide written standard operating procedures (SOPS).
9MandatorySolution must be completely hosted by Offeror within the United States.
10MandatorySystem must not require any installation of wiring to/from and/or within a patient rooms and installation must have minimum impact on the facility.
11MandatoryScope of Work: Replace 12 obsolete operating tables in the Operating Rooms. Work includes safe removal of existing Skytron tables; transport of removed equipment to a designated on-site location for pickup and disposal; supply and delivery of new operating table systems; handling, packaging, and mobilization; and compliance with applicable safety, environmental, infection-control, decontamination, and disposal requirements.
12MandatoryTechnical Requirements:
13MandatoryOperating Table Technical Specifications:

Quantity

- 12 Universal Powered Operating Tables, or comparable.

Performance Requirements

- Radiolucent carbon-fiber modular tabletop.

- Minimum patient capacity: 1,000 lb (454 kg) in normal surgical position.

- Minimum 800 lb at full powered articulation.

- Integrated powered or manually controlled kidney bridge; minimum 4 in. elevation.

Powered Table Motion

- Longitudinal slide: >=12 in.

- Height: approximately 23-48 in.

- Trendelenburg / Reverse Trendelenburg: +/-30 degrees.

- Lateral tilt: +/-20 degrees.

- Back section: +80 / -40 degrees.

- Leg section: +20 / -90 degrees.

- Removable head section.

Controls / Features

- Handheld wired/wireless controller.

- Table-mounted override controls.

- One-touch level return.

- Memory positions.

- Battery status indicator.

- Central locking brake.

- Emergency override and collision protection.

- Standard 25 mm accessory rails.

Electrical / Safety

- 100-240 VAC, 50/60 Hz.

- Rechargeable battery backup.

- FDA-cleared.

- IEC 60601-1 and applicable IEC 60601-2 compliant; AAMI and NFPA 99 compliant.

- Smooth, seamless, disinfectant-resistant surfaces.

14Or Comparable
15MandatoryOfferor must disclose any chronic issues with their system that impacts use and productivity
16MandatorySolution must comply with all applicable state, federal and local regulations.
17MandatoryOfferor must guarantee compliance with applicable regulations during the Contract term.
18MandatoryInstallation:
19Offeror must install the system as per the manufacturer specification, including assembly and installation of 12 tables, full functional testing, calibration, and commissioning, with coordination with hospital staff to minimize operational disruption.
20Offeror must provide an initial inspection certification report. Work must be performed by factory-trained, certified technicians and comply with hospital safety protocols, infection control, and regulatory standards, including cleanup and restoration of work areas.
21MandatoryWarranty Requirements:
22Offeror must provide two (2) years warranty on all aforementioned components.
23Warranty start from the date of the installation.
24Offeror must provide Preventive maintenance during warranty period.
25Preventive maintenance during warranty includes batteries, parts and labors
26Warranty covers parts and labor.
27Software and/or hardware upgrade included during warranty.
28MandatoryClinical In service:
29Offeror provides 40 hours of clinical in-service. The in-service should cover both day and night shifts.
30The in-service should accommodate basic training for general users, and advanced training for advanced users.
31MandatoryBiomedical Training:
32Three technical training sessions for biomedical engineering. Training must include connection, operation, maintenance, and troubleshooting of the system.
33Certification of achievement after successful completion of the training class.
34MandatoryExtended Service Requirements:
35Willingness to take part in Remi Equipment Maintenance Management Program
36System Support must be 24/7/365
37MandatoryOfferor must attend- in-person meetings as needed by SUNY-DMC over the term of the contract.
38MandatoryOfferor must describe solution enhancements planned over the next year and how often the solution is upgraded.
39MandatorySolution must provide source materials for ongoing education. Offeror must also provide all as-built drawings, operation manuals, warranty documents, and service schedules.