Exhibit B - Mandatory EquipmentV2.xlsx
XLSX spreadsheet 14 KB Posted
- Attached to
- IFB 27-190 – Operating Room Surgical Tables
- Solicitation number
- 2138970
- Issued by
- Kings County, New York
About this file
This is Exhibit B (Mandatory Requirements and Detailed Specifications) from an Invitation for Bid (IFB 27-190) issued by the State University of New York at Downstate Health Sciences University (SUNY-DHSU) for the procurement and installation of operating room surgical tables. The project requires the replacement of twelve (12) obsolete Skytron operating tables with new universal powered operating tables. The scope of work includes safe removal and disposal of existing equipment, delivery and installation of new systems, assembly, functional testing, calibration, commissioning, and coordination with hospital staff to minimize operational disruption. Bidders must submit evidence of comparable experience, provide three hospital references with similar scope operations, demonstrate a minimum five-year history in the operating room surgical table business, and commit to a two-week trial period prior to final clinical decision. The contract term includes a two-year warranty period commencing from the installation date, with preventive maintenance, parts, labor, and software/hardware upgrades included during the warranty period.
Offerors must provide forty (40) hours of clinical in-service training covering both day and night shifts with basic and advanced user training, three technical training sessions for biomedical engineering staff, and certification of achievement upon completion. The solution must be hosted entirely within the United States with 24/7/365 system support and willingness to participate in the Remi Equipment Maintenance Management Program. All operating tables must meet specific technical requirements including a minimum patient capacity of 1,000 pounds in normal surgical position and 800 pounds at full powered articulation, radiolucent carbon-fiber modular tabletop, powered articulation capabilities, handheld wireless controllers, FDA clearance, and compliance with IEC 60601-1, IEC 60601-2, AAMI, and NFPA 99 standards. Installation must be performed by factory-trained, certified technicians, and the offeror must provide as-built drawings, operation manuals, warranty documents, service schedules, and source materials for ongoing education.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Exhibit C - Cost Proposal Grid - IFB 27-190 -Surgical Tables.xlsx | XLSX spreadsheet | |
| Exhibit H - Question and Answer Form - Copy.xlsx | XLSX spreadsheet | |
| IFB 27-190 OR Surgical Tables_Final.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Sheet1
| IFB 27-190 - Operating Room Surgical Tables - Exhibit B - Mandatory Requirements and Detailed Specifications | ||||
| SUNY Downstate Health Siences University ( "SUNY-DHSU") | ||||
| # | Priority | Requirement/Specification Text | Indicate Whether You Satisfy This Requirement (Yes/No) | Indicate what page of your written proposal a detailed response to the requirement/specification can be found. |
| 1 | Mandatory | Offerors must submit evidence of its experience providing Operating Room Surgical Tables to institutions comparable in size and scope to SUNY-DHSU. | ||
| 2 | Mandatory | Offerors must submit the names and contact information of at least three (3) references from customers that are hospitals for which it has provided OR Surgical Tables and services of similar scale, purpose and scope of SUNY-DHSU. | ||
| 3 | Mandatory | Offeror has been in business providing OR Surgical Tables for no less than 5 years. | ||
| 4 | Mandatory | Offeror must provide peer-reviewed publications. | ||
| 5 | Mandatory | Offeror must be willing to provide two weeks trial period for at least one system prior to providing final clinical decision. | ||
| 6 | Mandatory | Offeror must have the ability to provide OR Surgical Tables , all maintenance and updates for the operation of the software system and all attendant services thereto, including, inter alia, implementation, set-up, configuration and training | ||
| 7 | Mandatory | Offeror must provide a detailed project plan, which includes a point-by-point written description of the proposed solution, including any explanatory drawings or screen shots, user acceptance testing scripts, and training workbooks. | ||
| 8 | Mandatory | Offeror must provide written standard operating procedures (SOPS). | ||
| 9 | Mandatory | Solution must be completely hosted by Offeror within the United States. | ||
| 10 | Mandatory | System must not require any installation of wiring to/from and/or within a patient rooms and installation must have minimum impact on the facility. | ||
| 11 | Mandatory | Scope of Work: Replace 12 obsolete operating tables in the Operating Rooms. Work includes safe removal of existing Skytron tables; transport of removed equipment to a designated on-site location for pickup and disposal; supply and delivery of new operating table systems; handling, packaging, and mobilization; and compliance with applicable safety, environmental, infection-control, decontamination, and disposal requirements. | ||
| 12 | Mandatory | Technical Requirements: | ||
| 13 | Mandatory | Operating Table Technical Specifications: |
Quantity
- 12 Universal Powered Operating Tables, or comparable.
Performance Requirements
- Radiolucent carbon-fiber modular tabletop.
- Minimum patient capacity: 1,000 lb (454 kg) in normal surgical position.
- Minimum 800 lb at full powered articulation.
- Integrated powered or manually controlled kidney bridge; minimum 4 in. elevation.
Powered Table Motion
- Longitudinal slide: >=12 in.
- Height: approximately 23-48 in.
- Trendelenburg / Reverse Trendelenburg: +/-30 degrees.
- Lateral tilt: +/-20 degrees.
- Back section: +80 / -40 degrees.
- Leg section: +20 / -90 degrees.
- Removable head section.
Controls / Features
- Handheld wired/wireless controller.
- Table-mounted override controls.
- One-touch level return.
- Memory positions.
- Battery status indicator.
- Central locking brake.
- Emergency override and collision protection.
- Standard 25 mm accessory rails.
Electrical / Safety
- 100-240 VAC, 50/60 Hz.
- Rechargeable battery backup.
- FDA-cleared.
- IEC 60601-1 and applicable IEC 60601-2 compliant; AAMI and NFPA 99 compliant.
- Smooth, seamless, disinfectant-resistant surfaces.
| 14 | Or Comparable | |
| 15 | Mandatory | Offeror must disclose any chronic issues with their system that impacts use and productivity |
| 16 | Mandatory | Solution must comply with all applicable state, federal and local regulations. |
| 17 | Mandatory | Offeror must guarantee compliance with applicable regulations during the Contract term. |
| 18 | Mandatory | Installation: |
| 19 | Offeror must install the system as per the manufacturer specification, including assembly and installation of 12 tables, full functional testing, calibration, and commissioning, with coordination with hospital staff to minimize operational disruption. | |
| 20 | Offeror must provide an initial inspection certification report. Work must be performed by factory-trained, certified technicians and comply with hospital safety protocols, infection control, and regulatory standards, including cleanup and restoration of work areas. | |
| 21 | Mandatory | Warranty Requirements: |
| 22 | Offeror must provide two (2) years warranty on all aforementioned components. | |
| 23 | Warranty start from the date of the installation. | |
| 24 | Offeror must provide Preventive maintenance during warranty period. | |
| 25 | Preventive maintenance during warranty includes batteries, parts and labors | |
| 26 | Warranty covers parts and labor. | |
| 27 | Software and/or hardware upgrade included during warranty. | |
| 28 | Mandatory | Clinical In service: |
| 29 | Offeror provides 40 hours of clinical in-service. The in-service should cover both day and night shifts. | |
| 30 | The in-service should accommodate basic training for general users, and advanced training for advanced users. | |
| 31 | Mandatory | Biomedical Training: |
| 32 | Three technical training sessions for biomedical engineering. Training must include connection, operation, maintenance, and troubleshooting of the system. | |
| 33 | Certification of achievement after successful completion of the training class. | |
| 34 | Mandatory | Extended Service Requirements: |
| 35 | Willingness to take part in Remi Equipment Maintenance Management Program | |
| 36 | System Support must be 24/7/365 | |
| 37 | Mandatory | Offeror must attend- in-person meetings as needed by SUNY-DMC over the term of the contract. |
| 38 | Mandatory | Offeror must describe solution enhancements planned over the next year and how often the solution is upgraded. |
| 39 | Mandatory | Solution must provide source materials for ongoing education. Offeror must also provide all as-built drawings, operation manuals, warranty documents, and service schedules. |