36C24419B0057-003.pdf
PDF 3 MB Posted
- Attached to
- 693-CSI-306 Install CT/SPECT in Room 220
- Solicitation number
- 36C24419B0057
About this file
36C24419B0057 S02 Siemmens Site Planning Drawing.pdf
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 36C24419B0057-0002000.docx | DOCX document | |
| 36C24419B0057-0002000.docx | DOCX document | |
| 36C24419B0057-0001004.pdf | ||
| 36C24419B0057-0001002.pdf | ||
| 36C24419B0057-0001000.docx | DOCX document | |
| 36C24419B0057-0001003.pdf | ||
| 36C24419B0057-0001001.pdf | ||
| 36C24419B0057-0001003.pdf | ||
| 36C24419B0057-0001000.docx | DOCX document | |
| 36C24419B0057-0001001.pdf | ||
| 36C24419B0057-0001004.pdf | ||
| 36C24419B0057-0001002.pdf | ||
| 36C24419B0057-004.pdf | ||
| 36C24419B0057-003.pdf | ||
| 36C24419B0057-002.pdf | ||
| 36C24419B0057-005.pdf | ||
| 36C24419B0057-001.docx | DOCX document | |
| 36C24419B0057-005.pdf | ||
| 36C24419B0057-004.pdf | ||
| 36C24419B0057-002.pdf | ||
| 36C24419B0057-001.docx | DOCX document | |
| 36C24419B0057-000.docx | DOCX document | |
| 36C24419B0057-000.docx | DOCX document |
Show all 23
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Healthcare
Design Construction Services Proposal No. 16-69-1071 Quote No. 1-96KNIL
110 Macalyson Court 919.468.7318 ph.
Cary, NC 27511 8 of 9 919-468-7321 fax
Siemens Preliminary Drawing
Drawing extract taken from Siemens preliminary drawing # 1600226RB dated January 21, 2016.
\M) "l
- X m'--
SCANNER
I
ROOM
1EO9A
RIGHT
PATIENT
LOAD
LEFT
PHS
PIVOT
THIS SET OF FINAL DRAWINGS IS REFLECTIVE OF THE
LATEST SALES CONFIGURATION. ANY CHANGES TO THIS
SALES CONFIGURATION MAY REQUIRE A REVISION TO
THIS PROJECT PLAN. IF REQUESTED, SIEMENS WILL
PRODUCE A REVISED SET OF FINAL DRAWINGS TO
REFLECT THE CHANGES, HOWEVER SIEMENS IS NOT
RESPONSIBLE FOR ANY CONSTRUCTION COSTS
ASSOCIATED WITH THE CHANGES THAT OCCUR FROM
THIS PLAN MODIFICATION.
VISUAL RADIATION INDICATOR
IS INCLUDED WITH THE SYMBIA
AND LOCATED ON THE PATIENT
BOOM ARM ACCORDING TO
IEC 60601-2-44 AND 21 CFR.
SIEMENS HIGHLY RECOMMENDS THE
CUSTOMER'S ARCHITECT DESIGNATES
SPACE FOR A HOT LAB, PATIENT
WAITING AREA, AND UPTAKE ROOM.
2 -
I
,,,,,---t
� v ,--__ 15 ltl----t20
IT IS THE RESPONSIBILITY OF THE
CUSTOMER/CONTRACTOR TO
PROVIDE A MEANS OF MOUNTING
THE PC TOWER OFF OF FINISHED
FLOOR FOR DAMAGE PROTECTION
AGAINST TIP-OVER, FLUIDS, IMPACT
ETC.
-__ j----------- COUNTER OR TABLE TO BE::::::-_u DESIGNED, SUPPLIED AND
CONTROL
1 _ _J _ _J ___ L...:s!�=1a=� i.J_� O
R
IEN
I
IO
N
:::3;:=... ________ �====....,_ _ ___:,ROOM
-'. I, I POINT
INSTALLED BY THE
CUSTOMER/CONTRACTOR.
I
TF (STEP DOWN TRANSFORMER) MA; @ g (,;;'\, _0 718•
BE LOCATED IN THE CEILING BY '-1'1
THE CUSTOMER/CONTACTOR.
----7 �-il!il9==,i� COLLIMATOR CARTS
H TO BE LOCATED PER
I I.!) CUSTOMER'S
DISCRETION.
____ _J
ARCHITECTURAL EQUIPMENT PLAN
ROOM MEASUREMENTS
ALL ROOM MEASUREMENTS AND ROOM DETAIL SPECIFICATIONS MUST BE
VERIFIED ON SITE PRIOR TO BEGINNING ANY CONSTRUCTION WORK.
I I
{ 23'-11"
NOISE LEVEL
SYSTEM COMPONENT DECIBEL LEVEL
(AT 3' -3" DISTANCE)
SYMBIA INTEVO EXCEL, INTEVO 2, 68
I NTEVO 6 AND I NTEVO 1 6 GANTRY
FRONT PHS (PATIENT TABLE) 60
UPS FOR IMS <45
1) NOISE DEPENDS ON THE ROOM TEMPERATURE AND THE
PROCESSOR LOAD.
5'-10 3/16"
- SYMBIA.N�
:i,_ WORKPLACEI I I ... � LOCATED AT ILJ ==a' CUSTOMER'S
L I nI�1scRET10
SCALE: 1 / 4" - 1 '-0"
ATTENTION:
-T--JIS [,:;.,\'iil\C 1:: [:E':;1,;1•.JE=: T:=: ')>,1F1=::;.H Tl) FEi•.Tl,F-C::: ,\hJ[I EOl_ll-,HEW RE'�'-IREl'/ C }IT':: -·FE'::E·rE[1 -IT 1·:; :;i::,:·,-:i1'/'-.•IC}.J[IE:·, T -F\T THE ::;i:J.,1Er\lS [1::;,,'-'iil'\J:::: 3E 11·-Ji:i=:F- --CIP.-\E[i 1.1,'IH- T-E c c·-_:.-p1_1c-1,::·\1
,\T -H:: Tllv'E o=- THC:IP rRE-'i\F-,i\Trn•J Sll•,JC:: B::,rn ---JESE =-/\::1,::,:;i:. i\F:E ::,_IB_.:CCT TC =::::x;1•,J [:,:c._.H::r\w; =-cc;,_ =s::FE:;:::\:::E 1-.,D[:1=r.c.TIC\, n -::\ A:;,:: l·-ICT - c :::: USEC =-:1R ,: :or·JsTc;;l_,[T[:1-.j l·L.Rl'(I':_:;:::::
-Tl s :::[T CT FV,\S R:F·TS[\-S;; CU/F'L[T[ :::,:T �:- :ns_::: ?l·D s ICII_,_[:, l'-IDT ][ �[PA�;,Ll_-:D.
EQUIPMENT
NO DESCRIPTION SMS WEIGHT
SYM (LBS)
(2) SYMBIA INTEVO 6 GANTRY W/COWMATORS @ 7,413
12) FRONT PHS 0
13) � SYMBIA INTEVO-ACC wrrH AOC - PRODUCTIVITY PACKAGE 0
4 UNDER THE FLOOR PHS CABLE STANDARD 0
5 PHS EXTENDED PIVOT 0 �
6 REAR PHS WITH SNAG 0
7 PATIENT BOOM SWING ARM 0
8 LINE CONNECTION BOX @
19 UPS FOR SPECT ®
110 � TRANSFORMER REQUIRED WITH UPS FOR SPECT (SPS) ®
11 TVSS SURGE PROTECTION UPS FOR SPECT 0
12 COLLIMATOR CART (EMPTY) (2) 0 '-'
(9 IMAGE CONSTRUCTION SYSTEM FOR SYNGO Ml
(ACQUISITION) WORKPLACE
® IMAGE RECONSTRUCTION SYSTEM FOR SYNGO Ml
(ACQUISITION) WORKPLACE
15 CONTROL AND KEYBOARD 0
16 1 B" MONITOR 0
17 SYNGO Ml (ACQUISITION) WORKPLACE UPS FOR IMS ® '-' STANDARD COMPONENT
� DEDICATED RECONSTRUCTION
STANDARD INTEVO 6
SYSTEM WORKPLACE - ®
@) DEDICATED RECONSTRUCTION- STANDARD INTEVO 6
SYSTEM WORKPLACE MONITOR 0
� DEDICATED RECONSTRUCTION SYSTEM WORKPLACE
KEYBOARD - STANDARD INTEVO 6
21 DVD @
2 2 SYMBIA.NET WORKPLACE CPU 8
I 2J SYMBIA.NET WORKPLACE KEYBOARD AND MONITOR 0
MAGNETIC FIELD PRECAUTIONS
THE PRESENCE OF MAGNETIC FIELDS IN THE VICINITY OF
EQUIPMENT MAY HAVE AN ADVERSE EFFECT. IT IS THE
CUSTOMER'S RESPONSIBILITY TO VERIFY THAT THE FOLLOWING
VALUES ARE NOT EXCEEDED.
MAXIMUM ALLOWABLE
MAGNETIC FIELD
DEVICES
1.0ml (10 GAUSS) COMPUTERS, MAGNETIC DISK DRIVES
0.2ml (2 GAUSS) SIEMENS CT PORTION OF SYMBIA
SIEMENS CT SCANNERS
0.15ml (1.5 GAUSS) COLOR MONITOR (FOR LIQUID CRYSTAL
DISPLAYS ONLY)
0.1 ml (1.0 GAUSS
AC OR DC)
SIEMENS HAS ESTABLISHED THE UPPER
LIMITS OF MAGNETIC FIELD EXPOSURE
FOR THE SYMBIA DETECTORS
MAGNETIC FIELDS SHOULD BE MEASURED PRIOR TO DELIVERY.
STATE AGENCY REVIEW
PRIOR TO SIEMENS EQUIPMENT INSTALLATION, APPROVAL OF
CONSTRUCTION OR STRUCTURAL MODIFICATIONS UTILIZING X-RAY FOR
DIAGNOSTIC OR THERAPEUTIC PURPOSES, MUST BE OBTAINED BY THE
CUSTOMER FROM THE APPROPRIATE STATE AGENCY, IF APPLICABLE.
FINISHED ROOM HEIGHT
SYMBIA INTEVO EXCEL, INTEVO 2,
INTEVO 6, INTEVO 16
SYMBIA INTEVO EXCEL, INTEVO 2,
INTEVO 6, INTEVO 16 WITH CEILING
MOUNTED COMPONENT OTHER THAN
RADIATION ON LAMP
MINIMUM 8' -0"
MINIMUM 8'-2"
MAXIMUM 12' -0"
CONSIDER THE WARNING LIGHT WILL BE PLACED ON TOP OF
THE PATIENT BOOM. ANY OTHER CEILING MOUNTED COMPONENT
MUST BE PLACED AS TO NOT COLLIDE WITH WARNING LIGHT.
-.-\L_ =·, l·A C:l·-1'311jf·._ '-':· SHC\,\11, ,Jf·J T -I':: [,pf'.,\1 r·.IG ,'-FE =--:.,_,:·,•/ =-l·J SH C:[, ·::1_ , :;::=-f<'E'::
-T--1[: DPt\\,' --.c; [(1E:::: l•J1T rp:,v1c,:: =;::!{,l-',Tl1:\ S-IEL[, ,:; F.E(1I_IIPEHEUS =-o:;, <-R.,'-'( /J-1 :i .-'-':-�.::(: /\TE[1
EIJ,_,lrl•iEW THE ·:._.':,T(r,1ER ': :;:::::re :-EL:: FCiP ('(J\l'::IJ_-lhlC 1,'.'I-H:,, PEC-':--E=;::::=: -:.,x q�,1--J F'I -✓s C[:T - :1 SF'[CII ( (AC . .'.\TIJl·J Pl!T[CTrn-1
2,512
415.3
I
LEGEND
BTU/HR DIMENSIONS (INCHES)
t-----,c--�-.-�---1
TO AIR w D H
24,574 93 84 1/2 90 1/2
31 1/8 97 1/2 23 3/16
31 1/8 97 1/2 23 3/16
1,365 29 1/2 11 3/4 32
1,024 10 28 3/8 17 7/8
8 6 10 47 3/8 32 5/8 47 1/2
2,389 8 22 18
• 8 22 18
18 3/8 2 5/8 14 13/16
5 19 17 1/4
• 8 20 3/4 18
18 3/8 2 5/8 14 13/16
1,400 8 21 17 1/2
• 17 5/8 18 5/16 17
REMARKS
6,826 BTU ON STANDBY.
WORST CASE WEIGHT 7,963
LBS. WITH (2) HIGH ENERGY
COLLIMATORS AT 275 LBS.
EACH
MAXIMUM HEIGHT 41"
WEIGHT CALCULATED WITH 1
SET LOW AND MEDIUM
ENERGY COLLIMATORS.
UNDER FLOOR
EXTENDED PIVOT - 45
DEGREES
ON FLOOR
ON FLOOR
ON FLOOR
CUSTOMER SUPPLIED PRIOR
TO INSTALLATION. SEE
POWER SCHEDULE
ON TOP OF UPS FOR SPECT
WORST CASE 1330 LBS.
WITH 1 SET HE AND 1 SET
ME
OFF FLOOR.
OFF FLOOR. *STU'S INCL.
WITH ICS
ON CUSTOMER'S COUNTER
ON CUSTOMER'S COUNTER
OFF FLOOR.
OFF FLOOR. *BTU'S INCL .
WITH ICS
ON CUSTOMER'S COUNTER
ON CUSTOMER'S COUNTER
ON CUSTOMER'S COUNTER
OFF FLOOR
*BTU'S INCL. WITH SNW CPU
SITE READINESS GUIDELINES
THE FOLLOWING GENERAL CONDITIONS ARE NECESSARY TO HAVE
THE STATUS OF "READY SITE":
1) PROPER POWER AVAILABLE AT SIEMENS EQUIPMENT POWER
CABINET LOCATION.
2) AIR CONDITIONING/HUMIDIFICATION SYSTEMS COMPLETE,
TESTED, AND FUNCTIONING PROPERLY ACCORDING TO SIEMENS
SPECIFICATIONS.
3) PROPER LIGHTING INSTALLED AND FUNCTIONING.
4) PLUMBING COMPLETE EXCEPT FOR ANY FINAL CONNECTIONS
TO SIEMENS EQUIPMENT. LINES MUST BE FLUSHED AND LEAKED
TESTED.
5) ALL CABLE TRAYS/DUCTS/CONDUITS CORRECTLY SIZED AND
INSTALLED IN THE CORRECT LOCATIONS ACCORDING TO THE
SIEMENS DRAWINGS.
6) ALL REINFORCEMENT PLATES INSTALLED AS REQUIRED.
7) ROOM FOR EQUIPMENT INSTALLATION AND IMMEDIATE VICINITY
IS DUST-FREE AND IS TO REMAIN SO FOR THE DURATION OF
THE INSTALLATION.
8) A SECURE AREA, APPROXIMATELY 1 o·x 10·, IS AVAILABLE AT
EQUIPMENT DELIVERY FOR PARTS AND INSTALLATION TOOLS.
9) CUSTOMER SUPPLIED CAMERAS AND PROCESSORS INSTALLED.
10) MODEM OUTLET OR ROUTER, VOICE AND DATA TELEPHONE
LINES INSTALLED.
11) WALLS TO BE PRIMED AND PAINTED, FLOORS TO BE TILED
CEILINGS TO BE INSTALLED.
IF THESE CONDITIONS ARE NOT MET, THE SIEMENS PROJECT
MANAGER AND THE DESIGNATED SIEMENS INSTALLATION
SUPERVISOR SHALL RESCHEDULE THE INSTALLATION START DATE.
NOTE: ADDITIONAL COST MAY BE INCURRED BY THE
CUSTOMER/CONTRACTOR WHEN THE SIEMENS SITE READINESS
GUIDELINES ARE NOT MET, AND DELIVERY DATES NEED TO BE
RESCHEDULED.
CASEWORK & ACCESSORY NOTES
1) ALL CASEWORK IS EITHER EXISTING OR IS TO BE DESIGNED.
DETAILED, FURNISHED AND INSTALLED BY THE CUSTOMER AND/OR
CONTRACTOR. FOLLOW DESIGN RECOMMENDATIONS INCLUDED
HEREWITH, AS THEY ARE ESSENTIAL FOR THE SUCCESSFUL
INSTALLATION & OPERATION OF THE SIEMENS EQUIPMENT.
2) ALL FURNITURE (CHAIRS, ETC.) FOR THE CONTROL ROOM ARE TO
BE PROVIDED BY THE CUSTOMER.
-·Fl-\. =-CT ''ih \J/',l-,F�': D/•\/ID I YH.,\JJ
E_· :,'":11:1 I) ,:,::,.2 'JT 1
o---+-----+---------- \/Mt.IL: ::::-T
ARCHITECTURAL NOTES
1) ALL PRELIMINARY EQUIPMENT LAYOUTS SUBMITTED BY SIEMENS
MEDICAL SOLUTIONS, INC. (SMS HEREAFTER) ARE BASED ON THE
RECOMMENDED SPACE NECESSARY FOR THE OPERATION AND
SERVICEABILITY OF THE EQUIPMENT BEING PROPOSED. SMS WILL NOT
SUBMIT AN EQUIPMENT LAYOUT THAT IS NOT IN THE BEST INTIEREST
OF BOTH THE CUSTOMER AND SMS. ALL EQUIPMENT LAYOUTS ARE
BASED EITHER ON AN ACTUAL SITE LOCATION SURVEY OR
ARCHITECTURAL DRAWINGS SUPPLIED TO SMS. SMS WILL NOT BE
RESPONSIBLE FOR ANY ALTERATIONS THAT ENCROACH WITHIN
DESIGNATED SAFETY AND SERVICE CLEARANCE ZONES AS INDICATED
ON DRAWINGS (IE. PIPE CHASES, VENTILATION DUCTS, CASEWORK, AND
SOFFITS, ETC.) MADE BY THE CUSTOMER OR REQUIRED BY A
CUSTOMER'S ARCHITECTURAL FIRM ONCE PRELIMINARY DRAWINGS HAVE
BEEN SUBMITTED AND APPROVED. DO NOT ALTER ANY SPECIFICATIONS
AND/OR DIMENSIONS WITHOUT CONTACTING AND RECEIVING WRITTEN
CONFIRMATION FROM SMS PROJECT MANAGER.
2) SMS IS NOT AN ARCHITIECTURAL OR ENGINEERING FIRM. DRAWINGS
SUPPLIED BY SMS ARE NOT CONSTRUCTION DRAWINGS. THEREFORE,
THESE DRAWINGS ARE TO BE USED ONLY FOR INFORMATION TO
COMPLEMENT ACTUAL CONSTRUCTION DRAWINGS AVAILABLE FROM A
CUSTOMER APPOINTIED ARCHITECTURAL REPRESENTATIVE OR A
CUSTOMER 'S ENGINEERING DESIGN GROUP. THE CUSTOMER'S
ARCHITECT AND GENERAL CONTRACTOR SHALL BE ULTIMATELY
RESPONSIBLE FOR COMPLIANCE WITH ALL APPLICABLE CODES AND
PROFESSIONAL DESIGN REQUIREMENTS.
3) THE CUSTOMER IS RESPONSIBLE FOR ALL ROOM AND AREA
PREPARATION COSTS, PROFESSIONAL FEES, PERMITS, REPORTS, AND
INSPECTION FEES.
4) EQUIPMENT WARRANTIES, EXPRESSED OR IMPLIED ON THE PART OF
SMS SHALL BE CONTINGENT UPON STRICT COMPLIANCE WITH THE
ARCHITECTURAL, STRUCTURAL, ELECTRICAL, MECHANICAL AND
RECOMMENDATIONS AND REQUIREMENTS CONTAINED IN THESE
DRAWINGS, UNLESS SPECIFIED OTHERWISE.
5) ALL DIMENSIONS SHOWN ARE TAKEN FROM FINISHED SURFACES
UNLESS SPECIFIED OTHERWISE.
6) THIS DRAWING DOES NOT PROVIDE RADIATION SHIELDING
REQUIREMENTS FOR X-RAY AND ASSOCIATED EQUIPMENT. THE
CUSTOMER IS RESPONSIBLE FOR CONSULTING WITH A REGISTERED
RADIATION PHYSICIST. ACTUAL PROTECTION REQUIREMENTS SHALL BE
SPECIFIED BY A REGISTERED RADIATION PHYSICIST AT CUSTOMER'S
ENGAGEMENT AND EXPENSE. RESPONSIBILITY FOR ALL INFORMATION AS
TO THE ROOM LOCATION, USE, AND NUMBER OF ANTICIPATED
EXAMINATIONS TO BE PERFORMED PER TIME PERIOD SHALL BE
PROVIDED TO THE PHYSICIST BY THE CUSTOMER. THE CUSTOMER
SHALL FURTHER TAKE ALL RESPONSIBILITY IN THE COMMUNICATION
AND COORDINATION OF ACTIVITIES OF THE RADIATION PHYSICIST AND
THE ARCHITECTURAL REPRESENTATIVE.
7) SMS SHALL BE RESPONSIBLE FOR SMS EQUIPMENT INSTALLATION
AND CALIBRATION, CONNECTION AND INSTALLATION OF SMS PROVIDED
CABLES, AND CONNECTION OF CONTRACTOR PROVIDED WIRES TO SMS
EQUIPMENT. IN THE EVENT THAT SPECIFIC TRADE RULES OR LICENSE
REQUIREMENTS PROHIBIT THIS, THE CUSTOMER SHALL INITIATE THE
SERVICES OF APPROVED OTHER CONTRACTORS AND PAY FOR
SELECTED, APPROVED PARTIES TO PERFORM THIS WORK WITH JOB
SUPERVISION TO BE PROVIDED BY SMS. CALIBRATION WHEN
ACCOMPLISHED OUTSIDE OF NORMAL INSTALLATION SEQUENCES DUE
TO CONTRACTOR OR TRADE RULE ACTIONS OR REQUIREMENTS SHALL
BE SUPPORTED BY, CHARGED TO, AND ACCEPTED BY THE CUSTOMER
AS AN ADDITIONAL INSTALLATION EXPENSE.
8) THE CUSTOMER SHALL VERIFY WITH SMS PROJECT MANAGER FINAL
INSTALLATION DRAWINGS THE LOCATIONS AND TRAVEL OF ALL
ANCILLARY EQUIPMENT TO BE CEILING OR WALL MOUNTIE□ (IE: O.R.
LIGHTS, MEDICAL GAS COLUMNS, PHYSIOLOGICAL MONITORING
INJECTORS, CRT PLATFORMS, SPRINKLER HEADS, SMOKE DETECTORS,
ELECTRICAL OUTLETS, HVAC GRILLES, SPEAKERS, AND GENERAL ROOM
LIGHTING, ETC.).
9) THE GENERAL CONTRACTOR/CUSTOMER SHALL BE RESPONSIBLE
FOR ALL FINAL PAINT, TOUCH-UP AND ANY COSMETIC OR TRIM WORK
WHICH NEEDS TO BE OR IS REQUIRED TO BE COMPLETED AFTER THE
INSTALLATION OF THE SMS EQUIPMENT AND ANY ASSOCIATED SUPPORT
APPARATUS.
ENVIRONMENTAL/POWER AUDIT
AS AN INDICATION OF OUR COMMITMENT TO QUALITY, SIEMENS
MAY, AT NO COST TO YOUR FACILITY, CHECK THE OPERATING
ENVIRONMENT AFTER SYSTEM TURNOVER TO DETERMINE IF THE
REQUIREMENTS FOR TEMPERATURE, HUMIDITY, POWER, AND
GROUNDING ARE MET AS PER SIEMENS' PUBLISHED
SPECIFICATIONS. SIEMENS WILL GENERATE A WRITTEN REPORT
DETAILING THE ENVIRONMENTAL AND ELECTRICAL CONDITION OF
THE SITE AFTER TURNOVER AND WILL SHARE THE REPORT WITH
YOU. IN THE EVENT WE IDENTIFY ANY ENVIRONMENTAL/POWER
DEFICIENCIES AT THE SITE, YOUR FACILITY WILL BE REQUESTED
TO CORRECT DEFICIENCIES WITHIN THIRTY (30) DAYS. SHOULD
ANY CORRECTIVE ACTIONS BE NECESSARY, AND UPON REQUEST,
SIEMENS WILL PROVIDE GUIDANCE IN AN EFFORT TO FACILITATE
RESOLUTION. PLEASE BE ADVISED THAT AFTER 30 DAYS
NOTICE ANY REPAIR OR MAINTENANCE SERVICES NECESSITATED
BY SEVERE DEFICIENCIES WILL FALL OUTSIDE YOUR WARRANTY
COVERAGE.
RESOURCE LIST (SMS USE ONLY)
DESIGNATION PG NUMBER
S'1MIII IN1EVO EXCEL MEVO t INTEVO 6 AND IN1EVO 16 NM02-001.891.11.05.02
DATIE
10/14
SYMBIA INTEVO
REV 2
SIEMENS
VA SALT LAKE CITY
& 06/27/15 UPDATE BACK<ll�D WltiS TD MAllll
ARCHITECTS BACKGRDUNO ORAMNG
& 06/05/15 R-101R3 \OOION DATED 01/16/15
I.PPROO Bl' TTIE CUSflVR FOO ANAIS.
SYM DATE DESCRIPTION
-ISSUE BLOCK-
500 FOOTHILL DRIVE, SALT LAKE CITY, UT 84148
-11c ._.sc c): r::·c==r:x:,1._.c-1:::1·.J �) Tl I C: T-1 I l'.I n.�.1, './,.- ITI GI.JT :-:.IHilH-.r:, A.Jlf---JR /Al U\I ''ii/I I
1-!_SLLI 1··.J 1··1.':J�LC.JI (J\ UI\JL)_ .:
F-LL E:<TElr 1J=- T-::: L,'\\\".
ALL RIGHTS ARE RESERVED.
SCALE;
AS NOTED
ROOM 1 E09 - SYMBIA INTEVO INTEVO 6
PROJECT #:
1403947
SHEET OF
DATE:
1 7
06/27/15
DRAWN BY:
M. YATZUS
SHEET:
A-101 z 0 -r u J r w z u r z r z w
J u w u z w w w sample site planning documents
| Project #693-CSI-306 Siemens Site Planning Info Room 220 |
| VA259-15-R-0246-A00004003 |
File details come from the government source that posted it.